Pyris 1 TGA 操作规范

Pyris 1 TGA 操作规范
Pyris 1 TGA 操作规范

1.0 目的 PURPOSE

1.1 本规范描述了PE Pyris 1热失重分析仪的操作,校正和维护方法。

This document contains instructions used to provide a uniform method of operating, calibrating, and maintaining the Perkin Elmer Pyris 1 Thermogravimetric Analyzer.

2.0 范围 SCOPE

2.1 本规范包括Perkin Elmer Pyris 1 TGA的校验,操作,维护程序。

This document covers the calibration, operation, and maintenance procedures for the Perkin Elmer Pyris 1 TGA.

3.0 职责 RESPONSIBILITIES

3.1 所有的实验室分析人员必须遵守本规范。

This instruction should be observed by all testing personnel.

3.2 分析人员使用前应先检查设备的校验日期,如果过期,则应向实验室主管或QC校验人

员汇报。

Visually check to make sure the calibration date on the instrument is not expired. If the instrument is expired, report it to a supervisor or the quality control calibration staff.

3.3 当仪器出现异常时应向实验室主管或QC校验人员汇报。

Malfunctioning instruments should be reported to both a supervisor and a QC

calibration staff.

4.0 安全防护 SAFETY PRECAUTION

4.1 按管理要求穿戴合适的个人防护用品。

It is necessary to wear all Personal Protective Equipment assigned by your

Supervisor.

4.2 预防措施Instrumental Precaution

注意:操作者在使用Pyris 1时必须遵守下面的预防措施:

NOTE: The following precautions must be observed when using the Pyris 1

Thermogravimetric Analyzer.

4.2.1 当Pyris软件出现异常时不要按下计算机的Reset键。如果要关闭异常的Pyris窗

口时应同时按下Ctrl+Alt+Del键,然后从任务管理其中关闭Pyris软件。

Never press the Reset button on the computer if the Pyris software appears to

malfunction. Press the Ctrl-Alt-Del keys simultaneously and select the Task

Manager. From the Task Manager, close the Pyris software.

4.2.2 在关闭PE Pyris 1设备和切断电源前,不要拆掉仪器上的任何面板。

Never remove the outer instrument cover or side panels on the PE Pyris 1 without shutting the instrument down and disconnecting its power cord from the power

source.

4.2.3 PE Pyris 1所使用的气体纯度不应低于99.9%。

EOI-0142/R2, Version 1.0

Only high-quality purge gases should be used with the PE Pyris 1. Minimum purity of 99.9% is recommended.

4.2.4 遵守PE Pyris 1和相关设备的开启和关闭程序。

Always observe the proper startup or shutdown procedures with the PE Pyris 1 and all related instruments.

4.2.5 通过天平室的气流流速始终控制在40-60cc/min, 建议使用惰性气体来作为天平室的

冲洗气,如:氮气、氩气。

Always maintain a positive purge flow (40 – 60 cc/min) through the balance

chamber at all times. An inert gas such as nitrogen or argon is recommended as a balance purge.

4.2.6 当加热炉的外套管移掉时禁止将加热炉抬升,否则将严重损坏PE Pyris 1的加热炉

和其他部件。

Do not raise the furnace when the clear Pyrex furnace tube has been removed.

This will cause severe damage to the furnace and other components of the PE

Pyris 1 analyzer.

4.2.7 当PE Pyris 1的加热炉在升温时,将其前面的塑料保护面板放下,以避免意外接触

高温的加热炉。

When the PE Pyris 1 standard or high temperature furnace is heating, keep the

protective plastic shield on the PE Pyris 1 in the down position. This will protect you from accidentally touching the furnace when it is hot.

4.2.8 在切断电源和冷却风扇停止旋转前,不要移动PE Pyris 1。

Do not attempt to move the PE Pyris 1 while it is on. Wait until the power is off and the furnace cooling fan has stopped spinning before moving the instrument.

4.2.9 在分析仪工作时禁止将其连接至PE Pyris 1的导线拔掉。

Do not remove cables from the PE Pyris 1 when the analyzer is turned on.

4.2.10 禁止触摸PE Pyris 1的加热炉开口,禁止向里面投入物体。

Do not touch the opening on the furnace lift mechanism or insert objects into the

opening.

5.0 定义 DEFINITIONS

5.1 TGA:热失重分析仪

TGA : ThermoGravimetric Analyzer

6.0 工作程序 INSTRUCTIONS

注意:在使用PE Pyris 1进行实验或者校正前,应先对热电偶进行检查。将温度设定至100°C后升温,如果热电偶校验正常,则显示的温度应非常接近100°C。

NOTE: Prior to performing any experiment or calibration procedure, be sure that the thermocouple currently installed on the PE Pyris 1 is functioning properly. The

thermocouple can be checked by programming the PE Pyris 1 to 100°C and

checking the temperature displayed in the status panel. The temperature should be close to 100°C if the instrument is calibrated.

6.1 设备的启动和测量

Start up and measurement

注意:在开机前确保动力气,天平气,参比气都已打开并调节到要求的流速。

NOTE: Make sure the Pneumatic gas, Balance gas and Reference gas are all open and adjusted to the requirement before the Analyzer turn on.

6.1.1 按下面的顺序打开系统:

Turn the system on following order:

计算机

Computer

分析仪

Analyzer

6.1.2 运行Pyris软件,然后按下相关设备的按钮。

Execute the Pyris manager, then click on each instrument button.

6.1.3 等待每个设备的绿灯都亮起。

Wait until the green light comes to each instrument button (instrument window

opens automatically).

6.1.4 将加热炉升起,待重量读数稳定后,按下“Zero”键去除样品盘的皮重。

Raise the furnace, after the sample weight reading is stabilized, then tare the

sample pan by pressing the Zero weight button.

6.1.5 降下加热炉,在支架的保护下取下样品盘,放入样品,再将样品盘放回原处。

Lower the furnace; remove and fill the sample pan with the specimen while

supporting the sample pan with platform.

注意:当取下或放回样品盘时,必须使用支架对样品盘进行保护。由于Pyris 1的天平非常敏感易损,禁止再没有支架的保护下取下或放回样品盘。

NOTE: Always use the platform to work on sample pan. The weight balance is

extremely sensitive, so do not change sample pan without platform support.

6.1.6 从“File”中打开适当的操作程序(每个程序可以根据特殊的要求进行修改)。

Open the appropriate method from the File menu (each method can be modified

according to specific condition requirement).

6.1.7 打开“Sample Info”栏,填写Sample ID, Operator ID, Comment, File Name,同

时检查初始状态栏并输入正确的系统参数。

Select the Sample Info tab and fill in the Sample ID, Operator ID, Comment, and

File Name. Also check initial state tab and program tab for proper parameter set up.

6.1.8 装入样品盘,升起加热炉。

Load the sample pan, then raise the furnace.

6.1.9 输入初始温度,然后点击“Go to temperature”键。当温度升至设定的初始温度

后,点击“Sample weight”键称量样品重量,再点击“Start/stop”键,系统进入自动运行状态。

Enter the initial temperature and press the "Go to temperature" button. When the

temperature reaches to initial temperature, measure the sample weight by pressing the Sample weight button, and run the analysis by clicking the Start/stop button

once.

6.1.10等待系统完成全部的测试。

Wait until the system finishes scanning.

6.1.11激活运行窗口。

Make the last run an active window.

失重百分比:Percent Weight Loss:

计算样品的失重百分比时,将图中的x轴设为温度(或者可以从“Display”菜单中

选”Rescale X”,再设定温度值来改变x轴的单位)y轴设为重量百分比(从

“Curves”中选择“Weight”,有时需要使用Rescale Tools中的Swap Y Axes功能将重量显示在左侧的x轴上)。

To calculate the percent weight loss of the specimen, be sure that the curve is

displayed with the x axis as Temperature (change by selecting Rescale X from the Display menu and selecting Temperature for the Set Axis Units value) and the y

axis as Weight percent (change by selecting Weight from the Curves menu; it

maybe required to use Swap Y Axes from the Rescale Tools bar to place Weight

on the left-hand axis).

从“Calc”菜单中选择“delta Y”,输入工作单上所要求的温度,点击

“Calculate”,将结果记录在工作单上。

Select delta Y from "Calc" menu, then enter the temperature parameter specified

on a traveler. Press "Calculate" and report Delta Y value on a traveler.

固含量分析:Filler Content Analysis:

当计算样品的固含量时,将图中的x轴设为时间(或者可以从“Display”菜单中

选”Rescale X”,再设定时间值来改变x轴的单位),y轴设为重量百分比(从

“Curves”中选择“Weight”,有时需要使用Rescale Tools中的Swap Y Axes功能将重量显示在左侧的x轴上)。

To calculate the filler content of the specimen, be sure that the curve is displayed

with the x axis as Time (change by selecting Rescale X from the Display menu and selecting Time for the Set Axis Units value) and the y axis as Weight percent

(change by selecting Weight from the Curves menu; it maybe required to use Swap Y Axes from the Rescale Tools bar to place Weight on the left-hand axis).

