Nitrogen generator model N2-BoraUser’s manualMATHESON TRI-GAS166 Keystone DriveMontgomeryville, PA 18936Phone: 215-641-2700Fax: 215-641-2714Email: ******************************IndexINDEX (1)INTRODUCTION (2)S COPE OF THE MANUAL (2)S PECIFICATIONS (2)N OTES ON FCC COMPLIANCE (2)C ORRECT USE (3)P ACKING LIST (3)DESCRIPTION (4)INSTALLATION (5)R ECEIVING THE GENERATOR (5)P LACING THE GENERATOR (5)S YMBOLS USED ON THE GENERATOR (5)G AS CONNECTIONS (5)INITIAL START-UP (6)START N2-BORA (6)O PERATION (6)MAINTENANCE (9)R OUTINE MAINTENANCE (9)R ETURNING THE UNIT (13)SPARE PARTS LIST (14)IntroductionScope of the manualThis manual provides operation and maintenance instructions for model N2-Bora nitrogen generator.SpecificationsSpecifications of the nitrogen generatorNitrogen flow rateSTP:Standard temperature and pressure(20°C,1bar)0-500cc/min at STP 0-750cc/min at STP 0-1300cc/min at STPMax outlet pressure5bar(73psi)Purity99.999%+(O2<10ppm) Weight18kgInput voltage 120or240V/50or 60HzFuse4A or2APressure accuracy0.1bar(±0.5%) Microprocessor controlled display Graphic display,128x64pixels Index of protection IP2xOperating conditions:-Temperature-Relative humidity 15°C to+40°C0-80%,non condensingOver voltage category IIPollution degree2Sound pressure level48dB(A)Case dimensions350x430x230mm(WxDxH)Notes on FCC complianceThis equipment has been tested and found to comply with the limits for a Class B digital de-vice,pursuant to part15of the FCC Rules.These limits are designed to provide reasonable protection against harmful interference in a residential installation.This equipment gener-ates,uses and can radiate radio frequency energy and,if not installed and used in accor-dance with the instructions,may cause harmful interference to radio communications.How-ever,there is no guarantee that interference will not occur in a particular installation.If this equipment does cause harmful interference to radio or television reception,which can be determined by turning the equipment off and on,the user is encouraged to try to correct the interference by one or more of the following measures:∙Reorient or relocate the receiving antenna.∙Increase the separation between the equipment and receiver.∙Connect the equipment into an outlet on a circuit different from that to which the re-ceiver is connected.∙Consult the dealer or an experienced radio/TV technician for help.WARNING!Any changes or modifications to this equipment not expressly approved by the manufac-turer may void the user's authority to operate the equipment.Correct useNitrogen generator is designed to produce nitrogen for laboratory use.The unit must only be operated for this purpose,according to the specifications and instructions provided in this manual.In particular,the following warnings must be observed at all times:∙Indoor use only∙Never operate the unit in below-zero temperatures.∙Only operate the unit in a room with sufficient ventilation(see“Placing the unit”).∙Always unplug the unit from the mains power supply before accessing the internal components for replacement.∙Only the parts described in the“Spare parts list”can be replaced by the user.Packing listList of items included in the shipmentQuantity Description1Nitrogen generator1Instruction manual1Power cableDescriptionThe nitrogen generator produces pure nitrogen through the filtration of compressed air.The key element of the generator is a carbon molecular sieve that is able to separate the nitro-gen molecules from the oxygen ones present in the compressed air stream.The two carbon molecular sieve tubes are self regenerating cyclic.Two compressors are an integral part of the generator and their purpose is to compress air up and force it into the tubes containing the carbon molecular sieve.The generated nitrogen gas is accumulated in a specific reservoir.The internal pressure is controlled by a pressure transducer.The outlet pressure is adjusted with a forward pressure regulator and by default set to 5bar.The nitrogen then passes through the outlet port at the rear.VALVESAIRINLEInstallationReceiving the generatorAll units have been carefully inspected before transport.Visual checks for damage and func-tional tests should be performed upon receipt.Any damage must be immediately noted and