CERMAX ISIssue 11, March 2003 Manual No.97046 43Pages1.1 (4)1.2 (5)1.3 ATEX (5)2.CERMAX IS (7)2.1 (7)2.2 (7)2.2.1 (7)2.2.2 (7)2.3 (8)2.4 (8)2.4.1 (8)2.4.2 (9)3. CERMAX IS (10)3.1 (10)3.2 (11)3.3 (11)3.3.1 (12)3.4 (12)3.4.1 (12)3.4.2 DEW POINT( ) (13)3.4.3 PPM(W)( ) ( ) (14)3.4.4 PPM(V) 1b/mmscf g/m3 %RH (14)3.5 (14)3.5.1 (15)3.5.2 (15)3.5.3 (15)3.5.4 (16)3.5.5 (16)3.5.6 (17)3.5.7 (18)3.6 (19)3.6.1 (Test No.) (20)3.6.2 (20)3.6.3 (20)3.6.4 (21)3.6.5 (21)3.6.6 (21)3.6.7 (21)3.6.8 (22)3.6.9 (22)3.6.10 (23)3.6.11 (23)3.6.12 (23)3.7 (24)3.7.1 (25)3.7.2 (26)3.8 (27)3.9 (29)4.1 (31)4.2 (31)4.3 (32)4.4 (32)5. (33)5.1 (33)5.2 CERMAX IS (33)5.2.1 (34)5.2.2 (34)5.3 (35)6. (38)1 (39)2 MICHELL (40)3 CERMAX IS (42)1.CERMAX IS MICHELL LCD -100 20 -120o C~+30o CCERMAX IS CERMAX IS CERMAX IS CERMAX IS DISPLAY 3.6CERMAX IS CERMAX IS ON ( 4.3 )1.1(NPL) UKAS -75 ~+20ºC( 0179) (NIST)MICHELL MICHELL MICHELL 5.01.2BS EN ISO 9001:1994 Baseefa (2001)B.S.I.&Baseefa(2001)1.3 ATEXCERMAX IS ATEX (94/9/EC) Baseefa (2001) Ltd (Notified body 1180) BS EN 50 014T bs en 50 020 C.E.N.E.L.E.C. Ex II 1G EEx ia IIC 155o C (T3) Baseefa 03ATEX0090XCERMAX IS ATEX (94/9/EC) ATEX worker protection Directive (1999/92/EC)( )(Flat Battery)CERMAX IS(Logging memory full or incorrect instrument setup.) CERMAX IS CERMAX IS/ (Sensor over/under range or inaccurate sensor measurement)1)2)3)4.2.CERMAX IS2.1CERMAX IS2.2CERMAX IS2.2.1CERMAX IS 4.2CERMAX IS SWAGELOK 1/8” 0.5~5Nl/min 30MPa(300barg) 90o2.2.2on/off (Default) (Previous)3. CERMAX IS2.3CERMAX IS field mode laboratory mode 3.7 3.7.10 015 2 2 -80CERMAX IS 2LCD2.4CERMAX IS2.4.1CERMAX IS 4CERMAX ISCERMAX ISCERMAX IS CERMAX IS2.4.2CERMAX IS RS232 3.6 3.7 3 CERMAX IS 9 D3. CERMAX IS3.110 USER SETUP DEFAULT( ) PREVIOUS( )DEW POINT,OVER ICENONEo C o FpsigDEFAULT/PREVIOUSCERMAX ISSENSOR ID C123 C45 CALIBRATED 14:04:99SOFTWARE VERSION B1.80.00 I.S.COPYRIGHT MICHELL INSTRUMENTS LTD.1aUSER SETUP: DEFAULT PREVIOUSBATTERY CHECK OK1b3.2(MAIN DISPLAY) (AUXILIARY DISPLAY)2Stable DP @ Pressure 10.2ºC10.2 °C DPWARNING - LOW BATTERY33.7DATE 14:04:97 TIME 14:25NO LOGGED DATA AVAILABLE43.33CERMAX ISMain Display Main option3.3.1MENU3.4.DP PPM(V)PPM(W)g/m 3Ib/mmscf %RH 3.4.1MAIN DISPLAY AUX DISPLAY LOGGING DP °C °F K OVER WATER PPM(V) PPM(W)5aMAIN DISPLAY AUX DISPLAY LOGGINGg/m 3lb/mmscf %RHMAIN DISPLAY AUX DISPLAY LOGGING DP °C °F K OVER WATER PPM(V) PPM(W)63.4.2 DEW POINT( )o C o F KMAIN DISPLAY AUX DISPLAY LOGGING DP °C °F K OVER WATER PPM(V) PPM(W)7 