Stability testing for IVDs

Stability testing for IVDs
Stability testing for IVDs

Stability testing for IVDs

Lisa A. Hornback

Stability is defined as “the extent to wh ich a product retains,

within specified limits and throughout its period of storage and

use (i.e., its shelf life), the same properties and characteristics

that it possessed at the time of manufacture.”1 According to

FDA regulations, IVD reagents and consumables marketed in

the U.S. must undergo testing to determine stability.

The regulatory requirements for stability testing of IVDs are

found in several places. For example, the Code of Federal

Regulations (CFR) states that “the basis for such instructions

(e.g., those needed to protect the stability of the product) shall

be determined by reliable, meaningful, and specific test

methods.”2 IVDs approved by the Center for Biologics

Evaluation and Research (CBER) are also subject to other CFR

requirements, which state that an application for a biological

license shall include “data establishing stability of the product

through the dating period.”3

While a significant number of guidance documents directly relate to the

requirements for stability testing of drugs, guidance regarding stability testing of medical devices in general and IVDs in particular is limited.1, 4, 5, 6 This article discusses the various aspects unique to stability testing of IVDs and offers suggestions on how to optimize the design of stability studies. This article also discusses effective responses to stability issues as they arise.

Design Control Elements

For new IVDs, manufacturers should first address stability requirements during the design control stage. At this stage, IVD manufacturers should determine the

device’s intended use and the critical design components that must be evaluated to assure that the labeled expiration dates are met. Manufacturers should also

consider user requirements and their impact on stability. IVD manufacturers should focus on user needs regarding product stability such as the following:

? Onboard stability (e.g., the maximum length of time IVDs can be loaded onto an instrument and still perform according to specifications). Shortened onboard

stability requires more user intervention to replace reagents and creates delays in work output.

? Labeled storage conditions. IVDs with multiple storage conditions, such as six months frozen or 24 hours thawed, require more user interaction, which could lead

Lisa A. Hornback is a compliance consultant at KMI/Parexel LLC (Alpharetta, GA). She can be reached at HERE .

to errors w hen trying to establish the “after opening/thawing” expiration.

? Preservative effectiveness for microbially controlled or sterile IVDs. The use of a preservative should control microbial growth in situations in which the user opens the device packaging m ultiple times during the product’s life. For many IVDs, microbial contamination can result in IVD performance issues due to microbial metabolism.

The user requirements related to stability are established as design inputs and are ultimately assessed during stability studies.

Using Accelerated Study Data

Accelerated stability studies are useful for predicting the shelf life of IVDs. Such accelerated studies subject IVDs to extreme conditions—typically elevated temperatures—to the extent that the device endures significant and measurable deterioration during the testing period. Mathematical extrapolations, such as the Arrhenius equation, are then used to calculate the IVD’s predicted shelf life.4 However, not all IVDs follow a predictable degradation rate. Some products will perform acceptably until they fail, in which case only real-time testing will suffice.

According to FDA’s Office of In Vitro Diagnostic Device Evaluation and Safety, accelerated stability studies are acceptable in the following situations:

? Establishing preliminary claims in new products only if there is sufficient correlation to an existing product.

? Supporting implementation of a change to an existing product.

The Center for Devices and Radiological Health (CDRH) recognizes the European standard EN 13640:2000, which provides guidance on not only conducting real-time and accelerated stability studies but also making calculations using the Arrhenius equation.7 Typically, CDRH requires premarket notification (510(k)) submissions to include a summary of the accelerated data, while premarket approval (PMA) submissions are expected to include the real-time study data and stability protocol. At the same time, CBER does not accept accelerated testing for either newly licensed IVDs or changes to existing IVDs. Only real-time stability data are accepted; they are also required for biologics license application (BLA) submissions, in which labeled storage and reconstituted stability data should be included.6

Overall, FDA expects all accelerated studies to be confirmed with real-time stability studies.4 However, prior to using accelerated data to support new IVDs or a significant change to existing IVDs, manufacturers should confer with the proper center at FDA about the agency’s requirements reg arding stability studies for specific IVD reagents.

Real-Time Stability Study Requirements

A real-time stability study includes an evaluation of those

factors that ultimately affect the expiration date of IVD

reagents. A number of IVD reagent stability factors should be

considered when designing a stability study program (see

Table I).1, 4, 5, 6

While originally established as a design verification

requirement, a stability study for IVD reagents has the same

elements as those dictated for stability testing of drugs

including the following:

? A written stability testing program designe d to assess the

stability characteristics of IVDs.

? A stability protocol with predefined acceptance criteria that can be correlated to the label claims.

? Testing multiple unique product lots. A stability study is required to use three product lots that are manufactured when the manufacturing process has been well defined and can be consistently executed.

