3M静电测试仪说明书
静电测试仪使用说明书

Hand-held static sensor locates and meas-ures static voltages, tests air ionizers.New from 3M,the 718 Static Sensor can help companies competing in the global high-tech marketplace prevent cost-ly losses due to electro static discharge (ESD) damage by playing a vital and valuable role in their own ESD control program. Easy to use,the hand-held 3M ™718 Static Sensor is designed to measure static voltages on objects and sur-faces arising from electrostatic charge buildups,and can help identify ESD trouble-spots — ensuring product relia-bility and customer satisfaction which translates into com-pany profits. As a bonus,when used in conjunction with the 3M TM Model 718A Air Ionizer Test Kit,the 718 can also be used to verify the operation of air ionizers.718 Static Sensor Features• Small-size,lightweight,conductive plastic housing • Membrane switches for Power,Range/Zero,and Hold functions.• Digital,LCD (liquid-crystal) display is easy to read and updates quickly.• Ranging system assists user in making quick and easy measurements• Measurements accurate to 5%• Output jack available for continuous measurements Convenient Size/Low Power Requirements The 718 is small enough to be carried in a pocket andweighs less than 5 oz. (142 g),including battery. The light-weight plastic housing is conductive,allowing a properly-grounded user to dissipate all electrostatic charges from the surface of the meter.Meter FunctionsThe meter is equipped with three membrane switches which control different functions. The POWER switch turns the instrument on and off. The RANGE/ZERO button performstwo functions; when pressed momentarily it switchesbetween the two measurement ranges of 0-2,000 volts and 0-20,000 volts,and if held for longer than 3 seconds,it resets the voltage display to 0 volts. The HOLD button allows the user to freeze a measurement on the LCD for later review.Ranging SystemIncluded with the 3M 718 Static Sensor is a ranging system consisting of two light-emitting diodes (LEDs) which each emit a circular red light onto the surface being measured for static. When the two lights intersect and form a single focused light,the measurement distance is the prescribed 1 inch (2.54 cm).AccuracyThe Model 718 Static Sensor is accurate to within ±5% of the displayed measurement,at a distance of one inch (2.54 cm) from the target. Accuracy will vary as the dis-tance between measured object and instrument changes from the one inch (2.54 cm) specification.Analog Output JackThe analog output jack located in the front of the unit pro-vides a convenient hook-up,via a 3/32 inch (2.5 mm)monophone jack,to a recorder/data acquisition console. The 3M 718 Static Sensor may then be used for remote monitoring or permanent