o F0o C OVER WATER0o C OVER WATERMAIN DISPLAY AUX DISPLAY LOGGING DP °C °F K OVER WATER PPM(V) PPM(W) 83.4.3 PPM(W)( ) ( )PPM(W)MAIN DISPLAY AUX DISPLAY LOGGING DP PPM(V)PPM(W) N 2 AIR CO 2 H 2 SF 69 PPM(W) 3.4.4 PPM(V) 1b/mmscf g/m 3 %RH PPM(V) 1b/mmscf g/m 3 %rh3.5%rh%rhAuxiliary OptionsMain Option NoneSet Temp Set Press SensorDisplay DPxPPM(V) xPPM(W) xg/m 3x x x lb/mmscf x xx %RHx xxNone. ( ) NONE( ) 21oCSET TEMP CERMAX IS %rhSET PRESS ( ) Set PressSENSOR. DISPLAY.AUX DISPLAY LOGGING MAIN DISPLAY NoneSet Temp Set Press10aAUX DISPLAY LOGGING MAIN DISPLAY Set Press Sensor Display10b ( ) 3.5.1AUX DISPLAY LOGGING MAIN DISPLAY NoneSet Temp °C °F K Set Set Press11 3.5.2%rh (Set temp)(Set Temp) (Set)AUX DISPLAY LOGGING MAIN DISPLAYNoneSet Temp °C °F K Set Set Press12 F 3.5.3(Set)Maximum allowable values Minimum allowable values +200°C -100°C +392°F -148°F 473K +173KAUX DISPLAY LOGGING MAIN DISPLAY NoneSet Temp°C °F K Set -25 Set Press13 -25 3.5.4%rh Set Press PSIG BARG barkPa kPa kPa(Set Press) (Set)AUX DISPLAY LOGGING MAIN DISPLAY NoneSet TempSet Press PSIG BARG kPa Set 100014 BARG 3.5.5Maximumallowable values Minimum allowable values 5880 PSIG 0 PSIG 400 BARG 0 BARG 40,000 kPa 50 kPaAUX DISPLAY LOGGING MAIN DISPLAY NoneSet TempSet Press PSIG BARG kPa Set 3817 381b/mmscf g/m 3 CERMAX IS 1b/mmscf CERMAX IS 1958 kPa bar0 50psig 10 psig 75~300 psig 25 psig 350 500 psig 50 psig 600 1000 psig 100 psig 1200 1500 psig300 psigCEREMAX 13psig bar kPa 10psig CERMAX IS 540psig 500psig 3.5.6( 3.5.4) AUXILIARY DISPLAY( ) (Sensor)%RHg/m 3 lb/mmscf AT PRESSAUX DISPLAY LOGGING MAIN DISPLAY Set Temp Set PressSensor At Press At Atm16 3.5.7(Auxiliary Display)(Sensor)%RHg/m 3 lb/mmscf PRESSAUX DISPLAY LOGGING MAIN DISPLAY Set Press SensorDisplay Press DP17DP @ Pressure 36.5°C10.2°C DPFIELD MODE CAL DUE 04:9818Set Pressure 550 PSIG 10.2°C DPFIELD MODE CAL DUE 04:98233.6CERMAX IS 1000 10000 PC 3.7(LOGGING)Test No. Interval SamplesStart/StopLOGGING MAIN DISPLAY AUX DISPLAY Test No. TextInterval.20aLOGGING MAIN DISPLAY AUX DISPLAYIntervalSamplesStart/Stop20b ( )3.6.1 (Test No.)0 TEST NO 0 Test No.3.6.215 /-TEXT3LOGGING MAIN DISPLAY AUX DISPLAYTest No.Text QUALITY TESTS 3THIS IS MY TESTInterval.213.6.3LOGGING MAIN DISPLAY AUX DISPLAYTest No.TextInterval 1 2 3 5 10 15 30 45 60 mins223.6.4999 10000LOGGING MAIN DISPLAY AUX DISPLAY Test No. TextInterval Samples 2523 3.6.53.6.6START( )LOGGING MAIN DISPLAY AUX DISPLAY Interval SamplesStart/Stop Stop Start Delay Start 