分光光度计常见故障

分光光度计常见故障
分光光度计常见故障

721型可见分光光度计故障维修

分光光度计作为一种理化分析仪器,广泛应用于工厂、矿山、医院以及科研单位的化验室,可在光谱区域360~800nm内对样品进行定性定量的分析。下面以最常见的721型可见分光光度计为例,分析它的常见故障与维修。

1、光路无光或能量档显示不到1 00% (其余部分正常) 该现象说明仪器电路无故障,只是光路系统需要调整。将波长盘转到580nm 处,应有一明显、完整的长方形黄色光斑,若没有或很模糊,应对光路进行调整。首先,按单色器入射狭缝高度,调整好光源灯高度,然后左右调整反射镜位置,使反射光进人人射狭缝。固定好反射镜后,继续调整反射镜背部螺丝,使进入单色器入射狭缝的光强为最大。

2.开机后光源灯不亮应先检查灯电压和光源灯的“通、断”情况。光源灯烧毁的现象较少,可能灯电压没有加上,此时应检查灯电源稳压部分。在灯稳压电路中,集成运放F004A为核心元件,在运放外围元件完好的情况下,调整运放的输入端电位器 W1,在输出端应有一电压变化,若没有,说明运算放大器损坏。然后,再检查功率放大管,若还没有解决,应向前检查整流块,否则,就是变压器故障。

3、旋转100%,光源灯亮度变化不大 (其余部分正常) ‘ 该现象也属于整机调整,仪器面板上100%旋钮均为细调,W1为粗调。将仪器零点调整好后,将面板上 100%旋钮调到最大,用万用表测量灯电压,旋转机内粗条电位器。当万用表显示的灯电压为最大时,停止调整,此现象即消失。

4、波长超差波长误差在检定结果上可分为两种情况:第一种情况为,误差方向一致,并呈线性,可通过转动波长度盘消减波长误差。第二情况为,误差方向不一致,有正差有负差。误差的变化可提示一个使误差呈线性的调整方向,然后转动波长度盘消除线性误差,依然超差的部分,再通过调节准直螺钉,使波长误差方向一致并呈线性,再配合移动波长度盘,使误差调到最小,如此反复调整。

对于第二种情况波长误差的调整方法通常为:使用干涉滤光片的标准峰值波长为438.2rim、553.5nm、638.5nm,调整准直镜,使580nm处样品室为黄光,测得峰值波长为420nm、550nm、640nm,最大最小峰值差为640nm-420nm=220nm,标准锋值差 638.5nm-438.2nm=200.3nm,调准直镜使红片峰值在650nm,测紫片为422nm,峰值差为228nm,说明调整方向不对,再用红片调整准直镜~q628nm,测紫片为423nm,峰值差为205nm,接近于标准峰值差,此时,逆时针转动波长度盘,由423nm至438nm,测得峰值为436nm、558nm、642nm,此时误差在合格范围内。

另外,如果用标准波长为466.2nm、554.Onm、676.3nm的干涉滤光片测得 721分光光度计的波长为476nm、554nm、664nm,最大最小峰值差为 188nm,而标准峰值差为210.1nm,需要扩展。用红片调至686nm,测另两片为567nm、483nm。此时顺时针转动波长度盘 1 9nm,测得值为474nm、557nm、679nm,再顺时针转 4nm,测得值为 467nm、550nm、672nm,误差合格。

由此可见,通过观察所测数据的变化,可使仪器的状态逐步得到改善,直至合格。一般情况下,当仪器波长范围需要扩展时 (即实测仪器两端的波长差小于标准波长差),可调准直镜螺丝,使峰值波长测得值高于标准值波长,然后顺时针移动波长度盘,若仪器波长需要缩小时 (即实测仪器两端的波长差大于标准波长差),可调整准直镜螺丝,使峰值波长测得值低于标准值波长,然后逆时针移动波长度盘,通过这种反复细致的调整,可是仪器处于良好状态。

721分光光度计电路故障维修

故障现象一:电源指示灯亮,光源灯不亮。

原理分析:根据其电源理图分析,光源灯不亮故障原因一般出现在光源灯稳压电路上。光源灯稳压电源是一个串联调整式稳压电路,采用集成运算放大器(5G23A)A1作放大环节,其输出电压为-3.7V至-11.55V连续可调。光源灯稳压电源电路的取样、基准电压均受放大器(5G23A)A1稳压电源电路控制。放大器稳压电源也是一普通的串联调整式稳压电源,输出六种电压。其中±12V电压供运算放大器使用,±8V供放大器输入级使用,±0.7V供调零使用。此外,+8V还作钨灯稳压电源基准电压及光电管阳极电压使用。

±8V、±0.7V、-12V等都不是独立电源,如果+12V电源出了问题,会影响其它两组电压(±8V和±0.7V)的正常工作,导致光源灯不亮。

故障处理:打开仪器的后面板,观察灯丝良好。打开电源开关,用三用表检查灯泡电压工作电压为OV,说明钨灯稳压电源电压没有送来。旋转调零选钮W3,发现表针可以调零,说明±0.7V调零电压正常,检测8V电源电压正常。可以断定放大器稳压电源电路工作正常,故障在钨灯稳压电源电路上,用三用表在线测量易损件,发现电源调整管3DDl5A可疑,焊下测量,发现其b-e级击穿,更换同型号的三极管,再测量其外围元件,没发现可疑之处,开机,工作正常。

故障现象二:各调节旋钮可以调节,但在测量过程中,电表指计不稳定。

原理分析:造成电表指计不稳定的因素很多,一般可以从以下几个方面考虑:

(1)光源灯电压不稳。调节W2钨灯工作电压应在-11.5±0.5V变化,而且输出稳定。在钨灯工作时,其电压波动应小于0.005V。(2)光门部件漏光,光电管、运算放大器受潮。这时检查暗盒下面的硅胶,看其颜色是否变红(粉),如变色,应及时将硅胶筒旋出,给以更换或烘烧。(3)光电管老化、损坏或放大器电路故障。放大器是由两个输入阻抗较高的结型场效应管3DJ6F作为输人级的,其两只管子是配对的,其特性相同,偏置对称,这样可以将输入的高阻抗信号变为低阻抗信号,有效地消除电压失调造成的漂移。