recording of electrostatic voltage readings.3M 718 Static Sensor SpecificationsDimensions 0.85" (H) x 2.4" (W) x 4.15" (L)2.2 cm (H) x 6.1 cm (W) x 10.5 cm (L)Weight4.5 oz. (128 g) with battery Power Requirements One 9-volt alkaline battery Measurement Ranges0 – 2 kV Low Range 0 - 20 kV High Range V oltage Display 3) digit liquid crystal display V oltage Output1/1000 of measured voltage @ low range1/10,000 of measured voltage @ high rangeDistance IndicatorLED targets. Aligned targets indicate 1 in. (2.54 cm) measurement distance Measurement Accuracy Within 5% of actual voltage Certifications UL,C-UL,CE,CB-scheme,NOM3718 Static Sensor718A Air Ionizer Test Kit718 Range Finder Unfocused718 Range Finder at 1" away3M 718 Static Sensor123。
防静电专用测试仪使用操作指导书

****有限公司工作指令文件修改记录表编号:SF008 0次修改保存期限:新版发行后1个月****有限公司部门工作指令文件编号 : WI.PNNPI.003页数:共 11 页版本号:第 C 版修改次数:第 0 次修改防静电专用测试仪使用操作指导书制订:审核:批准:生效日期:WI.PNNPI.003目录1. 目的2. 适用范围3. 职责4. 相关文件5 仪器使用指导5.1. 接地阻抗分析仪61-164CN的使用5.1.1设备功能简介、面板使用和其它。
5.1.2.电源插座地线测试步骤5.1.3.辅助地线测试步骤(这里指独立的防静电地线的测试).5.2 静电电压测试仪MODEL 520的使用5.2.1设备功能简介、面板使用和其它。
5.2.2静电电压测试步骤.5.3. 兆欧表MODEL 152的使用5.3.1设备功能简介、面板使用和其它。
5.3.2.表面点对点电阻测试步骤.5.3.3.表面点对地电阻测试步骤.5.3.4.体积电阻测试步骤5.4. 静电消除测试仪MODEL 156A使用。
5.4.1设备功能简介、面板使用和其它。
5.4.2. 静电消除测试仪MODEL 156A测试离子风机参数设置步骤。
5.4.3.静电消除测试仪MODEL 156A测试离子风机静电性能步骤。
5.5. 静电消除测试仪MODEL 158使用。
5.5.1设备功能简介、面板使用和其它。
5.5.2. 静电消除测试仪MODEL 158测试离子风机参数设置步骤。
5.5.3. 静电消除测试仪MODEL 158测试离子风机静电性能步骤。
5.6屏蔽袋测试仪ACL500的使用5.6.1设备功能简介、面板使用和其它5.6.2 . 屏蔽袋屏蔽性能测试步骤。
1. 目的规范接地阻抗分析仪61-164CN、静电电压测试仪MODEL 520、兆欧表MODEL 152、静电消除测试仪MODEL 156A、静电消除测试仪MODEL 1568、屏蔽袋测试仪ACL500等专用仪表的使用,以确保操作人员能正确操作相应测试设备。
3M Clean Trace NGi荧光检测仪系统及基座用户手册说明书

2.01 3M™ Clean‑Trace™ NGi 荧光检测仪的部件 2.02 一般特点 2.03 给电池充电 2.04 打开仪器开关 2.05 测量模式 2.06 插入待检品 2.07 采样棒传感器 2.08 省电模式 2.09 计算机通讯
3.0 配件
3.01 3M™ Clean‑Trace™ NG 荧光检测仪基座 3.02 3M™ Clean‑Trace™ NG 荧光检测仪软质携带箱
1.02 ATP 及表面卫生检测。 3M™ Clean‑Trace™ NGi 荧光检测仪加上配套的检测盒,便成为一种监测各种表面卫生状况的有效方法。 微生物和产品 残留物中的 ATP 都可被检测。 “总 ATP”的测量为总体清洁度提供了重要的借鉴。 表面清洁后存留的各种产品残留物, 将成为存留微生物的营养来源,还可能保护微生物免受消毒剂的作用。 由于使用 3M™ Clean‑Trace™ NGi 荧光检测仪可 迅速获得结果,故若结果不合格时,便可立即采取纠正措施。 重新清洁表面后再行使用,可确保良好的卫生条件及严 格的污染控制。 使用前应仔细阅读检测盒的说明书。
同样的,如果在“程序模式 (Program Mode)”下,在仪器样品室内没有 3M™ Clean‑Trace™ ATP 样品时进行检测,检测传感 器则会通过显示出错信息“未检测到采样棒 (No Swab Detected) ”,同时发出报警声来提醒用户样品室是空的。 而在“检测 样品 (Measure Sample) ”模式下,进行检测时则不会显示此出错信息。
现在便可使用随配的充电器为 3M™ Clean‑Trace™ NGi 荧光检测仪充电。 按图 2 所示将充电器连接器插入 3M™ Clean‑Trace™ NGi 荧光检测仪,再将充电器连接到交流电插座上,打开开关。 另外,也可使用可供选配的基座, 将 3M™ Clean‑Trace™ NGi 荧光检测仪插入到基座上,再将充电器连接器插入基座,然后将充电器连接到交流电插座 上。
静电测试仪的使用操作

1.綠
10 Ω
導体
2.綠
10 Ω
導体
3.綠
10 Ω
導体
4.黃
10 Ω
防靜电体
5.黃
10 Ω
防靜电体
6.黃
10 Ω
防靜电体
7.黃
10 Ω
防靜电体
8.黃
10 Ω
防靜电体
9.黃
10 Ω
防靜电体
10.紅
10 Ω
絕緣体
10
11.紅 (中心)1>210
絕 緣体
1.选择量程Range(LOW /HI).
2.檢查电池(置TEST於BATT位置,读数须大于700V,否则须更換电池.
3.归零(置TEST于ZERO位置).