reported.The generator must only be returned according to the shipping instructions pro-vided.Placing the generatorThe nitrogen generator must be placed on a flat,level,vibration-free,shock-free surface.Do not place the generator over a source of heat,as this may cause the device to overheat.The unit should not be in contact with any other objects on any side,and the air inlet must not be blocked.Leave at least30cm of free space at rear for ventilation.Do not operate the generator in a sealed or unventilated room.Do not operate the generator at below freezing temperatures.Operation is guaranteed at operating temperatures between+15and+40°C.WARNING!Normal precautions for any nitrogen supply should be taken when using the generator.DO NOT use in sealed or unventilated rooms.Nitrogen can cause suffocation. Symbols used on the generatorInitial start-upStart N2-Bora-Switch on the unit.-Push the START button.-The unit’s compressor will start to run and nitrogen will be produced immediately.-It will last 30minutes before the nitrogen produced will reach the stated purity.-The system is ready for use now.OperationThe operating status of the unit is shown on the main screen on the graphic display.The main screen has three options at the bottom,corresponding to the three buttons on the unit,which are used to run the various functions and access theconfiguration.-The button on the left is used to start the unit.-The button on the right is used to stop the unit.After swiching on the generator,the menu is show as above :Where xxxx represents the output pressure read.The range goes from 20to 9999mBar.Pressure N2=xxxx mBarStarted Stop-The button on the centre is used to enter the sub menu.*Set PN2AlarmRunning TimeSettingsExit↓SelectThe selected item is marked with a"*"and this can be changed by scrolling the middle key. Setting the N2Pressure AlarmSet PN2Alarm1000mBarOption ↓ ↔The value of the pressure alarm can be set by scrolling the values by using both↓and↔. Once the value is set,press"Option"button to select either"Save"the value or"Cancel"the whole operation.Remark:the default value after flashing the MCU is set to1000mBar;setting the PN2alarm does not require password;the delay is set to approx.six minutes.Running TimeThis submenu shows the total running time and the unit is hours.When4000working hours are passed,it’s time to replace some consumables,as indicated in the next chapter,“Main-tenance”.This data can be reset at any time as follows:1)Switch off the N2Generator2)Press the left button and maintain it pressed("Start button")3)Switch on the N2Generator4)Stop pressing the left button5)Enter"Running time"sub menu6)Enter password(not the default one,but a new one inserted before!)7)Type the"reset"buttonRemark:should the password be incorrect when pressing the "Reset"button,"Err"is shown just after the wrong code:if the correct code is set,pressing this key will reset this data and exit this sub menu.Setting sub menuSeveral tasks can set within this sub menu:start option,external remote configuration,switch on/off the buzzer in case of alarm and changing password.Start OptionWithin this sub menu,when "Auto start ON"is set,the N2Generator will automatically start when switching on the N2Generator.When "Auto Start OFF"is set,the N2Generator will be stopped when switching on the Generator.Remark:this option is disabled if external remote control is enabled.In this case,pressing "start"or "stop"key has no effect;the default setting after flashing the MCU is set to "Auto start ON";no password is required in this sub menu.External remote control mode ("Remote Relay Mode")Three choices are available :"Normally OPEN","Normally CLOSE"and manual mode."Normally OPEN":the N2generator automatically starts when the remote control is opened and it stops when this remote control is closed.The maximum delay time is 2seconds and the manual mode is disabled."Normally CLOSE":the N2generator automatically starts when the remote control is closed and it stops when this remote control is opened.The maximum delay time is 2seconds and the manual mode is disabled.If none of the option above is selected,the manual mode is enabled.This option is set by default after flashing the MCU.Remark:no password is required in