PERFORM OperatingDocumentUse and Maintenance of Tuttnauer Table-Top AutoclavePC-POD-CA-006-v04Revision History1.IntroductionPERFORM Centre has a Tuttnauer electronic table-top autoclave Model 2540 E. This table-top autoclave is designed for the sterilization of wrapped and unwrapped instruments as well as glassware and liquids. It is a manually operated device, with a control system based upon steam pressure. The operator can select a sterilization temperature from within a range of 212ºF - 273ºF (100ºC - 134ºC). Proper decontamination depends on temperature and contact time; steam must contact pathogens long enough for destruction.Note: This autoclave is intended to use only for laboratory research experiments. The autoclave is appropriate to sterilize surgical instrument for surgery or biopsy, please make sure to program at 121°C for 40 minutes. Do not use for sterilization of biological waste.1.1Training requirementsPrior to using the autoclave individuals should:•Read and sign this POD.•Undergo appropriate autoclave training and/or provide a proof of anexternal training to the clinical analysis supervisor prior to use.1.2General precautions•To avoid possible damage, do not leave the autoclave un-attended while in operation.•Never place flammables in the autoclaves•Never place materials that become TOXIC or corrosive when subjected to heat.•Personal Protective Equipment, clothes and other safety measures should be used at all times.•For proper sterilization do not overload the chamber. Only autoclavable products shall be used; please refer to the materials or instrumentsmanufacturer’s instructions for sterilization of unknown materials orinstruments. Recommended materials for autoclave are borosilicate glass,polypropylene, polycarbonate plastic and stainless steel.•Should the autoclave fail to reach the sterilizing temperature/pressure, always check first that the door is fully sealed. If not, then tighten the doorbolt further until completely sealed. Allow chamber pressure to reach 0psi before opening the door.•USE DI WATER ONLY. The impurities in tap water will create the need for more frequent cleaning and maintenance.•For cleaning, do not use steel wool, steel brush or bleach as this can damage the chamber and trays.•To avoid being burned by hot steam, do not place your face over the safety valve.1.3Relevant documents•Tuttnauer Operation & Maintenance Manual (Electronic TabletopAutoclaves Models 1730, 2340, 2540, 3140, 3850, 3870 E, EK, EA & EKA)2.Definition of T erms and Abbreviations3.Procedure3.1Autoclave Operation3.1.1 Preparation before autoclaving•Instruments to be sterilized must be clean and free from any residual matter, such as debris, blood, pads or any other material. Such substances may cause damage to the instrument or the sterilizer.•After ultrasonic cleaning or thorough cleaning, rinse under distillated water for 1 minute and pat dry.•Launder textile wraps prior to reuse, but do not use bleach.•Be sure that instruments of dissimilar metal (stainless steel, carbon steel, etc.) are separated. Carbon steel instruments should be bagged or placed on autoclavable towels and not directly on stainless steel trays (mixing will result in the oxidation of these metals).•Load items within the boundaries of the tray (not directly on the bottom) so that they do not touch the chamber walls, or fall off when the tray is inserted into the autoclave. The chamber walls are very hot, items that come into contact with the wall can be damaged.•Place a sterilization indicator in each tray or inside each wrapped pack. When using a paper / steam resistant plastic bag the plastic side should always be down.•All instruments must be sterilized in an open position (ex. scissors must be kept open).•Surfaces that are hidden because the item is in a closed position will not be exposed to the steam and will not be sterilized, loosen the caps of liquid containers. Use suitable wrapping when sterilizing clean equipment for reuse.•Disassemble or sufficiently loosen multiple-part instruments prior to packaging to permit the sterilizing agent to come into contact with all parts of the instrument.•Do not overload the sterilizer trays.•Empty canisters should be placed upside-down, in order to prevent accumulation of water.•Wrapped instruments should be packed in material which will allow steam penetration and promote drying, such as autoclave bag, autoclave paper, or muslin towels.•If autoclaving liquids use a heat-proof glass bottle and make sure contents are no more than 2/3 filled with the cap loosely closed to prevent pressure build-up.•If spotting is detected on the instruments the first step would be to use an ordinary eraser to remove the spot. If removal of the spot reveals pitting then the spot was most likely rust. It may also be an indication that the instruments were rinsed in tap water with a high content of minerals. These minerals when exposed to high temperature and steam will acceleratethe oxidation of the metal. One suggestion would be to final rinse the instruments in distilled water.•If the instruments exhibit a discoloration, this can be due to the mixing of carbon steel and stainless steel. When these two metals come into contact with each other electrolysis occurs that breaks down the metal.The best solution is to separately wrap the carbon steel to insulate it from other instruments or the trays.3.1.2Sterilization ProgramsThe autoclave offers three preset sterilization programs at a temperature of up to 134°C and one dry program. Cycle parameters can be customized to suit your needs.