从“Calc”菜单中选择“delta Y”,输入时间参数(时间应该包括右侧平直曲线部

分对应的时间),点击“Calculate”,将结果(100%-计算值)记录在工作单上。

Select delta Y from "Calc" menu, then enter the time parameter (the time

parameter should include flat region of the right hand side graph). Press

"Calculate" and report 100% - Delta Y value on a traveler.

6.1.12 将分析结果打印出来附在工作单后面。

Print the analysis report and attach it to the traveler.

6.2 维护要求

Maintenance Requirements

为了延长PE Pyris 1的加热炉的使用寿命,应定期对加热炉进行清洗,例如每5-10次测试后运行一次自动清洗程序,用棉棒和溶剂彻底清洗加热炉的内部,待其彻底干燥后再继续使用。

To prolong the life of the PE Pyris 1 standard and high temperature furnaces, it is

recommended that the automatic furnace cleaning procedure be performed routinely, such as every 5 – 10 runs. Thoroughly clean the inner surface of the furnace using Q-tips and proper solvent, and let it completely dry before use.

6.3 校验要求

Calibration Requirements

注意:PE Pyris 1的校验包含三项内容,为了确保测量的精确性,必须定期对设备进行校验。

NOTE: There are three calibration routines for the PE Pyris 1. Periodic calibration checks will be necessary to verify accurate calibration.

6.3.1 如果在一段时期内PE Pyris 1的操作环境没有改变,则其校验数据在此期间内有

效。即使当系统关闭后,前面的校验数据仍储存在电脑里,再次开机后校验依然有

效。某些条件会改变PE Pyris 1的温度校验或者称量校验结果,此时需要按下面的要求进行重新校验:

The PE Pyris 1’s calibration should remain unchanged for quite some time,

provided that there are no changes in the instrument's operating conditions. Even

when the system is shut off, the calibration values are stored so that the next time the instrument is turned on, it will still be calibrated. Some of the conditions that

could change either the temperature or weight calibration of the PE Pyris 1 and

require recalibration are as follows:

如果实验的温度范围改变,则需要重新校验。

If the operating temperature range of your experiments changes, recalibration may be necessary. Check the temperature calibration in the range of interest to

determine if the current calibration is still valid.

如果气体种类或者气体流速改变,则需要重新校验以保证结果的精确性。

If the purge gas type or purge gas flow rate is changed, the calibration should be

checked for highest accuracy.

如果更换新的加热炉,应重新进行温度校验。

If a new furnace is installed, the temperature calibration should be checked.

如果更换新的热电偶或者热电偶的位置有所改变,则应再次进行温度校验。

If a new thermocouple is installed or if the position of the thermocouple changes,

you should perform the temperature calibration again.

如果分析仪的位置(水平或者垂直)有所改变,则应进行称量校验。

If the analyzer is moved or releveled, you should perform the weight calibration

again.

6.3.2 PE Pyris 1的温度校验:

The temperature calibration routines for the PE Pyris 1 are:

打开Method Editor窗口,从File栏中选择Open Method,双击Method目录中的

Alumel.tgm文件,系统会显示出nickel校验操作窗口。

From the Method Editor, select Open Method from the File menu. Double click on the Alumel.tgm method file which is in the Method directory. The nickel calibration method is displayed on the screen.

打开Sample Info栏,输入Sample ID, Operator ID, Comment, , File Name信息。

Select the Sample Info tab and fill in the Sample ID, Operator ID, Comment, and File Name.

切一段长度1~2mm的细alumel条,放入样品盘中。

Cut a small piece of alumel approximately 1 – 2 mm long and place it in the sample pan.

点击操作窗口中的Raise Furnace,升起加热炉。

Click on the Raise Furnace button in the control panel.

点击操作窗口中的Zero Weight,分析仪会读取样品的毛重。

Click on the Zero Weight button in the control panel. The analyzer reads the weight of the sample.

当读数稳定后,点击OK,读数会显示在Sample Info页中的Enter Sample Weight 一栏中。

After the weight reading has stabilized, click on the OK button. The weight is entered into the Zero field of the Enter Sample Weight area on the Sample Info page.

将磁铁放在支架上移近加热炉。

Place the magnet onto the sample loading platform and move the platform next to the furnace tube.

调整支架的高度至读数显示在3~6mg之间,此读数是由样品的磁性产生的,当样品失去磁性时,读书会变小。

Adjust the height of the sample loading platform so that you get a weight reading between 3 and 6 mg. The apparent weight reading is a result of the magnetism of the sample. As the sample loses magnetism, the apparent weight reading decreases.

当读数稳定后,点击控制窗中的Sample Weight,读数会显示在Sample Info页中的Enter Sample Weight一栏中。

After the weight reading has stabilized, click on the Sample Weight button on the control panel. The weight is entered into the Weight field of the Enter Sample Weight area of the Sample Info page.

点击控制窗中的Start Method,开始测试。

Click on the Start Method button on the control panel.

实验结束后,从“alumel”中打开实验结果文件,将X轴设为温度(或者可以从“Display”菜单中选“Rescale X”,再设定温度值来改变x轴的单位)y轴设为重量百分比(从“Curves”中选择“Weight”,有时需要使用Rescale Tools中的Swap Y Axes功能将重量显示在左侧的x轴上)。

In Data Analysis, open the data file from alumel run. Be sure that the curve is

displayed with the x axis as Temperature (change by selecting Rescale X from the Display menu and selecting Temperature for the Set Axis Units value) and the y

axis as Weight (change by selecting Weight from the Curves menu; you may have to use Swap Y Axes from the Rescale Tools bar to place Weight on the left-hand

axis).

从“Math”菜单中选择“Derivative”,图中会显示出导数曲线,通过设定转折点来精确设定onset计算的左限。

Select Derivative from the Math menu. The first derivative of the curve is displayed.

Use the inflection point to accurately determine what the left limit should be for the onset calculation.

点击初始曲线。

Click on the sample curve to make it the focused or active curve.

从“Calc”菜单中选择“Onset”。

Select Onset from the Calc menu.

移动位于曲线上的两个红X标记,将左侧的标记移至一阶导数的转折点,将右侧的标记移至曲线过转折点平坦部分。

Move the two red X marks on the curve. Position the left limit marker at the

inflection point of the first derivative curve and place the right limit marker past the inflection point on a level portion of the curve.

点击Onset对话框中的Calculate键,屏幕会弹出Adjust Tangents对话框。

Click on the Calculate button in the Onset dialog box. The Adjust Tangents dialog

box is seen.

调节各点的切线直到它们相交。

Adjust the tangents of each point until they cross.

点击Adjust Tangents对话框中的Calculate键,系统会计算出样品的onset温度值(Onset X)并显示在屏幕上。

Click on the Calculate button in the Adjust Tangents dialog box. The onset

temperature is calculated for the sample and displayed on the screen (Onset X).

在Method Editor栏中,从View菜单中选择Calibrate。

In the Method Editor, select Calibrate from the View menu.

点击Calibration窗口中的Temperature栏,输入所测得的alumel的Onset值,点

击Use栏中的Check。

Click on the Temperature tab in the Calibration window. Enter the Measured Onset values for alumel and click in the check box in the Use column for reference

material.

点击Save and Apply键,将校验数据传至分析仪,保存校验文件。

Click on the Save and Apply button to send the calibration values to the analyzer

and save the current calibration file.

点击Close键退出校验窗口。

Select the Close button to exit the Calibration window.

6.3.3 PE Pyris 1的重量校验

The weight calibration routines for the PE Pyris 1 are

打开Pyris软件中的Instrument Viewer 或Method Editor,选择View菜单中的Calibrate。

While in the Instrument Viewer or the Method Editor in the Pyris software, select Calibrate from the View menu.

选择Calibration窗口中的Weight栏。

Select the Weight tab in the Calibration window.

在Reference Weight编辑栏内输入参照物的重量(100mg)。

Enter the weight of your reference material (100 mg) in the Reference Weight edit box.

注意:应使用NIST砝码,操作时需要戴上手套并且使用镊子。

NOTE: NIST traceable weights should be used. When handling the weights, always wear gloves and use forceps to hold.

点击Begin Calibration。

Select the Begin Calibration button.

如果垂线上有样品盘,则将盘中的样品清除;如果垂线下无样品盘,则在托盘上放置一空盘,随后装至垂线上。将托盘从样品盘下方移走,点击对话框内的OK。

If there is a sample pan already on the hang down wire, remove any sample from the pan. If there is no sample pan already in place, then place an empty sample pan on the sample loading tray and load the pan onto the hang down wire. Move the sample loading tray away from the sample pan. Click on OK in the dialog box. 当Read Zero栏中的0读数稳定后,点击对话框中的OK。

When the zero weight reading is stable in the Read Zero field, click on OK in the dialog box.

将校验参照物放进样品盘中,点击OK。

Place the reference weight in the sample pan. Move the sample loading tray away from the sample pan. Click on OK.

