静电测试作业指导书
ESD3000 - 模块化手持检测器 电静电ichtet 测试说明书

Accredited CalibrationQuality at EMC PARTNER is based on an ISO 9001 manage-ment system. This is the foundation for an ISO 17025 accredi-tation verified by the Swiss Calibration Service (SCS). SCS No. 146 is the accreditation number of EMC PARTNER AG. Locally accredited but recognized worldwide through affiliation with the ILAC organisationTHE ESD TESTING SOLUTIONPOWER AT YOUR FINGER TIPSElectrostatic Discharge (ESD) is the most common test procedure applied toany electronic or electrical equipment. Used extensively to test:›Commercial products›Industrial systems›Military electronics›Automotive electronics›Fuses and detonators›Railway electronics›Avionic equipment qualification›Telecom equipment›Electronic componentsThe easily changeable and fully calibrated discharge networks, quickly adaptESD3000 to the requirements of any new application.Light and easy to use.3EASIL Y EXPANDABLEESD3000 is designed with the user in mind. The modular design allows configuration ac -cording to customer wishes, addition of more networks when adding applications and easy extension from 16kV to 30kV using the same generator. Automatic network recognition means ESD3000 is always ready.ESD3000 + DMx ESD3000 + RM32 + DNxIncludedRechargable Long-life battteriesUp to 3 test tips (Sharp, Round, Corona)Charger & Firmware update cableHandy, robust carrying case4Discharge Modules (DMx)up to 16kV Air / 10kV Contact Discharges ESD3000 required Discharge Networks (DNx)up to 30kV Air- and Contact Discharges DM1 (150pF, 330 Ohm)DM2 (330pF / 2000 Ohm)DM4 (100pF / 1500 Ohm)DM6 (100pF / 1500 Ohm)DM7 (200pF / 0 Ohm)ESD3000 & RM32 required DN1 (150pF / 330 Ohm)DN2 (330 pF, 2000 Ohm)DN3 (150 pF, 2000 Ohm)DN4 (500 pF, 5000 Ohm)DN5 (500 pF, 500 Ohm)DN6 (330 pF, 330 Ohm)Available Extensions Available ExtensionsAVAILABLE MODULESCommercial & IndustrialIEC 61000-4-2 / ITU-T K.44ESD3000DM1 (150pF, 330 Ohm)ANSI / IEEE C63.16ESD3000 DM16A-C63H (150pF / 330 Ohm)ESD3000 DM16C-C63H (150pF / 330 Ohm)UL 991ESD3000DM32A-UL991 (100pF / 1000 Ohm)Component TestIEC 61340-3-1 / JEDEC 22-A114 / MIL-STD-750DESD3000DM6 (100pF / 1000 Ohm)IEC 61340-3-2 / JEDEC 22-A115ESD3000DM7 (200pF / 0 Ohm)AutomotivePSA B21 7110ESD3000DM2 (330pF / 2000 Ohm)ISO 10605 / SAEJSSI-IS/FORD AB/AC /GMW 3097ESD3000DN1(150pF / 330 Ohm)ESD3000DN2 (330pF / 2000 Ohm)ESD3000DN3 (150pF / 2000 Ohm)ISO 10605ESD3000DN6 (330pF / 330 Ohm)JASO D 001-94ESD3000DN32-CAR1 (150pF / 500 Ohm)Renault 32-10-001/D and 32-10-035/AESD3000DN32-CAR5 (330pF / 0 Ohm)MilitaryMIL-STD-883 / GR78-COREESD3000DM4 (100pF / 1500 Ohm)STANAG 4239 / ISO 14304 / MIL-STD-1512ESD3000DN4 (500pF / 5000 Ohm)MIL-STD-331C / MIL-DTL-23659D / STANAG 4239ESD3000DN5 (500pF / 500 Ohm)MIL-STD-1576ESD3000DN32-MIL2 (400pF / 150 Ohm)ESD3000DN32-MIL3 (500pF / 0 Ohm)5The original 30kV battery powered generator with a unique system of changeable modules. Flexibility and power combined.UNIQUE FEATURES Long life standard battery. Up to 8 hours on full battery charge. Readily available in any store.AA battery powered Add calibrated modules to maintain the system value. Unique modules and accessories for special