this sub menu.Pin out DB15Remote ControlDB9Remote Control open contact.open contact.1No alarm 6Alarm ControlStart /StopBuzzer alarm DB15Remote ControlUser can set the buzzer in case of alarm.This setting does not require password.Password settingThe password can be changed within this er needs to provide the old password before submitting a new one.If user forget the this password,the default one(04122)can be submitted.MaintenanceWith proper care and maintenance,your nitrogen generator should provide you with years of trouble-free operation.There are no adjustments to be made to the generator.The only rou-tine service operations are those described below.Nonetheless,the generator should be inspected approximately every3years.Contact your supplier.Routine maintenanceThe following section describes the maintenance operations required for the correct opera-tion of the nitrogen generator.CleaningThe internal components of the nitrogen generator do not need to be cleaned and should not be accessed by the user for cleaning.To clean the outside of the unit,only use a damp cloth (no detergents,acids or aggressive or abrasive substances).Service kits∙Every4000hours the user must replace the filters of the compressor.∙Every8000hours the user must replace the filters and the membrane of the com-pressor.∙Every24000hours the user must replace the carbon molecular sieve(CMS).The specific maintenance kits are available as spare part.These kits include all required parts:they are listed in the“List of spare part”.WARNING!Only qualified personnel should perform service on this product.Any damage done to this product as a result of improper maintenance procedure will void the warranty.4000h service kita)Exchange of the compressor’s inlet filters∙Switch off the instrument and disconnect from the power line ∙Open the hoodUnscrew the 2screws on rear of theunitUnscrew the 4screws on right side and on rear of theunitScrewsScrews∙Exchange the filtering media &Silencer∙Reassemble in reverse order.b)Exchange of the automatic purge filtering media∙Switch off the instrument and disconnect from the power line ∙Open the instrument∙Disconnect the 2tubes fixed at the top of the automatic purge (quick connectfittings)Filter &SilencerFilter &auto-matic purge valve∙Unscrew the 4screws on top of the automatic purge∙Exchange the filtering media (white cylinder)∙Reassemble in reverse order.8000h service kita)Exchange of the compressor’s inlet filter See 4000h service kitb)Exchange of the automatic purge filtering media See 4000h service kitc)Exchange of the membrane in the inside of each compressor ∙Switch off the instrument and disconnect from the power line ∙Open the instrument∙Unscrew the 4screws on top of the pump∙Replace the seals andmembranesScrewsScrewsFiltering MediaAttention:be very carefully to note the original position of the various parts.Reassemble in reverse order.Tighten screws a little at the time and crossover until tight.Start the generator and test for proper operationReturning the unitIn the event of any faults or damage,first notify the agent or distributor who supplied the unit.Please also provide full details of the problem,including the model and serial number.Instructions will then be provided for the service or the return of the unit.If the one year war-ranty has expired,or the fault is due to misuse of the unit,all repair and shipping costs are to be paid by the customer.All other costs are borne by the customer,except as otherwise ex-pressly agreed upon.WARNING!If the unit has to be transported,make sure to use suitable packaging.The unit should be transported in an upright position;this warning should be reported onthe outside of the packaging.Top view.Parts wereturned over 180°CCSpare parts listThe table below provides a list and description of the spare parts of the nitrogen generator. List of spare parts–N2-Borap/n DESCRIPTIONNB-4044000h service kit for N2-BORA-500(filters)NB-4054000h service kit for N2-BORA-750/ICP(filters)NB-4018000h service kit for N2-BORA-500(filters+compressor kit)NB-4028000h service kit for N2-BORA-750/ICP(filters+compres-sor kit)NB-40824000h CMS tubing kit completeIMPORTANT!The manufacturer reserves the right to change or modify its products without prior notice.。