? Evaluation of each stability attribute via a statistically valid sample size and testing intervals. The sample size should be sufficient to overcome the precision of the test method used, considering the cumulative effect of all elements of the test system (i.e., individual reagents and instruments). The test intervals should be chosen so that trends may be discerned from variability of the data.4 At a minimum, stability testing should continue to one time interval past labeled expiration.

? Control of material storage. For real -time stability testing, the IVD reagents should be stored under the conditions stated on the label (e.g., temperature, humidity, light protection).

? Testing IVDs in the same container -closure system as the marketed product. ? For IVDs that require reconstitution for use as directed in the labeling, testing IVDs for reconstitution as well as after they are reconstituted.

? Use of reliable, meaningful, and specific test methods.8

This last requirement presents the greatest challenge to the design of IVD stability studies. While stability testing of drugs is accomplished through direct analytical testing of the drug itself, the typical evaluation of IVD stability involves performing the assay and determining whether the control values conform to the specifications. This technique is used regardless of the IVD component being evaluated for stability: reagent, calibrator, or control. To allow for adequate evaluation of the IVD

components for stability, manufacturers should identify the control criteria for their particular system to detect and reduce the bias effect of the test system as a whole. These control criteria should consider the effects that changes in instruments,

Table 1. Factors to consider in real-time stability study (click to enlarge).

reagents, calibrators, or controls may have on the test system.

Monitoring Bias

There is a potential for failure to control the effects of bias on

stability results (see Figure 1). For illustration purposes,

assume the IVD component being tested for stability is a

control and the acceptance criteria are such that when the

assay is performed in accordance with the labeling instructions,

a control value is obtained that meets the label claims. In

Figure 1, the value at T3 reflects a change in the reference

reagent lot used during stability testing, which introduces a

bias into the stability degradation pattern. The net effect is a

downward shift in the control value, thus giving more time until

actual stability failure.

One method to monitor for bias is conducting parallel testing

when new reference materials are introduced into a stability study. Stringent

criteria should be created to assess the effects that changes in reference materials have on the stability profile. These criteria should consider the inherent variations in the test system itself, including the variation contributed by each IVD assay

component as well as the instrument. Comparing the values obtained by the original reference materials with the new reference materials will detect the introduction of bias created by the new reference materials. Any shift in assay values that is outside the expected precision of the test system should prompt an investigation.

Another effective control mechanism to evaluate stability data is conducting a linear regression analysis to detect changes in the slope of the stability degradation

pattern for each lot. This technique offers two unique opportunities for evaluating data:

? Detecting changes in the slope within the stability data obtained for a single lot may indicate that bias could have been introduced into the test system, which creates the potential for inaccuracy in evaluating the data.

? A regression analysis of stability data allows an estimation of time until stability failure, thereby allowing a more timely response to stability issues.

Ongoing Stability Study Monitoring

Once the stability studies necessary to support the labeled expiration date of a newly designed IVD have been completed, manufacturers must then address the need for postapproval or ongoing stability studies.

Unless stipulated as a postapproval requirement, FDA does not explicitly require

Figure 1. Effect of material change on stability profile (click to enlarge).

ongoing, or postapproval, stability study monitoring for IVDs. An FDA compliance policy guide (CPG) states that CBER-licensed IVDs generally are not required to conduct postapproval stability studies, with certain exceptions.5 While this CPG does not include CDRH-regulated IVDs within its scope, the scientific rationale regarding the need for postapproval studies in certain situations still applies; as such, CDRH-regulated IVDs should follow the same guidance.

According to the CPG, FDA investigators will cite CBER-licensed IVD manufacturers for not performing postapproval stability studies when:

? Performing the postapproval studies is required as a condition of the license.? A form ulation or manufacturing change has been made.

? There is a commitment to perform postapproval stability studies as part of a corrective and preventive action plan.5

Despite the limited requirements for postapproval stability studies as set forth in the CPG, ongoing stability studies are a useful and critical element of process monitoring as dictated by the CFR. One section states that, “manufacturers shall monitor production processes to ensure that a device conforms to its specifications.9 Another CFR section states that manufacturers shall analyze “quality data to identify existing and potential nonconforming product, or other quality problems.”9 It is up to IVD manufacturers to evaluate and determine the need for and extent of ongoing stability monitoring, taking into consideration the risk associated with the IVD reagent as well as regulatory commitments and requirements.4 Manufacturers should be prepared to discuss with FDA the scientific rationale behind their stability study programs, including both preapproval and postapproval study designs.

Investigating Stability Failures

Once a potential stability issue has been detected during either

premarket or postmarket stability studies, IVD manufacturers

should initiate a timely and thorough failure investigation. In

most cases, manufacturers have already released for

distribution the product lots with stability issues. Nonetheless,

an investigation must proceed with the knowledge that

stability failures on distributed products represent a high

regulatory and patient-care risk.

