The Agilent 8860 gas chromatograph, inheriting legendary Agilent expertise andproven quality in GC, generates reliable results with minimized complexity for routine analyses, run after run, day after day. The 8860 color touchscreen provides:• Access to setpoint and status information with the ability to turn setpoints on/off• Instrument configuration and flowpath•A signal plot, helping confirm that analyses are progressing as intendedThe browser interface is the most extensive interface to the 8860 GC. Optimized for a 10-inch tablet, you can use the browser interface on a tablet, PC, or even a phone. Access to advanced intelligence functions include:• Connectivity to check status or run diagnostics from anywhere within your network• Built-in self-guided diagnostics and maintenance capabilities • Method and sequence editing without the need for a data system •Easy access to logs and complete user documentationA browser interface optimized for mobile viewing on either iOS or Android phones provides status information including remaining run time and a static plot of the last 20 minutes of detector data.Agilent 8860 Gas Chromatograph SystemFull electronic pneumatics control (EPC), available for all inletsand detectors, automates control of gas pressures/flowsto predefined setpoints, and allows pressure and flow programming.Electronic pneumatics regulation (EPR) is available for select inlets and detectors. EPR provides a digital measurement and display of gas pressure/flow, and allows the user to manually adjust these gas pressures/flows, electronically, using the Browser Interface, avoiding the use of mechanical regulators. This provides constant pressure operation for the Split/Splitless inlet, and constant flow operation for the Packed Column inlet and the detectors with EPR. If desired, EPR actual pressure and flow can be saved as additional signals with any OpenLAB CDS software to record its operation for each GC run.The 8860 EPC and EPR are based on Agilent GC core 6th generation microchannel-based EPC architecture. Uniqueto Agilent, this design protects against gas contaminants such as particulates, water, and oils, and provides significant improvements in reliability and longevityover earlier generation GC designs from Agilent or other GC manufacturers. EPC and EPR are compensated for barometric pressure and ambient temperature changes, resulting in more stable retention times and detector baselines.Overall system performance** Using the 8860 GC with 8860 EPC (splitless), ALS, and Agilent Data System for analysis of tridecane (2 ng tothe column). Results may vary with other samples and conditions.• Retention time repeatability <0.06 %• Peak area repeatability <2 %Power requirements• Line voltage – 100/120/200/220/230/240 Volts ±10 % of nominal• Frequency – 50/60 Hz ±5%• 1,500 W at 100 V; 2,250 W at all other voltagesColumn ovenTable 1.Typical 8860 oven ramp rates.EPCAvailable on all inlets and detectorsEPRAvailable on S/SL and Packed Column inlets and FID, TCD, and ECD detectors. EPR allows the user to manually adjust pressure and total flow (for S/SL inlet) or flow only (for Packed Column inlet (PCI) and FID ,TCD, and ECD detectors) to a desired value using the Browser Interface. Makeup flow will not compensate for changes to column flow during oven temperature ramping.Inlets• Up to two inlets may be installed.• EPC pressure setpoint and control precision to 0.01 psi or0.069 kPa• Display resolution for EPR for pressure is 0.01 psi or0.069 kPa2Purged packed (EPC)• Electronic flow control• Septum purge• 400 °C maximum operating temperature• Maximum flow 100 mL/min• Adapters included for 1/8-inch packed columns• Adapters for 0.530-mm capillary columnsPacked column (EPR)• Constant flow operation• 400 °C maximum operating temperature• Maximum flow 100 mL/min• Adapters included for 1/8-inch packed columnsS/SL (EPC)• Electronic pressure/flow control• Septum purge• Turn top inlet sealing system• Includes leak/restriction, pressure decay, and split vent restriction test diagnostics (cap off septum purge and split vent)• Maximum operating temperature 400 °C• Pressure setting range 0–100 psi or 0–689.47 kPa • Maximum split ratio 12,500:1• Flow setting range 0 to 500 mL/min N2,0–1,250 mL/min H2or HeS/SL (EPR)• Constant pressure operation• Septum purge• Turn top inlet sealing system• Maximum operating temperature 400 °C• Pressure adjustable range 0 to 100 psi or 0 to 689.47 kPa • Maximum split ratio 12,500:1• Flow adjustable range 0–500 mL/min N2,0–1,250 mL/min H2 or HePCOC (EPC)• Maximum operating temperature 400 °C.Temperature programming in three ramps or tracking• Oven subambient control is not available• Pressure setting range 0–100 psig or 0–689.47 KPa• Electronic septum purge control• Automatic liquid injection supported directly ontocolumns ≥0.250 mm idDetectors• Up to three detectors may be installed.• A 3rd detector as FPD+ or TCD can be located on top inthe middle position, or a 3rd detector as TCD or ECD canbe located on the side of the GC.• EPC with electronic flow control is available for detectorgases for all detectors.