00:0024 3.6.7(Delay Start) (24 ) 8:00am 2 30 10:30 (Delay Start) 9 17:00(5:00pm) (Delay Start)Start/Stop(Delay Start)15LOGGING MAIN DISPLAY AUX DISPLAY Interval SamplesStart/Stop Stop Start Delay Start 14:3025 14:30(2:30pm) 3.6.8CERMAX ISStart/Stop STOPLOGGING MAIN DISPLAY AUX DISPLAY SamplesStart/Stop Stop Start Delay Start 14:3026 3.6.9/ ( )Date 14:04:97 Time 14:48NO LOGGED DATA AVAILABLE273.6.10( 3.7 PC )’None’ / (Stable/Responding) 0= (Stable) 1= (Responding) 3.6.114Enter Sample No:5 14:04:97 17% No Time Frostpoint°C PSIG 75 16:23 -15.38 0.0 76 16.24 -15.40 0.029 5 3.6.12Memory gaugeLOGGING MAIN DISPLAY AUX DISPLAYTest No. 5 TextInterval30Test Pt:5 QUALITY TEST 1 14:04:97 9% No Time Frostpoint°C PSIG 0 16:23 -23.39 0.0 1 16.24 -24.400.0313.7CERMAX IS RS232 9600baud 8data bits 1 start bits 1 stop bit no parity IDCERMAX IS COPYRIGHT MICHELL INSTRUMENT LTD. CERMAX IS UNIT ID FREDSERIAL NO.C123C45SOFTWARE VERSION R190.00HELPCLCK? READ THE REAL TIME CLOCK DATE=xx:xx:xx SET DATE TIME=xx:xx:xx SET TIME LANG=US US SETUP LANG=UK UK SETUPKDLY=xx SET KEY REPEAT DELAY IN 10mS UNITS (DEFAULT 100mS) KRPT=xx SET KEY REPEAT RATE IN 10mS UNITS (DEFAULT 100mS) ERAS CLEAR THE LOGGING MEMORY FIELD SPOT CHECK MODE LABUSELABORATORY MODEID= ENTER A MACHINE ID (MAXIMUM 20 CHARACTERS) LOAD DOWNLOAD LOGGED DATA TO TERMINAL 3.7.1CLCK? DATE IS 16:04:97 TIME IS 16:43:15 DATE=17:04:97 DATE IS 17:04:97TIME=10:30:00 TIME IS 10:30:00 LANG=UK or LANG=US LANGUAGE = UK or LANGUAGE = US FIELD UNIT CONFIGURED FOR SPOT CHECKLABUSE UNIT CONFIGURED FOR EXTENDEDMEASUREMENT MODECERMAX IS 10KDLY=10INITIAL KEY REPEAT DELAY SET TO 100 mS100CERMAX IS 10 100 KRPT=10 KEY REPEAT INTERVALSET TO100 mSPCLOAD DOWNLOADING LOGGINGFILES - PLEASE WAIT…DOWNLOAD SUCCESSFULCERMAX IS ERAS CLEARING THE LOGGINGMEMORY WILL TAKE AFEW SECONDS - PLEASEWAIT….LOGGING MEMORYCLEAREDID=FRED Unit ID is FREDLOCK OKUNLOCK OK3.7.2PC COM1 CERMAX IS RS232CERMAX ISWICDOWS 95 HYPER TERMINALHYPER TERMINAL CERMAX ISHYPER TRMCERMAX ISOKConnect COM1OKCOM1Baud 9600Data bits 8Parity NONEStop bits 1Flow control HardwareOKUNIT ID FREDTest Pt:0 MY TEST 15:04:97No Date Time LB/MMSCF PSIG0 15:04:97 14:59 2.30 1000.01 15:04:97 15:00 2.31 1000.0||8 15:04:97 15:02 2.28 1000.09 15:04:97 15:03 2.28 1000.03.8Cermax ISUKASTest HygrometerGenerated Dew point°CDew pointTemperature°CSensorTemperature°CCorrectionRequired°CExpandedUncertainty°C-39.89 -40.11 -20 +0.22 ±0.26-20.10 -20.31 0 +0.21 ±0.220.39 0.20 21 +0.19 ±0.18Generated