故障处理:首先对变色的硅胶进行烘烤,待变蓝后再装入硅胶筒。其次,进行光源灯电压测量,未发现异常。试机,电表指计仍旧不稳。将光电管上的两根引线焊下(排除光电管的影响),再试机,故障依旧。断定故障出现在放大器电路上,检测其周围元器件,未发现异常。怀疑是其输入级结型场效应管(BGl2、BGl3)对称性变差引起的,由于其输入级性能要求较高,为使其性能相同、偏置对称,更换两个型号相同、对称良好的3DJ6F管子后,开机故障排除。

计算机网络故障处理与维护方法(毕业论文)

五年制高职商贸信息专业毕业论文 计算机网络故障处理与维护方法 班级 姓名 学号 指导老师

目录 【摘要】 (1) 一、计算机网络故障的分类 (1) (一)计算机网络物理故障 (4) (二)计算机网络逻辑故障 (3) 二、计算机网络常见故障的处理 (1) (一)本地连接断开 (1) (二)本地连接收限制或无连接 (1) (三)本地连接正常,但浏览器无法连接网页 (1) 三、如何加强网络的维护 (1) (一)概括的说,应做到: (4) (二)具体来说,应该做到: (3) 四、结论 (8) 【参考文献】 (3)

计算机网络故障处理与维护方法 【摘要】 本文就网络中常见故障进行分类,针对各种常见网络故障提出相应的解决方法,并就如何加强网络的维护进行了概括论述。 网络出现故障是极普遍的事,其种类也多种多样,在网络出现故障时对出现的问题及时进行维护,以最快的速度恢复网络的正常运行,掌握一套行之有效的网络维护理论方法和技术是至关重要的。 【关键词】 网络故障分类处理维护 一、计算机网络故障的分类 计算机网络故障主要是指,用户在使用计算机网络过程中或网络在运行过程中出现的问题,导致计算机网络不能正常使用。通常计算机网络故障可以按照其故障的性质,分为物理故障和逻辑故障。 (一)物理故障: 物理故障也就是硬件故障,一般是指网络设备或线路损坏、接口松动、线路受到严重干扰,以及因为人为因素导致的网络连接错误等情况。出现该类故障时,通常表现为网络断开或时断时续。物理故障主要包括: (1)线路故障

线路故障的发生率在日常的网络维护中非常高,约占发生网络故障的60%~70%。线路故障包括线路的损坏和线路受到严重干扰。 (2)接口故障 接口故障通常包括插头松动和端口本身的物理损坏。如:双绞线RJ45接头的损坏。 (3)交换机或路由器故障 交换机或路由器故障在这里是指设备出现物理损坏,无常工作,导致网络不能正常运行的情况。 (4)网卡故障 网卡也称网络适配器,大多安装在计算机的主机部。通过主机完成配置和。网卡故障主要包括网卡松动、主机网卡插槽故障、网卡本身物理故障等。 (二)逻辑故障: 逻辑故障也称为软件故障,主要是指软件安装或网络设备配置错误所引起的网络异常。与硬件故障相比,逻辑故障往往要复杂得多。常见的网络逻辑故障有:主机逻辑故障、进程或端口故障、路由器故障等。 (1)主机逻辑故障 主机逻辑故障通常包括网卡驱动程序、网络通信协议或服务安装不正确、网络地址参数配置有误等。对计算机网络用户来讲,该类故障是十分常见的网络故障之一。 (2)进程或端口故障 进程或端口故障是指一些有关网络连接的进程或端口由于受到病毒或系统

智能仪表论文智能仪表的温度控制系统

智能仪表的温度控制系统 摘要:随着总线智能仪表技术的不断发展,智能化数字仪表功能和应用日益广泛。本系统是基于CPLD和A T89S52单片机设计采用专家PID控制的总线型温度控制系统。系统具有稳定度高、精度高和抗干扰能力强的优点,并且可以在工业生产过程中进行实时监控,具有将监控数据远程传输给控制终端的能力。应用实践证明,系统各方面均较为完善,具有很好的应用意义和市场价值。 关键词:温度控制;CPLD;PID控制;智能模糊算法 1 温控系统现状 智能仪表中的微处理器具有一定的数据存储和处理能力,在软件的配合下,智能仪表功能可以大大增强,用于温度测量的温度传感器如热电偶、热电阻,因其温度与热电势(或电阻)的关系是非常复杂的曲线关系,因此寻求合适的温度与热电势(或电阻)的关系式,以应用于温度测量及计算,是决定智能仪表温度测量精度高低的关键。 随着现代科学技术的迅速发展及工业控制中自动化要求的提高,对现场检测控制仪表的智能化程度的要求也越来越高,并且要求仪表具备较强的远距离通信的功能,智能仪表逐渐向数字化、网络化和智能化方向发展。在现代工业生产作业中,温度控制是各种工业生产过程中的重要因素。尤其是在钢铁、食品、化工、冶炼等行业的生产过程中,更加需要严密的温度控制系统。而且在这样的系统中通常是需要监测和控制多个温度参数并且需要将数据远程传输到控制终端。在以往的温度控制系统中,通常有以下的不足和缺陷:系统精确度不够,只能检测单个温度参数;温度控制仪表中检测使用电压较低,不能直接应用于控制的对象系统。基于以上的考虑,在设计系统的过程中增加了相应的功能,以便提高系统对整体效率和性能。系统采用AT89S52为核心控制器,利用A\D转换器和模糊智能算法实现四路温度监测和控制功能,并能通过远程通信传输到控制终端。 2 系统设计 系统主要组成模块:AT89S52 单片机、CPLD、信号输入、信号输出以及串口通信,如图1 所示。单片机电路:采集键盘的输入信号、串行端口的传输信号、液晶屏幕的显示信号、过零检测信号处理。CPLD模块:产生PWM控制信号,利用PWM输出的控制信号来控制加热器件的工作状态。功率控制电路模块:采用可控硅输出光耦的耦合形式,利用关断与导通的时间比值作为参数调节器件的功率。芯片采用MOC3081,是零触发双向可控硅模式芯片。这种设计方式可以减少后续功能器件对前端器件模块稳定性的影响。