applications. Extend testing capability Single key press switches polarity before and during the test.Electronic polarity changeErgonomic hand held tester without additional base unit. Program directly on generator.Compact solution6Technical Specifications7ESD3000 MODULES ESD3000M a i n f r a m e D M 1D M 2D M 4D M 6D M 7R M 32D N 1D N 2D N 3D N 4D N 5D N 6C A R 1C A R 5I N D 1M I L 2M I L 3A -C 63HC -63H U L 991IEC 61000-4-2✓✓✓ITU-T K20✓✓PSA B32 7110✓✓MIL-STD-883✓✓GR78-CORE ✓✓IEC 61340-3-1✓✓JEDEC 22-A114✓✓MIL-STD-750D ✓✓IEC 61340-3-2✓✓JEDEC22-A115✓✓✓ISO 10605✓✓✓✓✓✓DO-160 S25✓✓✓✓MIL-STD-461G CS118✓✓✓✓ GMW3100✓✓✓✓✓✓GMW3097✓✓✓✓STANAG4239✓✓✓✓ISO14304✓✓✓MIL-STD-1512✓✓✓✓MIL-STD-330C ✓✓✓✓JASO D 001-94✓✓✓Renault 32-10-001/D ✓✓ABD0100.1.2✓✓✓Special MIL ✓✓✓MIL-STD-1576✓✓✓ANSI C63.16✓✓✓UL991✓8ACCESSORIES FOR DIFFERENT MODULES ESD3000M a i n f r a me D M 1D M 2D M 4D M 6D M 7R M 32D N 1D N 2D N 3D N 4D N 5D N 6C A R 1C A R 5I ND 1M IL 2M IL3A -C 63HC -63HEARTH CABLE ✓IEC IECESD-VCP50✓IEC IECESD-TARGET2✓IECESD-TARGET2 DN ✓ISO ISO ISO ISOESD-VERI-V ✓✓O O O ✓O O ✓✓O O O ✓✓✓✓ESD-STAND ED2✓O O O O O O O O O O O O O O O O O O O ESD-HCP-AUTO ✓ISO ISODM-EXT ✓MIL MILSAFETY-S ✓MIL MILTC-MIG24 ED ✓MIL MILESD3000 CNH12✓O O O O O OOPTOLINK ✓O O O O O O O O O O O O O O O O O O O TEMA ✓O O O O O O O O O O O O O O O O O O O Overview ESD3000 | Modules | Accessories 91. ESD3000 MAINFRAME AND MODULES1.1. TECHNICAL SPECIFICATIONS ESD3000 MainframeTo be used with modules all modules from this document are compatible Construction fully portable, no additional basic unitControl LCD and 6 menu buttons, 1 trigger button Maximum voltage 16 kV or 30 kV, depending on modules chosen Voltage steps 100 V - testDischarge polarity positive, negative, alternatingDischarge modes contact discharge (CD) and air discharge (AD)Discharge frequency CD max. 20 Hz, (pulse every 0.05, 0.1, 0.2 … 99 s)Discharge frequency AD ≥ 30 Hz or more, for ex. with DM1, DN1, other Counter pre-selectable, 1 to 29999Discharge detection selectable, count pulses or count discharges Holding time ≥ 5 sTrigger manual, automatic, remote Ramps voltage, polarityReporting with TEMA automatic report sequence, amplitude,polarity Dimensions for ex. withDM1 and tip 340 x 130 x 70 mm Weight mainframe 730 g, w/o battery pack and module Weight equipped 1050 g, with battery pack, DM1 and AD tip Power supply 10 x AA rechargeable batteries or mainsIncluded 10 batteries, mains adapter, ground cable 2 m, 3 test tips (AD, AD2, CD), carrying case, RS232 cable, E3Loader for remote control Requires at least a DM module, or a DN module + RM32ESD3000DM1Standards IEC 61000-4-2 (latest), ITU-T K.20,Other standards MIL-STD-461G CS118, DO-160 Section 25Storage capacitor 150 pF ± 10 %Discharge resistor 330 Ω ± 10 %Voltage range AD 0.2 – 16 kV ± 5 %Voltage range CD 0.2 – 10 kV ± 5 %Current rise time into 2 Ω0.8 ± 25 %First I peak into 2 Ω CD 7.5 A @ 2 kV – 37.5 A @ 10 kV (± 15 %)Current at 30 ns 4 A @ 2 kV – 20 A @ 10 kV (± 30 %)Current at 60 ns 2 A @ 2 kV – 10 A @ 10 kV (± 30 %)Optional ESD-TARGET2Overview ESD3000 | Modules | Accessories10ESD3000DM2Standards PSA B32 7110 up to 16 kV / 10 kVStorage capacitor330 pF ± 10 %Discharge resistor2000 Ω ± 10 %Voltage range AD0.2 – 16 kV ± 5 %Voltage range CD0.2 – 10 kV ± 5 %Current rise time into 2 Ω0.7 – 1 nsFirst I peak into 2 Ω CD7.5 – 30 A ± 10 %RC time constant600 ± 130 nsESD3000DM4Standards MIL-STD-883, GR78-COREStorage capacitor100 pF ± 10 %Discharge resistor1500 Ω ± 10 %Voltage range AD0.2 – 16 kV ± 5 %Voltage range CD0.2 – 10 kV ± 5 %Current rise time into 2 Ω<********–8kVI peak into 2 Ω (CD) 2.66 A @ 4 kV, 3.33 A @ 5 kVESD3000DM6Standards IEC 61340-3-1, JEDEC 22-A114,MIL-STD-750DStorage capacitor100 pF ± 10 %Discharge resistor1500 Ω ± 10 %Voltage range CD0.25 – 8 kV ± 10 %Current rise time into 2 Ω2–*********–8kVRC time constant150 ± 20 nsESD3000DM7Standards IEC 61340-3-2, JEDEC 22-A115Storage capacitor200 pF ± 10 %Discharge resistor0 Ω ± 10 %Voltage range CD0.1 – 2 kV ± 5 %Current into short circuit 1. 7 A – 35 A ± 15 % into short circuit (< 1 Ω)Current waveform as per IEC 61340-3-2 into SC (< 1 Ω)Current ringing frequency11 – 16 MHz @ 0.1 – 2 kV into SC (< 1 Ω)ESD3000DM32A-UL991Standard UL991, paragraph 15.2.2Storage capacitor100 pF ± 10 %Discharge resistor1500 Ω ± 5 %Discharge mode AD only, consists of DM+ and DM- modulesVoltage range AD 4 – 30 kV ± 5 %Rise time at 4 kV, 20 kV< 5 nsRC const. at 4 kV, 20 kV150 ns ± 20 %Overview ESD3000 | Modules|Accessories1130 KVThese Extensions need the Module ESD3000RM32ESD3000DN1Standards IEC 61000-4-2,ISO 10605Other standards MIL-STD-461G CS118, DO-160 Section 25Storage capacitor150 pF ± 10 %Discharge resistor330 Ω ± 10 %Voltage range AD 1 – 32 kV ± 5 %Voltage range CD 1 – 30 kV ± 5 %Current rise time into 2 Ω0.8 ± 25 %First I peak into 2 Ω CD7.5 A @ 2 kV – 112.5 A @ 30 kV (± 15 %)Current at 30 ns 4 A @ 2 kV – 60 A @ 30 kV (± 30 %)Current at 60 ns 2 A @ 2 kV – 30 A @ 30 kV (± 30 %)Requires ESD3000RM32Included in ESD3000RM32Detachable module for faster tr (<700ps)Optional ESD-TARGET2 DN, ESD-HCP-AUTOESD3000DN2Standards ISO 10605, GMW3100, GMW3097Storage capacitor330 pF ± 10 %Discharge resistor2000 Ω ± 10 %Voltage range AD 2 – 30 kV ± 5 %Voltage range CD 2 – 30 kV ± 5 %Current rise time into 2 Ω0.7 – 1 nsFirst I peak into 2 Ω CD7.5 A @ 2 kV – 112.5 A @ 30 kV (- 0 / + 30 %) RC time constant600 ± 130 nsRequires ESD3000RM32Optional ESD-TARGET2 DNESD3000DN3Standards ISO 10605, GMW3100, GMW3097Storage capacitor150 pF ± 10 %Discharge resistor2000 Ω ± 10 %Voltage range AD 2 – 30 kV ± 5 %Voltage range CD 2 – 30 kV ± 5 %Current rise time into 2 Ω0.7 – 1 nsFirst I peak into 2 Ω CD7.5 A @ 2 kV – 112.5 A @ 30 kV (- 0 / + 30 %) RC time constant300 ± 60 nsRequires ESD3000RM32Optional ESD-TARGET2 DNOverview ESD3000 | Modules|Accessories12ESD3000DN4Standards STANAG4239, ISO14304MIL-STD-1512, MIL-STD-330Storage capacitor500 pF ± 10 %Discharge resistor5000 Ω ± 10 %Voltage range CD 2 – 30 kV ± 5 %I rise time into 2 Ω CD0.7 – 1 nsRequires ESD3000RM32Optional ESD3000 SAFETY-S, DM-EXT,TC-MIG24EDESD3000DN5Standards STANAG4239, MIL-STD-1512, MIL-STD-330Storage capacitor500 pF ± 10 %Discharge resistor500 Ω ± 10 %Voltage range CD 2 – 30 kV ± 5 %I rise time into 2 Ω CD0.7 – 1 nsRequires ESD3000RM32Optional ESD3000 SAFETY-S, DM-EXT,TC-MIG24EDESD3000DN6Standards ISO 