美国药典-溶出方法

合集下载

溶出度检查法美国药典USP-711

溶出度检查法美国药典USP-711

<711> DISSOLUTION溶出度(USP39-NF34 Page 540) General chapter Dissolution <711> is being harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. These pharmacopeias have undertaken to not make any unilateral change to this harmonized chapter.通则<711>溶出度与欧盟药典和日本药典中的相应部分相统一。

这三部药典承诺不做单方面的修改。

Portions of the present general chapter text that are national USP text, and therefore not part of the harmonized text, are marked with symbols to specify this fact.本章中的部分文字为本国USP内容,并没有与其他药典统一。

此部分以()标注。

This test is provided to determine compliance with the dissolution requirements where stated in the individual monograph for dosage forms administered orally. In this general chapter, a dosage unit is defined as 1 tablet or 1 capsule or the amount specified. Of the types of apparatus designs described herein, use the one specified in the individual monograph. Where the label states that an article is enteric coated and a dissolution or disintegration test does not specifically state that it is to be applied to delayed-release articles and is included in the individual monograph, the procedure and interpretation given for Delayed-Release Dosage Forms are applied, unless otherwise specified in the individual monograph.本测试用于检测药品口服制剂的溶出度是否符合各论中的规定。

溶出度检查法美国药典USP-711

溶出度检查法美国药典USP-711

溶出度检查法美国药典USP-711<711> DISSOLUTION溶出度(USP39-NF34 Page 540) General chapter Dissolution <711> is being harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. These pharmacopeias have undertaken to not make any unilateral change to this harmonized chapter.通则<711>溶出度与欧盟药典和日本药典中的相应部分相统一。

这三部药典承诺不做单方面的修改。

Portions of the present general chapter text that are national USP text, and therefore not part of the harmonized text, are marked with symbols to specify this fact.本章中的部分文字为本国USP内容,并没有与其他药典统一。

此部分以()标注。

This test is provided to determine compliance with the dissolution requirements where stated in the individual monograph for dosage forms administered orally. In this general chapter, a dosage unit is defined as 1 tablet or 1 capsule or the amount specified. Of the types of apparatus designs described herein, use the one specified in the individual monograph. Where the label states that an article is enteric coated and a dissolution or disintegration test does not specifically state that it is to be applied to delayed-release articles and is included in the individual monograph, the procedure and interpretation given for Delayed-Release Dosage Forms are applied, unless otherwise specified in the individual monograph.本测试用于检测药品口服制剂的溶出度是否符合各论中的规定。

美国药典溶出介质缓冲液的配制

美国药典溶出介质缓冲液的配制

美国药典溶出介质缓冲液的配制
美国药典配制法
1、在标准溶液配制标准指导下配制:0.2mol/L 盐酸溶液;0.2mol/L 氢氧化钠溶液
2、0.2mol/L邻苯二甲酸氢钾溶液:在水中溶解40.85g邻苯二甲酸氢钾,稀释制1000ml
3 、0.2mol/L磷酸二氢钾溶液:在水中溶解27.22g磷酸二氢钾,稀释制1000ml
4 、0.2mol/L硼酸氯化钾溶液:在水中溶解12.37g硼酸和14.91g氯化钾,稀释至1000ml
5、0.2mol/L氯化钾溶液:在水中溶解14.91g氯化钾,稀释至1000ml
6 、2N(当量浓度)的乙酸:在标准溶液配制标准指导下配制
标准缓冲液
盐酸缓冲液:将50ml氯化钾溶液置于200ml容量瓶中,加入一定量的盐酸溶液,用水定容至刻度。