当Read Value栏中的读数稳定后,点击OK,校验完成。

When the weight reading is stable in the Read Value field, click on OK. The calibration is complete.

点击Calibration窗口中的Save and Apply,保存校验数据。

Select the Save and Apply button in the Calibration window to send the new calibration values to the analyzer and save the calibration file.

进入下一个校验程序,或者点击Close关闭校验窗口并对参照物进行2次以上复测,检验校验结果正确与否。

Go to the next calibration procedure or select Close to close the Calibration window and begin using the new calibration values to check at least two more reference weights.

如果复测的结果正确(误差± 2%),则校验数据可以使用,反之,则应重新进行称量校验。

If the reading is correct (± 2%), begin using the calibration value. If not, repeat the weight calibration steps.

6.3.4 加热炉的日常校验:

The furnace calibration routines for the PE Pyris 1 are:

注意:在对加热炉进行校验前,应检查热电偶是否正常,样品盘内是否留有样品。

检查热电偶的方法是:升起加热炉,将程序设定到100°C(在Go To Temp栏中输入100,然后点击Go To Temp),然后检查Status Panel中测得的温度值是否处于或

低于校验范围的下限。完成检查工作最简单的方法是点击控制面板中的“Go To

Load”键,当提示选择校验的最低和最高温度时,选择一个包含操作范围的温度

值。当进行加热炉校验时,加热炉必须处于升起位置,手动抬升和夹紧是不够的,

必须选择控制面板中的“Raise Furnace”键。

NOTE: Before performing the furnace calibration of the PE Pyris 1, make sure that the thermocouple is functioning properly and that the sample pan is empty. To

verify thermocouple operation, raise the PE Pyris 1 furnace and program the

temperature to 100°C. Enter 100 in the Go To Temp entry field of the Control

Panel and click on the Go To Temp button. Check that the current sample

temperature displayed in the Status Panel is at or below the intended minimum

calibration temperature. The easiest way to accomplish this is to click on the Go To Load button on the Control Panel. When selecting minimum and maximum

temperatures for the furnace calibration, select the limits so that they encompass

the temperature range in which you plan to operate. The furnace must be in the

raised position to perform a furnace calibration. Lifting and clamping by hand is not sufficient. You must select the Raise Furnace button on the Control Panel.

注意:加热炉的校验必须在温度校验完成后进行。

NOTE: Furnace calibration must be performed after Temperature calibration.

在Instrument Viewer 或Method Editor中,从View栏中选择Calibrate。

While in the Instrument Viewer or the Method Editor, select Calibrate from the View menu.

如果温度校验的数据有效,则跳过温度校验。

If applicable, complete the Temperature calibration.

将样品盘清空。

Remove any sample from the sample pan in the stirrup.

在Calibration窗口中打开Furnace栏。

Select the Furnace tab in the Calibration window.

在Minimum栏内输入一个最小温度作为校验的起始温度,其数值应低于日常测量范围的下限。

In the Minimum field, enter a minimum temperature where the calibration will begin.

This should be below your normal operating region.

在Maximum栏内输入一个最大温度作为校验的结束温度,其数值应高于日常测量范围的上限。

In the Maximum field, enter a maximum temperature where the calibration will end.

This should be above your normal operating region.

注意:最大温度值减去最小温度值的差值应高于100°C,否则加热炉校验将无法进行。

NOTE: Maximum temperature minus minimum temperature must be greater than

100°C or the furnace calibration will not be performed.

点击Begin Calibration键。

Select the Begin Calibration button.

校验开始后,系统会显示出加热炉校验所需的估计剩余时间。

The approximate time remaining in the furnace calibration is displayed.

校验结束后,在Calibration窗口中选择Select the Save and Apply,保存校验数

据。

Select the Save and Apply button in the Calibration window to send the new

calibration values to the analyzer and save the calibration file.

点击Close,关闭校验窗口,完成校验。

Select the Close button to close the Calibration window and begin using the new

calibration values.

6.3.5 TGA专用100毫克标准砝码的校验

Calibration of specific 100mg Standard Weight for TGA

仪器生产商提供的TGA专用100毫克标准砝码必须每六个月校验一次。

The specific 100 mg Standard Weight for TGA provided by Perkin Elmer shall be

calibrated every six months.

用一校验过的天平称标准砝码六次,每次重量差异不应超过0.1mg。将结果记录在EOI-0142/R2。

Weight the specific 100 g Standard Weight on a calibrated balance six times, the

difference shall not be more than 0.1mg each time. Record the results in EOI-

0142/R2.

6.3.6 校验参照物

Calibration Reference Materials

TGA所需的校验参照物要符合ISO9000的规范要求。ISO规范中的校验参照物有两个等级:参考级和鉴定级。参考级参照物符合仪器校验的ISO规范,它可以直接从Perkin Elmer公司获得。Perkin Elmer通过每批产品的批号来提供所需的ISO文档编号,这个批号应填入原材料的ISO控制文件中。ISO指引30中将RM(参考级参

照物)定义为:材料或物质的一个或多个性质与被测物充分的相似,可以很好的用于仪器的校验、测试方法的评估或者物质的定量分析。

Perkin Elmer Thermal Analysis calibration materials are ISO 9000 compatible.

They are called Reference Materials. ISO has two classifications for calibration

materials: Reference Materials and Certified Reference Materials. Reference

Materials (RMs) comply with ISO for instrumentation calibration. They are available from Perkin Elmer and Perkin Elmer provides documentation with each RM that

cross references the material’s lot code. This declaration should be filed with all

other material certificates and documents in your ISO file. ISO Guide 30 defines a RM as a “Material or substance one or more of whose property values are

sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to

materials.”

鉴定级参照物(CRMs)可以通过连续的对比追溯至国家或国际标准,并不是所有的校验物都是鉴定级参照物。鉴定级参照物非常昂贵并且不是适用于所有的热分析技

术,如果不能确定应使用何种参照物,应将参考级参照物与鉴定级参照物进行对

比,与QA部门协商决定。如果最终决定使用鉴定级参照物,可以向全国标准技术

研究所订购。

Certified Reference Materials (CRMs) are traceable to national or international

standards through an unbroken chain of comparisons. Not every calibration

material is a CRM. CRMs can be very expensive and are not available for all

thermal analysis techniques. If you are not sure what material to use, then check

with Quality Assurance Department and discuss the differences between RMs and CRMs. If you find that you need CRMs, you may order them from the National

Institute of Standards and Technology (301 975-6776).

磁力校验参照物是用于对加热炉进行Curie温度校验,它可以从PEXPRESS处整套购得。建议每次进行校验时使用新的参照物,这样可以避免磁滞和氧化对校验结果

产生影响。每个参照物的附属文件中指出了其磁力标准的转移温度。

Magnetic Calibration Reference Materials are used to perform the Curie point

temperature calibration on the standard furnace. They are available from PE

XPRESS in kit form and separately. It is recommend that to use fresh reference

materials each time to calibrate instrument. This way one can avoid the effects of

hysteresis and oxidation on the Curie point of the reference materials. The

transition temperature for each of the magnetic standards is indicated in the

documentation included with the reference material.

7.0 参考文件 REFERENCE

8.0 记录 RECORD

9.0 附录 APPENDIX

实验室常见仪器使用标准及操作规程SICOLAB

超净工作台 标准: A.根据环境的洁净程度,可定期(一般2~3个月)将粗滤布拆下清洗予以更换; B.定期(一般为一周)对超净工作台环境进行灭菌,同时,经常用纱布沾上酒精或丙酮有机溶剂将紫外杀菌灯外表面揩擦干净,保持表面清洁,否则会影响杀菌效果; C.当加大风机电压不能使操作风速达到0.32m/s时,必须更换高效空气过滤器; D.更换高效空气过滤器时可打开顶盖,更换时应注意过滤器上的箭头标志,箭头指向即为层气流流向; E.更换高效空气过滤气后,应用尘埃粒子计数器检查四周边框密封是否良好,调节风机电压,使操作平均风速保持在0.32~0.48m/s范围内,再有用Y09-4型尘埃粒子计数器检查洁净度。操作规程: A.使用工作台时,应提前1小时开机,同时开启紫外灭菌灯,处理操作区内表面积累的微生物,三十分钟后关闭杀菌灯; B.新安装的或长期未使用的工作台,使用前必须对工作台和周围环境先用超净真空吸尘器或用不产生纤维的工具进行清洁工作台,再采用药物灭菌法和紫外线灭菌法进行灭菌处理; C.操作区内不允许存放不必要的物品,以保持操作区的洁净气流流型不受干扰; D.操作区内应尽量避免作明显扰乱气流流型的动作; E.操作区内的使用温度不得大于60℃。 冰箱 使用标准: A.在冰箱接入电源之前,请仔细核对冰箱的电压范围和电源电压是否相等; B.冰箱必须有干燥的接地,如果电气线路没有接地,那么必须请电工将冰箱单独接地; C.不可将汽油、酒精、胶粘剂等易燃、易爆品放入冰箱内,以免引起爆炸; D.冰箱不能在有可燃性气体的环境中使用,如发现可燃气体泄漏,千万不可拔去电源插头,关闭温控器或灯开关,否则会产生电火花,引起爆炸; E.切勿用水喷洒冰箱顶部,以免使电气零件受损,发生危险; F.清洁保养及搬动冰箱时必须切断电源,并小心操作,不让电气元件受损; G.冰箱应放置在平坦、坚实的地面上,如放置不平,可调节箱底平脚; H.冰箱应放置在通风干燥,远离热源的地方,并避免阳光直射; I.冰箱在一般使用时,会结霜,当结霜特别严重时,可关机或关掉电源进行人工化霜,必要时可打开柜门加速霜层融化; J.当冰箱搁置不用时或长时间使用箱内出现异味时,必须进行清理; K.不可用酸、化学稀释剂,汽油、苯之类物品清洗冰箱任何部件; L.箱内不要放熟的食品,热的仪器必须冷却至室温后,再放入箱内; 操作规程: A.首次通电或长时间不用重新通电时,由于箱内外温度接近,为迅速进入冷藏状态,可把温度调至最冷处,待冰箱连续运行2~3小时,箱温降低后,再将温度调至适当位置; B.在使用中,不要经常调动温度控制器; 电子天平 操作规程: A.将电子天平置于稳固、平整的台面上; B.插上电源,打开开关;