4.测试:由远及近(大约离物体距10inches处开始)靠近被测物体,当
sensor距物体约1英寸时,此刻读数即为测量值.
6
表面电阻及接地电阻测试仪
用途:
1.量测ESD防护材质的表面阻抗值.
2.量测点到接地点电阻.
3
12
7
操作步驟(量测ESD防护材质的表面电阻)
1.使用专用清洁液清洁测量工作表面或材质. 2.將表面阻抗测试仪放置于被测物表面,按下测试鍵:
1)被测物是导体则LED导体指示灯亮; 2)防静电体则LED防静电体指示灯亮; 3)被测物非导体的防静电导体则绝缘体指示灯发亮.
8
操作步驟(测点到接地点电阻)
1.清除被测工作表面. 2.將引线插入仪器,放置工作表面,引线接地点连接,按下测试
鍵. 3.则对应阻值的指示灯会发亮.
9
表面与量測点到接地点电阻测试仪
規格: 表面电阻系数:1*10 ~1*10 ohms/sq. 点到接地点电阻:1*10 ~1*10 ohms. LED指示粉(由左至右)
SMX-IN-C0063 3M701型静电表面阻抗测试仪操作指引

或WEIGHT需要維護或清洗. 試引線,接住“TEST”5秒,從歐姆刻度上讀取電阻值,用完后將儀表關閉;
文 號版 本日 期 b. 靜電台面質量鑒定
將開關設在要求的“SURFACE TEST”檔,把WEIGHT放在待測台面上,並連上測
5.4 靜電台面阻值參數及測試
5.4.1 一般性說明
靜電臺面的測試主要包括以下三個方面:
管制印章
5.3.3.4 表面測試(靜電阻值測量)
東 莞 新 勁 電 子 有 限 公 司
3M701型靜電表面阻
抗測試儀操作指引0第 4 共 9 頁31/Oct/2006 a. 靜電台面周期性能測試
文件名稱
SMX-IN-C0063 附圖:
c. 靜電台面材料評估
5.4.2 周期性能測試
5.4.2.1 在正常的溫濕度(溫度:15-30℃,濕度:30-70%RH)條件下,靜電台面對防靜電接地的電阻值測量 防靜電接地的測量,確保整個靜電臺面防護系統能正常運作.
(注:防靜電接地在此是指連接靜電台面的接地線或其它接地導體,接地可能是電气接地, 物接地,或其它合適的接地).電阻對地測試是一項重要的測試,可在靜電台面進行,因為它 保靜電台面使用正常及某一時間存在有泄放靜電.這種測量包括從靜電台面,經導體或地PCAR 2001年B 2002年C…月份(10月:A,11月:B,12月:C)流水號。
静电测试仪说明书(参考Word)

静电场测试仪是一非接触式手提静电场测量仪,是测量物体表面静电值和测量静电消除设备的离子平衡度。
外壳为防静电材质,并有接地扣,确保测量结果正确可靠,显示有数字和图形,清晰明确,在暗黑环境下, 更可着亮面板显示灯光。
有离子平衡测量板,测量静电消除设备的残余电压,可测范围+/-200V。
电场测量范围 +/-20Kv 2 LED确定25mm工作距离可配离子平衡测量板(+/-200V)单按归零重置单按显示锁定闲置自动关机电池效能显示功能:有效检测ESD工作场所设备,工作台面,工具的带静电电荷。
特性:液晶/电平显示检测数据,测试数据保持,数字清零显示,高/低档量程。
输出:红色LCD显示正电荷、蓝色LCD显示负电荷测量范围:(0~±1.49kv低范围),正负1kv-20kv(高范围)。
测试精度:±10%电源:9V精度:低范围精度为:正负0.1kv ;高范围精度为:正负1kv输入电压:9V,6F22 manganese battery静电场可测:+/-(0.00-20.0kV)(自动显示调整)标准测距:25mm+/-1mm离子平衡可测:+/-(0-200V)测量距离根据:ESD-STM3.1-2000 lonization 标准LCD显示更新:每秒5次反应时间:1秒工作温湿度:10?40 deg C/RH:0?60%R图形显示:正:红色/ 负:蓝色声响报警:开机/闲置自动关机/超出可测范围自动关机:闲置约5分钟测量模式切换静电场:单按Power掣,有一beep声离子平衡(请先安装离子平衡测量板)同时按Zero和Power,有二beeps声重量/体积0.13kg/11.5cm*6.5cm*2.5cm(连电池)(注:素材和资料部分来自网络,供参考。
请预览后才下载,期待你的好评与关注!)。
M静电测试仪

M静电测试仪 This manuscript was revised on November 28, 2020718静电测试仪使用指南安全须知3M718静电测试仪便携,可手持的设计便于测量静电电荷的电压数。
主要用于测量物体表面产生的0~20千伏静电电压。
任何其他使用行为都可能损坏仪器的能效或造成操作不安全。
3M718空气离子测试装置是与718静电测试仪配套使用装置。
主要通过中和抵消金属盘表面静电时间来检验空气离子发生器。
任何其他使用行为都可能损坏仪器的能效或造成操作不安全。
警告·718静电测试仪和718A充电器使用9V直流碱性电池。
使用任何其他电源都可能损害仪器。
·718静电测试仪和718A空气离子测试装置没有用户服务机构。
无论任何原因切勿自行拆卸仪器。
任何未经授权的行为都将终止保修协议。
·718和718A并不被设计为危险环境中使用,如任何可能有火源存在的环境。
安装与操作本仪器之前请阅读理解所有的安全须知信息1.0说明3M718静电测试仪为手携式仪器,用于定位及测量静电电荷。
他能定位ESD严重区域,是工程人员管控静电的得力助手。
本产品与3M718i空气离子发生器(另外供应)配套使用时,可校验空气离子发生器。
718静电检测仪采用电池供电有多种量测特征:使用范围:可选择测量0~2千伏或0~20千伏范围自动归零:凸起按键的设置方便归零。
无需旋钮或拨号。
HOLD功能:使得用户能够“冻结”屏显的数值,以便之后的评估。
自动关机:保护电池电量,20秒未操作即自动关机。
2.0电量需求及电池的安装2.1718和718A充电器均需要9伏碱性电池供应直流电。
2.2给两个仪器安装电池时:2.2.2卸下电池的外壳,找到内部指示相应的正负方向对准。
2.3718静电检测仪带有低电量指示标志,一旦电池电压低于大约6.5伏的时候,仪器的屏幕就会显示BAT字样。
此时,718不能输出精确的结果,应更换新电池。
静电测试仪操作指南

5.2注意事项
1)测试仪如果经过大约5万次操作后,声音讯号听不到,而绿色“OK”灯仍然闪亮,请更换电池。
2)测试OK后,请记录在“每日员工静电测试记录表”上。
3)每位带静电环的人员必须确保手挽带紧贴自己的手臂。
4)带有绳静电环的人员离开岗位时,必须取下静电绳暗扣,严禁离岗时将其静电绳一同带走。
1.目的:
为确保操作者能够正确使用静电测试仪,特制订本操作指南。
2.范围:
适用于本司操作者对静电测试仪的正确操作方法
3.职责:
静电测试仪操作者应按本指南进行具体操作。
4.相关资料:
4.1 《静电测试仪使用说明书》
4.2 《不合格品处理控制程序》
4.3《标示和可追溯性控制程序》
4.4《供应商监察控制程序》:
操作程序
5.1.操作方法
1)在测试仪后的电池槽内放入两个9V电池。
2)插入香蕉夹以便使测试仪连接到地线夹。
3)戴上静电手环/带,手环/带须紧贴着手腕。
4)把地线接到手环/带上。
5)按下测试仪器正面测试钮,可看到发光二极管显示灯亮和听到声音讯号。
A.若绿色“OK”灯亮,则接地系统安全。
B.若红色“LO”灯亮,需检查手挽带是否在手腕上戴紧,再检查手挽带与地线的导线是否断裂。
5)静电绳上的夹子,必须使其牙纹紧夹静电接地线。
6.质量记录
6.1《每日员工静电测试记录表》
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