酸性邻苯二甲酸缓冲液:将50ml邻苯二甲酸氢钾溶液置于200ml容量瓶中,加入一定量的盐酸溶液,用水定容至刻度。

中性邻苯二甲酸缓冲液:将50ml邻苯二甲酸氢钾溶液置于200ml容量瓶中,加入一定量的氢氧化钠溶液,用水定容至刻度。

磷酸二氢钾缓冲液:将50ml磷酸二氢钾溶液置于200ml容量瓶中,加入一定量的氢氧化钠溶液,用水定容至刻度。

碱性硼酸缓冲液:将50ml硼酸氯化钾溶液置于200ml容量瓶中,加入一定量的氢氧化钠溶液,用水定容至刻度。

醋酸缓冲液:将规定量的三水乙酸纳置于1000ml容量瓶中,加入规定量的乙酸溶液,用水定容至刻度,混匀。

美国药典溶出度试验方法的建立与验证指导原则的解读

美国药典溶出度试验方法的建立与验证指导原则的解读
的相容性。
小结
➢注重实操,关注细节。 ➢认真验证,数据为优。 ➢逻辑缜密,有理有节。 ➢宽容有度,得心应手。
重点关注
1、溶出的目的 2、造成溶出差异的因素
➢ 样品(是我们想知道,想找出的) ➢ 试验(是不想要的,应降到最低)
3、目测检视 4、方法学验证
谢谢!
溶出度与释放度的方法学验证
准确度
溶液稳 定性
专属性
验证 项目
耐用性
线性范 围
精密度
测定方法与验证-方法学验证
专属性
胶囊
辅料与其他 活性成分
沉降篮
干扰来自何方?如何除去干扰?
测定方法与验证-方法学验证
专属性应注意的问题
• 空白包括:其他活性成分(复方制剂中)、 辅料、包衣、油墨、沉降篮、胶囊壳、装 置(桨、杆、篮)等。
自动取 样
对比验证 内容
试验设计
取样应注意的问题
➢ 手动取样与自动取样要进行比较验证 ➢ 自动取样应注意日常性能检查和维护 ➢ 自动取样的装置要注意对溶出杯中流体力学的
干扰。 ➢ 自动取样的验证还包括:
残留药物的扣减 药物的吸附 洗涤或循环洗涤
试验设计
滤膜吸附及验证
滤过与离心
滤膜吸附的考察
滤膜吸附的验证
美国药典溶出度试验 方法的建立与验证指
导原则的解读
2017.6.30
涵盖内容
▪ 总体评价 ▪ 溶出介质 ▪ 溶出仪器 ▪ 实验设计 ▪ 测定方法与验证
总体评价
限度 范围
区分 力
总体 评价
稳定 性
变异 范围
总体评价-限度
▪ 限度范围应考虑的问题
1、多批次的考量 2、具有代表性 3、具有针对性(针对重点药品) 4、考虑样品的稳定性