Stability failure investigations should include an impact

analysis to determine the scope of product lots potentially

affected by the problem. IVD manufacturers should place all

lots on hold until they determine the scope of the stability issue.

Manufacturers should report any immediate action taken to

control nonconforming products, such as recalls, corrections,

or removals, to FDA as required in accordance with the

regulations.9, 10

IVD manufacturers can identify the product lots at risk for

premature stability failure by using various techniques,

including statistical tools, prior to determining the root causes. For example, assume a stability failure has occurred prior to the labeled expiration date. For purposes of this example, the decay rate for the lots at risk for stability failure is calculated to be 20 ng/ml/month, which is significantly higher than what is typically seen for this IVD.

Through linear regression and by using the upper stability specification limit at expiration, a model can be created to determine the release specification range that is most likely to fail to meet the labeled expiration date. By determining this

specification at the point that intercepts the y-axis, this minimum value can then be compared with the release date for previous product lots to determine the lots most susceptible to stability failure (see Figure 2).

A root cause analysis should focus on the variables that could potentially affect product stability. Primary suspects are raw materials and manufacturing changes. For temperature-sensitive products, IVD manufacturers should examine product temperature control during manufacturing and handling. Manufacturers should also look at manufacturing activities that subject raw materials or in-process products to elevated temperatures (e.g., outside of labeled storage conditions) for any period of time, including time spent in thawing or adjusting materials to hit the target potency.

Once IVD manufacturers identify the root causes, appropriate global corrective actions must be assigned and implemented. For example, issues with raw material control may result in additional specifications. Identifying adequate corrective

Figure 2. Correleation of stability failure to release data. In this example, based on the data obtained from the stability failure lot, all product lots with a T0 (release) value of >220 ng/mL are at risk for stability failure (click to enlarge).

actions must also consider the potential gaps in the manufacturer’s quality system that allowed the stability failure to occur. For example, design issues may not have adequately addressed raw material control. In addition to a raw-material specification change, the proper corrective action in this case would be to modify the design control process to include the knowledge gained from the stability failure.

After implementing the corrective actions, an evaluation of the effectiveness of the corrective actions should be conducted through verification and/or validation activities. For stability issues, this would involve activities such as placing additional product lots manufactured after the corrective action into the stability program and potentially adding test intervals to better monitor the product and/or degradation rate. The success of the corrective actions should be directly measurable. In other words, to say the measure of success of the corrective actions is that no more lots will fail stability is not adequate. The simple reason is that it would take a stability failure that equates to nonconforming products in the field to prove the corrective action was not adequate.

Conclusion

Meeting the regulatory requirements regarding stability testing of IVDs represents a challenge to IVD manufacturers. Manufacturers must consider special considerations unique to IVDs to assure that the stability study design provides accurate data to support labeled product expiration dates. A robust IVD stability program initiated as early as possible during product development, in conjunction with prompt and thorough responses to identified stability issues, assures that IVDs will continue to meet the user needs throughout their labeled expiration.

References

1. Shelf Life of Medical Devices (Rockville, MD: Division of Small Manufacturers Assistance, Center for Devices and Radiological Health, 1991).

2. “In Vitro Diagnostics for Human Use,” Code of Federal Regulations, 21 CFR Part 809.

3. “Licensing,” Code of Federal Regulations, 21 CFR Part 601.

4. Stability Testing of In Vitro Diagnostic Reagents, EN 13640:2002 (Brussels: European Committee for Standardization, 2001).

5. Section 280.100, “Stability Requirements for Licensed In Vitro Diagnostic Products,” Compliance Policy Guide (Rockville, MD: Office of Regulatory Affairs, U.S. Food and Drug Administration, 2000).

6. Guideline for the Manufacture of In Vitro Diagnostic Products (Rockville, MD: Office of Compliance, Center for Devices and Radiological Health, 1994).

7. IVD Standards Web Page (Rockville, MD: Office of In Vitro Diagnostic Device Evaluation and Safety, Center for Devices and Radiological Health, 2004 [cited 19 January 2004]); available from Internet: CLICK HERE.

8. “CGMP for the Manufacture, Processing, Packing, or Holding of Drugs and Finished Pharmaceuticals,” Code of Federal Regulations, 21 CFR Part 211.

9. “Quality Systems Regulation,” Code of Federal Regulations, 21 CFR Part 820.

10. “Medical Devices; Reports of Corrections and Removals,” Code of Federal Regulations, 21 CFR Part 806. Copyright ?2004 IVD Technology

相关主题
相关文档
最新文档