横河压力变送器常见故障处理方法

2、典型故障的处理方法 2.1 对测量超限的处理方法通过研究分析,发现此类故障通常与以下因素有关:① 仪表操作使用不当以抚顺石油一厂酮苯装置 C-101液位控制系统(LICA-1201)为例,如图1所示,由于仪表始终在高液位(100%以上)运行,或仪表始终在低液位(5%以下)运行,都有可能使仪表指示为超限。因此,要求工艺操作人员应能根据工艺流程及工艺控制要求正确判断出是仪表故障还是工艺操作不当。所以,需要工艺人员和仪表维护人员密切配合,保证工艺介质在仪表所能测量范围内,避免使操作人员误认为仪表故障。 图1 C-101 液位控制系统工艺图②仪表量程选择不当在对该厂酮苯装置中EJA 智能双法兰变送器测量量程检查时,均发现变送器量程存在设计计算错误,如对LICA-1201等变送器在DCS工程师站上检查它们的量程时,发现双法兰量程无迁移,这是造成仪表测量不准及超限的重要原因,如图2所示。 图2 塔101 量程计算参数图原设计采用量程为0~19.71kPa,无量程迁移,因此测量结果在仪表量程之外,出现测量超限情况。实际上对此台仪表应按下面的方法进行量程计算:已知:仪表可测范围,介质比重,毛细管硅油比重。求仪表量程。求解方法:仪表的量程是指当液位由最低升到最高时,液面计上所受的压力,故量程为:当液面最低时,液面计正、负压室的受力为:液面计迁移量为: =-2.65=-2.65×1.07×9.81 =-27.82kPa P+>P-,故为负迁移。按上述计算修改量程后,仪表运行即正常。因此,只有按正确的计算方法及引用迁移量来进行计算才能保证仪表量程的准确。 2.2 安全柵不配套造成仪表无输出及测量不准由于智能变送器要求使用与之配套的安全柵,当用了未取得与智能变送器配套许可证的安全柵后,大部分都会出现这样那样的问题,其主要故障有:①安全柵电

常见网络故障处理方法

故障处理分析方法 【互联网专线】 一、客户所有电脑不能上网,网管正常 1、可以远程登陆路由器可能的故障原因及解决思路: 1.有可能是客户侧路由器配置口插入网线。 2.内网中ARP病毒。 3.交换机故障,查看是否上电 4. 查看交换机连接到路由器的网线是否有松动 (1)客户无法ping通网关(192.168.1.1) 查看是否获取到正常的IP地址(192.168.1.X),或者禁用网卡再启用,还是不行就重启下电脑。 (2)客户可以ping通网关 可以尝试重启下电脑,因为有可能是客户浏览器问题导致。 (3)本地连接受限 1. 查看下路由器DHCP是否有配置,如果有就叫客户禁用网卡再启用, 2. 还是不行就重启下电脑。 3. 如果还是受限就叫客户手动配置IP地址(192.168.1.100以后的没有人用的)跟首选DNS (218.207.217.241或219.141.136.10) 2、无法远程登陆路由器可能的故障原因及解决思路: 1.查看网管是否掉点,查看系统附件是否有路由器配置附件,打开查看是否有配置远程登 入 2.尝试PING客户侧外网IP地址,请到跳板上PING,因为部分集团有配置禁PING,只 有跳板才可以PING通。 3.查看设备是否有上电。 假如无法远程登陆路由器没必要判断以下三种情况(除非可以PING通客户侧外网IP)(1)客户无法ping通网关 xxxxx (2)客户可以ping通网关 xxxxx (3)本地连接受限 xxxx 二、客户部分电脑不能上网 1. 基本为客户侧内网问题,假如对方懂网络就直接跟他说我们可以连接到你们那边的路由器,外网一切正常,请查看下内网问题。(内网的问题基本为客户自己解决) 2. 可以与客户沟通,将设备重启

常见网络故障处理方法完整版

常见网络故障处理方法 Document serial number【NL89WT-NY98YT-NC8CB-NNUUT-NUT108】

NO: 常见网络故障处理方法 作业指导书 (第A版) 编制人: 审核人: 批准人:

版本修订记录

目录 3、交换机常见故障及解决方 法 (7) 1光缆链路的主要故障 一般分为两步: 收发器暂时不要和交换设备连接。我们先使用两台笔记本电脑连接收发器,两台电脑之间互Ping。待测试好了以后再连接交换设备。一台笔记本电脑ping 另外一台电脑的IP 地址,例如,PC1 Ping PC2, 命令为Ping –t –l 65000。如果丢包少于5%,则比较正常,如果丢包较多,则需要仔细检查。 收发器连接交换设备以后,我们建议仍然使用Ping 的命令来测试,例如PC1 PingPC2, 命令为Ping –t –l 1500, 数据包长度一般不是65500,因为不同的交换机或路由器对包长的限制不同。但是1500 字节的数据包应该很少丢包,否则需要仔细检查。 故障现象: 1光缆熔接不良(有空气) 2光缆断裂或受到挤压 3接头处抛光不良 4接头处接触不良 5光缆过长 6核心直径不匹配

7填充物直径不匹配 8弯曲过度(弯曲半径过小) 2光纤故障排除方法 首先看光纤收发器或光模块的指示灯和双绞线端口指示灯是否已亮 2.1.1如收发器的光口(FX)指示灯不亮,请确定光纤链路是否交叉链接光纤跳线 一头是平行方式连接;另一头是交叉方式连接。 2.1.2如A收发器的光口(FX)指示灯亮、B收发器的光口(TX)指示灯不亮,则故障在A收发器端: 一种可能是:A收发器(TX)光发送口已坏,因为B收发器的光口(FX)接收不到光信号;另一种可能是:A收发器(TX)光发送口的这条光纤链路有问题(光缆或光线跳线可能断了)。 c、双绞线(TP)指示灯不亮,请确定双绞线连线是否有错或连接有误请用通断测试仪检测;(不过有些收发器的双绞线指示灯须等光纤链路接通后才亮。) d、有的收发器有两个RJ45端口:(To HUB)表示连接交换机的连接线是直通线;(To Node)表示连接交换机的连接线是交叉线(接单机); e、有的发器侧面有MPR开关:表示连接交换机的连接线是直通线方式;DTE开关:连接交换机的连接线是交叉线方式。 2、光缆、光纤跳线是否已断 a、光缆通断检测:用激光手电、太阳光、发光体对着光缆接头或偶合器的一头照光;在另一头看是否有可见光如有可见光则表明光缆没有断。 b、光纤连线通断检测:用激光手电、太阳光等对着光纤跳线的一头照光;在另一头看是否有可见光如有可见光则表明光纤跳线没有断。 3、半/全双工方式是否有误 有的收发器侧面有FDX开关:表示全双工;HDX开关:表示半双工。 4、用光功率计仪表检测 光纤收发器或光模块在正常情况下的发光功率:多模:-10db--18db之间;单模20公里:-8db--15db之间;单模60公里:-5db--12db之间;如果在光纤收发器的发光功率在:-30db--45db之间,那么可以判断这个收发器有问题。 1. TXLINK灯不亮; 答:造成该故障的原因有二,一为接错双绞线,本收发器和光纤头及指示器同侧的RJ45口接PC机用交叉双绞线,接HUB或SWITCH用平行双绞线;二为通过双绞线所连的电口不是100M速率。 2. FXLINK灯不亮; 答:原因一:光纤线接错,正确接法为TX-RX; 原因二:传输距离太长或中间损耗太大,超过本产品的标称损耗,解决办法为采取办法减小中间损耗或是更换为传输距离更长的收发器; 3.五灯全亮或指示器正常但无法传输; 答:一般关断电源重启一下即可恢复正常; 4.光纤正常连接后FXRX灯常亮;