Dew point°C Instrument Display°C-40.1 -40.2-20.1 -20.10.2 0.1Cermax IS(Calibration Correction)(Calibration Correction)Ref DP( ) ( )Cermax IS DP(Cermax IS ) Cermax IS ) +20o C~-100o C dp 13 Ref DP Cermax IS DP-20o C~+10o C Ref DP Cermax IS DP3.9CERMAX IS1.2.CERMAX IS 2.4.112 141.2. CERMAX IS LOW BATTERY WARNING MICHELL3.4.4.4.14.2CERMAX ISMICHELLCERMAX IS 1/8” PTFE0.5~5L/min4.34.2CERMAX IS 15~90 CERMAX ISON RESPONDINGRESPONDING STABLE( )84.4CERMAX IS MICHELLCERMAX IS IS5.5.1CERMAX IS UKASCermax IS Cermax ISCERMAX ISCERMAX IS CERMAX IS 5.25.2 CERMAX ISCERMAX ISMICHELL CERMAX IS5.2.1CERMAX IS1. RS2322. 8 363.CERMAX IS4. CERMAX IS 3 3 365.6. 1/8” CERMAX IS7. CERMAX IS MICHELL5.2.21. 1/8”2.3. 1/8”4.5. CERMAX IS6.85.3EPROM EPROM EPROM EPROM5.3.1 2.15.3.2 IC145.3.3 EPROM IC145.3.4 85.3.5SETUPMARKET=USLANG=US5.3.6 CERMAX ISMICHELL36 Cermax IS6.MICHELL3161/8” SWAGELOK240 64 LCD4 40 LCD-100o C~+20 o C-120o C~+30o C-60~+20o C, 1o C-100~60o C; 2o C0.1 o C 3o C , o F, Kppm(v)ppm(w)gm-3lb/mmscf0~20mA 4~20mA10,000 RS2326V2412 14250mm 300mm 150mm3kgIP65 NEMA4-20~+50o C-40~+70o C30MPa0.5~5L/min1MICHELL CERMAX IS MICHELL2 MICHELL Michell Instruments LtdNuffield Close, Cambridge, CB4 1SS, UK Tel: +44 (0)1223 434800Fax: +44 (0)1223 434895email: i nfo@web: Michell Instruments GmbH Industriestrasse 27, D-61381 Friedrichsdorf GermanyTel: +49 6172 5917-0Fax: +49 6172 591799email: info@michell.deweb: www.michell.deMichell Instruments SARL4 rue du Docteur Heulin, 75017 Paris FranceTel: +33 (0)1 42 26 01 24Fax: +33 (0)1 42 26 01 37email: i nfo@michell.frweb: www.michell.frMichell Instruments BVKastanjelaan 18M, 2982 CM RidderkerkThe NetherlandsTel:+31 (0)180 466460Fax:+31 (0)180 415623email: info@michell-instruments.nlweb: www.michell-instruments.nlMichell Italia SRLVia Capecelatro 10, 20148 MilanItalyPhone: +39 (0) 024 047194Fax: + 39 (0) 024 *******E-mail: info@michell.itWeb: www.michell.itMichell Instruments Shanghai OfficeUnit A9112, Jia Hua Shang Wu Zhong Xin 808 Hong Qiao Lu, Xu Hui, Shanghai 200030 P R ChinaPhone: +86 (0)21 6448 0152Fax: +86 (0)21 6447 9417E-mail: michellasia@Michell Japan KKMusashino Center Building, 1-19-18 Nakacho, Musashino Tokyo 180- 0006JapanPhone: +81 422 502600Fax: +81 422 521700E-mail: NokeiKWN@web: 3 CERMAX IS1) 1 2L/mina) CERMAX IS b) CERMAX IS,2) Field Mode( ) Lab Mode( ) USER SETUP:DEFAULT FIELD MODE LAB MODE3)RESPONDINGDEW POINT AT ATMOSPHERIC CERMAX IS4) 3 0.4 STABLE 5~105)MICHELL +86 0 21 64479417 64480152 michellasia@。