3740 Wrist strap +shoe testerOperating Instructions MO 0688-0201February, 2001II. Declaration of ConformityThe 740 Wrist Strap and Shoe Tester is in conformity with the regulations of the EMC-directive 89/336/EWG. The following standards have been taken into account while designing and manufacturing the instrument: EN 60204-1/85,EN 60204-1/91, EN 61010 (Safety),EN 50082-1/92 and EN 50082-2/92.I. GeneralPersonnel grounding devices such as wrist straps and conductive footwear are the primary method to minimise charge generation on the human body. They must be able to drain this charge as rapidly as it is generated.For this reason, it is important to test wrist straps and conductive footwear on a regular basis to ensure they are functioning properly and that the test results can be recordedGround cord 4 mm studContact plate for wrist strap teste.g. - data transfer to PC, electrical door opening system, acoustic signal, etc.selection for wrist strap and shoe testshoe electrode On / Off switchRmax III. DescriptionThe 740 wrist strap and shoe tester is an electronic test instrument which is easy to operate and designed to be wall mounted. An AC/DC -transformer is supplied. The 740 switches off automatically after non-use for approximately 30minutes.The unit consists of two measurement circuits with an open circuit test voltage of 20 V for wrist strap and shoe test applications. For wrist strap and footwear testing, a minimum thresholdresistance of 750 k Ω is set internally. The desired maximum allowable resistance level can be selected separately for each measurement. If the resistance of the tested wrist strap or footwear is within the selected range, the green "o.k."-LED will illuminate. The red "<" -LED alerts theoperator that the resistance of the tested device isunder 750 k Ω. The red ">"-LED indicates that theselected maximum resistance is exceeded.The 740 can be connected to external devices such as computers (for data recording), electrical door opening systems, counter or other evaluating components.The 740 wrist strap and shoe testerconsists of:-3.5" - diskette containing instruction manual and 740 daily log sheet - Base unit-Data output connector plug-Wall mounting kit, 3M Dual Lock ™ fastening system and template-Cover for wrist strap plug-in jack -AC/DC - transformer -Ground cordAccessories:-Shoe electrodes type 741 (single electrode) or 741D (dual electrode) (to be ordered separately).IV. OperationConnect the AC/DC - transformer (DC 24-30 V,150 mA) to the 740 tester.Note: If you notice a delay in the response time of the LED illumination please connect the ground cord supplied to the ground socket of the 740.Please ensure proper earth ground connection!Switch the unit on. All LEDs will flash forapproximately 1 s to check their function. Selectthe desired upper maximum resistance level forboth the wrist strap and shoe test by using a smallscrew driver.The 741/741D shoe electrode must be connected if footwear testing is required.V. Wall MountingThe enclosed 3M Dual Lock™ fastening discsmust be screwed to the wall using the attachedtemplate approximately 1500 mm (5 feet) abovethe floor. Drill three holes of 5mm (0.2 in)diameter at marked locations. Make sure the areaon the 740 in which the Dual Lock™ fasteningstrips are placed is free of dust and dirt. Attach the3M Dual Lock™ fastening strips to the rear of the740 tester to the bottom and top in parallel; seedrawing for locations.For securing on the wall, press the 740 firmlyagainst the discs. For removal of the 740, lift thetop corners with both