•Unwrapped cycle is use for unwrapped instruments and materials. Correct loading of the autoclave is essential to successful sterilizing. Default settings: sterilization at 134ºC, 3 min (no dry).•Wrapped cycle. This cycle is use for wrapped instruments and porous loads. All instruments must be sterilized in an open position. Default settings: sterilization at 134ºC, 7 min (dry time 60 min).•Glassware cycle. For non-medicinal liquids in open bottles and glassware. Default settings: sterilization at 121ºC, 30 min (drying is not allowed).•Drying cycle. Select the accessory drying program to continue the heat assisted drying process.3.1.3Fill the Water ReservoirNote: Daily before operation, check the water level in the reservoir and add DI water when required. Once a week or after 20 cycles (the shorter period)replace the water in the reservoir. Enter the date and time for each use in the log book.•Ensure that the drain valve is in a CLOSED position.•Remove the water reservoir cover.•Pour 350-400 mL distilled water or mineral free water into the reservoir through the opening on top of the autoclave, until it reaches the base ofthe safety valve holder. Under no circumstances fill any higher than thebase of the safety valve holder. For the empty reservoir pour 3 liters DIwater.•For proper operation make sure the water level is above the coils of the cooling coil.3.2Preventive3.2.1Daily maintenanceClean door gasket with a mild detergent, water and a soft cloth or sponge. The gasket should be clean and smooth.3.2.2Weekly1.Clean the air jet. To ensure that the temperature inside the chamber rises properly, it is necessary to keep the air jet clean.2.After 20 cycles or as required, clean and descale the chamber, copper tubes and the reservoir using Chamber Brite.3.Put a few drops of oil on the 2 door pins and door tightening bolt screw shaft and bearing.4. Clean the outer parts of the autoclave with a soft cloth. The water sensor has to be cleaned with a damp cloth once per week to ensure that the water level in the chamber is properly reported to the microprocessor at all times during the cycle. The water sensor is located in the rear of the chamber.3.2.3Periodically1.Once every month clean and check the safety valve (see section 9.6 of user manual).2.Replace the door gasket every 12 months, or as needed (see section 9.5 of user manual).3.Once a year inspect the locking device for excessive wear.4.Drain the reservoir approximately after 20 cycles.5.Clean strainer (see section 9.8 of user manual)3.2.4Draining the reservoir (section 9.3 of user manual)Caution:Before draining, ensure that the electric cord is disconnected and there is no pressure in the autoclave.The drain valve is located on the front left side of the autoclave after the door is opened. The function of the drain valve is to drain the water reservoir.1.Connect the silicone hose, supplied with the autoclave, to drain into a bucket.2.Turn drain valve counter clockwise to the open position.3.Fully drain the reservoir4.With a quart of tap water flush out the reservoir5.Turn drain valve clockwise to the close position.6.Connect the electric cord to power source.7.Fill the reservoir with 3 liters distilled water to just below the safety valve (see section 6.5 of user manual)8.Turn on the main power switch.9.The autoclave is now ready for use.3.2.5Cleaning the Air Jet(Located in the water reservoir)A dirty air jet is the number one cause of failed cycleThe elimination of air from the sterilization chamber during heat up is critical to the proper operation of the autoclave. If the air jet is clogged, an error message “Low Heat” will appear.The air jet consists of a small orifice with a clean out wire inserted in it (wire is permanently installed and will not come out). It is required that the air jet be cleaned once per week or more often if necessary, to remove any accumulated dirt and debris. It is preferred to clean the air jet when the unit is running a cycle and under pressure. This is so that any loosened debris will be blown away, however, it can be done while the unit is idle.1.Remove the water reservoir cover.2.Clean the hole of the jet by manipulating the air trap wire back and forth 10 times. It is important to clean the hole of the air trap.3.2.6Replacing the Door GasketPull off the gasket from the door groove. Install the new gasket as described in the manual page 44. Caution: This gasket is designed with a trapezoidal cross section. The gasket should be placed with the widest side towards the door.3.2.7Checking the Safety Valve(Located in the water reservoir)In order to prevent the safety valve from becoming blocked, it is necessary to allow the steam pressure to escape through the valve. This procedure should be done every month as follows:1.Run a sterilization cycle with a sterilization temperature of 273ºF (134o C) according to the manual.2.Allow a pressure of approximately 30 psi (260 kpa) to build up in the chamber.3.Turn the unit off.4.Remove the water reservoir cover. This next step will expose you to HOT STEAM so be careful. To avoid being burned by hot steam, do not place your face over the safety valve.5.Pull the ring of the safety valve using a tool, i.e. screwdriver, hook etc. and open the safety valve for 2 seconds then release. Be careful not to burn your hands.6.Turn the unit back on an press the STOP key to abort and vent the cycle.7. Wait until the pressure decreases to zero, only then can the door be opened.。