网络维护及常见故障排除

网络维护及常见故障排除 第一章排除故障过程 1.1识别故障现象 在你排故障之前,也必须确切地知道网络上到底出了什么毛病,是不能共享资源,还是找不到另一台计算机,如此等等。知道出了什么问题并能够及时识别,是成功排除故障最重要的步骤。为了与故障现象进行对比,你必须知道系统在正常情况下是怎样工作的,反之,你是不好对问题和故障进行定位的。 识别故障现象时,应该向操作者询问以下几个问题: (1)当被记录的故障现象发生时,正在运行什么进程(即操作者正在对计算机进行什么操作)。 (2) (2)这个进程以前运行过吗? (3)(3)以前这个进程的运行是否成功? (4) (4)这个进程最后一次成功运行是什么时候? (5)(5)从那时起,哪些发生了改变? (6)带着这些疑问来了解问题,才能对症下药排除故障。 1.2对故障现象描述 当处理由操作员报告的问题时,对故障现象的详细描述显得尤为重要。如果仅凭他们的一面之词,有时还很难下结论,这时就需要你亲自操作一下刚才出错的程序,并注意出错信息。对此在排除故障前,可以按以下步骤执行: (1)收集有关故障现象的信息; (2)对问题和故障现象进行详细描述; (3)注意细节; (4)把所有的问题都记下来; (5)不要匆忙下结论。

第二章网络故障原因 2.1 网络连通性故障 1、故障表现 连通性故障通常表现为以下几种情况: ①计算机无法登录到服务器; ②计算机无法通过局域网接入Internet; ③计算机在"网上邻居"中只能看到自己,而看不到其他计算机,从而无 使用其他计算机上的共享资源和共享打印机 ④计算机无法在网络内实现访问其他计算机上的资源; 2、故障原因 以下原因可能导致连通性故障: ①网卡未安装,或未安装正确,或与其他设备有冲突; ②网卡硬件故障; ③网络协议未安装 ④IP地址设置错误; ⑤网线、信息模块故障; 3、排除方法 ①确认连通性故障 当出现一种网络应用故障时,如无法接入Internet,首先尝试使用其他网络应用,如查找网络中的其他计算机,或使用局域网中的Web浏览等。如果其他网络应用可正常使用,如虽然无法接入Internet,却能够在"网上邻居"中找到其他计算机,或可ping到其他计算机,即可排除连通性故障原因。如果其他网络应用均无法实现,继续下面操作。 ②用ping命令排除网卡故障 使用ping命令,ping本地的IP地址或计算机名,检查网卡和IP网络协议是否安装完好。如果能ping通,说明该计算机的网卡和网络协议设置都没有问题。 因此,应当检查网线和交换机的接口状态,如果无法ping通,只能说明TCP/IP协议有问题。这时可以在计算机的"控制面板"的"系统"中,查看网卡是否已经安装或是否出错。如果在系统中的硬件列表中没有发现网络适配器,或网络

智能电磁流量计常见故障分析及解决

智能电磁流量计常见故障分析及解决智能电磁流量计是一种速度式仪表。除可测量一般导电液体的体积流量外,还可用于测量强酸强碱等强腐蚀液体和泥浆、矿浆、纸浆等均匀的液固两相悬浮液体的体积流量。广泛应用于石油、化工、冶金、轻纺、造纸、环保、食品等工业部门及市政管理,水利建设、河流疏浚等领域的流量计量。常见故障,有的是由于仪表本身元器件损坏引起的故障,有的是由于选用不当、安装不妥、环境条件、流体特性等因素造成的故障,如显示波动、精度下降甚至仪表损坏等。它一般可以分为两种类型:安装调试时出现的故障(调试期故障)和正常运行时出现故障(运行期故障)。 (1)调试期故障调试期待故障一般出现在仪表安装调试阶段,一经排除,在以后相同条件下一般不会再出现。常见的调试期故障一般由安装不妥、环境干扰以及流体特性影响等原因引起。 1)安装方面通常是智能电磁流量计传感器安装位置不正确引起的故障,常见的如将传感器安装在易积聚气体的管系最高点;或安装在自上而下的垂直管上,可能出现排空;或传感器后无背压,流体直接排入大气而形成测量管内非满管。 2)环境方面通常主要是管道杂散电流干扰,空间强电磁波干扰,大型电机磁场干扰等。管道杂散电流干扰通常采取良好的单独接地保护就可获得满意结果,但如遇到强大的杂散电流如电解车间管道,有时在两电极上感应的交流电势峰值可高达尚需采取另外措施和流量传感器与管道绝缘等。空间电磁波干扰一般经信号电缆引入,通常采

用单层或多层屏蔽予以保护。 3)流体方面被测液体中含有均匀分布的微小气泡通常不影响电磁流量计的正常工作,但随着气泡的增大,仪表输出信号会出现波动,若气泡大到足以遮盖整个电极表面时,随着气泡流过电极会使电极回路瞬间断路而使输出信号出现更大的波动。 低频方波励磁的电磁流量计测量固体含量过多浆液时,也将产生浆液噪声,使输出信号产生波动。 测量混合介质时,如果在混合未均匀前就进入流量传感进行测量,也将使输出信号产生波动。 电极材料与被测介质选配不当,也将由于化学作用或极化现象而影响正常测量。应根据仪表选用https://www.360docs.net/doc/e31888172.html,或有关手册正确选配电极材料。 (2)运行期故障运行期故障是智能电磁流量计经调试并正常运行一段时期后出现的故障,常见的运行期故障一般由流量传感器内壁附着层、雷电打击以及环境条件变化等因素引起。 1)传感器内壁附着层由于电磁流量计常用来测量脏污流体,运行一段时间后,常会在传感器内壁积聚附着层而产生故障。这些故障往往是由于附着层的电导率太大或太小造成的。若附着物为绝缘层,则电极回路将出现断路,仪表不能正常工作;若附着层电导率显著高于流体电导率,则电极回路将出现短路,仪表也不能正常工作。所以,应及时清除电磁流量计测量管内的附着结垢层。 2)雷电打击雷击容易在仪表线路中感应出高电压和浪涌电流,使