hands.VI. Wrist Strap TestPut the wrist band on with ground cord attachedand insert the ground cord into the wrist strapplug-in jack. Depress the metal contact plate andhold it. One of the indicator-LEDs will illuminate.An illuminated green LED indicates that the wriststrap performs within the resistance range of750kΩ to the upper selected value of 2, 5, 10 or35 MΩ. If a red "<"-LED is on, the currentlimiting resistor in the ground cord is bypassedand the cord has to be replaced. The red ">"-LEDindicates that the selected resistance range isexceeded. Check to see if the high resistance is inthe cord, the wrist band or in the contact betweenwrist band and the operator’s skin. To check theresistance of the ground cord, leave the plug endof the ground cord attached to the tester anddisconnect the snap end from the wrist band.Connect ground cord snap to the stud in the handsymbol or press it onto the contact plate anddepress the plate until one of the LEDs isilluminated.TemplateIf the green LED illuminatesnow, the ground cord can beused.If the red ">"-LED is lit,replace the ground cord.In some cases, high contact resistance between theskin and wrist band will cause the tester to show ared ">" - LED condition. This resistance may becaused by dry skin or the presence of hair in thewrist area. The use of a skin lotion isrecommended to solve this problem.If a red ">"- LED condition still exists, replace thewrist band.VII. Shoe testEither the 741/741D shoe electrode must be connected to the 740 shoe jack.a) when the 741 single shoe electrode is used:If the red ">"-LED is illuminated, clean the soles of the shoes or check shoe grounding system (heel ground or toe ground strap) and retest. If, after cleaning, the red ">"-LED remains lit, replace it by an appropriate device. If the red "<"-LED is illuminated, the minimum resistance of the shoes or shoe grounding system is under 750 k Ω.An electrical hazard to personnel may exist at the work place.b) when the 741D dual shoe electrode is used:The dual shoe electrode allows differentiation between the right and left shoe; giving the right shoe a preference in indication. This means, if both shoes or just the right one are out of range,you will get identical indications as it would be with the single shoe electrode. If just the left shoe is out of range, the corresponding red LED will flash. Consequently, in case of a steadily lit red LED, you shall clean or remove right shoe first and retest. If the right shoe is in the "ok"-range then, you will obtain information aboutperformance of the left shoe and you can react accordingly.The shoe test does not influence the wrist strap test, therefore the shoes can be tested while the wrist strap is still connected to the 740 or the wrist strap can be tested while the operator is standing on the 741 shoe electrode. Both metal plates shall not be pressed at the same time.Switch the 740 on. Stand on the applied shoe electrode. If you are using shoe straps with a 1M Ω - resistor on both feet to be tested with the 741 shoe electrode, you have to test the straps one after the other, to avoid a red "too low" -indication. Care must be taken not to put the non-tested foot on ESD - protective flooring to avoid a bypass to ground. Depress the metal contact plate for shoe test until one of the indicator - LEDs lights up. The green LED indicates that theresistance of the person through the