高纯氮气说明书

高纯氮气说明书

7、用户不要自行将电解池拆卸打开(用户无法自行修理),以免影响整机使用。

8、仪器如需搬运时,把储液桶中的电解液用吸液管吸干净,然后盖好上盖,以免在运输中残留的电解液外溢,将整个仪器腐蚀,造成无法修复的后果9、如仪器停机一个月或一个月以上时间,请把电解液抽出。

四、仪器的工作原理与特点:1、工作原理:本仪器根据电催化法进行空气分离原理制成。

其中电解池是利用燃料电池的逆过程设计而成。

当作为压力稳定且纯净的原料空气进入到电解池中,空气中的氧在阴极被吸附而获得电子并与水作用生成氢氧根离子并迁移到阳极,最后在阳极处失去电子析出氧气,因此空气中的氧不断被分离只留下氮气。

再经过后级过滤、稳压、稳流处理从而得到高纯度的氮气。

2、仪器特点:①自动排空:仪器增加了自动排空装置,免除了手动排空的繁琐及误操作,以保证开机后自动排空30分钟左右。

②程序控制:仪器的控制系统采用专用芯片。

使全部工作过程均由程序控制完成。

自动恒压、恒流,氮气流量是根据气相色谱仪的用气量来实现0-500ml/min全自动调节。

③工艺先进:电解池采用北京怡丰瑞普科技有限公司最新膜分离技术,催化层使用了贵金属催化物,使电解池催化能力强、效率高、产气量大、氮气纯度高。

电解池出厂前经过100小时以上的高压、大电流的老化实验,使电解池的性能和工作状态极为稳定。

④浸泡式电解池:工作时电解池散热性能良好,不需强制循环。

电解效率高,并且在电解池的进气和出气装有单向阀,使电解池内在任何情况下都保证有0.15MPa的压力。

从根本上解决了过液现象。

⑤三级催化:除电解池中两级催化外另有第三级催化,催化剂选用新型贵金属,使含氧量小于3PPM。

⑥产氮气湿度低:采用了超高分子量渗透膜分离技术及有效的除湿装置,因而降低了原始湿度,并能在停机后自动排出水分。

采用了金属聚合物除湿及两级吸附,使氮气的纯度大大提高。

⑦操作方便:免运输钢瓶之劳,省搬运钢瓶之苦,使用时只需打开电源开关即可输出流量稳定的高纯氮气。

氮气发生器资料Nitroflow Lab membrane N2 Generator

氮气发生器资料Nitroflow Lab membrane N2 Generator

一、原理及构造•1、原理•把自带压缩机产生的压缩空气通过膜组分离成氮气和氧气。

•分离过程示意图如下:• A 压缩空气进口 B 中空的纤维管 C 氮气出口•F 快速透过S 慢速透过2分离过程说明•环境空气中氮气含量大约为78.1%,氧气含量大约为20.9%,氩气含•量大约1%,还有微量的二氧化碳、水蒸气和惰性气体。

•压缩空气流过中空管束的外围,氮气透过管壁的速度远远慢于空气中•的气体气体,这样,扩撒快的气体迅速透过管壁释放到大气中,这样•只有渗透慢的氮气通过管束被收集起来使用3空气压缩机氮气压缩机5构造-外部-背面M 主电源开关N 电源接入口O 氮气出口阀P 透气网7氧气传感器氮纯度调节阀压力控制阀取样流量控制阀8进气碳吸附器触摸屏二、安装和调试1、安装1.1、环境条件-环境温度在10-35℃相对干净的室内,相对湿度在31℃时<80%,在40℃时<50%,通风良好,没有阳光直射。

1.2 、因为设备能产生热量,所以要有良好的通风。

安装四面离墙间隙大于50cm。

1.3、电源-230V.50HZ,电压波动在10%的范围内。

1.4、客户可以直接接使用设备到氮气机上,建议在氮气机和使用设备之间加装一个储气罐,以保证稳定的氮气流量和纯度。

92、制氮流程图N1 空气进口 C 进口碳吸附器N2 氮气出口LP 空气压缩机RKV 启动阀M 气体分离膜PI1 膜压力显示FCV 流量控制阀HP 氮气压缩机PSV1/2 压力释放阀PCV 氮气压力释放阀PI2 氮气压力显示V1 止回阀V4 减压阀V2 出口球阀10113、开机调试•3.1、确认所有的连接是正确的和紧固的。