医院网络维护常见网络故障及排除技巧

龙源期刊网 https://www.360docs.net/doc/e31888172.html, 医院网络维护常见网络故障及排除技巧 作者:王唯 来源:《科学与信息化》2018年第04期 摘要随着近两年来社会经济以及网络信息技术的不断发展,医院信息化建设的进程也越来越快。网络维护作为医院信息化建设过程中非常重要的内容,通过对医院网络系统运行过程中所存在的故障进行排除,不仅可以有效提高医院网络的运行效率,与此同时也能大大提高医院经济效益。医院作为信息资料十分丰富的场所,网络信息技术在医院管理过程中的应用有效提高了信息数据的传送以及保存效率,同时也为医院工作人员大大节省了工作时间和精力。但随着医院信息化建设的不断发展,网络故障问题也频繁出现,这些故障不仅造成了信息数据的丢失,并且给一些不法分子的病毒入侵提供了便利,基于这些情况,相关医院管理人员应该要加强对各部门工作程序的管控,通过科学、合理的方式对网络故障进行有效排除,从而促进医院信息化发建设的良好发展。 关键词医院网络维护;常见故障;排除技巧 Abstract with the continuous development of social economy and network information technology in the past two years, the process of hospital information construction has been getting faster and faster. Network maintenance as a very important content in the process of hospital information construction, through the hospital network system of faults in the operation process,not only can effectively improve the efficiency of the operation of the hospital network, at the same time also can greatly improve the hospital economic benefit. Hospital as a place of information is very rich, the network information technology application in the process of hospital management improve the efficiency of data transmission and preservation, as well as hospital staff work greatly saves time and energy. But with the continuous development of hospital informationization construction,network fault problem also appeared frequently, these failures not only caused the loss of information data, and gives some criminals network intrusion is convenient, based on these conditions, related to hospital managers should strengthen the control of various departments work procedures, through the scientific and reasonable way to effectively eliminate of network failure,so as to promote the development of the construction of hospital informatization. Keywords Hospital network maintenance; Common failures; Rule out skills 随着网络信息技术的不断进步,医院信息化建设也有了良好的发展趋势。但根据相关调查结果显示,我国部分医院管理人员还没有正确地意识到网络维护的重要性,在医院网络运行过程往往会出现各种故障问题,这些问题在一定程度也严重影响医院信息化建设的发展。因此,相关医院管理人员应该要加强对内部网络的维护和管理,通过对网络运行过程出现故障进行有效排除,从而确保医院网络系统能够安全、稳定运行。另外,医院网络管理过程中相关维护人员应该要充分掌握好各种常见网络故障的排除方法,这样便可以及时、高效地解决各种突发故

基于单片机的智能控制仪表简单设计

智能控制仪表课程设计 ----基于51单片机地智能控制仪表简单设计 学校:红河学院 专业:电气工程及其自动化 姓名:和红昌 学号:201005050354 班级:10级电气叁班 指导老师:牛林

第1章引言 仪器仪表是人类认识世界地工具,人们借助于各种仪器仪表对各种物理量进行度量,反映其大小与变化规律. 随着人类认识能力地提高与科学技术不断进步,仪器仪表技术得到了飞速发展.50 年代以前,仪器仪表多为指针式,其理论基础是机电学. 从50 年代起,电子技术特别是数字技术地发展,给仪表行业带来了生机,各种数字式仪表相继问世,许多传统地指针式仪表相继被淘汰,数字仪表使仪表外观耳目一新,数据表达能力与总体性能都大幅提高. 70 年代中期,随着微处理器地出现以及单片机地兴起与应用,设计者将计算机特有地许多优点引入仪表设计,随之产生了一代崭新地智能仪表,使仪表逐渐由数字型向智能化发展,其功能也由单一显示功能转变为具有信息处理、传输、存贮、显示、控制等功能,使仪表性能产生了质地飞跃.,品种繁多. 目前,我国仪器仪表有13 大类,1 300 多个产品. 其中自动化仪表及控制系统是和国民经济各产业部门关系最为密切地一类产品,其传感变送单元与主控装置及I/O接口均正朝智能化方向发展.在本设计中采用以单片机作为仪表核心控制器件,可以利用A/D转换芯片对标准信号进行采集、转换,将输入地模拟量转换成单片机能够检测地数字量进行分析和监测控制,同时可以利用键盘显示电路将相关数据进行显示.与此同时通过所查阅地资料我还了解到随着测量技术地发展和微处理器地广泛应用,单片机系统地电路越来越复杂,而系统地可靠性问题也越来越突出,一般地单片机系统在工业现场等恶劣地环境下容易死机,因此系统在这些场合要保证能够稳定地工作就必须外加监视电路,在设计中采用了美国集把关定时器、电压监控和串行EEPROM三项功能于一体地专用集成芯片 X5045.该芯片地应用将有利于简化单片机系统地结构,增强功能、降低系统地成本,尤其是大大地增加了系统地可靠性.X5045中地看门狗对系统提供了保护功能.当系统发生故障而超过设置时间时,电路中地看门狗将通过RESET信号向CPU作出反应.X5045提供了三个时间值供用户选择使用.它所具有地电压临控功能还可以保护系统免受低电压地影响,当电源电压降到允许范围以下时,系统将复位,直到电源电压返回到稳定值为止.本次毕业设计旨在掌握智能控制仪表地设计方法,同时掌握在开发系统下实现部分软件地仿真方法. 第2章控制系统地硬件设计 硬件组成智能仪表地硬件方框图如图2.1 图2.1 智能控制仪表地原理框图