立式压力蒸汽灭菌器筒的使用、清洁、维护保养操作规程

立式压力蒸汽灭菌器筒的使用、清洁、维护保养操作规程

陈家温一.目的:建立立式电热压力蒸汽灭菌器的使用、清洁、维修保养规程,确保其操作规范化。

二.依据:《药品生产质量管理规范》2010版《立式电热压力蒸汽灭菌器使用说明书》三.适用范围:适用于立式电热压力蒸汽灭菌器的操作。

四.责任:立式电热压力蒸汽灭菌器的操作者。

五.内容:1.灭菌原理:立式电热压力蒸汽灭菌器是根据湿热杀灭微生物的原理设计而成的.利用密封容器内安装的浸入式电加热器,加热介质水所产生的高压饱和水蒸汽,以热力对细胞壁和细胞膜的损伤及对核酸的作用,使微生物蛋白质发生凝固致使其死亡。

2.立式压力灭菌器筒使用方法:2.1 正确放置灭菌器:为了使灭菌器能正确地工作,灭菌器必须放置在结实的水平地面上,勿使本立式电热压力蒸汽灭菌器筒直接撞伤或受到剧烈震动。

2.2 利用立式电热压力蒸汽灭菌器筒的脚轮,踩下刹车脚轮,固定好设备的位置,防止设备移动。

检查电源线、排水管是否连接完好.检查手动放汽阀是否关闭、压力表显示示数是否为“0MPa”.2.3基本操作流程2.3.1自动加水工作流程开启容器盖→取出灭菌网篮→装载→密封→加入适量纯化水→开启电源→设置工作参数→启动→自动运行→工作结束→开启容器盖→卸载→关闭电源注:密封时,仔细检查密封圈安装状态:密封圈应完全嵌入。

确认密封圈上无异物,无裂痕,若有异物,请以软布清除密封圈上的异物.2.3.1.1启动长按“启动”键,启动程序,设备开始工作.长按“启动”键后,有“滴”的声响,系统开始正常工作,进入自动控制灭菌过程. 2.3.1.2自动运行设备按所设定的参数功能自动运行,电加热器启动,开始加热。