footwear is in the range between 750 k Ω and the desired upper maximum level (10, 35, 50 or 100 M Ω).VIII. Data Output ConnectorIf you are going to use the data output connector you must follow a test sequence. It must be wrist strap test first and then the shoe test. If this sequence is not strictly kept, the output at the pins described on page 7 will be mixed up .The 740 will provide digital signals (high/low) while testing the personnel grounding devices. These digital signals can be used for data recording (Example 1) and authorisation control equipment such as electrical door opening systems (Example 2). The output of the 740 will give you permanently +5 V at pin 3 and DC-return at pin 6. The remaining pins 1, 2, 4 and 5 (open collector) will be high(max.+30 V/20 mA have to be provided externally) or low (DC-return) while pressing on the contact plates either for wrist strap or shoe tests. The chart on page 7 shows all of the possible test results and the corresponding output levels.Example 1: Output connection for use with a computer Example 2: Output connection for use with a controlling deviceComputer input Computer groundData output connector pin status when testing wrist strap/shoeThe sequence for testing must be wrist strap test first and then the shoe test.Shaded pin indicators will read "hi" if the sequence is reversed or only the shoe tests are performed.Data Output Connectorpin 1, 2, 4, 5-high or low pin 3-+5 V pin 6-DC-returnIX. SpecificationsDimension of base unit :138 mm x 190 mm x 53 mm (5.25 x 7.5 x 2) inches Weight :450 g (15.75 oz)Power supply :external AC/DC - transformer,DC 24-30 V/150 mAAccuracy :± 10% of 2, 5, 10, 35 and 50 M Ω-ranges ± 20% of 100 M Ω-range Output :open collector at pin 1,2, 4, 5(max. +30 V/20 mA)DC-return at pin 6, +5 V at pin 3Measurement voltage :DC (20±1) V (open circuit)750 k ΩᕡᕣᕢX. Calibration ProcedureThe following procedure can be used to determine if the 740 operates within its specifications.Please note: The tester has no adjustable components.Equipment needed:1.Resistors750 k Ω - 120 M Ω, tolerance ± 1%2. 2 leadsas required to connect the reference resistora) calibration of wrist strap test circuit:Connect the reference resistor to the 740 as shown in the figure below. Switch the 740 unit on. Select the 2 M Ω-range and use the reference resistors as indicated in the test table and press contact plate A.b) calibration of shoe test circuits:The 740 Tester can either operate with the single or dual shoe electrode. They are identified by the wiring in the 3-pin plug according to DIN 41 524,type HIRSCHMANN MAS 3100, that connects the shoe electrode to the 740. For the single shoe electrode, the cord is connected to pin 1 while pins 2 and 3 are shunted and grounded.The dual shoe electrode uses the same plug but has got the following connections:pin 1: right shoe electrode pin 2: openpin 3: left shoe electrode.Reference Resistance LED -resistor range settings Indication *750 k Ω 2 M Ω - rangeRed 900 k ΩGreen 1.8 M ΩGreen >2.2 M ΩRed 4.5 M Ω 5 M Ω - range Green >5.5 M ΩRed 9.0 M Ω10 M Ω - range Green >11.0 M ΩRed 31.5 M Ω35 M Ω - rangeGreen >38.5 M ΩRed*This example is used aboveThe LEDs will indicate as shown below if the 740performs within specifications. Repeat this procedure for 5, 10 and 35 M Ω-ranges.11.1 MΩII) calibration of the dual shoe test circuits: The 740 Tester measures the resistance of