10605, GMW3100Storage capacitor330 pF ± 10 %Discharge resistor330 Ω ± 10 %Voltage range AD 2 – 30 kV ± 5 %Voltage range CD 2 – 30 kV ± 5 %Current rise time into 2 Ω0.7 – 1 nsRC time constant300 ± 130Requires ESD3000RM32Optional ESD-TARGET2 DN, ESD-HCP-AUTOSPECIALESD3000DN32-CAR1Standard JASO D 001-94Storage capacitor150 pF ± 10 %Discharge resistor500 Ω ± 10 %Voltage range AD 2 – 30 kV ± 10 %Voltage range CD 2 – 30 kV ± 10 %I rise time into 2 Ω CD0.8 ± 25 %First I peak into 2 Ω AD< 5 nsRC time constant300 ± 60 nsRequires ESD3000RM32Optional ESD-TARGET2 DNOverview ESD3000 | Modules|Accessories13ESD3000DN32-CAR5Standard Renault 32-10-001/DApplication testing airbag initiatorsStorage capacitor330 pF ± 10 %Discharge resistor0 Ω ± 10 %Source impedance Rt< 30 ΩVoltage range AD 2 – 16 kV ± 5 %Voltage range CD 2 – 16 kV ± 5 %Requires the module is already integrated in RM32 ESD3000DN32-IND1Standard ABD0100.1.2, IEC801-2Storage capacitor150 pF ± 10 %Discharge resistor150 Ω ± 10 %Voltage range AD 2 – 30 kV ± 5 %Voltage range CD 2 – 30 kV ± 5 %I rise time into 2 Ω CD0.7 – 1 nsRC time constant30 ns ± 30 %Requires ESD3000RM32ESD3000DN32-MIL2Standards special military requirementStorage capacitor400 pF ± 10 %Discharge resistor150 Ω ± 10 %Voltage range AD 2 – 30 kV ± 5 %Voltage range CD 2 – 30 kV ± 5 %I rise time into 2 Ω CD0.8 ± 25 %Requires ESD3000RM32ESD3000DN32-MIL3Standard MIL-STD-1576Storage capacitor500 pF ± 10 %Discharge resistor0 Ω ± 10 %Voltage range CD 2 – 30 kV ± 5 %Requires ESD3000RM32ESD3000DM16A-C63HStandard ANSI C63.16 (AD requirement)Storage capacitor150 pF ± 10 %Discharge resistor330 Ω ± 10 %Voltage range AD 2 – 16 kV ± 5 %Requires-Overview ESD3000 | Modules|Accessories14ESD3000DM16C-C63HStandard ANSI C63.16 (CD requirement)Storage capacitor150 pF ± 10 %Discharge resistor330 Ω ± 10 %Voltage range CD 2 – 16 kV ± 5 %Requires-1.2. POWER, CLIMATIC CONDITIONS, SHIPPING WEIGHT, DIMENSIONSESD3000 mainframeMains adapter95 – 250 V (50/60 Hz)Power consumption ON < 20 VA, standby < 5 VATemperature range15 – 35 °CHumidity30 – 60 % non-condensingAir pressure86 – 106 kPaShipping weight 2.4 kg plus selected modulesCarrying case dimensions46 x 41 x 17 cmIncluded in deliveryUser manual with conformity declarationCalibration certificate factory calibration for selected modules2. ACCESSORIES FOR ESD30002.1. TECHNICAL SPECIFICATIONSEARTH CABLEApplication connection of HCP or VCP to ground planeImpedance 2 x 470 kΩLength 2 mConnectors 2 x banana plugsOverview ESD3000 | Modules|Accessories15ESD-VCP50Application indirect ESD application as per standardSpacer in between10 cm wooden spacerCoupling plane50 x 50 cmApplication points one on each sideDimensions50 x 50 x 10 cmWeight8 kgIncluded 2 m earth cable (with 2 x 470 kΩ)ESD-TARGET2Standard IEC 61000-4-2 latestApplication current target for calibration of ESD generator Input impedance 2 ΩInput voltage max. 10 kV CDFrequency range± 0.5 dB up to 1 GHz, ± 1.2 dB up to 4 GHzCurrent range0 – 50 A standard, could be extendedTransfer function0.2 V / 1 A with 20 dBDiameter70 mmThickness40 mmWeight398 g including attenuatorFixing8 x M3 screws, included in deliveryIncluded20 dB att., 50 Ω coax. cable (1 m) with BNC out ESD-TARGET2 DNStandard IEC 61000-4-2 latestApplication current target for calibration of ESD generator Input impedance 2 ΩInput voltage max. 30 kV CDFrequency range± 0.5 dB up to 1 GHz, ± 1.2 dB up to 4 GHzCurrent range0 – 120 A standard, could be extendedTransfer function0.2 V / 1 A with 20 dB, 0.02 V / 1 A with 40 dB Diameter70 mmThickness40 mmWeight398 g including attenuatorFixing8 x M3 screws, included in deliveryIncluded40 dB att., 50 Ω coax. cable (1 m) with BNC out ESD-VERI-VApplication target for ESD DC voltage measurementInput impedance20 GΩ || 3 pFInput voltage range0 – 32 kVOutput voltage range0 – 1.6 VOutput connector BNCDimensions17 cm height, 5.5 cm diameterWeight433 gIncluded earth conductorOverview ESD3000 | Modules|Accessories16ESD-STAND Ed2Application stand for supporting ESD gun, fixed point testHeight50 – 180 cm, adjustablePosition360˚ adjustableDimensions64 x 17 x 12 cm (packed)Weight 4 kgIncluded cable holder for calibrationESD-HCP-AUTOApplication coupling plane as per ISO 10605 annex FLength1500 mmInsulation distance50 mmTest level8 – 20 kV, as per standardDimensions180 x 120 x 5 cmWeight 2.2 kgIncluded EARTH CABLEESD3000DM-EXTApplication 1 m extension cable for ESD3000 modulesCompatibility DM and DN modulesCable length 1 ± 0.05 mWeight0.5 kg including accessoriesCan be used with DN modulesESD3000 SAFETY-SStandard MIL-STD-1512Application safety switch for testing explosive devicesVoltage max. 30 kV CDDimensions20 x 16 x 12 cmWeight 2 kg including accessoriesIncluded mains adapterTo be used with ESD3000DM-EXT, TC-MIG24 EDTC-MIG24 ED for testing explosive devicesApplication test cabinetVoltage insulation max. 36 kVEUT dimensions max. 30 x 30 x 20 cmEUT weight max. 5 kgSafety circuit door interlockSignal lamps red and green lamps built-inControl of signal lamps for example with ESD3000 SAFETY-SDimensions47 x 43.5 x 25.4 cmWeight8 kgIncluded connection cable 25pol/ESD3000 SAFETY-SOverview ESD3000 | Modules|Accessories17ESD3000 CNH12Application Magnetic field loop for ESD3000 test system Loop diameter12 cmLoop current allowed> 100 ADischarge mode selection contact discharge (CD)Current at 15 kV setting50 A, with RM32 and DN1Current at 30 kV setting108 A, with RM32 and DN1Dimensions33 x 13 x 1 cmWeight0.2 kg including accessoriesRequires ESD3000RM32, ESD3000DN1or other DNESD3000-OPTOLINKApplication(insulating) optical cable ESD3000 - computer Length10 mInterface RS232, requires USB adapterBaud rate9600 bpsWeight0.2 kgIncluded tranceiver with mains adapterRequires USB-RS 232 ADAPTERUSB-RS232 ADAPTERApplication adapter between RS232 and USB interfac-esAdapter type Passive, no power supplyInput RS232 maleOutput USBCable length0.35 mWeight0.2 kgTEMAApplication control software for ESD3000 systemLicense 1 license for 1 generatorCapabilities with ESD3000Sequences, reporting, remote controlCompatibility Windows XP, 7, 8, 10Requires ESD3000-OPTOLINK, USB-RS 232 ADAPTEROverview ESD3000 | Modules|Accessories18THE EMC PARTNER PRODUCT RANGE Find further brochures on our website /brochures or contact your local representative for a hardcopy.LIGHTNING TESTSImpulse test equipment and accessories