XPDM露点仪使用规程及注意事项

XPDM露点仪使用规程及注意事项

XPDM露点仪使用规程及注意事项一.本仪器的测量范围为(-120℃~20℃),在使用该仪器的过程中,避免高压气体(仪器测试压力可达120PSI/0.8公斤),避免高温气体(仪器操作温度为-10℃至50℃),避免传感器暴露在空气中超过1-2分钟(在传感器处于测量状态时,不要断开样气使传感器暴露在空气中,测量完成后,最好立即推回干燥存储室,将红色拉杆推回CRY状态)请避免下列气体,否则保修失效1.避免油污污染的气体:避免油污污染的气体。

因为油污可能损坏电抛光表面,使得测量过程减慢或直接损坏仪器。

2.避免腐蚀性气体避免腐蚀性气体,这里指含有氯气,氨气,HCL或SO2的气体。

因为该类气体会损坏传感器。

当湿度含量非常低时,可监测含有SO2的气体。

3.避免含有冷凝水的气体:1)避免测量湿度过大的气体。

因为该气体可能因已经饱和而含有冷凝水2)避免压力过高的气体。

因为结构原因,可能使得压力过高的气体在仪器内部形成冷凝水,因而,此时最好选用压力调节阀选件。

二.测量过程中,切记以下两条:1.测量前,要用所测气体导气管路,处掉管路中的冷凝水和污物。

具体时间根据所测露点值不同而不同。

2.测量后,一定要将红色拉杆推回到原位置(使传感器回到干燥腔室中)。

3.本仪器的测量范围为(-120℃~20℃)。

三.校准过程中(新仪器不需要校准,自己校准只能校准单点,出厂仪器已经进行多点校准,一般更换传感器需要校准,最好送厂家进行校准),切记以下两条:1.切记不要将校准球中的水滴滴入仪器。

2.校准时的环境温度应为21~27℃之间。

操作过程操作前,首先选定合适的导气管路,只能使用不锈钢管,铜管或PTFE管,切忌使用塑料管、橡胶管、乳胶管等等。

然后分析所测气体的污染和压力情况,看是否选用压力调节阀和过滤器。

最后确认气体是排空还是放到某回路中,以免危险。

操作步骤:1.打开测试点阀门反复开关几次,目的是排清阀门处的冷凝水、灰尘或油污。

2.接上导气管和露点仪,先不要急于测试,打开气路阀门数分钟,吹干导气管路,然后保持气体约2.升/分钟左右的流量。

便携式露点仪简易步骤

便携式露点仪简易步骤

便携式露点仪简易步骤
便携式露点仪简易步骤(MDM300)
1、简易步骤
1.1将待测气体阀门开大吹扫3分钟以上
1.2正确连接气路
1.3通过仪表的样气流量控制在0.2~1.2 标准升/分钟,建议
1.4用位于仪表顶部的电源开关开机。

1.5三分钟的初始化程序将自动运行,屏幕上显示倒计时间。

该过程中,传感器被加热,显著增强水分子的流动性从而加速和样气的平衡时间。

状态显示区域将显示内置传感器初始化中(Initializing internal sensor)。

1.6初始化程序完成后,仪表进入测量模式。

在开始的几分钟内,通常会看到显示的湿度值飞快下降,过冲后再回到最终的稳定数值上。

该响应是传感器初始化后的正常现象。

1.7仪表将显示测量值并给出直观指示,说明传感器开始对变化的湿度产生响应(状态栏为测量进行中(Measurement in progress) )以及一段时间后的最终测量读数(状态栏为测量完毕(Measurement OK))。

1.9当仪表的状态栏指示测量完毕(Measurement OK),其显示的读数就表示了样气中的真实湿气水平。

2、注意事项
2.1正确安装气路,防止漏气,影响结果。

2.2流量严格控制在0.2~1.2 标准升/分钟,防止损坏传感器。

2.3及时对仪器充电,否则电用完仪器会自动锁定。

2.5被测样气中不能含有油污或颗粒污染物。

2.6为了保证测量结果准确性,管路只能使用聚四氟乙烯或者不锈钢管和高质量的接头。

氮气专用便携式露点仪安全操作及保养规程

氮气专用便携式露点仪安全操作及保养规程

氮气专用便携式露点仪安全操作及保养规程1. 引言便携式露点仪是一种用于测量氮气中露点温度的应用设备。

为了确保设备的正常运行和使用安全,本文将详细介绍氮气专用便携式露点仪的安全操作和保养规程。

2. 安全操作指南2.1 使用前的准备在使用便携式露点仪之前,务必仔细阅读用户手册,并确保具备以下操作条件:•确保便携式露点仪处于良好的工作状态,无任何损坏或异常;•检查仪器的电源电压和其他电气参数,确保与实际使用环境的要求相匹配;•确保便携式露点仪的相关附件和连接线均齐全并在良好状态下;•了解设备性能和功能,对各部分的使用方法有所了解;•保持工作区域干燥、通风良好,并远离易燃材料。