(若为自动进水,程序启动后,先行自动加水约2分钟.而后,电加热器开始工作。

)灭菌室内的蒸汽与冷空气被排出。

(达到102℃后,排气电磁阀将延时约2分钟后自动关闭,停止排汽。

)灭菌室没温度达到设定值,灭菌计时开始,灭菌时间窗口显示剩余灭菌时间。

(在此期间,灭菌室内温度与压力恒定;本设备安全阀整定压力为0.24MPa,当灭菌室内压力过高时,会自动启动。

高压灭菌器的操作说明书

高压灭菌器的操作说明书

高压灭菌器的操作说明书高压灭菌器操作说明书第一部分:产品概述高压灭菌器是一种用于医疗设备、实验室器具以及食品饮料等领域的消毒设备。

其基本原理是通过高温高压的环境,达到杀灭病菌和微生物的目的。

本操作说明书将详细介绍高压灭菌器的使用方法和注意事项。

第二部分:安全须知在使用高压灭菌器之前,请仔细阅读以下安全须知,并严格遵守,以确保个人安全以及设备的正常使用。

1. 在操作高压灭菌器时,请确保自己已经穿戴好相应的防护装备,包括手套、防护眼镜和防护服等。

2. 严禁打开或更换设备的内部部件,除非有合格的维护人员指导。

3. 避免将易燃物质或易爆物质放置在高压灭菌器附近的区域。

4. 在操作过程中严禁将手伸入高压灭菌器的内部。

5. 使用结束后,请确保关闭高压灭菌器的电源开关,并将设备从电源中拔出。

第三部分:操作步骤1. 准备工作a. 将高压灭菌器放置在平稳的工作台面上,并确保周围没有其他杂物。

b. 将高压灭菌器的电源插头插入电源插座,并确认电源指示灯亮起。

2. 加载物品a. 打开高压灭菌器的门,并将待处理的物品放置在设备内部的物品托盘上。

b. 注意不要过度放置物品,确保每个物品之间有足够的空间,以便于蒸汽充分进入。

3. 设置操作参数a. 使用高压灭菌器的控制面板,设置灭菌时间和温度。

b. 建议根据待处理物品的特性和灭菌要求,选择合适的操作参数。

4. 启动高压灭菌器a. 关闭高压灭菌器的门,确保密封良好。

b. 按下设备上的启动按钮,高压灭菌器开始工作。

5. 等待和结束a. 根据所设定的操作参数,等待灭菌时间的结束。

b. 听到设备发出的结束提示音后,可以关闭高压灭菌器的电源开关。

第四部分:注意事项1. 在操作过程中,不要随意打开高压灭菌器的门,避免受热蒸汽烫伤。

2. 灭菌结束后,待处理的物品会变得烫手,建议使用热防护手套进行取出。

3. 高压灭菌器的内部可能残留高温高压的环境,请在操作结束后等待一段时间再进行清理和维护。

高压蒸汽灭菌登记表范本

高压蒸汽灭菌登记表范本

高压蒸汽灭菌登记表范本在实验室中进行高压蒸汽灭菌时,需要记录灭菌过程的相关信息,以便于追踪和查询。

本文提供了一个高压蒸汽灭菌登记表的范本,供读者参考使用。

下面是本店铺为大家精心编写的4篇《高压蒸汽灭菌登记表范本》,供大家借鉴与参考,希望对大家有所帮助。

《高压蒸汽灭菌登记表范本》篇1一、高压蒸汽灭菌登记表基本信息1. 表号:2. 实验日期:3. 实验时间:4. 实验人员:5. 灭菌物品名称:6. 灭菌物品数量:7. 灭菌方式:高压蒸汽灭菌8. 灭菌温度:9. 灭菌时间:10. 灭菌人员:二、高压蒸汽灭菌过程记录1. 灭菌前物品状态:2. 灭菌物品包装情况:3. 灭菌物品放入灭菌器时间:4. 灭菌器开启时间:5. 灭菌器关闭时间:6. 灭菌物品取出时间:7. 灭菌物品放置地点:8. 灭菌器型号:9. 灭菌器使用情况:三、高压蒸汽灭菌结果记录1. 灭菌物品名称:2. 灭菌物品数量:3. 灭菌结果:4. 灭菌时间:5. 灭菌温度:6. 灭菌人员:四、高压蒸汽灭菌注意事项1. 灭菌物品应清洗干净后再进行灭菌。