常见仪表常见故障及处理办法

仪表常见故障检查及分析处理 一、磁翻板液位计: 1、故障现象:a、中控远传液位和现场液位对不上或者进液排液时液位无变化;b、现场液位计和中控远传均没有问题的情况下,中控和现场液位对不上; 2、故障分析:a、在确定远传液位准确的情况下,一般怀疑为液位计液相堵塞造成磁浮子卡住,b、现场液位变送器不是线性; 3、处理办法:a、关闭气相和液相一次阀,打开排液阀把内部液体和气体全部排干净,然后再慢慢打开液相一次阀和气相一次阀,如果液位还是对不上,就进行多次重复的冲洗,直到液位恢复正常为止;b、对液位计变送器进行线性校验。 二、3051压力变送器:压力变送器的常见故障及排除 1)3051压力变送器输出信号不稳 出现这种情况应考虑A.压力源本身是一个不稳定的压力B.仪表或压力传感器抗干扰能力不强C.传感器接线不牢D.传感器本身振动很厉害E.传感器故障 2)加压变送器输出不变化,再加压变送器输出突然变化,泄压变送器零位回不去,检查传感器器密封圈,一般是因为密封圈规格原因(太软或太厚),传感器拧紧时,密封圈被压缩到传感器引压口里面堵塞传感器,加压时压力介质进不去,但是压力很大时突然冲开密封圈,压力传感器受到压力而变化,而压力再次降低时,密封圈又回位堵住引压口,残存的压力释放不出,因此传感器零位又下不来。排除此原

因方法是将传感器卸下看零位是否正常,如果正常更换密封圈再试。 3)3051压力变送器接电无输出 a)接错线(仪表和传感器都要检查) b)导线本身的断路或短路 c)电源无输出或电源不匹配 d)仪表损坏或仪表不匹配 e)传感器损坏 总体来说对3051压力变送器在使用过程中出现的一些故障分析和处理主要由以下几种方法。 a)替换法:准备一块正常使用的3051压力变送器直接替换怀疑有故障的这样可以简单快捷的判定是3051压力变送器本身的故障还是管路或其他设备的故障。 b)断路法:将怀疑有故障的部分与其它部分分开来,查看故障是否消失,如果消失,则确定故障所在,否则可进行下一步查找,如:智能差压变送器不能正常Hart远程通讯,可将电源从仪表本体上断开,用现场另加电源的方法为变送器通电进行通讯,以查看是否电缆是否叠加约2kHz的电磁信号而干扰通讯。 c)短路检测:在保证安全的情况下,将相关部分回路直接短接,如:差变送器输出值偏小,可将导压管断开,从一次取压阀外直接将差压信号直接引到差压变送器双侧,观察变送器输出,以判断导压管路的堵、漏的连通性 三、雷达液位计:

网络常见故障维修

网络常见故障处理方法 1.网络突然中断 网卡IP地址的设置错误 右键点击网上邻居,在弹开菜单中选择属性,然后继续右键点击本地网络,在弹开菜单中选择属性,进入本地连接属性栏,之后双击INTERNET协议( TCP/IP),可才看到自己的IP地址,子网掩码,DNS等相关设置。 在公司内部每个人的地址多不相同,如果你的地址与别人的相同,就会造成 IP地址冲突,导致上不去网络;而且配置与网关给的配置不一样,也有上不去网络 的可能。 网卡误操作被禁用 网卡被禁用后,在右下角将没有连接提示,需要用右键点击网络邻居,在弹开菜单中选择属性,然后继续右键点击本地网络,在弹开菜单中选择启用。 网线接触不良,网卡插错或插的不严实 这类故障通常因为设备的老化或者网络头的磨损导致的,这类故障要彻底解决的话需要更换交换机或者更换网络头。 交换机停止工作 通常是有人不小心碰到了电源,导致设备断电。通常这种情况发生,会导致掉电设备上所有的用户多会中断与网络的连接。 电脑中了恶性病毒 电脑在中病毒后,通常情况是系统运行速度变慢,上网速度也变的缓慢;当电脑中一些恶性病毒后,病毒会对一些常用端口发病毒包,从而导致电脑上不去网或 者一些软件无法正常使用。 2.网络正常,邮件收发有问题

邮件服务器设置错误 邮件服务一般需要用户设置SMTP服务器,POP3服务器以及用户名和密码;其中SMTP服务器是发件服务器,邮件的收发多是通过该服务器来发送,POP3服务器是收 件服务器,你收到的邮件多是从POP3服务器上传送到本地的。如果用户更改设置后,发现邮件能收不能发,或者邮件能发不能收,只需要查看响应的服务器设置就可以。 电脑操作系统故障导致邮件收发出现问题 ???此类故障主要因为电脑配置,系统稳定性所导致。当电脑配置教低,系统稳定性又很差时,电脑经常出现各种故障,有时出现突然邮件收发出现问题;一般此类问题解决方法就是重启操作系统。 邮件提供商的配置问题 当您在一个地方使用邮件服务很正常,换到另外一个地点后,邮件服务突然出 现问题,而上网正常,很又可能是邮件提供商或者当地的网络提供商对网络进行了一些安全设置,所以这时你需要联系邮件提供商或者当地的网络提供商来处理此类问题。 3.网络时断时续,很不稳定 ???当网络配置完后,发现网络时断时续,首先我们需要查看我们的路由等设备配置,查看网络设备配置上是否有任何问题。 ???如果网络时断时续是网络正常运行一段时候后才发生的,那我们需要查看路由设备的CPU利用率等相关数据,以此来确定是否问题来源于内部网络病毒。 ???如果确认上面那些多没有问题,那我们需要联系我们的网络提供商,一起配合检查线路。 4.网络故障查询经常使用的命令 Ping命令的使用技巧 ???Ping是个使用频率极高的实用程序,用于确定本地主机是否能与另一台主机交换(发送与接收)数据报。根据返回的信息,我们就可以推断TCP/IP参数是否设置得正确以及运行是否正常. Ping命令的常用参数选项: ???ping IP –t 连续对IP地址执行Ping命令,直到被用户以Ctrl+C中断。