[精品]对乙酰氨基酚片溶出度测定

[精品]对乙酰氨基酚片溶出度测定

[精品]对乙酰氨基酚片溶出度测定对乙酰氨基酚片是一种非处方药,用于缓解疼痛、退烧、减轻关节炎和牙痛等。

在药品质量控制方面,溶出度测定是测定药品释放活性成分的关键方法之一,可用于评估药品的有效性和一致性。

本实验旨在采用美国药典(USP)提供的方法,测定对乙酰氨基酚片的溶出度。

1.实验原理溶出度是指药品中活性成分从药片或胶囊中释放出来的比例。

药品的溶出度对药效产生影响,因此,需要对它进行检测以确保药品质量达到标准。

常用的溶出度测定方法包括旋转篮法和流动池法。

旋转篮法适用于药片和胶囊,有利于模拟胃肠道的运动。

而流动池法适用于注射剂和眼药水等液体制剂。

本实验采用的是旋转篮法,将对乙酰氨基酚片放入篮中,放到模拟胃肠道的容器中,并在旋转的条件下采集样品,测定其对乙酰氨基酚溶出度。

2.实验步骤2.1 准备a.先将篮子和箔纸放在烘箱中烘干,预热至50℃左右。

b.称取一定量的对乙酰氨基酚片粉末,将其压制成药片。

c.提取篮子和箔纸,稍稍冷却,然后将篮子置于箔纸上,放回烘箱中预热至50℃。

d.取一定量的模拟胃液,将其预热至37℃。

2.2 实验操作a.取一个已称好药片,将其放入篮子中,然后盖上篮子盖子,确保药片合适地固定在篮子中。

b.用钢丝将篮子悬挂在模拟胃液的容器中,保证篮子不接触容器底部。

c.启动旋转器,设置旋转速度为50rpm,开始测定。

d.每个时间点收集一个样品。

将采样器插入模拟胃液中,确保其到达药片所在的位置(通常为时间的1/3、2/3和3/3),并将样品收集到样品瓶中,以供后续分析。

e.持续测定至90分钟,并在每个时间点记录剩余的药片和溶出度数据。

f.一旦测定完成,将篮子和样品在干燥的环境下过夜。

2.3数据处理a.逐一记录采集的样品的时间和对乙酰氨基酚溶出量。

b.将数据绘制成输药时间与对乙酰氨基酚溶出量的关系图。

c.计算药物的平均溶出度,这是从剩余药物的总质量中计算出来的。

3.实验注意事项a.对乙酰氨基酚片的制备需准确称取药材,不得有误差。

美国药典溶出介质缓冲液的配制

美国药典溶出介质缓冲液的配制
2.0
2.1
2.2
0.2mol/L盐酸溶液(ml)
85.0
67.2
53.2
41.4
32.4
26.0
20.4
16.2
13.0
10.2
7.8
酸性邻苯二甲酸缓冲液
将50ml邻苯二甲酸氢钾溶液置于200ml容量瓶中,加入一定量的盐酸溶液,用水定容至刻度
pH值
2.2
2.4
2.6
2.8
3.0
3.2
3.4
3.6
3.8
4.0
0.2mol/L盐酸溶液(ml)
49.5
42.2
35.4
28.9
22.3
15.7
10.4
6.3
2.9
0.1
中性邻苯二甲酸缓冲液
将50ml邻苯二甲酸氢钾溶液置于200ml容量瓶中,加入一定量的氢氧化钠溶液,用水定容至刻度
pH值
4.2
4.4
4.6
4.8
5.0
5.2
5.4
5.6
5.8
0.2mol/L NaOH溶液(ml)
8.1
11.6
16.4
22.4
29.1
34.7
39.1
42.4
44.5
46.1
碱性硼酸缓冲液
将50ml硼酸氯化钾溶液置于200ml容量瓶中,加入一定量的氢氧化钠溶液,用水定容至刻度
pH值
8.0
8.2
8.4
8.6
8.8
9.0
9.2
9.4
9.6
9.8
10.0
0.2mol/L NaOH溶液(ml)
3.9
6.0
5.30

中_美_英_日四国药典溶出度研究方法比较(1)

中_美_英_日四国药典溶出度研究方法比较(1)

为 900 ml 。小杯法尚未有介质体积的规定 ,一般选
择 100~250 ml 。虽然国际上并不建议使用有机溶
媒 ,中国药典上仍有个别药物 ,如吲达帕胺片以稀乙
四个国家药典中 ,唯有我国药典桨法中存在转 速过高的问题 ,有 4 个药品转速大于 100 r/ min ,国 内文献中一些药品的溶出度试验也存在同样的问 题 ,尤其是儿童和老年人服用的制剂 ,应考虑到其特 殊的生理状况 ,采用低转速 。建议不应为了符合药 典标准而选用不适宜的转速 ,最好从制剂的制备工 艺入手来提高溶出度 。四个国家药典转速选择情况 见表 2 。
试验次数 试验片 (个) 数
判断标准
《美国药典》
试验片 (个) 数
判断标准
每片 (个) ≥Q ;6 片
1
6
(个) 平 均 值 ≥Q , 不小于 Q - 10 % ,
6
每片 ( 个 ) ≥Q + 5%
片 (个) 数 ≤2
对照品 ,可消除样品与对照品组分不同而产生的误 差 ,可减少辅料和杂质的干扰 。目前 ,对于将该法应
本草天工科技有限公司 南昌 330000)
摘要 :文章对中国 、美国 、英国和日本药典中溶出度的研究方法进行了比较和讨论 ,为药物溶出度检查合理的选择测定条件 ,
更好地发挥溶出度在药品质量控制中的作用提供参考 。
关键词 :溶出度 ;药典 ;中国 ;美国 ;英国 ;日本
中图分类号 : R 921 文献标识码 :A
在溶出方法方面 ,各国越来越趋于统一 ,流室法
典》15 版) 中口服固体制剂的溶出度检测方法进行 已与篮法 、桨法一样 ,为日本和一些欧洲国家所接
比较 。
受 。《美国药典》在此方面较为领先和创新 ,不但大

阿齐沙坦片体外溶出方法

阿齐沙坦片体外溶出方法

阿齐沙坦片体外溶出方法阿齐沙坦片是一种常用的降压药物,它属于类似素拮抗剂,通过抑制血管紧张素Ⅱ对血管的收缩作用,从而达到降压的效果。

在临床上,对药物的体外溶出性能进行研究是非常重要的,下面将介绍一种常用的阿齐沙坦片体外溶出方法。

一、研究目的研究阿齐沙坦片的体外溶出性能,确定药物的释放速率和溶出度。

二、实验仪器和试剂1. 仪器:UV/VIS分光光度计、振荡器、离心机。

2. 试剂:阿齐沙坦片标准品、甲醇。

三、实验步骤1. 准备样品溶液:取一定量的阿齐沙坦片标准品,用甲醇溶解并稀释至合适浓度的溶液。

2. 准备模拟体液溶液:按照美国药典(USP)推荐的生理缓冲液pH 6.8来配制模拟体液溶液。

3. 准备试剂:将模拟体液溶液和甲醇按照适当比例混合,得到合适的模拟体液溶液。

4. 液体驱动:将准备好的模拟体液溶液注入振荡器中,并进行液体驱动,设置转速和温度,使其与人体模拟体液相近。

5. 开始测试:将阿齐沙坦片标准品制备的样品溶液加入试剂中,开始测试。

6. 反应时间:根据实验要求,确定每次测试的反应时间。

7. 收集样品:分别在规定的时间点(如0.5小时、1小时、2小时、4小时等)离心收集样品。

8. 检测:用UV/VIS分光光度计测定样品的吸光度,并据此计算阿齐沙坦片的溶出度。

四、结果评价根据实验数据,计算出不同时间点的吸光度值,并转化为溶出度百分数。

根据溶出度与时间的关系,得出阿齐沙坦片的体外溶出曲线。

根据美国药典(USP)提供的溶出度标准,判断阿齐沙坦片的溶出速率是否符合要求。

五、实验注意事项1. 实验过程中,要保持模拟体液溶液的温度和转速恒定。

2. 收集样品时,要注意严格按照规定的时间收集,避免误差的产生。

3. 在测定样品吸光度时,要选择适当的波长,确保准确性。

4. 实验结束后,要及时清洗仪器,并妥善保存实验数据。

六、结论通过上述实验方法,可以明确阿齐沙坦片的体外溶出性能,得出药物的溶出速率和溶出度。

这有助于判断药物的释放特性,从而指导临床用药的安全性和有效性。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