•3.2、接上合适的电源,打开在设备背面的主开关。

•3.3、这时所有的开机和设定都是在操作显示屏上完成。

•不要用尖锐的器具操作触摸屏。

•3.4、屏幕显示3分钟的延时后,压缩机启动。

•注意:必须在所有的外壳都安装好的情况下启动设备,否则会破坏设备热交换系统,导致设备自动停机或损坏。

氮气发生器使用方法

氮气发生器使用方法

氮气发生器使用方法氮气发生器是一种用于产生高纯度氮气的设备,广泛应用于实验室、工业生产和医疗领域。

正确的使用方法可以确保设备的正常运行和氮气的高纯度输出。

下面将介绍氮气发生器的使用方法。

一、设备准备1. 将氮气发生器放置在通风良好的地方,远离易燃易爆物品,并保持设备周围的清洁。

2. 检查设备的电源线是否连接良好,确保设备接地可靠。

3. 确认设备的压力表显示为零,并检查压力表是否正常工作。

二、氮气发生器操作1. 打开氮气发生器的电源开关,并等待设备启动。

2. 调节设备的压力控制阀,使之逐渐增加到设定值。

一般来说,氮气发生器的输出压力范围为0.5-10bar,根据具体需求进行调节。

3. 打开氮气发生器的进气阀门,使氮气发生器开始吸入空气。

注意,进气空气应为干燥无油的空气,否则会影响氮气的纯度。

4. 当压力表显示达到设定压力时,说明氮气发生器已经正常工作,开始产生高纯度氮气。

5. 当需要停止使用氮气时,首先关闭氮气发生器的进气阀门,然后调节压力控制阀,将输出压力调至零,最后关闭氮气发生器的电源开关。

三、设备维护1. 定期清洁氮气发生器的进气过滤器,防止灰尘和杂质进入设备。

2. 检查氮气发生器的压力表和压力控制阀是否正常工作,如有异常及时更换或修理。

3. 对氮气发生器进行定期的保养和维修,以确保设备的长期稳定运行。

四、注意事项1. 使用氮气发生器时,应遵循相关的安全操作规程,确保人身安全。

2. 在操作氮气发生器时,应注意设备的工作状态,确保设备正常运行,避免发生故障或事故。

3. 当设备长时间不使用时,应将氮气发生器的电源开关关闭,避免不必要的能源浪费。

以上就是氮气发生器的使用方法。

通过正确的操作和维护,可以保证氮气的高纯度输出,满足实验、工业和医疗领域的需求。

在使用过程中,务必注意安全,并定期进行设备的保养和维修,以延长设备的使用寿命。

希望以上内容能对您有所帮助。

氮氢空三气一体发生器安全操作规定

氮氢空三气一体发生器安全操作规定

氮氢空三气一体发生器安全操作规定概述氮氢空三气一体发生器是一种将氮气、氢气和空气混合后发生爆炸反应的设备,主要用于实验室等科研领域。

由于其使用过程中存在较大的安全风险,因此需要严格的安全操作规定。

操作前的准备工作在操作之前,需要进行以下准备工作:1.仔细阅读说明书和安全操作规定,了解设备的使用方法和注意事项;2.检查设备的所有部件是否完好无损,特别注意气体进口和出口是否正常;3.检查设备是否处于可靠的地面状态,以防止设备移动或翻倒。

操作步骤以下是氮氢空三气一体发生器的安全操作步骤:1.打开气源阀门:先打开氮气源的阀门,然后打开氢气源的阀门,最后打开空气源的阀门;2.调整流量:根据需要设定氮氢空三气的比例,然后逐渐调整气体流量,直到设备工作稳定;3.启动装置:按照说明书操作,启动装置。

在启动前要确保周围无明火或其他易燃物品;4.注意观察:在设备工作期间要时刻观察设备的运行情况,如出现异常现象需立即停止操作;5.关闭装置:操作结束后,先关闭空气源的阀门,再关闭氢气源的阀门,最后关闭氮气源的阀门;6.排空操作:在排出管内残留气体时,要确保周围无明火或其他易燃物品,防止受到气体爆炸的伤害。