2.2 正确操作步骤以下是氮气专用便携式露点仪的正确操作步骤:步骤一:接通电源并启动设备。

验证设备的电源电压是否与实际使用环境的电源电压匹配。

步骤二:连接测量设备。

使用合适的连接线将便携式露点仪与被测氮气管路连接。

步骤三:设定测量参数。

根据实际需求,在便携式露点仪的控制面板上设定合适的测量参数,如露点温度的单位、报警阈值等。

步骤四:开始测量。

按下启动按钮,便携式露点仪开始对氮气中的露点温度进行测量。

步骤五:记录测量结果。

根据需要,将测量结果记录在纸质表格或电子文档中,以备后续分析和参考。

步骤六:关闭设备。

测量结束后,及时关闭设备电源,并拔除连接线。

2.3 注意事项在操作便携式露点仪时,请注意以下事项:•使用正确的氮气管路连接方式,确保连接紧密并无泄露;•避免将便携式露点仪置于高温、潮湿、腐蚀性气体等有害环境中;•不要将手指或其他物体触碰测量传感器,以防污染或损坏;•定期校准设备以确保测量结果的准确性和可靠性;•如有异常情况出现,应立即停止使用,并联系相关维修人员进行排除故障。

3. 设备保养规程为了延长便携式露点仪的使用寿命并确保其正常运行,需要进行定期的设备保养。

3.1 定期清洁定期清洁便携式露点仪是保持设备良好状态的重要措施。

清洁步骤如下:1.使用软布蘸取适量的无纺布清洁剂,轻轻擦拭设备表面,确保去除灰尘和污渍;2.不要使用含有酸性或腐蚀性物质的清洁剂,以免对设备造成损害;3.注意保持设备的显示屏、按钮和连接线的清洁,并防止其沾染污物。

便携式露点仪安全操作规定

便携式露点仪安全操作规定

便携式露点仪安全操作规定在使用便携式露点仪进行测试之前,请务必按照以下安全操作规定进行操作。

这些规定将确保您和他人的安全,并防止仪器损坏。

1. 使用正确的电源便携式露点仪通常由电池供电。

请使用同样品牌和型号的电池,以确保设备能够持续稳定地工作。

如果您不确定哪种电池适合您的露点仪,请查看用户手册或者向制造商询问。

2. 避免电击露点仪可能面临电击风险。

在操作设备之前,请确保您的手部干燥,并避免手部接触另一个接地的表面。

要在使用仪器前执行接地最佳实践,例如在操作时使用静电消除器或接地垫。

3. 避免潮湿环境便携式露点仪通常用于测试水汽。

因此,避免在潮湿的环境中使用地址,这样可以保证精度,并延长设备的寿命。

如果您必须在潮湿环境下使用露点仪,请使用适当的保护措施,例如遮阳伞或人造覆盖物。

4. 避免高温、寒冷环境露点仪需要工作在设计温度范围内,以确保测试结果准确可靠。

在高温或寒冷环境下使用设备可能会损坏测试结果的准确性。

请在使用设备之前检查当前实验室或测试环境的温度,并确保仪器设计温度范围内。

5. 避免对设备造成物理损害便携式露点仪通常较为脆弱,必须小心使用。

在使用设备时,必须避免将其摔落或碰撞,并始终将设备放置在稳定的表面上。

如果您必须搬运设备,请遵守设备的操作手册中提供的有关设备搬运和运输的指示。

6. 保证清洁保证设备的清洁非常重要,因为灰尘和污垢可能会影响设备的性能。

用干净的棉布轻轻擦拭露点仪,并使用符合要求的溶液或化学剂进行清洁。

7. 确保良好的卫生条件露点仪常用于医疗、食品、制药和化工行业等领域。

因此,请在测试之前保证良好的卫生条件,在使用前使用适当的消毒剂和工具,消毒测试前需要接触试剂的所有区域。

8. 遵守设备选配规范在操作设备时,请遵守设备选择规范,不得依据个人喜好或偏见任意选择或更改设备。

此外,请严格遵守制造商提供的有关设备使用和保养的指示。

如果您有任何疑问,请联系制造商的客户服务人员或技术支持部门以获取更多信息。

xpdm便携式露点仪分析操作指导书

xpdm便携式露点仪分析操作指导书

XPDM便携式露点仪分析操作指导书----------------------------------------------------------------------------------- 一、引用标准:本操作指导书依据Q/HS LH 1047-2008 气中水含量测定法(露点仪法) 标准和XPDM便携式露点仪分析仪器操作手册制定。