纯化水系统操作、维护保养标准操作规程

1 目的 建立纯化水系统使用的标准操作规程及维护保养标准操作规程。 2 范围 纯化水系统的操作及维护保养。 3 职责 纯化水系统操作人员、维护人员按本规程操作,设备办对本规程有效执行承担监督检查责任。 4 内容 4.1 纯化水系统工作流程 4.1.1 本系统采用反渗透系统生产纯化水,其工作流程如下图表示 4.1.2 水质要求 4.2 标准操作规程 4.2.1 打开原水进水按钮,观察原水罐中是否有半桶以上的水,如果没有达到半桶以上,待其水位达到后,在进行下面操作

4.2.2 检查砂罐是否按照要求进行定期反洗(三天/次)。 4.2.3 检查炭罐是否按照要求进行定期反洗(三天/次)。 4.2.4检查混床是否按照要求进行定期维护(每周/次),将混床的“上进水”开,“下排水”开。 4.2.5制水前,检查各阀门的位置、管道是否正常,确定无异常后可以两种方案开启纯水系统: 4.2. 5.1将纯化水机开关钮(自动/停止/手动)调节至“自动”状态,纯水机自动运行。 4.2. 5.2将纯化水机开关钮(自动/停止/手动)调节至“手动”状态,开启原水泵启动按钮、一级高压泵启动按钮、二级高压泵启动按钮。 4.2.6观察仪器上RO 电导率的变化情况,直至电导率到达10μs/cm 以下,数值稳定后,计时10min.时间到后关闭中转水箱下排水。当RO 电导率大于10μs/cm 需更换反渗透膜; 4.2.7在RO 水位超过1/3时,开启EDI 启动按钮。观察仪器上EDI 电导率的变化情况,直至电导率到达3μs/cm 以下,数值稳定后,计时10min.时间到后关闭中转水箱下排水。当RO 电导率大于3μs/cm 需更换反渗透膜; 4.2.8观察压力表和流量计是否在合格范围内

橘子的种植操作规程

有机橘子生产技术规程 1品种选择 应选择抗病、抗逆性好且品质优良的橘子品种树苗。 2栽培方式 采取保护地栽培方式。在使用塑料薄膜时,只允许选择聚乙烯、聚丙烯或聚碳酸酯类产品,并且使用后应从土壤中及时清除,禁止焚烧,禁止使用聚氯类产品。 3植株管理 有机橘子树的管理同常规橘子生产,如根据品种特性、树龄、产量等确定结果母枝的剪留强度及更新方式,进行合理的修剪;在橘子生长季节,采用抹芽、定枝、新梢摘心、副梢处理等夏季修剪措施对树体进行整形控制,增强通风透光,以减轻病害发生。 4花果管理 采用疏花、疏果等常规方式对橘子进行处理,以控制产量、提高果实的品质。需要特别强调的是,绝不使用任何激素如赤霉素、CPPU等对果穗进行拉长或膨大处理。 5土、肥、水管理 5.1土壤管理 5.1.1中耕与深翻橘子生长季节及时中耕松土,保持土壤疏松,松土深度10cm ~20cm;每年果实采收后结合秋施基肥进行全园深翻,将栽植穴外的土壤全部深翻,深度30cm~40cm。 5.1.2生草与覆草有机橘子应提倡生草覆草技术,这样既有利于保墒

和保持土壤肥力,减轻日灼、气灼等生理病害的发生,又体现了生物多样性,为天敌提供了良好的栖息地。有机橘子园区进行生草时,一方面可以直接利用橘子园区的草资源,对高杆杂草加强管理,使其不影响橘子的生长,另一方面可以在4月份前后,在橘子行间种植不含转基因的白三叶草。覆草时间一般在7月份前后,将其刈割后覆盖在树根周围。 5.2施肥管理 5.2.1肥料要求生产前期可购买认证过的有机肥;持续有机橘子生产区应制定土壤有机培肥计划,如在自身橘子种植园区,结合“园区生草-养殖业”等进行绿肥或堆肥。绝对不能使用化学肥料、不能使用含有转基因的物质如转基因豆粕或经任何化学处理过的物质作为肥料,使用菜籽饼、大豆,应当按照相关要求进行充分腐熟和无害化处理。补充钾肥可用草木灰,补充磷肥可使用高细度、未经化学处理的磷矿粉。在施用磷矿粉时应与农家肥经充分混合堆制后使用。 5.2.2施肥在生长季节培肥的基础上,以施基肥为主,秋季施入,每66 6.7m2施入1000kg~1500kg充分腐熟的有机农家肥。采用沟施,在行间挖条状沟; 5.3水分管理 5.3.1补水时期一是萌芽到开花期,当土壤湿度过低时,定期补水。 5.3.2补水方法以采用滴灌法为宜。水质在符合GB5084规定的基础上,应加强有机橘子生产周边水质的监控,以免由于水质受污染而影响有机橘子生产。

实验室岗位安全操作规程

实验室岗位安全操作规程 1、在实验室工作当中存在化学灼伤、气体中毒、火灾触电、机械伤害等风险。 2、实验人员在实验室内必须穿戴工作服装,取样人员必须佩戴好相应的劳保用品与安全防护用具。 3、所有药品、标样、溶液都应有标签,绝对不要在容器内装入与标签不相符的物品。 4、不准穿拖鞋进入实验室,注意保持实验室的清洁卫生。 5、有毒、腐蚀、易燃、易爆的物品应妥善保管。贮存和使用应遵守?化学危险物品安全管理条例?。 6、实验室内严禁吸烟、饮水、进食。 7、开启易挥发液体试剂之前,先将试剂瓶放在自来水流中冷却几分钟,开启时,瓶口不要对人,最好在通风厨中进行。 8、易燃溶剂加热时,必须在水浴或沙浴中进行。 9、装过强腐蚀性、可燃性、有毒或易燃物品的器皿必须由使用者亲自洗净。 10、实验时若发现仪器设备出现故障或异常情况(如:有异味、冒烟等)时,应立即关闭电源开关,拨掉电源插头,并及时向实验室管理人员报告。 11、取下正在沸腾的溶液时,应用瓶夹摇动后取下以免溅出伤人。 12、玻璃棒、玻璃管、温度计等插入或拔出胶管时,均应垫有棉布且不可强行插入或拔出,以免折断刺伤人。

13、实验完毕,要关闭设备的电源、关好通风橱、整理好仪器设备,并打扫卫生。 14、配制药品或试验中能放出HCN、NO2、H2S、SO2、NH3及其它有毒和腐蚀性气体时,应在通风厨中进行,并带好必要的劳保用品。 15、实验室内应备有急救药品,消防器材和劳保用品。 16、化验室内应保持空气流通,环境清洁、安静。 17、易燃性气体不可与有助燃性的气体放到一个气瓶间,气瓶间内一定要有相应的防爆防倾设施。 18、样品的取样、接受、贮存和处置等要符合国家和公司的相关规定。 19、对实验产生的废液、废油、废物要分类存放并定期处置,禁止随意倾倒和储存。 20、实验室使用及存储的化学药剂或化学危险品都应备有相对应的化学品安全技术说明书(即MSDS),包括电子版和纸质版,并存放于实验室工作人员易于查找阅读的地方。同时实验室工作人员在使用化学药剂(特别是危险化学品)之前要对MSDS进行阅读学习,了解其危险特性及应急措施。 21、化学烧伤事故应急措施:当浓酸溅到眼睛或皮肤上时,应立即用大量清水冲洗,再用0.5%的碳酸氢钠溶液清洗;当强碱溅到眼睛或皮服上时,应迅速用大量清水冲洗再用2%的稀硼酸溶液清洗眼睛或用1%的醋酸溶液清洗皮肤。 当酸和碱滴溅到眼睛或皮肤上时,除经过上述处理外,还应马上送往医院进行救护。

仪表工作者安全技术操作规程(标准版)

( 操作规程 ) 单位:_________________________ 姓名:_________________________ 日期:_________________________ 精品文档 / Word文档 / 文字可改 仪表工作者安全技术操作规程 (标准版) Safety operating procedures refer to documents describing all aspects of work steps and operating procedures that comply with production safety laws and regulations.