• EPR with constant flow operation is available for detectorgases for FID, TCD, and ECD.Flame ionization detector (FID)• Maximum operating temperature 425 °C• MDL <3 pg carbon/s as tridecane• Linear dynamic range >107 with N2carrier and 0.29-mmid jet• Maximum data acquisition rate 500 Hz• Full range digital data path enables peaks to be quantifiedover the entire 107 concentration range in a single runThermal conductivity detector (TCD)• Maximum operating temperature 400 °C• MDL <800 pg tridecane/mL using He carrier (MDL maybe affected by laboratory environment)• Linear dynamic range 105 (±10 %)• Single-filament TCD can provide rapid baselinestabilization from turn on with low drift, and does notrequire a separate reference gas or manual potentiometeradjustment• The TCD can be mounted as a third detector on the leftside of the GC3Electron capture detector (ECD*)• Maximum operating temperature 400 °C• Makeup gas types argon/5 % methane or nitrogen • Radioactive source <15 mCi 63Ni• MDL <9 fg/mL lindane• Dynamic range >104 with lindane• 500 Hz maximum data acquisition rate• Equipped with hidden anode and high-velocity flows for contamination resistance• The ECD can be mounted as a third detector on the left side of the GCNitrogen phosphorous detector (NPD)• MDL <0.3 pg N/s, <0.04 pg P/s withazobenzene/malathion mixture• Selectivity 25,000 to 1 gN/gC, 75,000 to 1 gP/gC with azobenzene/malathion mixture• Dynamic range >104 N, >104 P withazobenzene/malathion mixture• Data acquisition rate up to 100 Hz• Blos bead standard (white ceramic bead not supported) Maximum operating temperature 400 °CFlame photometric detector (FPD) +(Plus)• Single wavelength• MDL <4.5 pg S/s, <120 fg P/s with methylparathion • Dynamic range >103 S, 104 P with methylparathion • Selectivity 106 g S/g C, 106 g P/g C• Data acquisition rate up to 500 Hz• Maximum operating temperature 400 °CSCD (Model 8355)• Highest sensitivity and selectivity for sulfur-containing compoundsNCD (Model 8255)• Highest selectivity for nitrogen-containing compounds Supports integrated and standalone SCD, and standalone NCD. See Agilent Sulfur Chemiluminescence Detector and Nitrogen Chemiluminescence Detector Specification Guide for more information.Mass spectrometer• Agilent 5977B Series MSD (stainless-steel source only)Other EPC DevicesThe 8860 has six total slots for EPC devices. If two inlets and two detectors are used, then two auxiliary slots are available. If two inlets and three detectors are used, then one auxiliary slot is available. Purged GC capillary flow devices are not supported.Auxiliary EPC module• Three channels of pressure control• Psig (gauge) pressure control• Forward pressure regulated 0–100 psig• Maximum of one auxiliary EPC module per GC Pneumatics control module (PCM)• Two channels for operation• PCM may be located in either/both inlet positions and in auxiliary position• Maximum of three PCMs per GC• First channel- Pressure or flow control (psig (gauge))- Forward pressure regulated 0–100 psig- Forward flow regulated 0–200 mL/min• Second channel- Pressure control- Forward pressure or back pressure regulated- Psig (gauge) and psia (absolute) pressure control Data communications• Two analog output channels (1 V and 10 V output available) as standard• Remote start/stop• LAN• Binary code decimal input for a stream selection valve Environmental conditions• Ambient operating temperature 15 to 35 °C• Ambient operating humidity 5 to 90 % (noncondensing)• Storage extremes –40 to 70 °C• Operating altitude 4,600 m4Safety and regulatory certificationsSafety standards• Canadian Standards Association (CSA) C22.2 No. 61010-1• Nationally Recognized Test Laboratory (NRTL):ANSI/UL 61010-1• International Electrotechnical Commission (IEC): 61010-1, 61010-2-010, 61010-2-081• EuroNorm (EN): 61010-1Electromagnetic compatibility (EMC) and radio frequency interference (RFI) regulation conformity• CISPR 11/EN 55011 Group 1, Class A• IEC/EN 61326-1• AS/NZS CISPR 11• This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 duCanada. ISM 1-A.Designed and manufactured under a quality system registered to ISO 9001. The Declaration of Conformity is available.Other specifications• Height 49 cm• Width 58 cm (68 cm with side mounted detector)• Depth 54 cm• Average weight 51 kg• Four internal 24-volt connections (up to 150 mA)• Two on/off contact closures (48 V, 250 mA max)• Support for up to any combination of three valves total:• Up to three heated gas sample valves• Up to two liquid sampling valves• Six independent heated zones, not including oven (two inlets, two detectors, and two auxiliary). Third detectorcan use any available auxiliary zones.• 400 °C Maximum operating temperatures forauxiliary zone Optional automated sample injectors and samplersSupports one 7693A autoinjector with capacity for 16 sample vialsorSupports one 7693A autoinjector and automatic sampler tray with capacity for 150 sample vials(Heater/mixer/bar code reader not supported)orSupports one 7650A autoinjector with capacity for 50 sample vialsorSupports one PAL3 AutosamplerSoftwareAgilent software choices are available designed to help you make the most of every run, and every workday, from:• OpenLAB CDS Workstation* and OpenLAB CDS Workstation Plus*• OpenLAB CDS VL Workstation* and OpenLAB CDS VL Workstation Plus*• OpenLAB CDS ChemStation Edition* or EZChrom Edition* • OpenLAB CDS ChemStation VL* or EZChrom VL* * Includes Agilent Retention Time Locking (RTL) supportfor 8860 GC with EPC* Optional RTL databases/libraries are not supportedwith 8860 GC• OpenLAB CDS EZChrom Compact• DA Express (Optional Data Analysis accessible from 8860 GC)For more informationFor more information on our products and services, visit our Web site at /chem5/chemDE44386.2271064815This information is subject to change without notice.© Agilent Technologies, Inc. 2018, 2020, 2021 Printed in the USA, July 9, 20215994-0453EN。