二、操作步骤:1.正常分析(无需校正)步骤1.1 按电源键打开仪器,等1分钟使仪器读数稳定。

1.2 使用和上下键选择所需的工程单位(一般选择 .XXX PPM )。

DEWPOINT PSI1.3 当传感器仍处于干燥室中时,注意仪器显示的读数(湿度应低于要测气体的估计湿度)。

1.4 打开仪器连接口,选择合适的连接管路连接仪器和样气源,打开阀门把样气引进仪器。

等一分钟,使测试室和样气管路干燥。

1.5 堵住仪器出气口。

测量室内的气压足以推动活塞并使传感器进入测量室。

如果压力不够,可用手慢慢拉动位于仪器前部的传感器传动装置(即仪器显示屏右侧的红色旋钮),同时仍堵住样气出口。

1.6 观察仪器读数变化。

如果露点逐步增加,读数将在3分钟内稳定;如果露点逐步下降,将需要更长的稳定周期。

当读数基本上稳定后,读取读数并利用传动装置把传感器推回到干燥室干燥以备下次使用。

2.当管道有压力,需要压力校正,则需按如下步骤进行:2.1打开电源,短按一下 POWER ,出现 -101.1。

℃DEWPOINT2.2短按 PRESSURE 一下,出现 -101.1℃CORRECT DEWPOINT PSI 。

2.3长按 PRESSURE ,出现 SET 14.7 ,此为标准值,一般不改。

CORRECT PSI2.4 短按键一下,交替闪烁出现 SET CRS (14.77) 。

PSI2.5用和上下键调节14.7(原始数值)到所需的压力数值(14.7=101.325KPa)。

(计算公式为 14.7 X ,将管道的101.325 管道的压力值压力值(即X)代入,即可求出X值)。

XPDM中文手册

XPDM中文手册

操作手册XPDM型便携式露点仪Xentaur目录1.介绍 (2)2.警告 (2)3.原理 (3)4.操作XPDM (4)4.1样气连接 (4)4.1.1 连接接口 (4)4.1.2 连接管路 (4)4.1.3采样条件 (4)4.2 测量过程 (4)5.用户接口 (5)5.1显示惯例 (5)5.2状态模式 (5)5.2.1湿度测量状态 (6)5.2.2观察传感器温度 (6)5.2.3自动校准状态模式 (6)5.2.4观察串口数据 (7)5.3修改参数 (7)6.选件 (7)6.1 RS-232C接口 (7)6.2 电流模拟输出 (8)6.3 外接电源操作 (8)7.自动校准 (9)8.更换干燥剂筒及电池 (9)9.特殊信息,报警及错误提示 (10)附录A: 状态模式下用户操作过程的流程图 (11)附录B: 修改参数状态下用户操作接口的流程图 (11)附录C: 选择lbs, ppmv工程单位时仪器读数与4-20mA输出的关系 (11)附录D: 计算LBS与PPMV时的不确定性 (11)附录E: 电池寿命 (11)XPDM是电池操作的便携式露点仪,由微处理器控制,内置用于传感器干燥的分子筛干燥剂。

仪器可用于需要快速测量的所有场合。

标准设备:i带有传感器,干燥剂筒及锂电池的XPDM便携式主仪器。

i用于不锈钢导管安装的2个VCO接口及用于刺软管安装的2个接口i肩带选件:i VCO到VCO安装接口i灵活的不锈钢软管i压力调节装置i凝结过滤器i墙式变送器(带有工厂安装的I/O选件)。