仪表工作者安全技术操作规程(标准版) (1)一般安全规定 1.仪表工应熟知所管辖仪表的有关电气和有毒有害物质的安全知识。 2.在一般情况下不允许带电作业。 3.在尘、毒作业场所,须了解尘、毒的性质和对人体的危害,采取有效预防措施。 4.进入塔、槽等罐内作业应按罐内作业安全规定执行。 5.非专责管理的设备,不准随意开停。 6.仪表工工作前需仔细检查所使用工具和各种仪器以及设备性能是否良好,否则不能开始工作。 7.检修仪表时事前要检查各类安全设施是否良好,否则不能开始检修。

8.现场作业需要停表或停送电时必须与操作人员联系,得到允许,方可进行。电气操作由电气专业人员按制度执行。 9.仪表检修时,应将设备余压、余料泄尽,切断水、电、气及物料来源,降至常温,并悬挂“禁止合闸”及“现在检修”标志,必要时要有专人监护。取下保险。 10.使用电烙铁,不准带电接线,应接好再送电,严禁在易燃易爆场所使用易产生火花的电动工具,必须使用时要办理动火证。 11.仪表及其电气设备均须有良好的接地。 12.任何仪表和设备,在未证实有无电之前均应按有电对待。凡尚未弄清接线端的接线情况时,都应以高压电源对待。 13.仪表、电气及照明设备等的导线不得有破损、漏电情况。 14.仪表电源开关与照明或动力电源开关不得共用,在防爆场所必须选用防爆开关。 15.仪表及其附属设备,送电前应检查电源、电压的等级是否与仪表要求相符合,然后检查绝缘情况,确认接线正确、接触良好后,方可送电。

水处理系统质量控制标准操作规程

编号:CZ-GC-09877 ( 操作规程) 单位:_____________________ 审批:_____________________ 日期:_____________________ WORD文档/ A4打印/ 可编辑 水处理系统质量控制标准操作 规程 Standard operating procedures for quality control of water treatment system

水处理系统质量控制标准操作规程 操作备注:安全操作规程是要求员工在日常工作中必须遵照执行的一种保证安全的规定程序。忽视操作规程 在生产工作中的重要作用,就有可能导致出现各类安全事故,给公司和员工带来经济损失和人身伤害,严重 的会危及生命安全,造成终身无法弥补遗憾。 一、水处理系统的运行与保养 (一)水处理间应该保持干燥,水、电分开。每半年应对水处理系统进行技术参数校对。此项工作由生产厂家或本单位科室专业技师完成。 (二)水处理设备应该有国家食品药品监督管理局颁发的注册证、生产许可证等。每一台水处理设备应建立独立的工作档案,记录水处理设备的运行状态,包括设备使用的工作电压、水质电导度和各工作点的压力范围等。 (三)水处理设备的滤砂、活性炭、阴阳离子树脂、反渗膜等需按照生产厂家要求或根据水质情况进行更换。 1、石英砂过滤器根据用水量每周反洗1~2次。一般每年更换1次。 2、活性炭过滤器反洗的周期为1~2次/周,建议每年更换1次。

3、树脂软化器阳离子交换树脂一般每1~2年更换1次。 4、再生装置其再生周期为每2天再生1次。 5、精密过滤器过滤精度为5~10μm,一般2个月更换1次。 6、反渗透膜每2~3年更换1次。 (四)每天应对水处理设备进行维护与保养,包括冲洗、还原和消毒,每次消毒后应该测定消毒剂的残余浓度,确保安全范围,保证透析供水。 (五)做好维护保养记录。 二、透析用水的水质监控 (一)电导率正常值约10μs/cm。 (二)纯水的pH值应维持在5~7的正常范围。 (三)细菌培养应每月1次,要求细菌数<200cfu/ml;采样部位为反渗水输水管路的末端。透析机每台透析机每年至少检测1次。 (四)内毒素检测至少每3个月1次,要求细菌数<200cfu/ml,内毒素<2EU/ml;采样部位同上。每台透析机每年至少检测1次。 (五)化学污染物情况至少每年测定1次,软水硬度及游离氯

水稻种植技术操作规程完整

绿色水稻种植技术操作规程 为确保生产出绿色食品优质大米,特制定此技术规程。对水稻生产基地的规管理和技术指导,严格控制各种污染。 一、水稻栽培管理 (一)育秧 1、品种选择 选择适宜地区的优质高产新品种,不带病菌、虫源、无破粒、无秕粒、优质高产新品种。 经两年实验示,选出绥粳4,龙洋16,180,旱香74,旱香52等优质品种。这些品种抗病性强、生产性好、米质优良(一级米)、适口性好。 2、种子处理 晒种 3月25日——3月30日,室外晒种1—2天,增加种子活力。 盐水浸种,浸种4—5天后捞出控干进行催芽,有50%种子露白后方可播种。

3、营养床土配制 根据当地条件选用疏松、肥沃、富含有机质、偏酸、无草籽的腐殖土或旱田土,风干过筛,最好加15-20%草碳土,按照每平方米苗床需要20kg原土,加入适量75%浓硫酸使PH值达到5-6.0,制成营养土。 4、秧床整理 选取无盐碱、土地肥沃、平坦的壤土为育秧地块, 田要泡透使土质松软。秧床设置宽1.5米、长15米,秧沟宽30厘米,将秧床精细整理,达到平整一致,无作物根茬、杂草。 5、播种 4月5日——4月12日,首先在浇透水的置床上铺有孔地膜或软盘,然后铺撒3cm厚的营养土,刮平,浇透水,盘育苗每盘播70克左右,合理密植,稀播育壮秧。 6、秧田管理 温度管理 出苗前密封保温,棚温控制在30℃左右,苗出齐后及时撤去地膜。2叶1心时,温度控制在20—25℃左右,低温大床苗,严防高温烧苗和徒长。降温主要靠通风大小来调节。 水分管理

6、整地 11月上旬,水稻机收后,秸秆全部还田深翻,亩秸秆还田量在600公斤以上。 按照绿色食品优质米要求,四月下旬在耙地时施用em 菌发酵好的农家肥2吨/亩,施底肥磷酸二胺5公斤/亩,农家肥要施均匀,可以增加土壤有机质含量,改进土壤理化性状。 一次性深施,这样不仅节省施肥次数,而且有利于提高有效分蘖率和有效穗数及穗大粒饱,减少垩白率,提高稻米品质。 4月25日——5月1日泡田。 耙地 春季耙地以旱耙为主,泡田后水耙找平为辅,水耙地以寸水不露泥为准。 (二)插秧 5月9日——5月28日,当气温稳定在13℃时,就可开始插秧,插前3—5天整好地耙细耙平,寸水不露泥,沉淀适宜时,就可以插秧,插秧时要保证插秧质量,浅插,插秧

实验室操作规范

实验室安全操作规范 一、穿着规定 1、进入实验室,必须穿戴工作服。 2、进行危害物质、挥发性有机溶机、特定化学物质或其它毒性化学物质等化学药品及生物样品操作,必须要穿戴防护具(例如防护手套、口罩、防毒面具等)。 3、进行实验中,严禁戴隐形眼镜(防止化学药剂溅入眼内而腐蚀灼伤眼睛)。 4、需将长发及松散衣服妥善固定,且实验室内不得穿拖鞋。 5、操作高温(低温)实验,必须戴防高温(低温)手套。 6、实验服严禁穿回办公区。 二、饮食规定 1、严禁在实验室内吃东西。 2、食物禁止储藏在实验室的冰箱或储藏柜内。 三、试剂存储及操作相关规定 1、未进检测室时对实验过程进行预习,掌握操作过程及原理,弄清所有检测仪器及试剂的特性。估计可能发生危险的实验,操作时注意防范。 2、实验开始前检查仪器是否完好,装置是否正确稳妥。 3、实验进行时应经常注意仪器是否漏液、漏气、碎裂,反应是否正常等。 4、操作危险性试剂时,必须严格遵守操作守则,严禁自行更改实验流程。 5、使用试剂时,要首先确认容器上标示的名称是否为需要的实验试剂。确认药品是否为危害品,有无警告标识。 6、使用挥发性有机溶剂、强酸强碱、腐蚀性试剂、有毒试剂等必须在通风橱内进行操作。 7、有机溶剂、固体化学药品、酸性和碱性化合物均需分开存放,挥发性的化学药品必需放于通风良好的试剂柜内保存。 8、避免独自一人在实验室做危险实验。 9、做危险性实验时必须经领导批准,有两人以上在场方可进行,节假日和夜间严禁做危险性实验。 10、做有危害性气体的实验时,必须在通风橱里进行。 11、进行实验操作时(蜗旋、挥干、吹干等),严禁容器口对着人。 12、绝对不允许任意混合各种试剂以免发生危险。不能用手接触药品,不要用鼻子去闻试剂的气味,不得品尝任何试剂的味道。 13、检测剩余的实验试剂不得随意丢弃或带出实验室,放回指定容器或按规定处理。 14、实验操作完毕必须仔细认真清洗或擦拭可重复使用的实验器材,放置于规定位置。 四、用电相关安全规定 1、实验室内电气设备的安装和使用管理,必须符合安全用电要求,大功率实验设备用电必须使用专线,严禁与照明线共用,谨防因超负荷用电着火。 2、实验室内不准乱拉乱接电线。 3、实验室内的用电线路和配电盘、板、箱、柜等装置及线路系统中的各种开关、插座、插头等均应经常保持完好可用状态,空气开关功率必须与线路允许的容量相匹配。室内照明器具都要经常保持稳固可用状态。 4、实验室内仪器设备凡本身要求安全接地的,必须接地;要定期检查线路。 5、实验室内不得使用明火取暖,严禁抽烟。 6、手上有水或潮湿的,禁止接触电器用品或电器设备。 7、实验室内的鉴定人员必须掌握本室仪器、设备的性能和操作方法,严格按照规程操作。