2. 灭菌物品应放置在灭菌器的中间,避免接触灭菌器的壁面。

3. 灭菌器的盖子应紧贴灭菌物品,确保蒸汽能够到达物品的各个部位。

4. 灭菌过程中应定期检查灭菌器的压力表和温度计,确保灭菌器正常运行。

5. 灭菌结束后应等待灭菌器冷却后再打开盖子,避免烫伤。

6. 灭菌物品在取出后应及时放置在无菌环境中,避免再次污染。

以上是一个高压蒸汽灭菌登记表的范本,可以根据实际需要进行修改和补充。

在实验室中进行高压蒸汽灭菌时,应认真填写登记表,确保灭菌过程的可追溯性。

《高压蒸汽灭菌登记表范本》篇2以下是一份高压蒸汽灭菌登记表的范本,您可以根据需要进行修改和调整。

高压蒸汽灭菌登记表日期:年月日时间:起始时间:分秒结束时间:分秒操作人员:姓名操作工号设备名称:高压蒸汽灭菌器设备型号:工作压力:kPa工作温度:摄氏度灭菌物品:包装方式:包装材料:包装尺寸:长 x 宽 x 高灭菌时间:分钟灭菌批次:批次号操作步骤:1. 检查设备是否正常运行,确认温度和压力达到要求。