both shoes almost at the same time. The test result for the circuit of the right shoe has got priority. The calibration can be either performed with two reference resistors for both circuits to be checked at the same time or -corresponding to the single shoe test - with one reference resistor for one circuit to be checked and an appropriate bridge resistor (1MΩ e.g.) for the other circuit. For completion of calibration, the test has to be repeated with connections of the resistors vice versa. Follow the instructions for the single electrode calibration test accordingly.For the operation with two reference resistors proceed as follows: Connect one reference resistor between contact 1 of the socket and the stud and the other one between contacts 2 and 3. Select the 10 MΩ-range and set the reference resistors as indicated in the test table and press contactplate B. The LEDs shall light as shown below. Repeat this procedure for 35, 50 and 100 MΩ-ranges.Right shoe Left shoeelectrode electrodecircuit circuit SignalOK OK green steadily high OK>red steadilylow OK<red steadilyOK high>red flashingOK low<red flashing high low>red steadilylow high<red steadily high high>red steadilylow low<red steadilyI) calibration of the single shoe test circuit: Connect a reference resistor to the 740 between contact 1 of the socket and the stud as shown below. Contacts 2 and 3 of the socket shall be bridged via an appropriate resistor (1MΩ e.g.) Select the 10 MΩ-range and set the reference resistors as indicated in the test table and press contact plate B.The LEDs shall light as shown below. Repeat this procedure for 35, 50 and 100MΩ-ranges.Shoe electrodecircuit SignalOK green steadilyhigh>red steadilylow<red steadily3M 740 Wrist Strap + Shoe Tester Logbook11 / 11MO 0688-0201Distributed by:SJM Eurostat UK Ltd.Countess Avenue Stanley Green Trading Estate Cheadle Hulme, Cheshire SK8 6QS United Kingdom Tel: +44 (0) 161 485 5002Fax: +44 (0) 161 485 46783M France Electronic Handling & Protection Boulevard de l ’Oise 95006 Cergy Pontoise Cedex T él.: +33 (0) 130316809Fax: +33 (0) 130316181SA au capital de 52500000francs RC Pontoise b 542078555, ape 246 C 3Magnab Eurostat Skvasta Industriby 16BN 61192 Nyk öping Sweden Tel: +46 (0) 155 20 26 80Fax: +46 (0) 155 26 98 34Important notice to purchaser "Warranty-3M expressly warrants that for a period of twelve (12) months from the date of purchase,3M Static Control Products will be free of defects in materials (parts) and workmanship (labor).Defects occurring during the warranty period will be repaired or products will be replaced at 3M ’s option and expense, if 3M receives notice during the warranty period. Defective products must be returned to 3M with proof of purchase date.Warranty exclusions: the foregoing express warranty is made in lieu of all other productwarranties, express and implied, including fitness and merchantability. The express warranty will not apply to defects or damage due to accidents,neglect, misuse, alterations, operator error, or failure to properly maintain, clean, or repair products.Limit of liability-in no event will 3M or seller be responsible or liable for special, incidental, orconsequential losses or damages, whether based in tort or contract. Fulfillment of 3M ’s warranty obligations will be customer ’s exclusive remedy and 3M ’s and seller ’s limit of liability for any breach of warrant or otherwise."。