电气系统操作规程标准版本

文件编号:RHD-QB-K9031 (操作规程范本系列) 编辑:XXXXXX 查核:XXXXXX 时间:XXXXXX 电气系统操作规程标准 版本

电气系统操作规程标准版本 操作指导:该操作规程文件为日常单位或公司为保证的工作、生产能够安全稳定地有效运转而制定的,并由相关人员在办理业务或操作时必须遵循的程序或步骤。,其中条款可根据自己现实基础上调整,请仔细浏览后进行编辑与保存。 1、根据工作,先开启电机或接通总电源,明确各开关或按钮的功能(进、退、左、右档位),按先低后高、先点动后连续的原则操作。 2、当设备在运行过程中出现异常现象时,必须立即停车,查找原因,排除故障,严禁设备带病运行。 3、电气系统出现故障需进行检修时,必须切断电源,严禁带电作业。 4、发电机在工作时,操作和维修人员必须穿绝缘鞋才允许进入设备。 5、电气系统的各种故障必须由专业电工或专职

维修人员检修处理,严禁其它人员擅自进行检修和改动。 6、操作人员应严格按照操作规程进行各项操作,严禁擅自违反、省略、更改操作程序,以确保设备正常运行及现场人员的安全。 7、操作人员必须听从现场施工指挥人员的统一指挥进行相应的操作,严禁凭个人想象或主观臆断盲目操作。 8、非操作人员严禁擅自进入操作室随意启动或操作各种控制按钮。 9、各项动作运行开始前,操作人员必须首先鸣响电铃,以示现场人员各就各位,准备进入工作状态。 10、起升机构一旦出现溜钩,一定要查清原因,处理后方可继续使用,以免发生溜钩失控事故。

11、避免在雷雨天气使用设备,遇到雷雨天气时,首先停掉设备总电源,同时工作人员不要逗留在设备上。 12、架桥机金属结构应采取保护接零,同时采取雷电,防感应电措施,接零保护每班检查一次。 13、运梁车和架桥机集中操作控制处须配备必要的干粉灭火装置,遇有火警,能及时处置。 14、设备启动前,应注意检查所有电气设备是否安全,各急停按钮是否回到工作位置,各个操作开关是否在零位。 15、当发电机启动后,应使用试电笔检测设备金属外壳是否有漏电现象,设备上的漏电保护开关应每月校验一次。 16、设备上不得使用电焊,如果必须使用电焊时,应与电气人员联系,并将微电箱、驾驶室内的所

有机水稻种植操作规程

有机水稻种植操作规程 (2012-06-05 19:26:26) 转载▼ 标签: 杂谈 格林有机公司 一、有机水稻栽培技术规程 1 有机稻产地环境技术条件 1.1 产地环境要求,有机稻生产基地应远离城区、工矿区、交通主干线、工业污染源、生活垃圾场等。基地的环境质量应符合以下要求: ①土壤环境质量符合GB15648-1995中的二级标准。 ②农用灌溉水符合GB5084的规定。 ③环境空气质量符合GB3095-1996中的二级标准和GB9137的规定。 1.2 产地要求,产地周边5Km以内无污染源,上年度和前茬作物均未施用化学合成物质;稻农技术好,自觉性高;土壤具有较好的保水保肥能力;土壤有机质含量 2.5%以上,pH值6.5~7.5;光照充足,旱涝保收。 1.3 灌溉水要求,稻田灌溉水水质符合GB5084的规定,水源为洮儿河水系,察尔森水库节制供水,水源充足、水质纯净、渠系配套,并且是分水后第一灌溉区,在灌区上游没有工业区和其他污染源。 2 育苗 2.1 品种选择 2.1.1 必须选择经过认证的达标的,同时经过转换的种子。通过穗选,连续两年的有机栽培获得有机稻种。 2.1.2 选择抗逆性强、抗病虫性强、熟期适中的、适口性好的优质品种。 2.1.3 禁止使用转基因种子。 2.2 种子处理 2.2.1 盘育秧每亩用芽籽4.0公斤。

2.2.2 晴天中午晒种2-3天,可杀菌并提高种子活力。 2.2.3 用黄泥选种,即50公斤水加10公斤过筛黄泥搅拌成浆(放新鲜鸡蛋漂出即可),放入稻种,去除秕粒,捞出好籽清洗一遍。 2.2.4 常规浸种催芽,常温水浸种5-6天,捞出放在30℃条件下破胸,80%种子露白时,降至25℃催芽,芽出齐后散温凉芽4-6小时即可播种。 2.2.5 温汤浸种催芽,将种子放入恒定的53℃水中提温半个小时(可提高种子活力和杀菌),然后进行常规催芽。 2.3 置床处理及播种 2.3.1 选择河淤土或山根腐殖土做盘土,去除杂草,破碎过细筛。 2.3.2 置床翻深10-15厘米,清除杂草,打碎坷垃,整平压实,碎土块搂至四周。 2.3.3 盘土配制,选择猪粪、羊粪和牛粪的混合腐熟农家肥(农家肥腐熟可杀死杂草种子、虫卵及病菌),农家肥与床土比例是1:5-10,破碎过细筛(6-8毫米);将食用白醋用水稀释到 pH值3左右的酸化水,用稀释的酸化水浇拌盘土,使盘土pH值达到4.5-5.5之间。 2.3.4 先将置床浇透水,待水渗下后,置床上摆盘,盘盘紧靠,装满盘土,浇透清水即可。 2.3.5 气温稳定通过5-6℃即可播种(4月20日-25日)。盘育苗每盘播芽籽100克,播均播匀,将种子轻压三面入土后覆土,再将地膜覆上压牢。 2.4 苗床管理 2.4.1 出苗前,棚内保持20-28℃温度。 2.4.2 秧苗出齐,顶膜立针时,及时撤掉地膜。 2.4.3 管水,床土发白变干时要及时浇水,浇水时间为早揭膜后和晚扣膜前。随气温增高和通风口加大,增加补水,每次浇透,切忌大水漫灌。结合浇水使用 2-3次酸化水(pH值5左右食用白醋稀释的酸化水),保持盘土的酸性,利于育壮苗抗病。 2.4.4 炼苗,出苗撤膜后,白天晴朗高温时,开始通风炼苗;1.5叶前小通风,即两端和背风面通风;1.5-2.5叶期,两侧适当通风;2.5叶后加大通风,使棚内温度保持在25℃左右,日均气温稳定在12℃以上时,昼夜通风炼苗,并逐步揭膜,防止失水,青枯死苗。 2.4.5 追肥,苗期结合浇水浇农家肥浸出液1-3次。

实验室安全标准操作规程

编号:SM-ZD-85177 实验室安全标准操作规程Through the process agreement to achieve a unified action policy for different people, so as to coordinate action, reduce blindness, and make the work orderly. 编制:____________________ 审核:____________________ 批准:____________________ 本文档下载后可任意修改

实验室安全标准操作规程 简介:该规程资料适用于公司或组织通过合理化地制定计划,达成上下级或不同的人员之间形成统一的行动方针,明确执行目标,工作内容,执行方式,执行进度,从而使整体计划目标统一,行动协调,过程有条不紊。文档可直接下载或修改,使用时请详细阅读内容。 1职责 1.1实验室主任对安全全面负责。经常进行安全督察,组织安全检查,负责处理安全事故。 1.2实验员负责水、电线路、消防器材的配置和设施安全检查。 1.3各科实验老师负责本科的化学药品、水电气、门窗的安全。 1.4实验员负责试剂、药品,特别是有毒有害,易燃、易爆物质的管理。 2. 工作程序 2.1 安全操作规范 2.1.1检测人员在工作中要严格按照操作规程,杜绝一切违章操作,发现异常情况立即停止工作,并及时登记报告。 2.1.2禁止用嘴、鼻直接接触试剂。使用易挥发、腐蚀性