压力蒸汽灭菌器的维护保养

压力蒸汽灭菌器的维护保养

压力蒸汽灭菌器的维护保养压力蒸汽灭菌器是医疗机构中常用的设备,它可以通过高温高压的蒸汽对物品进行灭菌处理,是防止交叉感染的必备工具。

然而,如果长时间使用且缺乏维护保养,设备的灭菌效果会受到影响,甚至可能存在安全隐患。

下面介绍一下正确的维护保养方法。

灭菌器的清洁维护在使用灭菌器之前,需要对设备进行清洁消毒以及对零部件进行维护,确保设备的正常运行。

首先,要对灭菌装置内面进行清洗,建议选择专业的清洗剂(如双氧水),要注意清洗时间以及清洗的次数,一般情况下每半年进行一次清洗即可。

此外,还要注意清洗装置内的蒸汽向导管道和真空压力表,避免灰尘或杂质影响设备的正常运转。

检查设备的工作状态在日常使用过程中,需要不断检查设备的工作状态,以确保设备的工作正常。

首先要检查门和密封垫是否完好,如果发现有磨损或漏气的情况需要及时更换。

其次,要检查蒸汽点和真空表是否正常。

最后,还需要对设备进行调试,确保设备的所有功能正常。

维护灭菌器的电器设备灭菌器的电器设备包括控制系统、电动阀门等,需要对其进行定期的维护保养。

一般情况下,每年需要对电器设备进行一次大维护,检查电缆、接线、端子等是否损坏,同时还要进行润滑等维护保养工作。

提高设备的使用寿命除了以上的维护保养方法,我们还可以通过如下几种方式来延长设备的使用寿命。

首先,要定期对设备进行检查和校正,以保障设备的运转和灭菌效果。

其次,要在使用过程中减少设备的负载,避免过度使用,从而延长设备的使用寿命。

最后,还要注意对设备进行适当的保险措施,增强设备的安全性和稳定性。

在日常维护保养过程中,压力蒸汽灭菌器的使用人员需要学习掌握相关知识,对设备进行适当的维护保养,并定期进行检查和校正,以确保设备的运转和灭菌效果。

同时,还需要注意对设备进行适当的保险措施,确保设备的安全性和稳定性。

高压灭菌器的使用说明书

高压灭菌器的使用说明书

高压灭菌器的使用说明书使用说明书一、产品介绍高压灭菌器是一种用于彻底灭菌的设备,采用高压蒸汽的方式,能够有效杀灭细菌和病毒,确保物品的卫生与安全。

本使用说明书将详细介绍高压灭菌器的使用方法,帮助用户正确、安全地操作设备。

二、安全须知1. 使用高压灭菌器前,请仔细阅读本说明书,并确保已了解所有的安全须知。

2. 请确保设备的电源线不被压到或绊倒,以防止设备短路或跌倒造成人身伤害。

3. 在使用高压灭菌器时,必须保持设备周围的通风良好,避免出现过热现象。

4. 请勿在高压灭菌器工作过程中打开设备的盖子,以免受伤或烫伤。

5. 高压灭菌器仅适用于灭菌用途,禁止将其用于其他任何用途。

6. 请勿将高压灭菌器放置在易燃、易爆或潮湿的环境中,以免造成设备故障或危险。

7. 使用高压灭菌器时,请勿将手指或其他物品插入设备内部,以防触电或其他伤害。

8. 高压灭菌器需要定期进行维护和保养,请按照说明书中的要求进行操作。

三、准备工作1. 首先,确保高压灭菌器连接正常的电源,并使设备处于关闭状态。

2. 接下来,将待灭菌物品全部放入高压灭菌器的灭菌篮中,并确保物品之间有足够的间隙,以保证蒸汽能够顺利进入。

3. 将灭菌篮放入高压灭菌器内,并确认篮子处于适当的位置。

四、操作步骤1. 打开高压灭菌器的盖子,确保蒸汽能够顺利进入设备。

2. 根据物品的需求,设定适当的灭菌时间和温度。

建议根据物品类型选择不同的灭菌程序,以确保最佳的灭菌效果。

3. 确认设备设置无误后,关闭高压灭菌器的盖子。

4. 打开高压灭菌器的电源开关,启动设备。

5. 高压灭菌器开始工作后,将显示屏上的进度条进行监控,确保灭菌过程的顺利进行。

6. 当灭菌过程完成后,高压灭菌器将发出蜂鸣声,同时显示屏上会显示相应的提示信息。

7. 此时,可以关闭高压灭菌器的电源开关,停止设备的工作。

等待设备冷却后,即可打开设备的盖子取出灭菌好的物品。

五、维护保养1. 每次使用高压灭菌器后,请用干净的布或纸巾擦拭设备的外部,并定期进行清洁和消毒,保持设备的整洁。

(完整word版)高压蒸汽灭菌器期间核查记录表

(完整word版)高压蒸汽灭菌器期间核查记录表
B法:采用标准菌株对照法进行核查。经过一个灭菌周期,在无菌条件下,取出标准试验包或指示菌片,投入溴甲酚紫葡萄糖蛋白胨培养基中,56℃培养48h,观察颜色变化。
4பைடு நூலகம்结果:
评定标准
A指示卡达到标准黑色以上,表示符合灭菌条件;低于标准黑色则表示不符合灭菌条件。
B溴甲酚紫葡萄糖蛋白胨培养基颜色未变,表示符合灭菌条件;颜色变黄,则表示不符合灭菌条件。
核查结果
核查人:审核人:年月日
高压蒸汽灭菌器期间核查记录表
检测室名称
核查日期
设备名称
设备编号
规格型号
核查依据
环境温度
环境湿度
核查内容
核查步骤
1.外观及标志:
2.功能性检查:
3.核查方法:
A法:采用压力蒸汽灭菌化学指示卡,将此指示卡印有指示剂的一端,放入物品包装的中间;按灭菌操作的常规,预热、彻底排除空气,当锅内温度达到121℃后,保留20分钟以上(不同物品所需灭菌时间,请按有关规定执行),灭菌完毕将指示卡抽出,观察颜色变化。