711 DISSOLUTIONThis general chapter is harmonized with the corresponding texts of the European Pharmacopoeia and/or the Japanese Pharmacopoeia. These pharmacopeias have undertaken not to make any unilateral change to this harmonized chapter.Portions of the present general chapter text that are national USP text, and therefore not part of the harmonized text, are marked with symbols () to specify this fact.This test is provided to determine compliance with the dissolution requirements where stated in the individual monograph for dosage forms administered orally. In this general chapter, a dosage unit is defined as 1 tablet or 1 capsule or the amount specified. Of the types of apparatus described herein, use the one specified in the individual monograph. Where the label states that an article is enteric-coated, and where a dissolution or disintegration test that does not specifically state that it is to be applied to delayed-release articles is included in the individual monograph, the procedure and interpretation given for Delayed-Release Dosage Forms is applied unless otherwise specified in the individual monograph. For hard or soft gelatin capsules and gelatin-coated tablets that do not conform to the Dissolution specification, repeat the test as follows. Where water or a medium with a pH of less than 6.8 is specified as the Medium in the individual monograph, the same Medium specified may be used with the addition of purified pepsin that results in an activity of 750,000 Units or less per 1000 mL. For media with a pH of 6.8 or greater, pancreatin can be added to produce not more than 1750 USP Units of protease activity per 1000 mL.USP R EFERENCE S TANDARDS11—USP Chlorpheniramine Maleate Extended-Release Tablets RS. USP Prednisone Tablets RS.APPARATUSApparatus 1 (Basket Apparatus)The assembly consists of the following: a vessel, which may be covered, made of glass or other inert, transparent material1; a motor; a metallic drive shaft; and a cylindrical basket. The vessel is partially immersed in a suitable water bath of any convenient size or heated by a suitable device such as a heating jacket. The water bath or heating device permits holding the temperature inside the vessel at 37 ± 0.5 during the test and keeping the bath fluid in constant, smooth motion. No part of the assembly, including the environment in which the assembly is placed, contributes significant motion, agitation, or vibration beyond that due tothe smoothly rotating stirring element. An apparatus that permits observation of the specimen and stirring element during the test is preferable. The vessel is cylindrical, with ahemispherical bottom and with one of the following dimensions and capacities: for a nominal capacity of 1 L, the height is 160 mm to 210 mm and its inside diameter is 98 mm to 106 mm; for a nominal capacity of 2 L, the height is 280 mm to 300 mm and its inside diameter is 98 mm to 106 mm; and for a nominal capacity of 4 L, the height is 280 mm to 300 mm and its inside diameter is 145 mm to 155 mm. Its sides are flanged at the top. Afitted cover may be used to retard evaporation.2 The shaft is positioned so that its axis is not more than 2 mm at any point from the vertical axis of the vessel and rotates smoothly and without significant wobble that could affect the results. A speed-regulating device is used thatallows the shaft rotation speed to be selected and maintained at the specified rate given in the individual monograph, within ±4%.Shaft and basket components of the stirring element are fabricated of stainless steel, type 316, or other inert material, to the specifications shown in Figure 1. A basket having a gold coating of about 0.0001 inch (2.5 µm) thick may be used. A dosage unit is placed in a dry basket at the beginning of each test. The distance between the inside bottom of the vessel and the bottom of the basket is maintained at 25 ± 2 mm during the test.Figure 1. Basket Stirring ElementApparatus 2 (Paddle Apparatus)Use the assembly from Apparatus 1, except that a paddle formed from a blade and a shaft is used as the stirring element. The shaft is positioned so that its axis is not more than 2 mm from the vertical axis of the vessel at any point and rotates smoothly without significant wobble that could affect the results. The vertical center line of the blade passes through the axis of the shaft so that the bottom of the blade is flush with the bottom of the shaft. The paddle conforms to the specifications shown in Figure 2. The distance of 25 ± 2 mm between the bottom of the blade and the inside bottom of the vessel is maintained during the test. The metallic or suitably inert, rigid blade and shaft comprise a single entity. A suitable two-part detachable design may be used provided the assembly remains firmly engaged during the test. The paddle blade and shaft may be coated with a suitable coating so as to make them inert. The dosage unit is allowed to sink to the bottom of the vessel before rotation of the blade is started. A small, loose piece of nonreactive material, such as not more than a few turns of wire helix, may be attached to dosage units that would otherwise float. An alternative sinker device is shown in Figure 2a. Other validated sinker devices may be used.Figure 