强、有毒物质必须带防护手套,并在通风橱内进行,中途不许离岗。 2.1.3在进行加热、加压、蒸馏等操作时,操作人员不得随意离开现场,若因故须暂时离开,必须委托他人照看或关闭电源。 2.1.4各种安全设施不许随意拆卸搬动、挪作他用,保证其完好及功能正常。 2.1.5操作人员要熟悉所使用的仪器设备性能和维护知识,熟悉水、电、燃气、气压钢瓶的使用常识及性能,遵守安全使用规则,精心操作。 2.2 有毒有害物质的管理 2.2.1化学试剂、药品中凡属易燃易爆,有毒(特别是剧毒物品)、易挥发产生有害气体的均应列为危险物品,严格分类,加强管理,专人负责。 2.2.2建立详细帐目,帐、物、卡相符,专人限量采购,入库检查。 2.2.3危险物品、易燃易爆物品单独存放,有毒物品放入专用加锁铁柜内,注意通风。

实验室仪器标准操作规程的编写规范

适用范围 本规范适用于本单位仪器设备标准操作规程的编写。 操作规程编写的基本要求 1格式要求 1.1文件名称及编码 1.1.1文件名称命名方式为“生产厂家(或者品牌)简称+型号+仪器名称+标准操作规程”,例如:安捷伦1200高效液相色谱仪标准操作规程。 1.1.2编码方式为“ZY+序号-5.5-2013”,例如:ZY12-5.5-2013。 1.1.3字体均为宋体小四,并加粗。 1.2实施日期 1.2.1实施日期格式参照“2013.01.01”,字体均为宋体小四,并加粗。 1.3规程内容格式 1.3.1全部字体均为宋体小四,行距为1.5倍,内容小标题加粗。规程中的层次划分及编号,一律采用阿拉伯数字,层次的划分,一般不超过4部分,且首行不缩进。 2内容要求 2.1依据仪器设备使用说明、培训教程和上级已发布的标准操作规程编写,且与有关规定相协调。 2.2文字表达应准确、简明、通俗易懂,逻辑严谨,避免产生不易或不同理解的可能。 2.3规程中的图样、表格、数值、公式和其他技术内容应正确无误。 2.4规程中的术语、符号、代号应统一,与有关标准相一致,同一术语应表达同

一概念。 3规程一般由以下部分构成 3.1操作程序 3.1.1操作前准备 3.1.1.1核对和确认所用设备与检测目的所要求的一致性。 3.1.1.2识别设备计量检定状态标志和技术状态标志应符合规定;需要时,应标明对设备进行检定(自验)与检定周期的要求。 3.1.1.3设备使用技术条件和工作环境条件的检查与确定。 3.1.1.4安全与劳动保护条件的检查与确认。 3.1.1.5设备正确停机状态的检查与确认。 3.1.1.6检查并加油、加水、加燃料(必要时),达到规定要求。 3.1.2开机与运行 3.1.2.1接通电源,观察指示信号的要求与规定。 3.1.2.2必要时,仪器的预热或预置的要求与规定。 3.1.2.3校准或调零的要求与规定。 3.1.2.4设备自身系统误差的测定。 3.1.2.5必要时(如长时间停机后初次运行时)开机空运行,观察运行的稳定性。 3.1.2.6对样品的要求与规定 3.1.2.7上样和运行的操作步骤,要求与规定 3.1.2.8读取检测数据的方法。 3.1.2.9必要时(如长期停机,初次运行)在首次运行前,采用非检验试样做试验,观察设备运行的稳定性、可靠性。 3.1.3停机与保养 3.1.3.1检测完毕,停机状态的要求与规定。 3.1.3.2切断电源,观察指示信号的要求与规定。

排水系统操作规程标准范本

操作规程编号:LX-FS-A93284 排水系统操作规程标准范本 In The Daily Work Environment, The Operation Standards Are Restricted, And Relevant Personnel Are Required To Abide By The Corresponding Procedures And Codes Of Conduct, So That The Overall Behavior Can Reach The Specified Standards 编写:_________________________ 审批:_________________________ 时间:________年_____月_____日 A4打印/ 新修订/ 完整/ 内容可编辑

排水系统操作规程标准范本 使用说明:本操作规程资料适用于日常工作环境中对既定操作标准、规范进行约束,并要求相关人员共同遵守对应的办事规程与行动准则,使整体行为或活动达到或超越规定的标准。资料内容可按真实状况进行条款调整,套用时请仔细阅读。 1.0排污泵 1.1检查水泵电源是否正常。 1.2合上排污泵主电源回路的空气开关和控制电源开关。 1.3手动运行 1.3.1将控制箱面板上转换开关转到“手动”位置。 1.3.2按控制箱面板上“运行”按钮,水泵启动,污水开始排出。 1.3.3按控制箱面板上“停止”按钮,水泵停止运行,污水停止排出。

1.4模拟自动运行 1.4.1将控制箱面板上转换开关转到“自动”位置。 1.4.2手动提升控制杆浮球,触动高位触点,排污泵启动,污水开始排出。 1.4.3手动压下控制杆浮球,触动低位触点,排污泵停止运行。 1.5自动运行 1.5.1将控制箱面板上转换开关转到“自动”位置。 1.5.2当排污井内水位上升至排污井的3/4位置时,浮球触动高位触点,排污泵自动启动,污水面开始下降。 1.5.3当污水面下降至排污井1/4位置时,浮球触动低位触点,排污泵自动停止运行,污水停止

有机甘蓝种植操作规程

有机甘蓝种植规程 一、立地条件 1、气候条件 甘蓝为半寒性蔬菜,喜温和冷凉的气候。 2、土壤条件 甘蓝对土壤的适应性较强,以中性和微酸性土壤较好,且可以忍耐一定的盐碱性。甘蓝是喜肥和耐肥作物,对于土壤营养元素的吸收量比一般蔬菜作物要多,栽培上除选择保肥保水性能较好的肥沃土壤外,在生长期间还应施用较大量的肥料。 二、种子及其处理 1、品种选择 有机转换期内第一年可使用未经禁用物质处理的常规种子,以后年份必须使用有机种植区内繁育的种子。 2、种子质量 种子在播前进行发芽试验,种子质量应符合GB16715、3——1999中二级以上要求。 3、种子处理 将选好的种子先用冷水浸湿,再用45℃热水不断搅拌浸烫10分钟。

三、栽培技术措施 1、选地及整地 选择排水良好,土质肥沃的土壤。秋深耕,春耙糖,改善土壤耕作层的理化性质,提高地温。达到土质疏松、平坦,耕后耙平作畦,要求畦平埂直,每1.7米开厢起沟。 2、施肥 春季播种或定植前每亩撒施有机肥料300Okg。 3、播种 将处理好的种子用温水淘洗干净,在室温下浸种4小时后拌细土均匀撒播,播种后盖lcm厚细土,以不见种子为宜。每亩用种5Okg。 4、中耕 生育期内可以根据生长情况进行中耕,中耕可以保墒以提高地温。 5、浇水 甘蓝应变保持田块湿润,灌水后及时排除余水叶球包心后,气温下降少灌水。以防止叶球开裂。 6、追肥 进入保心期是甘蓝生长最快的时期,要求肥料充足,避免脱肥、干旱。莲座生长期和结球前中期应追施肥料。做到小肥提苗、中肥促叶、大肥功球。 7、病虫害防治 病虫防治应以预防为主,结合物理生物治疗为辅的方法进行。不能使用化学药剂和基因工程制品。 黑腐病:叶片边缘呈“v”字形黄色干枯。预防方法主要是加强管理,减少伤口,适时播种,与非十字花科轮作,亦可用Bt乳剂700倍液。

实验室安全标准操作规程(正式)

编订:__________________ 单位:__________________ 时间:__________________ 实验室安全标准操作规程 (正式) Standardize The Management Mechanism To Make The Personnel In The Organization Operate According To The Established Standards And Reach The Expected Level. Word格式 / 完整 / 可编辑

文件编号:KG-AO-5509-66 实验室安全标准操作规程(正式) 使用备注:本文档可用在日常工作场景,通过对管理机制、管理原则、管理方法以及管理机构进行设置固定的规范,从而使得组织内人员按照既定标准、规范的要求进行操作,使日常工作或活动达到预期的水平。下载后就可自由编辑。 1职责 1.1实验室主任对安全全面负责。经常进行安全督察,组织安全检查,负责处理安全事故。 1.2实验员负责水、电线路、消防器材的配置和设施安全检查。 1.3各科实验老师负责本科的化学药品、水电气、门窗的安全。 1.4实验员负责试剂、药品,特别是有毒有害,易燃、易爆物质的管理。 2. 工作程序 2.1 安全操作规范 2.1.1检测人员在工作中要严格按照操作规程,杜绝一切违章操作,发现异常情况立即停止工作,并及时登记报告。