for aircraft, military and tel -ecom applications. Complete solutions for RTCA / DO-160 and EURO -CAE / ED-14 for indirect lighting on aircraft systems, MIL-STD-461 tests CS106, CS115, CS116, CS117, CS118 and Telecom, ITU-T .K44 basic and enhanced tests for impulse, power contact and power induction.COMPONENT TESTSImpulse generators for testing; varistors, gas discharge tubes (GDT),surge protective devices (SPDs), X / Y capacitors, circuit breakers,electricity meters, protection relays, insulation material, suppressor diodes, connectors, chokes, fuses, resistors, emc-gaskets, cables, etc. EMISSION MEASUREMENTSMeasurement of Harmonics and Flicker in 1-phase and 3-phaseelectrical and electronic products according to IEC /EN 61000-3-2 and 61000-3-3 . HARCS Immunity software adds interharmonic tests, voltage variation according to IEC/EN 61000-4-13, -4-14.IMMUNITY TESTSTransient Test Systems for all EMC tests on electronic equipment. ESD, EFT, surge, AC dips, AC magnetic field, surge magnetic field, commonmode, damped oscillatory and DC dips. According to IEC and EN 61000-4-2, -4, -5, -8, -9, -10, -11, -12, -13, -14, -16, -18, -19, -29.SYSTEM AUTOMATIONA full range of accessories enhance the test systems. Test cabinets, test pistols, adapters and remote control software, simplify interfacing with the EUT. Programmable PSU, EMC hardened for frequencies from 16.7Hz to 400Hz. PS3-SOFT-EXT complies with IEC / EN 61000-4-14 and -4-28.SERVICEOur committment starts with a quality management system backing up our ISO 17025 accreditation. With the SCS number 146, EMC PARTNERprovide accredited calibration and repairs. Our customer support teamare at your service!。
静电环测试器校准作业指导书

校准环境: 23℃±5℃;55%20% RH
标准器选用标准:FLUKE5500A多功能校正器
一、校正项目:电阻值。
LOW GOOD HIGH
红灯900KΩ(±10%)绿灯35MΩ(±10%)红灯
二、校正前应准备事项:
1.阅读静电环测试器相关资料;如:使用手册,说明书…等。
2.校正用设备及待校设备必需先在仪校室热机30min以上。
3.上述两值即为量测值,须符合规格。
四、校正后之处理:
1.拆处连接线关闭电源,并将各仪器收归定位。
2.判定该校验仪器是否合格:
3.判定标准:
±10%
3.确认校正环境在合格的状态下。
三、校正方法:
1.将标准件的Grounding(黑)线接至待校件的Grounding端。
2.以标准件的Ω输出端(红棒)按下静电环测试器的push钮,将标准件设定输出39MΩ,观察待校件是否亮HIGH红灯。再向下调整标准件的输出值,直到待校件亮GOOD绿灯,记录此值。继续向下调整输出,直到待校件亮LOW红灯,记录此值。
静电测试仪作业指导书

显示屏右下角显示"Hi"字样,无红光,测试如右图距待测试物25mm测试。
4.1:插头或鳄鱼夹必须与地线相接;
4.2:测试员若接触LED或IC时需佩戴静电环,其它的不用佩戴静电环(如图一不需佩戴静
电环,
4
注意 事項
图4.3二:需除佩静戴电静
效4.4果:确当认静:电除
0~20KV档
值超过200V
作业。
佩戴静电环
SOE
淳安電子股份有限公司 Shun On Electronic
Co.,LTD
生產部
Production Department
静电测试仪作業指導書
文件編号File No.:SOETPD-
版本/Version:1.2
發行日期/Edit Date: 机密等級Class:一般/Normal 頁数Page:1/1
接地鳄鱼夹
检验 每班作业前除尘员对除尘机进行静电效果确认,并将确认值填写在《设备日、周保养卡》上
5
要求
图二测试时可能会接触产品中的LED或IC
5
检验 要求 。
图二测试时可能会接触产品中的LED或IC
資產編號
設備名稱
静电测试仪
工序名稱
所有除静电工序
車間/部門
生产部
作業序号 作業名称
1
作业目 的
1.1:防止产品因静电造成不良;
2
治工具 及物料
2.1:待测静电材料、静电测试仪;
3.1:将鳄鱼夹接地;
作業內容
工藝簡圖(仅供参考)
3.2:打开电脑"POWER"档;
3.3:选择档位尽量先选择0~20KV档位,若测试值很小,无法显示则选测0~200V;
SMT静电测试作业指导书

2.测量:
2.1 将10欧的按钮按下,再按“MEAS”按钮,观察其指针
旋转状况,当指针超过满刻度时,再使用100欧档进行测
量;一般情况下,测量地线静电线只用10欧档,规格为10
欧以下;
2.2 当用10欧档进行测量,发现有大于10欧时有两种可 能:
图3
⑴在车间外打的桩不合格,这种情况较多,只要把桩
的距离和与线的连接调整好以后即可;
制订日期
页码 相关 图片
பைடு நூலகம்
2018/5/1 第1页,共1页
图1
三、接地电阻测量(图3) 1.接线:
相关 图片
将导线绿、黄、红分别接在仪表上的E、P、C三极(分别有
颜色标识),然后把黄、红两条线拉到车间外与大地相连的
两个铁桩上,注意:两个桩之间的距离必须要5~10米之
间,把绿色一条线接在需要测试的地在线;
XX电子科技有限公司
SMT静电测试作业指导书
文件编号 文件版本
XXX-QPAENG012