安全提示使用氮氢空三气一体发生器时需要注意以下安全提示:1.操作时必须佩戴安全防护装备,如护目镜、手套、防护服等;2.操作时必须保持清醒,切勿在疲劳或饮酒后操作设备;3.气源管道和设备必须进行防静电处理,以防止静电引起火灾或爆炸;4.在操作时,应时刻注意周围环境,防止引发火灾或爆炸事故;5.操作人员必须经过专业培训和合格考试,具备操作和应急处理能力。

应急措施在设备使用过程中,若出现意外情况或突发情况,应立即采取以下应急措施:1.断开气源:立即关闭气源阀门,切断气体供应;2.疏散人员:立即疏散人员,远离现场,保持安全距离;3.立即通知:立即向管理人员报告情况,并立即采取措施控制风险;4.等待救援:在采取措施后,等待救援人员到场处置。

氮气发生器的分类及应用

氮气发生器的分类及应用

氮气发生器的分类及应用氮气是一种广泛应用于工业、医疗、实验室等领域的气体。

在许多应用场合中,氮气被用作惰性气体、氧化剂、脱氧剂等。

为了满足不同行业领域的需求,氮气发生器应运而生。

本文将介绍氮气发生器的分类及应用。

氮气发生器分类氮气发生器根据其工作原理和使用范围可以进行分类,下面分别介绍:根据工作原理分类1.压缩空气氮气发生器:该类氮气发生器主要是通过压缩空气将氮气从空气中分离出来,它们通常需要额外的空气净化和压缩单元,因此需要较大的安装空间。

2.膜式氮气发生器:该类氮气发生器是利用扩散膜将空气中的氮气从其他气体中分离出来,其工作原理与反渗透膜类似。

该类氮气发生器能够在相对较小的安装空间内生成氮气,但需要经常更换扩散膜。

3.分子筛氮气发生器:该类氮气发生器是利用特殊的化学反应,将空气中的氧气和水蒸气从其他气体中分离出来。

它们在生产纯氧和脱除湿气方面表现出色,但需要高温和高压环境下才能制造氮气。

根据使用范围分类1.工业氮气发生器:该类氮气发生器主要应用于工业领域,例如金属冶炼、氮化、电子工业和食品行业。

2.医用氮气发生器:该类氮气发生器主要应用于医疗领域,例如麻醉、急救、氧气疗法和氧气呼吸。

医用氮气发生器需要满足严格的卫生标准和医疗设备证明要求。

3.实验室氮气发生器:该类氮气发生器主要应用于实验室,例如基础研究、化学合成和分析测试。

氮气发生器应用氮气发生器在不同领域都应用广泛,下面将分别介绍不同领域的应用:工业领域工业领域中氮气常用于金属加热、氮化处理和电子组件制造。

•金属加热:金属加热通常需要使用氮气作为惰性气体防止氧化反应,同时可以提高加热效率。

•氮化处理:氮化处理需要在高温下进行,可以使用氮气作为气源,其含氮量高,使氮元素更容易被物质吸收。

•电子组件制造:电子组件的制造需要在洁净的环境下进行,使用氮气可以有效防止杂质的污染,同时保持高质量的制造。

医疗领域医疗领域中氮气常用于麻醉和氧气呼吸。

•麻醉:氮气是一种惰性气体,在麻醉中使用,可以减少病人的意识反应,从而减轻病人的疼痛感。

BCHP高纯度氮气发生器SPN-300A SPN-500A使用说明书

BCHP高纯度氮气发生器SPN-300A SPN-500A使用说明书

BCHP高纯度氮气发生器 SPN-300A / SPN-500A使 用 说明书北京中惠普分析技术研究所B C H P 受控文件敬告:1. 在使用仪器前请您认真阅读此说明书!2. 本仪器必须配有空气源为原料气才能工作。