093 GC-8860MSD-5977B气相色谱质谱联用仪操作维护程序(作业指导书)

093 GC-8860MSD-5977B气相色谱质谱联用仪操作维护程序(作业指导书)

********有限公司作业指导书********-03-A093GC-8860/MSD-5977B气相色谱-质谱联用仪操作维护程序编写人:审核人:批准人:持有人:2020年5月15日批准2020年5月15日起实施********有限公司********有限公司文件编号:********-03-A093 页码:第1页共10页作业指导书版号:第1版第0次修订标题:GC-8860/MSD-5977BGC气相色谱-质谱联用仪操作维护程序颁布日期:2020-05-15GC-8860/MSD-5977BGC气相色谱-质谱联用仪操作维护程序1操作方法软件光盘安装顺序如下:1)安装MassHunter GC/MS Acquisition Software,安装完成后,桌面产生Agilent GCMS 配置图标,双击图标进入仪器配置界面如下,双击所要配置的仪器1:2)根据下图的提示进行对应的设置:********有限公司文件编号:********-03-A093 页码:第2页共10页作业指导书版号:第1版第0次修订标题:GC-8860/MSD-5977BGC气相色谱-质谱联用仪操作维护程序颁布日期:2020-05-152)安装MH Qualitative Analysis B.10.00 出现定性分析图标。