常见网络故障的分析及排除方法

常见网络故障的分析及排除方法 【摘要】计算机网络是一个复杂的综合系统,网络故障十分普遍,故障种类也极其繁杂。本文在对具体的网络故障分析基础上,给出了相应的排除方法。 【关键词】网络故障;常见故障;分类诊断;物理故障;逻辑故障 一、网络故障的分类 网络故障的成因无非是硬件和软件两个方面。按照网络故障的性质,网络故障可划分为物理故障与逻辑故障两类。物理故障也叫硬件故障,是指由硬件设备所引发的网络故障。在硬件故障中线路故障、端口故障、集线器或路由器故障及主机物理故障是较为常见的几种故障。 逻辑故障又称为软故障,表现特征为网络不通,或者同一个链路中有的网络服务通,有的网络服务不通。究其根源,是由于设备配置错误或者软件安装错误所致。路由器逻辑故障、主机逻辑故障、病毒故障是几种常见的逻辑故障。 二、排除故障的具体方法 排除故障的方法是不外乎从软件设置和硬件损坏两个方面来考虑: ㈠物理故障及排除方法 1、线路故障最普遍的情况是线路不通,是网络中常见的故障。线路损坏或线路受到严重电磁干扰时最容易引发该故障。诊断此故障时,若线路很短,最直接的方法是将该网络线一端插入一台能够正常连入局域网的主机的RJ45插空内,另一端插入正常的集线器端口中,然后在DOS环境下,使用PING命令在本主机上检测线路另一端主机(或路由器)的端口能否响应,用TRACEROUTE命令检查路由器配置是否正确,根据检测结果进行判断;若线路稍长,不方便移动,可使用网线测试仪器进行线路检测;若线路太长,或线路由电信供应商提供,则需要与提供商协同检查线路,确认是否线路中间出现了故障。 对于存在严重电磁干扰的检测,可以使用屏蔽性能很强的屏蔽线在该线路上进行通信测试,若通信正常,表明存在电磁干扰。若问题依旧,可排除电磁干扰故障。 2、端口故障分为插头松动及端口本身的物理故障。此类故障一般会直接影响到与其相连的其他设备的信号灯状态。信号灯较直观,通过信号灯大体上可以判断出故障的发生范围及有可能存在的因素。检测时,首先应检查RJ45插头是否松动或检查RJ45接口是否制作完好,然后查看集线器或交换机的接口,如果某个接口存在问题,可以更换接口后再进行验证是否真的存在端口故障。 3、路由器或集线器故障会直接导致网络不通。这类故障也是网络上一种常见的故障,故障的现象与线路故障很相近,在诊断此种故障时,必须用专门的诊断工具来收集路由器的端口流量、路由表、路由器CPU温度、负载及路由器的内存余量、计费数据等数据。检测时,可采用替换排除法,用通信正常的网线和主机来连接路由器或集线器,若通信正常,表明路由器或集线器没有故障;反之则应调换路由器(或集线器)的端口来确认故障;很多情况下,路由器(或集线器)的指示灯表明了其本身是否存在故障,正常的情况下对应端口的指示灯为绿色指示灯。通过以上测试后,若问题依旧,可断定路由器或集线器上存在故障。 4、主机物理故障包括网卡物理故障,网卡插槽故障,网卡松动及主机本身故障。对于网卡插槽故障和网卡松动的诊断可通过更换网卡插槽来进行。如果更换插槽仍不能解决故障,可将网卡放到其他正常工作的主机上测试,若正常通信,是主机本身故障,若无法工作,是网卡物理物理故障,更换网卡故障可排除。

网络常见故障诊断排除论文

浅谈网络常见故障的诊断与排除 摘要:网络故障极为普遍,网络故障的种类也多种多样,要在网络出现故障时及时对出现故障的网络进行维护,以最快的速度恢复网络的正常运行,掌握一套行之有效的网络维护理论、方法和技术是关键。就网络中常见故障进行了分类,并对各种常见网络故障提出了相应的解决方法。 关键词:网络故障;网络维护;分类;解决办法 中图分类号:tp393.06 文献标识码:a 文章编号:1007-9599 (2011) 22-0000-01 network fault diagnosis and troubleshooting sun jie (china shipbuilding industry corporation 715 institute,lianyungang 222006,china) abstract:network failure is extremely common,but also a variety of types of network failure,network failure in a timely manner to maintain a network failure,the fastest to restore the normal operation of the network,an effective network control maintenance theory,methods and techniques is the https://www.360docs.net/doc/e31888172.html,mon faults on the network have been classified in a variety of common network problems and corresponding solutions proposed. keywords:network failure;network

智能仪器设计论文

引言 我国目前中小型企业在整个工业产业中占相当大的比例,这些企业的监控模式主要为模拟控制系统加以常规仪表为主的数据采集系统。这种监控模式存在着检修维护工作量大、没有可靠的历史记录等缺点。而且常规模拟仪表也进入老化淘汰期,设备可靠性明显降低,某些仪表的备品备件也得不到保障,因此中小型企业监控系统的技术改造工作已势在必行。 数据采集系统是从一个或多个信号获取对象信息的过程。随着微型计算机技术的飞速发展和普及,数据采集监测已成为日益重要的检测技术,广泛应用于工农业等需要同时监控温度、湿度和压力等场合。数据采集是工业控制等系统中的重要环节,通常采用一些功能相对独立的单片机系统来实现,作为测控系统不可缺少的部分,数据采集的性能特点直接影响到整个系统。 数据采集系统可以采集的工业运行数据包括电气参数和非电气参数两类。其中电气参数主要有电流、电压、功率、频率等模拟量,断路器状态、隔离开关位置、继电保护动作信号等开关量以及表示电度的脉冲量等。而非电气参数种类较多,既可以是采集某些工业中的各种温度、压力、流量等热工信号,也可有水电厂中的水位、流速、流量等水工信号,还可以采集诸如绝缘介质状态、气象环境等其它信号。 本次设计中数据采集系统是基于单片机的测量软硬件来实现灵活的测量显示系统,它主要完成数据信息的采集、A/D转换、标度变换、数据显示及实现报警系统。随着计算机技术的飞快发展和普及,以数据采集系统为核心的设备也迅速在国内外得到了广泛的应用,现代工业生产和科学研究对数据采集的要求也越来越高。