2.1.2禁止用嘴、鼻直接接触试剂。使用易挥发、腐蚀性强、有毒物质必须带防护手套,并在通风橱内进行,中途不许离岗。 2.1.3在进行加热、加压、蒸馏等操作时,操作人员不得随意离开现场,若因故须暂时离开,必须委托他人照看或关闭电源。 2.1.4各种安全设施不许随意拆卸搬动、挪作他用,保证其完好及功能正常。 2.1.5操作人员要熟悉所使用的仪器设备性能和维护知识,熟悉水、电、燃气、气压钢瓶的使用常识及性能,遵守安全使用规则,精心操作。 2.2 有毒有害物质的管理 2.2.1化学试剂、药品中凡属易燃易爆,有毒(特别是剧毒物品)、易挥发产生有害气体的均应列为危险物品,严格分类,加强管理,专人负责。 2.2.2建立详细帐目,帐、物、卡相符,专人限量采购,入库检查。 2.2.3危险物品、易燃易爆物品单独存放,有毒

实验仪器标准操作规程

实验仪器标准操作 规程

微波炉标准操作规程 1. 操作程序: 1.1插上电源,打开柜门,放进需要加热的物品,关上柜门。 1.2选择火力,融化培养基一般用中高火即可。 1.3选择时间:融化培养基须注意边加热边拿出观察培养基融化情况,防止培养基爆沸。其它物品加热时间视情况调整时间。1.4加热完毕,将定时器调至0位置,小心拿出加热物品,关闭柜门。 2.微波炉操作注意事项: 2.1 微波炉上不得堆放杂物,微波炉周边不应放置其它电子产品。 2.2炉内未放入加工原料时,不要通电工作,不可使微波炉空载运行,否则会损坏磁控管。 2.3 凡金属的餐具,竹器,塑料,漆器等不耐热的器皿,有凹凸状的玻璃制品,均不可放入炉内使用。瓷制碗碟不能镶嵌有金属边的容器,不能放入炉内。 2.4袋密封包装及金属罐的食品不能直接放入炉内,以免爆炸伤人。 2.5 一定要关好炉门,确保连锁开关和安全开关的闭合。 2.6炉子关掉后,不宜立即取出加热物品,因为此时炉内尚有余热,物品还可继续加工,应过1分钟后再取出物品。

2.7定期检查炉门四周和门锁,如有损坏,密闭不严,应停止使用。以防微波泄漏,不一把脸贴近微波炉观察窗,防止眼睛因微波辐射而损伤,也不宜长时间受到微波照射,以防止引起头晕,目眩,乏力,消瘦,脱发等症状,使人体损伤。 2.8如设备发生问题,第一时间告知主管,设备停止使用,报维修部维修。 2.9光波烹调为烧烤专用,一般情况下不使用光波烹调和组合烹调按钮。 WD-A型电子连锁传递窗标准操作规程 1 .操作程序: 1 .1电子连锁传递窗任何状态下只能打开一扇门,以达到两边空气的严格隔离。 1 .2接通传递窗电源红色电源指示灯亮,操作人员可经过两个按键操作菌灯或开门。 1 .3按开门键(红色钮),一侧门开,货物进入传递窗内,关窗门(据需要开启灭菌灯)。 1.4另一侧门按红色键,取出物品,关闭窗门(若灭菌完需先关闭紫外灯)。 1 .4停电应急开启:传递窗两侧都设有小孔,停电时只要在小孔内插入细铁丝往中间拨动即可开锁。 2维护和保养与清洁 2.1每次使用完毕均需要进行紫外照射舱体。

废气处理系统标准操作规程【最新版】

废气处理系统标准操作规程 1.目的: 规范公司厂区废气处理设施的标准操作规程。 2.适用范围 适用于厂区捕尘系统的标准操作规程。 3.职责 全厂区捕尘系统的操作人员必须按照该规程进行操作。 4.内容 4.1. 捕尘系统的工作原理。 捕尘系统从产尘房间经过集气罩,将含尘空气经过风机吸入主管道内。在主管道内含尘空气经过捕尘滤袋进行捕尘操作,经过过滤的空气经过捕尘风机排放到房顶以上。

4.2. 捕尘系统的工艺流程图 4.3. 捕尘系统操作规程。 4.3.1. 操作前检查各台设备是否完好。 4.3.2. 检查所有设备传动装置是否有松动现象。 4.3.3. 检查所有设备接电是否良好。 4.3.4. 检查所有设备接地是否良好。 4.3. 5. 电动风机开关检查风机转向是否正确。 4.3.6. 检查过滤器是否完好。

4.3.7. 检查不使用房间集气罩风阀是否关闭。 4.3.8. 启动捕尘器风机,风机运行稳定后。进行下一步操作。 4.3.9. 打开需要使用房间集气罩风阀。 4.3.10. 打开需要使用房间独立风机。 4.3.11. 关闭时现关闭使用房间集气罩风阀。 4.3.12. 再关闭使用房间独立风机的开关。 4.3.13. 待产尘房间都关闭后,再关闭捕尘风机。 4.4. 捕尘系统的巡检工作。 4.4.1. 捕尘系统的日常检查 每日对捕尘系统进行巡回检查工作,主要检查内容包括独立风机运行是否正常、是否有异常声响。捕尘器风机运行是否良好、是否有异常声响。捕尘过滤器是否完好、是否存在漏洞。每天对捕尘系统检查两次,分别为上班(工作前)检查一次、下班(工作后)检查一次。

实验室安全操作规程

实验室安全操作规程 1、化验室安全操作规程 (1)实验前应做好准备,必须对所用药品与设备性能有充分的了解,熟悉每个具体操作中的安全注意事项。 (2)实验前必须熟悉实验室及其周围的环境和水龙头、电闸门的位置 (3)实验时应保持安静,思想要集中,遵守操作规程,切勿粗心大意,马马虎虎,更不准在实验室内开玩笑。 (4)严禁在实验室内饮食或煮食,或者把食具带劲实验室。 (5)每次实验完毕后应把手洗净,并检查水、电、气等安全措施完善后才能离开实验室(6)每个实验室都必须备有灭火器或砂土,并尽可能放在显眼的地方,能同事备有消防用的消防栓或水缸则更好 (7)实验室应保持空气流通,并设有专用的卫生箱,以供及时治疗的需要。常备药品有:红汞药水:供一般破伤使用 酒精:轻微的灼烧伤可用进过酒精的脱脂棉擦拭 5%硼酸氢钠溶液:受酸性物灼伤可用作冲洗 3%硼酸溶液:受碱性物灼伤可用作冲洗 还需要备有碘酒:紫药水及绷带和药棉 2、火和电的安全预防 (1)在使用中电气动力时,必须事先检查电开关,马达以及机械设备各部门是否安置妥善(2)开始工作时和停止工作时,必须将开关彻底扣严和拉下 (3)在更换保险丝时,要按负荷量,不得加大或以铜丝代替使用。 (4)严禁用湿手、湿布或铁柄毛刷等去清扫或擦拭电闸刀、点插销等,防止触电 (5)凡电气动力设备过热时,应立即停止运转 (6)定碳、定硫电炉或其他高温炉,其硅碳棒露出部分应设有安全罩,严禁将安置妥善的安全罩随意撤掉,以免发生触电事故。 (7)禁止洒水在电气设备和电线路上,以免漏电 (8)凡使用110伏以上电源装置,仪器的金属部分必须安装地线 (9)电热设备,例如马弗炉、烘箱、电炉和电热板等,所用电源的导线应经常注意检查其各接触处是否妥当,导线有无损坏和被腐蚀等 (10)马弗炉、烘箱等用电设备,使用时必须要有人负责照管,以防发生事故 (11)马弗炉需放在水泥等不燃物砌成的坚固台子上,不要靠近木板墙或木质门窗。(12)使用易燃物时,必须在距离火源较远的地方进行,绝不可靠近火源,尤其是乙醚着火的危险性极大,用时必须小心,用完后的剩余部分也应及时的存放到专门的安全地方。(13)绝不可以将氧气钢瓶存放在靠近电源的地方,并需防止强烈震动,气体出口活门处绝不可涂油和与有机物接触,以免发生爆炸的危险。 3、化学药品的安全预防 (1)距度星的药品,例如KCN、AS2O3等等,必须制定保管使用规则,并严格遵守,及时工作人员很少,也不可例外的有所忽视。这类药品不能与一般药品同样的存放和任意使用,即使用过后的余量已经很少也应及时送保管员及时查收,不应任意放在工作台上。 (2)内服有毒药品,如氰化物、铅化物、汞及汞化物、络酸盐、氧化砷钡盐等,应装在坚固的瓶中保管,禁止入口,与手接触用后要洗手。 (3)接触皮肤有毒的药品,例如氰化物、氟氢酸、溴水、过氯酸等,要装在严禁坚固的瓶中保管,使用时要特别小心,不得与皮肤接触。 (4)呼吸有毒药品(及有毒气体和蒸气)如氰化氢、氮的氧化物、氯化氢、硫化氢、溴、

相关文档