A/01
一、表面阻抗测量(图1) 1.将表面阻抗仪平放在待测桌面上;
相关 图片
二、静电电压测量(图2) 1.按住仪表上ON按钮3秒钟,松开按钮之后测量仪打开;
2.用手指按在仪表上表面的红色按钮上;
2.再按住ZERO按钮,即ON按钮,直到显示值为零;
3.观察手指按下时指示灯的显示位置,当在106~1010范围 内时,表示被测面的阻抗合格; 4.当被测桌面不在106~1010范围内时,须更换桌面静电皮 。
3.把测量仪侧面发光处正对被测面,然后靠近,拉远观察 光环的晃动状况,当光环不晃动时(晃动很小时),按下 HOLD按钮,此时显示的值即是被测面的静电电压; 4.按住HOLD按钮3秒钟之后,测量仪关闭电源; 5.标准为小于70V。
静电环测试作业指导书

测试人 5.1.6 (注员H意I必G事H须
6.0
项: 当第一次测
6.2.1 试不为合不格合应格更
换防新静的电防手静
6.2.2 环不能套
測試儀 HIGH燈
亮或 LOW燈 亮表示
(LOW灯亮表 示阻值偏低:不合
图七
警告:禁止 将防静电手
注意:当静电手环测试NG时,必须交到拉长处,更换OK的静电手环!
制订: 李付中
测试步
5.0
骤将防
5.1.1 静电手
图一
5.1.2
将测试 仪的夹
图四
注意: 此 處必須 相互夾
5.1.3
用手指 按防压静测电
手环测
图五
测试仪中 绿灯亮表 示测试OK
_
XXX有限公司
静电环测试 作业指导书
编 号 XX-XX-XXX 制作日期
页码
2/2
版次
A/0
如果亮
5.1.5 红直灯到OK
后才能
图六
审核:
批准: _
XXX有限公司
静电环测试 作业指导书
编 号 XX-XX-XXX 制作日期
页码
1/2
版次
A/0
1.0
目的
树立人
员的防危静
2.0
害适,用进范 围适用
3.0
于使本用公工 具:静4.0、责任Fra bibliotek4.1
生产单 位负:责
4.1.1 自负用责静
4.1.2 4.2
生品产管线 单位:
4.2.1 负责各区域 静电防护措施实施的确认及监督。
静电检测作业指导书

1.0目的
1.1.为静电防护外围环境设施检测办法提供一个系统的管制方式,以求完善管理我司静电防护
设备的检测系统
2.0范围
2.1.适用于本公司保护ESD敏感性电子组件及产品所使用的所有相关材料及设备的测试指导
3.0测量方法
3.1.静电接地线的测量:
用万用表一表笔接车间内第一线的ESD接地线,另一表笔接第二线的ESD接地线,依次进
行测试车间内的ESD接地线;参照图如下:
3.2.设备、仪器软、硬接地的测量:
用万用表一表笔接到地线上,另一表笔接到设备、仪器的金属外壳,测试其电阻值,参照下图左所示,接地电阻值:R≤10Ω;
3.3.电烙铁、焊接设备漏电压测量:
用万用表一表笔接触烙铁头,另一表笔接地,量测其电压,其电压范围为:U≤20V
4.0相关文件
4.1ANSI/ESD-S20.20-2009
4.2SJT 10694-2006
4.3GJB3007-1997
编制:日期:批准:日期:
IQC防静电测试作业指导书

IQC防静电测试作业指导书测试项目测试频率接收标准仪器抽样水准防静电手链插孔接地电阻1次/周RTG﹤4Ω万用表100%工作台表面接地电阻1次/周1×106Ω﹤RTG﹤1×109Ω兆欧表每个独立的ESD台面相关金属设备接地电阻1次/月RTG﹤4Ω万用表100%防静电保护地板表面对地电阻1次/月1×105Ω~1×109Ω兆欧表每个独立的ESD地面ESD椅子接地电阻1次/月RTG﹤1×109Ω兆欧表100%ESD工作服点对点及袖对袖电阻1次/月1×105Ω~1×1011Ω兆欧表20%耗散型材料表面电阻率105Ω/sq~1012Ω/sq表面电阻率测试仪100%耗散型材料表面电阻104Ω﹤R﹤1011Ω兆欧表100%防静电手链测试次/天N/A防静电鞋/手链测试仪100%防静电鞋测试次/天N/A防静电鞋/手链测试仪100%防静电手套电阻率105Ω/sq~1011Ω/sq表面电阻率测试仪防静电手套摩擦电压﹤100V静电压测试仪手指套表面电阻104Ω﹤R﹤1011Ω兆欧表手指套摩擦电压﹤100V静电压测试仪1次/月1次/月100%1次/月20%5.3防静电手链插孔接地电阻(RTG)测试:a、将万用表量程拨杆拨至电阻量程MΩ/Ω档;b、将万用表的两表笔跨接于防静电手链插孔和接地点上;c、万用表显示屏上显示稳定后,其读数就是该插孔的对地电阻(RTG)值;d、将显示屏上的读数记录到《防静电手链插座和ESD桌面测试记录》里;e、使用自动报警手腕带接地孔,出现报警时,需要确认ESD手腕带或手腕带接地孔接地是否良好。
若仍然出现报警,则通知相关部门进行维修。
注: 所有手链插孔的要求请依据下图所示:通用ESD手腕带接地孔 ESD自动报警手腕带接地孔5.4ESD桌面对地电阻测试:a、将兆欧表的一极接连至接地点,另一极放在工作台面;b、待读数稳定后读出对地电阻值并记录;5.5静电地板对地电阻测试a、将兆欧表的测试接触器一端放在需测试地板的表面.另一端夹在接地插座上, 将仪器测试档拔到SURFACE TEST-100V 档, 按下TEST 测试键. (如图)b、待读数稳定后读出对地电阻值并记录;c、手指套表面电阻测试如下图:5.11防静电鞋/防静电手带测试:5.11.1防静电鞋/防静电手链两用测试仪说明1、Selective switch/ 转换开关2、OK/NG indicators / 合格与不合格指示灯3、“ Battery” LED indicator / “ 电池” 指示灯4、Metal contact plate(hand touch) / 金属触盘(手触)5、Jack / 插孔6、Metal contact plate(Footwear) / 金属触盘(鞋触)ESD人体综合测试仪5.11.2 防静电手链测试:a、带上防静电手链,将手链另一端插入测试插孔中. 并将转换开关切换到750K-10M;b、按下金属触盘(手触)并压紧。
静电线(ESD)作业指导书

工程名SMT 页版本A1文件名静电线(ESD)作业指导书文件编号.ESD 分布图 管理员进行联络相关技术员维修。
3.5>彩煌三厂公司在亚南电子里静电主线是南亚共用的,标准也大多相同只是SMT特点做相应的修改3.6>静电皮点每月一次,点检测试阻值显示应小于106--1011Ω1. 检测时万用表必须按检测要求进行档位设置,万用表笔必须与检测点进行紧密接触。
2. 必须在〈静电线检测记录表〉内如实填写所检测的数据。
3. 检测时发现异常时必须立即通知上级管理员进行确认。
3. 静电线主线连接到各生产线静电线回路通断检测3.1>将数显多功能万用表检测档位设置于测试电阻阻值的200Ω档位。
3.2>将数显多功能万用表红表笔与静电线主线牵出的测试线接触,黑表笔与静电线主线连接 的支路上所有生产线路静电线回路检测点分别接触。
3.3>确认万用表所检测的两点间测试阻值分别应小于5Ω,反之静电线主线所连接到相应生产线的静电3.4>将万用表检测状况填写于〈静电线检测记录表〉内,当检测阻值大于5Ω时,必须立即通知上级2.1>将数显多功能万用表检测档位设置于测试电阻阻值的2k Ω档位。
2.2>将万用表红表笔与静电线主线《绿线和黄线》分别接触,黑表笔与设备接地线检测点接触。
2.3>确认万用表所检测的两点间测试阻值显示应小于R<10-120Ω,反之静电线主线所连接 到生产线的所有支路静电线与设备接地线部份有短路现象。