高压灭菌器使用方法说明 灭菌器如何做好保养

高压灭菌器使用方法说明 灭菌器如何做好保养

高压灭菌器使用方法说明灭菌器如何做好保养高压灭菌器是能产生高蒸汽压和高温,对物体进行灭菌的装置。

那么高压灭菌器实在应如何使用呢?下面将为您说明:1.首先在使用前应打开排气阀、消毒阀、干燥阀及给水阀,加RO水或蒸馏水制玻璃管绿线处。

2.然后再关闭全部阀,设定灭菌时间(15~30分钟左右)。

3.打开电源开关,将外锅开始加热加压(此时电源开关及加热灯亮,外锅压力表开始升压)。

4.将待灭物放入内锅,关紧锅门。

待外锅压力表达 1.2 kg/cm2时,打开消毒阀。

5.待内锅压力表达,温度上升至121时,再打开定时开关,自动计时约15~30分钟后鸣叫,说明灭菌完成。

当灭菌完毕后,不可放气减压,否则瓶内液体会猛烈沸腾,冲掉瓶塞而外溢甚至导致容器爆裂。

应待灭菌器内压力降至与大气压相等后才可开盖。

(1)不要将灭菌锅安装在有易燃气体或易燃液体的地方,确认电源所供电压与仪器所示电压相同,蓝线为零线,棕线为火线,如绿线为地线,务必要使仪器接地良好.(2)对不同类型、不同灭菌要求的物品,不要放在一起灭菌,以免顾此失彼,造成损失。

(3)每次使用前,必需检查灭菌锅内的水量,确保有充分的水,使水位高于电热管(4)对液体灭菌时,应将液体灌装在耐热的玻璃瓶中,以不超过3/4体积为好。

要用玻璃纸和纱布包扎瓶口,如用橡皮塞的,应插入针头排汽,不要使用未打孔的橡胶和软木塞.(5)摆放灭菌物品时,各种包裹不应过大、不应过紧,以免阻拦蒸汽透入,影响灭菌效果。

严禁堵塞安全阀的出气孔,必需留出空位保证其畅通放气,否则安全阀因出气孔堵塞不能工作,造成事故。

(6)易燃和易****如苯类等,禁用高压蒸汽灭菌法。

锋利器械如刀、剪不宜用此法灭菌,以免变钝。

(7)己灭菌的物品应做记号,以便识别,并需与未灭菌的物品分开放置,以免弄错。

(8)安全阀应定期检查其牢靠性。

每次灭菌前,应检查安全阀的性能是否良好,以防锅内压力过高,发生爆炸。

工作压力超过0.165 MPa时需要更换合格的安全阀。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
高压灭菌器使用维护记录本
高压灭菌器使用维护记录
设备编号:22018年月
日期
开机时间
使用情况
压力
温度
灭菌时间
日保养
关机时间
操作者
1
正常□
异常□
是□
2
正常□
异常□
是□
3
正常□
异常□
是□
4
正常□
异常□
是□
5
正常□
异常□
是□
6
正常□
异常□
是□
7
正常□
异常□
是□
8
正常□
异常□
是□
9
正常□
异常□
是□
10
正常□
异常□
是□
26
正常□
异常□
是□
27
正常□
异常□
是□
28
正常□
异常□
是□
29
正常□
异常□
是□
30
正常□
异常□
是□
31
正常□
异常□
是□
日保养
仪器运行正常标准
1、清洁、(75%酒精)消毒仪器外部2、用纱布擦净门胶圈、容器内壁、提篮3、去除容器内壁的水垢4、检查安全阀5、检查逸气阀排气是否通畅
1、电源正常2、仪器检正常
3、显示屏正常4、逸气阀排气通畅
5、高压灭菌效果合格
是□
11
正常□
异常□
是□
12
正常□
异常□
是□
13
正常□
异常□
是□
14
正常□
异常□
是□
15
正常□
异常□是□ຫໍສະໝຸດ 16正常□异常□
是□
17
正常□
异常□
是□
18
正常□
异常□
是□
19
正常□
异常□
是□
20
正常□
异常□
是□
21
正常□
异常□
是□
22
正常□
异常□
是□
23
正常□
异常□
是□
24
正常□
异常□
是□
25
正常□
异常□
相关文档
最新文档