i可替换的干燥剂筒。

i便携式皮箱。

2.注意事项z保修权限: 除了工厂培训的专业人员外,其它人不要拆卸XPDM采样槽/活塞件。

如果采样槽/活塞件被用户拆卸,则取消保修。

z避免污染样气流: XPDM可提供温度低达-100℃露点的快速湿度测量。

因此所有与样气流接触的表面都是由电抛光不锈钢制成。

这使仪器缩短了与气样达到平衡的时间,测量过程因此加快。

便携式露点仪技术支持安全操作及保养规程

便携式露点仪技术支持安全操作及保养规程

便携式露点仪技术支持安全操作及保养规程1. 引言便携式露点仪(Portable Dew Point Meter)是一种常用于测量空气中水蒸气含量的仪器。

它由于其便携性和准确性而在许多领域得到广泛应用,如环境监测、工业制造、气候控制等。

为了确保使用便携式露点仪的操作安全和仪器保养,本文档提供了相关的技术支持和操作规程。

2. 技术支持2.1 准确性校准为了确保便携式露点仪的准确性,定期进行校准是非常重要的。

校准的频率取决于使用环境和要求。

校准可以通过以下步骤完成:•使用标准校准气体进行零点校准。

•使用不同的湿度标准校准气体进行斜率校准。

2.2 电池维护便携式露点仪通常使用电池供电。

为了确保仪器的长时间运行和准确性,以下事项需要注意:•定期检查电池电量,确保电池能够正常工作。

•如果电池电量不足,请及时更换新电池。

•在长时间不使用仪器时,建议将电池取出,以防止电池漏液。

2.3 仪器存放正确存放便携式露点仪可以延长其使用寿命和保持准确性:•避免将仪器存放在高温、潮湿或有尘埃的环境中。

•注意防止仪器受到震动或碰撞。

2.4 其他技术支持•运行时发现任何异常情况或问题时,请立即停止使用,并联系厂商进行更进一步的故障排除。

•遵循仪器的使用说明书和相关安全指南进行操作。

3. 操作规程3.1 前期准备在使用便携式露点仪之前,需要进行一些准备工作:•确保仪器已经校准,并且电池电量充足。

•准备标准校准气体和湿度标准校准气体。

3.2 正确操作按照以下步骤使用便携式露点仪:1.打开仪器的电源开关。

2.等待仪器启动,确保显示屏显示正常。

3.使用仪器附带的橡皮管连接仪器和被测气体源。

4.打开被测气体源,确保气体能够流通到仪器。

5.观察仪器显示屏上的数据,记录露点温度。

3.3 后期处理使用完便携式露点仪后,需要进行一些处理工作,以确保仪器的性能和准确性:•关闭被测气体源和仪器的电源开关。

•将橡皮管从仪器和被测气体源中拆除。

•清洁仪器外壳,防止灰尘或杂质进入仪器内部。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
技术参数: 介绍:XPDM型露点仪是电池供电的便携式仪表,可以在- 100℃~+20 ℃的范围内快速准确地测量气体的露点温度。 应用领域:包括石化,天然气,干燥气和压缩气,发电机冷却氢气,变压器和高压开关绝缘气,焊接气以及船舶和航空用的氧气。 原理与结构:内芯为一高纯铝棒,表面氧化成氧化铝薄膜,其外涂一层多空的金膜,该金膜与内芯之间形成电容,由于氧化铝薄膜的吸水特性,当水蒸汽分子被吸入其中时, 导致电容值发生变化,检测并放大该电容信号即可得到露点温度。 技术指标: 露点传感器单元: 原理:超薄的氧化铝电容原理 量程:- 100℃~+20 ℃ 分辨率:0.1℃ 准确度:±3℃ 重复性:±0.5℃ 响应时间:达到63%时用时90秒 显示:带有背光灯的三位半液晶显示 压力修正:可编程压力修正键 操作温度:传感器:- 30 ~+50 ℃ 电子器件:- 10 ~+50 ℃ 采样流速:100米/秒 节电方式:无操作时电源自动定时关闭 校准方法:自动校准,提供+20 ℃的饱和湿气 机械: 外壳:氧化铝压制而成 尺寸:63.5 x 108 x 190立方毫米 可外接压力调节阀或过滤器。 测试室:电抛光不锈钢材料 工作压力:标准200Kpa 最大690Kpa绝压 气体连接:1/4”VCO输入和输出 隔离:传感器,外壳和电源隔离 电源:9V层叠电池 传感器情况:不使用时,传感器处于干燥室并保持约-80℃的露点。当使用时,传感器进入测试室,使得响应速度加快。
file:///D|/webtopdf/便携式露点仪XPDM露点水分测定仪、露点仪.htm[2010-1-8 22:38:35]
© 2005-2009 必和国际贸易(香港)有限公司 版权所有,并保留所有权利。 上海市长乐路989号2006室,邮编:200031,电话:021-60896520,13661956095,hXPDM露点水分测定仪、露点仪
仪器描述
仪器说明
仪器标签
体积小,重量轻,携带方便 量程范围为- 100℃~+20 ℃ 每次测量时间不超过10分钟 多参数显示
主要特点: ★ 便携式:仪器体积小重量轻,携带方便 ★ 量程广:量程范围为-100℃~+20℃ ★ 响应快:每次测量时间不超过10分钟 ★ 自动校准:无须任何其它设备,可定期对仪器自动校准 ★ 多参数显示:可显示温度,湿度,包括:露点,PPMv,克/升,LBS等 ★ 互换性:传感器可任意互换 ★ 输出信号:可选:4~20mA模拟输出和RS 232接口