2.4>将万用表检测的是否OK填写于〈静电线检测记录表〉内,当检测阻值R<10-120Ω时,必须立即 通知上级管理员进行联络ESD担当进行维修。
1.1>将数显多功能万用表检测档位设置于测试电阻阻值的200Ω档位。
1.2>将万用表红表笔与静电线主线分别接触绿线和黄线,黑表笔与ESD主接地点牵引出的测试线接触。
1.3>确认万用表所检测的两点间《绿线和黄线》测试阻值显示应小于1Ω,反之静电线主线断。
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400 版次 A.0 制定日期 2013-1-7-
静电测试作业指导书
变更页码 N/A
文件页次 第 1 页共 7 页
文件版本 A.0
备注
制定/变更内容摘要 初版发行
制 本文件加盖发行章与 发行日期方可生效 制定部门 工程部
定 制 定 审 核
审
核 批 准
副本文件部门分发明细 品质部 销售部 采购部 物控部 生产部 工程部 人事部 仓 库 计划部 市场部 后勤部 I S O 总经办 其 他
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400 一、目的: 1.1.为静电防护外围环境设施检测办法提供一个系统的管制方式,以求完善管理我司静电防护 设备的检测系统 1.2.定义 ESD 材料或相关设施的测试方法,测试周期,及标准范围,并将测试数据记录于 《OVALE 静电测试记录表》中(编号 OVALEQR-K-46A) 二、 适用范围: 适用于本公司保护 ESD 敏感性电子组件及产品所使用的所有相关材料及设备的测试指导 三、名词解释: 3.1.ESD:Electro Static Discharge,静电释放。
概念中需要特别注意的地方,就是释放,它 是一个动态的过程,是指静电消散的过程。
3.2.EPA:ESD Protect Area,静电防护区域。
四、测量方法: 4.1.静电鞋的测量: 4.1.1 用表面电阻测试仪的两只夹子分别夹于同一只静电鞋的不同位置,测量其表面电阻, 如图所示:
静电测试作业指导书
文件页次 第 2 页共 7 页
文件版本 A.0
静电鞋表面电阻为 3*105Ω≤R≤1*1011Ω
4.2.静电衣﹑帽表面电阻的测量:
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400
静电测试作业指导书
文件页次 第 3 页共 7 页
文件版本 A.0
4.2.1.静电衣表面电阻测试时,用表面电阻测试仪一只夹子夹住静电衣一袖内壁,另一只夹 子夹接至同一袖外壁,按下测试仪,待仪表稳定后,读取测量数值;再分别依法测试两前胸 ﹑一口袋﹑一衣袖;测量如图所示:
静电衣﹑帽表面电阻为:1*105Ω≤R≤1*109Ω 4.3.静电手环的测量: 4.3.1.有线静电环的测量:用表面电阻测试仪的一只夹子夹住静电环的扣的金属扣,另一只 夹子夹住静电环线带的夹子,测量其电阻,如下图左所示:
4.3.2.无线静电环的测量:用表面电阻测试仪的一只夹子夹住静电环的扣的金属扣,另一只 夹子夹住静电环扣的螺丝,测量其电阻,如上图右所示: 静电环电阻范围为:3*105Ω≤Rg≤1*107Ω
4.4.手套、指套表面电阻的测量: 使用表面电阻测试仪,对手套﹑指套的表面电阻进行测量,如下图所示:
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400
静电测试作业指导书
文件页次 第 4 页共 7 页
文件版本 A.0
手套、指套表面电阻为: 1*105Ω≤Rp≤1*109Ω 4.5.静电接地线的测量: 用万用表一表笔接车间内第一线的 ESD 接地线,另一表笔接第二线的 ESD 接地线,依次进 行测试车间内的 ESD 接地线;参照图如下: 第一线 ESD 接地线 万用表笔一
万用表
万用表笔二 第二线 ESD 接地线 静电接地线电阻:R ≤10Ω 4.6.设备、仪器软、硬接地的测量: 4.6.1.硬接地:用万用表一表笔接到地线上,另一表笔接到设备、仪器的金属外壳,测试其 电阻值,参照下图左所示;
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400
静电测试作业指导书
文件页次 第 5 页共 7 页
文件版本 A.0
4.6.2.软接地:用表面电阻测试仪一只夹子夹住设备、仪器的金属外壳,另一夹子夹住设备 仪器下面的静电皮,测量其阻值,如上图右所示。
硬接地电阻值:R ≤10Ω 软接地电阻范围:3*105Ω≤Rs≤1*109Ω 4.7.工作台面静电皮的测量: 4.7.1.工作台面静电皮表面电阻的测量:表面电阻测试仪放置到台面静电皮上按下测试仪, 或用表面电阻测试仪两只夹子分别夹于同一块静电皮不同位置,待仪表稳定后读数;至少测 试 3 个不同位置的值,如下图左所示:
4.7.2.工作台面静电皮系统电阻的测量:表面电阻测试仪一只夹子夹于静电皮上,另一只夹 子夹于静电接地线上,测量其电阻值,如上图右所示。
工作台面静电皮表面电阻为:3*105Ω≤Rp≤1*1010Ω 工作台面静电皮系统电阻为:3*105Ω≤Rs≤1*1010Ω 4.8.周转箱、盒、盘等中间运输容器的表面电阻的测量: 表面电阻的测量方法是将表面电阻测试仪的两只夹子分别夹于器具表面极的不同地方(如 工作台面表面电阻的测试方法),至少选取 3 个测试点,得出两点之间的电阻值。
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400
静电测试作业指导书
文件页次 第 6 页共 7 页
文件版本 A.0
一般要求表面电阻为:1*105Ω≤Rp≤1*109Ω 4.9.传送带的测量: 4.9.1.传送带表面电阻的测量:表面电阻测试仪放置到传送带表面上按下测试仪,或用表面电 阻测试仪两只夹子分别夹于传送带不同位置,待仪表稳定后读数;至少测试 3 个不同位置的 值,如下图左所示:
4.9.2. 传送带系统电阻的测量:表面电阻测试仪一只夹子夹于传送带上,另一只夹子夹于静电 接地线上,测量其电阻值,如上图右所示。
传送带表面电阻为:1*105Ω≤Rp≤1*1011Ω 传送带系统电阻为:1*105Ω≤Rs≤1*1011Ω 4.10.测试夹具静电电压的测量: 用普通的防尘布或擦拭布,或其他不防静电的泡棉之类物件,在测试夹具上摩擦数次,再用 静电测试仪,对准测试夹具,测量其静电电压,如下图所示。
测试夹具静电电压为:U≤200V 4.11.电烙铁、焊接设备漏电压测量:
深圳市欧凡尔科技有限公司
文件编号
OVALE/WI-K-400
静电测试作业指导书
文件页次 第 7 页共 7 页
文件版本 A.0
用万用表一表笔接触烙铁头,另一表笔接地,量测其电压,其电压范围为:U≤20V
4.12.静电刷子、清洁布等辅助工具的静电电压测量: 将辅助工具摩擦后,再用静电测试仪测量其静电电压,如图所示:
辅助工具静电电压范围为:U≤200V 4.13. 能产生静电之材料表面静电压的测试:(试验参考) 先用塑料摩擦一下被测试物体,将静电测试仪对准待测试材料,读取数值,小于 150V 表示 合格,若读取数值无法降至 150V 以下时,表示不合格,需及时做出改善动作,改善 OK 后 再次检测。
一般要求 EPA 内所有材料/设备表面的安全静电压小于或等于+\-150V 五、参考文件 5.1. OVALEQR-K-46A《OVALE 静电测试记录表》
。