2. Paddle Stirring ElementFigure 2a. Alternative sinker. All dimensions are expressed in mm.Apparatus 3 (Reciprocating Cylinder)NOT ACCEPTED BY THE JAPANESE PHARMACOPOEIAThe assembly consists of a set of cylindrical, flat-bottomed glass vessels; a set of glassreciprocating cylinders; inert fittings (stainless steel type 316 or other suitable material), and screens that are made of suitable nonsorbing and nonreactive material and that are designed to fit the tops and bottoms of the reciprocating cylinders; and a motor and drive assembly to reciprocate the cylinders vertically inside the vessels and, if desired, index the reciprocating cylinders horizontally to a different row of vessels. The vessels are partially immersed in a suitable water bath of any convenient size that permits holding the temperature at 37 ± 0.5 during the test. No part of the assembly, including the environment in which the assembly is placed, contributes significant motion, agitation, or vibration beyond that due to the smooth, vertically reciprocating cylinder. A device is used that allows thereciprocation rate to be selected and maintained at the specified dip rate given in the individual monograph within ±5%. An apparatus that permits observation of the specimens and reciprocating cylinders is preferable. The vessels are provided with an evaporation cap that remains in place for the duration of the test. The components conform to the dimensionsshown in Figure 3 unless otherwise specified in the individual monograph.Figure 3. Apparatus 3 (reciprocating cylinder)Apparatus 4 (Flow-Through Cell)The assembly consists of a reservoir and a pump for the Dissolution Medium; a flow-through cell; and a water bath that maintains the Dissolution Medium at 37 ± 0.5. Use the specified cell size as given in the individual monograph.The pump forces the Dissolution Medium upwards through the flow-through cell. The pump has a delivery range between 240 and 960 mL per hour, with standard flow rates of 4, 8, and 16 mL per minute. It must deliver a constant flow (±5% of the nominal flow rate); the flow profile is sinusoidal with a pulsation of 120 ± 10 pulses per minute. A pump without pulsation may also be used. Dissolution test procedures using a flow-through cell must be characterized with respect to rate and any pulsation.The flow-through cell (see Figures 4 and 5), of transparent and inert material, is mounted vertically with a filter system (specified in the individual monograph) that prevents escape of undissolved particles from the top of the cell; standard cell diameters are 12 and 22.6 mm; the bottom cone is usually filled with small glass beads of about 1-mm diameter with one bead of about 5 mm positioned at the apex to protect the fluid entry tube; and a tablet holder (see Figures 4 and 5) is available for positioning of special dosage forms, for example, inlay tablets. The cell is immersed in a water bath, and the temperature is maintained at 37 ± 0.5.Figure 4. Apparatus 4, large cell for tablets and capsules (top), tablet holder for the large cell (bottom). (All measurements are expressed in mm unless noted otherwise.)Figure 5. Apparatus 4, small cell for tablets and capsules (top), tablet holder for the small cell (bottom). (All measurements are expressed in mm unless noted otherwise.)The apparatus uses a clamp mechanism and two O-rings to assemble the cell. The pump is separated from the dissolution unit in order to shield the latter against any vibrations originating from the pump. The position of the pump should not be on a level higher than the reservoir flasks. Tube connections are as short as possible. Use suitably inert tubing, such as polytef, with about 1.6-mm inner diameter and chemically inert flanged-end connections.APPARATUS SUITABILITYThe determination of suitability of a test assembly to perform dissolution testing must include conformance to the dimensions and tolerances of the apparatus as given above. In addition, critical test parameters that have to be monitored periodically during use include volume and temperature of the Dissolution Medium, rotation speed (Apparatus 1 and Apparatus 2), dip rate (Apparatus 3), and flow rate of medium (Apparatus 4).Determine the acceptable performance of the dissolution test assembly periodically.The suitability for the individual apparatus is demonstrated by the Performance Verification Test. Performance Verification Test, Apparatus 1 and 2— Test USP Prednisone Tablets RS according to the operating conditions specified. The apparatus is suitable if the results obtained are within the acceptable range stated in the technical data sheet specific to the lot used and the apparatus tested.Performance Verification Test, Apparatus 3— Test USP Chlorpheniramine Maleate Extended-Release Tablets RS according to the operating conditions specified. The apparatus is suitable if the results obtained are within the acceptable range stated in the technical data sheet specific to the lot used.Performance Verification Test, Apparatus 4— [To come.]PROCEDUREApparatus 1 and Apparatus 2IMMEDIATE-RELEASE DOSAGE FORMSPlace the stated volume of the Dissolution Medium (±1%) in the vessel of the specified apparatus given in the individual monograph, assemble the apparatus, equilibrate