3. 本仪器要以纯净并且压力恒定的空气为原料气。

切勿在缺电解液的情况下运行!以免造成电解池无法修复的后果。

4. 运输前请务必将电解液倒净!以免在运输途中腐蚀机箱及内部元件,造成仪器报废。

5. 若产品更新,仪器内部结构略有变化,恕不另行通知! 注意:1. 本说明书适用于SPN-300A型和SPN-500A型。

2. 在加电解液时请缓慢将电解液倒入储液桶内。

要经常观察电解液水位线,降到下限须立即加入纯净水或二次蒸馏水。

切勿缺水运行!3. 更换完过滤器中的填充物后,务必要将过滤器上盖拧紧,并将过滤器和过滤器底座拧紧,以免造成漏气。

4. 本仪器标称流量为最大产气量,不能长时间使用在最大流量,建议用气量不要大于最大产气量的1/2,否则会影响气体纯度,缩短电解池使用寿命!企业通过ISO9001:2008质量管理体系认证产品通过国家分析仪器质量监督检验中心检测产品通过国家标准物质检测中心检测产品荣获部级科技进步二等奖产品中国专利号:94222810.3B C H P 受控文件目 录一、公司介绍 (1)二、产品概述 (1)三、仪器各部位名称 (2)四、仪器的安装与使用 (2)五、仪器的工作原理与特点 (5)六、主要技术参数 (7)七、仪器的成套性 (7)八、制造商的保证 (8)九、附图1 气路连接示意图 (8)十、附图2 仪器的故障原因与排查方法…………………………………9 B C H P 受控文件一、公司介绍 北京中惠普分析技术研究所成立于1994年,是目前国内规模最大的气相色谱仪器配套气源发生器的生产厂家之一,产品涵盖高纯度氢气、高纯度氮气、低噪音空气源等各种流量单体机及各种相关组合机。

产品系列多,品种全,有多种流量和纯度可供选择。

制氮机制氮设备氮气发生器安全操作及保养规程

制氮机制氮设备氮气发生器安全操作及保养规程

制氮机、制氮设备、氮气发生器安全操作及保养规程前言制氮机、制氮设备、氮气发生器是化工、医疗、生物制药、电子、化妆品等行业广泛使用的设备之一。

在安全使用制氮机、制氮设备、氮气发生器的同时,正确的保养也是不可忽视的。

本文将从安全操作和保养方面进行详细介绍。

安全操作规程设备准备在使用前应检查设备,确认各部位紧密,防护措施完好。

检查后应将设备操作手册翻阅一次,获取操作注意点,了解设备操作细节。

设备操作•在操作前应将压力表、控制器等仪器设备校准好,确保设备工作的准确性,安装好仪表,保障操作人员安全;•操作人员应全程留意设备运转情况,注意观察压缩机头温度、排水量、的调整等指标,及时调整;•在使用过程中如果设备出现故障,应立即停机,并查找故障原因,找到准确的解决方法之后才能再次开机;•在设备运行过程中严禁喷雾等活动,以免引起火灾;•操作人员在操作期间应保持注意力集中,无论工作时间多长,都不能出现疏漏、懈怠。

设备的保养保养设备可以延长设备寿命,保障生产效率。

以下为设备保养内容。

•每次使用完成后应将设备清理干净,包括设备的所有部位。

严禁用水冲洗,应使用清洁布擦拭干净,以保障设备的完好性;•定期检查滤网、气密性、散热器等,来保证设备性能的稳定;•在设备长时间未使用时,应关闭设备电源,并进行防尘防潮处理;•定期进行设备保养,养护设备的标准是每半年一次,清洗除尘、造成氧化损坏其他杂物的时间段则根据使用频率来确定。

保障操作人员安全氮气具有较多的危险性,如凝固、凝缩、转化等,容易引发突发环境变化等意外事件。

操作人员在操作时要注意以下事项:•保持工作间空气流通情况,确保操作人员的呼吸安全;•操作过程中要穿着符合相应行业安全标准的工作服、劳保品,如静电服、耳罩、安全鞋等;•操作人员在操作过程中应保持冷静清醒,避免过度疲劳导致事故的产生。