第1章数据采集系统概述 1.1 数据采集系统发展概况 数据采集系统起始于20世纪50年代,1956年美国首先研究了用在军事上的测试系统,目标是测试中不依靠相关的测试文件,由非熟练人员进行操作,并且测试任务是由测试设备高速自动控制完成的。由于该种数据采集测试系统具有高速性和一定的灵活性,可以满足众多传统方法不能完成的数据采集和测试任务,因而得到了初步的认可。大约在60年代后期,国外就有成套的数据采集设备产品进入市场,此阶段的数据采集设备和系统多属于专用的系统。20世纪70年代中后期随着微型机的发展,诞生了采集器、仪表同计算机溶为一体的数据采集系统,由于这种数据采集系统的性能优良,超过了传统的自动检测仪表和专用数据采集系统,因此获得了惊人的发展。从70年代起,数据采集系统发展过程中逐渐分为两类,一类是实验室数据采集系统,另一类是工业现场数据采集系统。就使用的总线而言,实验室数据采集系统多采用并行总线,工业现场数据采集系统多采用串行数据总线。20世纪80年代,随着计算机的普及应用,数据采集系统得到了极大的发展,开始出现了通用的数据采集与自动测试系统。该阶段的数据采集系统主要有两类,一类以仪器仪表和采集器、通用接口总线和计算机等构成,第二类以数据采集卡标准总线和计算机构成。20世纪90年代至今,在国际上技术先进的国家,数据采集技术已经在军事、航空电子设备及宇航技术工业等领域被广泛应用。由于集成电路制造技术的不断提高,出现了高性能、高可靠性的单片数据采集系统。 1.2 数据采集系统的应用 数据采集系统的硬件设备又叫数据采集器,根据数据采集器的使用用途不同,数据采集器大体上可分为两类:在线式数据采集器和便携式数据采集器。在线式数据采集器又可分为台式和模块式,台式、便携式数据采集器大部分由交流电源供电,模块式数据采集器大部分由直流电源供电,一般是非独立使用的。在采集器与计算机之间由电缆联接构成数据采集传输系统,一般不脱机单独使用。数据采集器的应用涉及到众多的领域,下以介绍数据采集器及系统的几种典型应用。

2020年常见的网络故障分析与处理

常见的网络故障分析与处理 如今,计算机网络技术飞速发展,在社会生活和工作中的重要性日趋凸显。它给人们带来了极大的便利,但是同时,层出不穷、种类繁多的网络故障也给人们带来了很多的烦恼。本文将简单介绍一些常见的网络故障及其处理方法。 一、计算机网络故障分类 网络故障按性质分为物理故障和逻辑故障;按对象分为线路故障、路由器故障和主机故障。 物理故障主要有网卡、网线、交换机、路由器等故障。 逻辑故障主要有网络协议、网络设备配置等故障。 二、物理故障分析与处理 1、线路故障 线路故障的发生率很高,包括线路损坏和线路受到严重电磁干扰。 处理方法:观察网线两端口处是否松动、磨损或断开,如发现问题则处理;否则继续处理如下:若线路短,可将网线一端插入正常RJ45插座内,另一端插入正常HUB端口,然后用Ping命令检查线路与网管中心服务器端口是否连通。若线路长,或网线不方便移动,就用网线测试器测量网线的好坏。目前,大约有70%的网络故障是由此类故障引起的。 2、集线器(或路由器)故障 集线器(或路由器)物理损坏,导致网络不通。

处理方法:替换排除法,用通信正常的网线将主机和集线器连接,如能正常通信,则可确定集线器正常;否则更换一个端口排查是 端口故障还是集线器故障,如更换端口后能正常通信,则确定为端口故障;如若始终不能正常通信,则可能是集线器故障。更换一个能正 常使用的集线器,如能正常通信则确定为集线器故障。 3、主机物理故障 包括网卡松动,网卡物理故障,主机网卡插槽故障。 处理方法:网卡松动可重新拔插固定网卡;主机网卡插槽故障可更换一个网卡插槽;网卡物理故障可将其插到正常工作的主机上测试,若仍无法工作,即更换网卡。 四、逻辑故障分析与处理 1、路由器逻辑故障 包括路由器端口参数设定有误,路由器路由配置错误、路由器CPU利用率过高和路由器内存余量太小等。 逻辑故障最常见的是配置错误导致网络异常或故障。配置错误 可能是路由器端口参数设定有误,或路由器路由配置错误以至于路由循环或找不到远端地址,或路由掩码设置错误等。比如,掩码为255.255.255.0,IP地址为192.168.0.1的主机就ping不通IP地址为192.168.1.1的路由器。 处理方法:该故障可用ping命令或用traceroute命令(路由跟踪程序)查看远端地址中哪个节点出现问题,对该节点参数进行检查

常见网络故障排查

计算机网络故障及其维修方法 目标: 1.常见计算机网络故障检测、分析能力;掌握计算机网络故障维修方法; 2.会配置小型计算机网络系统;了解常见计算机网络故障原因;了解计算机网络故障处理方法; 3.能利用所学知识和经验(灵活性)创造性地解决新问题。 内容: 一、了解常见计算机网络故障原因 (一)硬件故障 硬件故障主要有网卡自身故障、网卡未正确安装、网卡故障、集线器故障等。 首先检查插上计算机I/O插槽上的网卡侧面的指示灯是否正常,网卡一般有两个指示灯“连接指示灯”和“信号传输指示灯”,正常情况下“连接指示灯”应一直亮着,而“信号传输指示灯”在信号传输时应不停闪烁。如“连接指示灯”不亮,应考虑连接故障,即网卡自身是否正常,安装是否正确,网线、集线器是否有故障。 1.RJ45接头的问题 RJ45接头容易出故障,例如,双绞线的头没顶到RJ45接头顶端,绞线未按照标准脚位压入接头,甚至接头规格不符或者是内部的绞线断了。

镀金层厚度对接头品质的影响也是相当可观的,例如镀得太薄,那么网线经过三五次插拔之后,也许就把它磨掉了,接着被氧化,当然也容易发生断线。 2.接线故障或接触不良 一般可观察下列几个地方:双绞线颜色和RJ-45接头的脚位是否相符;线头是否顶到RJ-45接头顶端,若没有,该线的接触会较差.需再重新压按一次;观察RJ-45侧面。金属片是否已刺入绞线之中?若没有,极可能造成线路不通;观察双绞线外皮去掉的地方,是否使用剥线工具时切断了绞线(绞线内铜导线已断,但皮未断)。 如果还不能发现问题,那么我们可用替换法排除网线和集线器故障,即用通信正常的计算机的网线来连接故障机,如能正常通信,显然是网线或集线器的故障,再转换集线器端口来区分到底是网线还是集线器的故障,许多时候集线器的指示灯也能提示是否是集线器故障,正常对应端口的灯应亮着。 (二)软件故障 如果网卡的信号传输指示灯不亮,这一般是由网络的软件故障引起的。 1.检查网卡设置 普通网卡的驱动程序磁盘大多附有测试和设置网卡参数的程序。分别查验网卡设置的接头类型、IRQ、I/O端口地址等参数,若有冲突.只要重新设置(有些必须调整跳线),一般都能使网络恢复正常。

相关文档
最新文档