the Dissolution Medium to 37 ± 0.5, and remove the thermometer. Place 1 dosage unit in the apparatus, taking care to exclude air bubbles from the surface of the dosage unit, andimmediately operate the apparatus at the specified rate given in the individual monograph.Within the time interval specified, or at each of the times stated, withdraw a specimen from a zone midway between the surface of the Dissolution Medium and the top of the rotating basket or blade, not less than 1 cm from the vessel wall. [NOTE—Where multiple sampling times are specified, replace the aliquots withdrawn for analysis with equal volumes of fresh Dissolution Medium at 37 or, where it can be shown that replacement of the medium is not necessary, correct for the volume change in the calculation. Keep the vessel covered for the duration of the test, and verify the temperature of the mixture under test at suitable times. ]Perform the analysis as directed in the individual monograph using a suitable assaymethod.3 Repeat the test with additional dosage form units.If automated equipment is used for sampling or the apparatus is otherwise modified, verification that the modified apparatus will produce results equivalent to those obtained with the standard apparatus described in this general chapter is necessary.Dissolution Medium— A suitable dissolution medium is used. Use the solvent specified in the individual monograph. The volume specified refers to measurements made between 20and 25. If the Dissolution Medium is a buffered solution, adjust the solution so that its pH is within 0.05 unit of the specified pH given in the individual monograph. [NOTE—Dissolvedgases can cause bubbles to form, which may change the results of the test. If dissolved gases influence the dissolution results, dissolved gases should be removed prior to testing.4 ] Time— Where a single time specification is given, the test may be concluded in a shorter period if the requirement for minimum amount dissolved is met. Specimens are to be withdrawn only at the stated times within a tolerance of ±2%.Procedure for a Pooled Sample for Immediate-Release Dosage Forms—Use this procedure where Procedure for a Pooled Sample is specified in the individual monograph. Proceed as directed in Procedure for Apparatus 1 and Apparatus 2 in Immediate-Release Dosage Forms. Combine equal volumes of the filtered solutions of the six or twelve individual specimens withdrawn, and use the pooled sample as the test specimen. Determine the average amount of the active ingredient dissolved in the pooled sample.EXTENDED-RELEASE DOSAGE FORMSProceed as directed for Immediate-Release Dosage Forms.Dissolution Medium— Proceed as directed for Immediate-Release Dosage Forms.Time— The test-time points, generally three, are expressed in hours.DELAYED-RELEASE DOSAGE FORMS NOT ACCEPTED BY THE JAPANESE PHARMACOPOEIAUse Method A or Method B and the apparatus specified in the individual monograph. All test times stated are to be observed within a tolerance of ±2%, unless otherwise specified. Method A—Procedure (unless otherwise directed in the individual monograph)—ACID STAGE— Place 750 mL of 0.1 N hydrochloric acid in the vessel, and assemble the apparatus. Allow the medium to equilibrate to a temperature of 37 ± 0.5. Place 1 dosage unitin the apparatus, cover the vessel, and operate the apparatus at the specified rate given in the monograph.After 2 hours of operation in 0.1 N hydrochloric acid, withdraw an aliquot of the fluid, and proceed immediately as directed under Buffer Stage.Perform an analysis of the aliquot using a suitable assay method. The procedure is specified in the individual monograph.BUFFER STAGE— [NOTE—Complete the operations of adding the buffer and adjusting the pH within 5 minutes. ]With the apparatus operating at the rate specified in the monograph, add to the fluid in the vessel 250 mL of 0.20 M tribasic sodium phosphate that has been equilibrated to 37 ± 0.5. Adjust, if necessary, with 2 N hydrochloric acid or 2 N sodium hydroxide to a pH of 6.8 ±0.05. Continue to operate the apparatus for 45 minutes, or for the specified time given in the individual monograph. At the end of the time period, withdraw an aliquot of the fluid, and perform the analysis using a suitable assay method. The procedure is specified in the individual monograph. The test may be concluded in a shorter time period than that specified for the Buffer Stage if the requirement for the minimum amount dissolved is met at an earlier time.Method B—Procedure (unless otherwise directed in the individual monograph)—ACID STAGE— Place 1000 mL of 0.1 N hydrochloric acid in the vessel, and assemble theapparatus. Allow the medium to equilibrate to a temperature of 37 ± 0.5. Place 1 dosage unit in the apparatus, cover the vessel, and operate the apparatus at the rate specified in the monograph. After 2 hours of operation in 0.1 N hydrochloric acid, withdraw an aliquot of thefluid, and proceed immediately as directed under Buffer Stage.Perform an analysis of the aliquot using a suitable assay method. The procedure is specified in the individual monograph.BUFFER STAGE— [NOTE—For this stage of the procedure, use buffer that previously has been equilibrated to a temperature of 37 ± 0.5. ] Drain the acid