结束语通过使用制氮机、制氮设备、氮气发生器,并正确的进行保养,可以保证设备的性能及安全,并延长设备的使用寿命。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

色谱联用氮气发生器使用说明
一、仪器的工作原理:
本仪器(AYN-500/300)氮气发生器根据电催化法进行空气分离原理制成。

其中电解池是利用燃料电池的逆过程设计而成。

当压力稳定且纯净的原料空气进入到电解池中,空气中的氧在阴极被吸附而获得电子并与水作用生成氢氧根离子并迁移到阳极,最后在阳极处失去电子析出氧气,因此空气中的氧不断被分离只留下氮气。

再经过后级过滤、稳压、稳流处理从而得到高纯的氮气。

二、仪器的安装与使用:
(一)、启用前的准备:
1、将仪器从包装箱内取出,检查有无因运输不当而损坏,核对仪器备件、合格证是否齐全。

2、加电解液:
①首先取出备件中的氢氧化钾150g,全部倒入容器内,然后加入去离子水或桶装纯净水1000ml充分搅拌等待电解液冷却后备用。

②用一根Φ3的气路管,一端与氮气发生器背后的空气进气口连接。

另一端与空气源的空气出口连接,不得漏气。

③打开储液桶上盖,将冷却后的电解液全部倒入储液桶内,盖上储液桶上盖浸泡25分钟(储液桶的容量为1000ml)。

④首先打开空气源的开关,这时空气源的压力在上升,(要求输入空气的压力不低于0.4MPa)氮气发生器压力随着空气源的上升也缓慢上升,当空气源的压力上升到0.4MPa,氮气发生器的压力也上升到
0.4MPa。

(氮气压力比空气压力略低一点)
三、仪器的自检(勿与气相色谱仪联机)
1、将氮气的电源插头与电源插座连接,打开氮气发生器的开关,此时氮气发生器压力为0.4MPa,数字流量显示较大超过500以上,3分钟内应稳定在500-550之间。

2、拧紧氮气发生器出口螺帽,不能漏气。

3、氮气发生器大约在3分钟内数字流量显示逐渐降至为030以下,说明仪器系统工作正常,自检合格。

四、仪器的使用:(与气相色谱仪联机)
1、旋下氮气发生器出气口的螺帽,此时仪器前面板上的竖向数值显示为500-550之间。

2、把气路管的一头首先连接仪器背后的氮气出气口上拧紧,气路管另一端连接色谱仪,这时仪器可以使用了。

3、仪器使用时应注意氮气发生器的流量显示是否与气相色谱仪的用气量一致,如氮气发生器的流量显示超出气相色谱仪的实际用量较大时,应停机检漏。

其方法参照仪器的故障原因与排除方法进行调整,再用自检检查合格后方可使用。

4、本仪器过滤器中装有变色硅胶和分子筛。

使用过程中透过观察窗口观察过滤器中的变色硅胶下端是否变色,如变色需要更换或再生。

下层的分子筛在仪器使用三个月后烘干再生。

更换或再生的方法为:首先把系统放气,(关掉电源,卸下空气进气口即可放气),使系统压力为常压状态再旋下过滤器,拧开过滤器上盖,更换过滤材料后拧紧过滤器上盖,将过滤器装到底座上拧紧。

不能漏气。

也可自检的方法观察数显是否回030以下。

5、仪器使用一段时间后,电解液会逐渐减少,当电解液接近下限时应及时补水,加液时不要超过上限。

6、当仪器使用半年后更换电解液。

氮气发生器使用氢氧化钾的浓度为15%左右。

7、氮气发生器切勿在未接空气源时空载运行,否则会造成整个仪器报废。

8、用户不要自行将电解池拆卸打开(用户无法自行修理),以免影响整机使用。

9、仪器如需搬运时,把储液桶中的电解液用吸液管吸干净,然后盖好上盖,以免在运输中残留的电解液外溢,将整个仪器腐蚀,造成无法修复的后果。

10、如仪器停机一个月或一个月以上时间,请把电解液抽出。