from the vessel, and add to the vessel 1000 mL of pH 6.8 phosphate buffer, prepared by mixing 0.1 N hydrochloric acid with 0.20 M tribasic sodium phosphate (3:1) and adjusting, if necessary, with 2 N hydrochloric acid or 2 N sodium hydroxide to a pH of 6.8 ± 0.05. [NOTE—This may also be accomplished by removing from the apparatus the vessel containing the acid and replacing it with another vessel containing the buffer and transferring the dosage unit to the vessel containing the buffer. ]Continue to operate the apparatus for 45 minutes, or for the specified time given in the individual monograph. At the end of the time period, withdraw an aliquot of the fluid, and perform the analysis using a suitable assay method. The procedure is specified in the individual monograph. The test may be concluded in a shorter time period than that specified for the Buffer Stage if the requirement for minimum amount dissolved is met at an earlier time.Apparatus 3 (Reciprocating Cylinder)NOT ACCEPTED BY THE JAPANESE PHARMACOPOEIA IMMEDIATE-RELEASE DOSAGE FORMS Place the stated volume of the Dissolution Medium in each vessel of the apparatus, assemble the apparatus, equilibrate the Dissolution Medium to 37 ± 0.5, and remove the thermometer. Place 1 dosage-form unit in each of the six reciprocating cylinders, taking care to exclude air bubbles from the surface of each dosage unit, and immediately operate the apparatus asspecified in the individual monograph. During the upward and downward stroke, thereciprocating cylinder moves through a total distance of 9.9 to 10.1 cm. Within the time interval specified, or at each of the times stated, raise the reciprocating cylinders and withdraw a portion of the solution under test from a zone midway between the surface of theDissolution Medium and the bottom of each vessel. Perform the analysis as directed in the individual monograph. If necessary, repeat the test with additional dosage-form units.Dissolution Medium—Proceed as directed for Immediate-Release Dosage Forms under Apparatus 1 and Apparatus 2.Time—Proceed as directed for Immediate-Release Dosage Forms under Apparatus 1 and Apparatus 2.EXTENDED-RELEASE DOSAGE FORMSProceed as directed for Immediate-Release Dosage Forms under Apparatus 3.Dissolution Medium—Proceed as directed for Extended-Release Dosage Forms under Apparatus 1 and Apparatus 2.Time—Proceed as directed for Extended-Release Dosage Forms under Apparatus 1 and Apparatus 2.DELAYED-RELEASE DOSAGE FORMSProceed as described for Delayed-Release Dosage Forms, Method B under Apparatus 1 and Apparatus 2 using one row of vessels for the acid stage media and the following row of vessels for the buffer stage media and using the volume of medium specified (usually 300 mL).Time—Proceed as directed for Immediate-Release Dosage Forms under Apparatus 1 and Apparatus 2.Apparatus 4 (Flow-Through Cell)IMMEDIATE-RELEASE DOSAGE FORMSPlace the glass beads into the cell specified in the monograph. Place 1 dosage unit on top of the beads or, if specified in the monograph, on a wire carrier. Assemble the filter head,and fix the parts together by means of a suitable clamping device. Introduce by the pump the Dissolution Medium warmed to 37 ± 0.5 through the bottom of the cell to obtain the flow rate specified in the individual monograph and measured with an accuracy of 5%. Collect theeluate by fractions at each of the times stated. Perform the analysis as directed in the individual monograph. Repeat the test with additional dosage-form units.Dissolution Medium—Proceed as directed for Immediate-Release Dosage Forms under Apparatus 1 and Apparatus 2.Time—Proceed as directed for Immediate-Release Dosage Forms under Apparatus 1 and Apparatus 2.EXTENDED -RELEASE DOSAGE FORMSProceed as directed for Immediate-Release Dosage Forms under Apparatus 4.Dissolution Medium —Proceed as directed for Immediate-Release Dosage Forms under Apparatus 4.Time —Proceed as directed for Immediate-Release Dosage Forms under Apparatus 4.DELAYED -RELEASE DOSAGE FORMSProceed as directed for Delayed-Release Dosage Forms under Apparatus 1 and Apparatus 2, using the specified media.Time —Proceed as directed for Delayed-Release Dosage Forms under Apparatus 1 and Apparatus 2.INTERPRETATIONImmediate-Release Dosage FormsUnless otherwise specified in the individual monograph , the requirements are met if the quantities of active ingredient dissolved from the dosage units tested conform to Acceptance Table 1. Continue testing through the three stages unless the results conform at either S 1 or S 2. The quantity, Q, is the amount of dissolved active ingredient specified in the individual monograph , expressed as a percentage of the labeled content of the dosage unit; the 5%, 15%, and 25% values in Acceptance Table 1 are percentages of the labeled content so that these values and Q are in the same terms. Acceptance Table 1Immediate-Release Dosage Forms Pooled Sample — Unless otherwise specified in the individual monograph, the requirements are met if the quantities of active ingredient dissolved from the pooled sample conform to the accompanying Acceptance Table for a Pooled Sample.Stage NumberTested Acceptance CriteriaS 16Each unit is not less than Q + 5%.S 26Average of 12 units (S 1 + S 2) is equal to or greater than Q, and no unit isless thanQ 15%.S 312Average of 24 units (S 1 + S 2 +S 3) is equal to or greater than Q , not morethan 2 units are less than Q 15%, and no unit is less than Q 25%.Continue testing through the three stages unless the results conform at either S 1 or S 2. The quantity, Q is the amount of dissolved active ingredient specified in the individual monograph, expressed as a percentage of the labeled content.Acceptance Table for a Pooled SampleExtended-Release Dosage FormsUnless otherwise specified in the individual monograph , the requirements are met if the quantities of active ingredient dissolved from the dosage units tested conform to Acceptance Table 2. Continue testing through the three levels unless the results conform at either L 1 or L 2. Limits on the amounts of active ingredient dissolved are expressed in terms of the percentage of labeled content. The limits embrace each value of Q i , the amount dissolved at each specified fractional dosing interval. Where more than one range is specified in the individual monograph , the acceptance criteria apply individually to each range. Acceptance Table 2Stage Number Tested Acceptance Criteria S 16Average amount dissolved is not less thanQ + 10%.S 26Average amount dissolved (S 1 + S 2) is equal to or greater than Q + 5%.S 312Average amount dissolved (S 1 + S 2 + S 3) is equal to or greater than Q.Level NumberTested CriteriaL 16No individual value lies outside each of the stated ranges and no individualvalue is less than the stated amount at the final test time.L 26The average value of the 12 units (L 1 + L 2) lies within each of the statedranges and is not less than the stated amount at the final test time; none ismore than 10% of labeled content outside each of the stated ranges; andnone is more than 10% of labeled content below the stated amount at thefinal test time.L 312The average value of the 24 units (L 1 + L 2 + L 3) lies within each of the statedranges, and is not less than the stated amount at the final test time; not morethan 2 of the 24 units are more than 10% of labeled content outside each ofthe stated ranges; not more than 2 of the 24 units are more than 10% oflabeled content below the stated amount at the final test time; and none ofthe units is more than 20% of labeled content outside each of the statedranges or more than 20% of labeled content below the stated amount at thefinal test time.Delayed-Release Dosage FormsNOT ACCEPTED BY THE JAPANESE PHARMACOPOEIA .Acid Stage — Unless otherwise specified in the individual monograph , the requirements of this portion of the test are met if the quantities, based on the percentage of the labeledcontent, of active ingredient dissolved from the units tested conform to Acceptance Table 3. Continue testing through all levels unless the results of both acid and buffer stages conform at an earlier level. Acceptance Table 3Buffer Stage — Unless otherwise specified in the individual monograph , the requirements are met if the quantities of active ingredient dissolved from the units tested conform to Acceptance Table 4. Continue testing through the three levels unless the results of both stages conform at an earlier level. The value of Q in Acceptance Table 4 is 75% dissolved unless otherwise specified in the individual monograph . The quantity, Q specified in the individual monograph is the total amount of active ingredient dissolved in both the Acid and Buffer Stages , expressed as a percentage of the labeled content. The 5%, 15%, and 25% values in Acceptance Table 4 are percentages of the labeled content so that these values and Q are in the same terms. Acceptance Table 4Level NumberTested CriteriaA 16No individual value exceeds 10% dissolved.A 26Average of the 12 units (A 1 +A 2) is not more than 10% dissolved, and no individual unit is greater than 25% dissolved.A 312Average of the 24 units (A 1 + A 2 +A 3) is not more than 10% dissolved, andno individual unit is greater than 25% dissolved.Level NumberTested CriteriaB 16Each unit is not less than Q + 5%.B 26Average of 12 units (B 1 + B 2) is equal to or greater than Q, and no unit isless thanQ – 15%.B 312Average of 24 units (B 1 + B 2 + B 3) is equal to or greater than Q, not morethan 2 units are less than Q – 15%, and no unit is less than Q – 25%.1The materials should not sorb, react, or interfere with the specimen being tested.2If a cover is used, it provides sufficient openings to allow ready insertion of the thermometer and withdrawal of specimens.3Test specimens are filtered immediately upon sampling unless filtration is demonstrated to be unnecessary. Use an inert filter that does not cause adsorption of the active ingredient or contain extractable substances that would interfere with the analysis.4One method of deaeration is as follows: Heat the medium, while stirring gently, to about 41, immediately filter under vacuum using a filter having a porosity of 0.45 µm or less, with vigorous stirring, and continue stirring under vacuum for about 5 minutes. Other validated deaeration techniques for removal of dissolved gases may be used.Auxiliary Information—Please check for your question in the FAQs before contacting USP. Array USP34–NF29 Page 278Pharmacopeial Forum: Volume No. 35(3) Page 719。

相关文档
最新文档