3)安装MH Quantitative Analysis B.10.00出现定量分析图标。

2、开机步骤1)打开氦气钢瓶,输出压力设置到0.5Mpa。

2)打开气相色谱电源,用手推住质谱侧板上凸起的方块(如果没按住,侧板缝隙太大,真空泵无法把侧板吸住),打开质谱电源,直至侧板被吸住。

注:对于早期5977A型号的真空泵软启动是开启状态的,所以需要较多的时间启动,转速会分阶段上升,转速升到100%大约需要20分钟。

3、调谐调谐MSD 包括优化一些仪器参数,如透镜电压和灯丝电流等,这使MSD 能够获得最佳质谱灵敏度和良好的分离度。

安捷伦8860气相色谱仪技术参数

安捷伦8860气相色谱仪技术参数

安捷伦8860气相色谱仪技术参数
安捷伦8860气相色谱仪的主要技术参数如下:
1. 柱箱温度范围:室温+150℃~400℃
2. 柱箱温度控制精度:±0.1℃
3. 柱箱温度升温速率:0.1℃/min ~ 40℃/min
4. 检测器类型:FPD(火焰光度检测器)、FID(火焰离子检测器)、TCD(热导检测器)、ECD(电化学检测器)、PID(光离子化检测器)等
5. 检测器灵敏度:FPD:1ppm,FID:1ppm,TCD:1%,ECD:1ppm,PID:1ppm
6. 载气:高纯度氮气或氢气
7. 进样方式:自动进样器、手动进样器、分流/不分流进样器等
8. 数据处理系统:Agilent MassHunter B.07.00软件
9. 分析时间:根据样品种类和仪器配置而定
10. 分析效率:根据样品种类和仪器配置而定
11. 稳定性:在24小时内,漂移应小于0.1%
12. 噪音水平:在24小时内,噪音应小于0.01%
13. 电源:220V/50Hz或110V/60Hz,功率:200W
14. 重量:约100kg
以上是安捷伦8860气相色谱仪的主要技术参数,该仪器具有高精度、高效率、高稳定性等特点,适用于多种样品的气相色谱分析。

安捷伦8860气相色谱仪的操作规程

安捷伦8860气相色谱仪的操作规程

安捷伦8860气相色谱仪的操作规程
操作规程如下:
1. 准备工作:
- 检查仪器是否处于正常工作状态,确认气源、电源等是否正常。

- 校准色谱柱并设置方法参数,确保准备好所需的试剂和样品。

2. 打开仪器:
- 打开色谱仪仪器主电源开关。

- 打开仪器的液晶显示屏,检查仪器是否正常开机。

3. 设置方法参数:
- 在仪器的操作界面上选择要使用的方法。

- 设置流速、温度、时间等参数,按照实验要求进行设定。

4. 准备样品:
- 根据所需测试的样品,准备好合适的样品溶液。

- 注射样品溶液到进样器中,注意避免气泡的产生。

5. 进行试验:
- 将进样器连接到色谱柱,并选择合适的进样方式。

- 在仪器操作界面上点击开始按钮,开始进行试验。

6. 数据分析:
- 在试验完成后,仪器会自动记录下各个峰的保留时间和峰
面积。

- 使用数据分析软件对试验结果进行分析,生成相应的色谱
图和数据报告。

7. 关闭仪器:
- 在试验完毕后,关闭仪器的运行。

- 清洗进样器和色谱柱,并将色谱柱存放在适当的条件下。

注意事项:
- 操作过程中,应注意安全措施,避免发生火灾、溶剂溢出等
意外情况。

- 操作时应严格按照仪器操作手册和实验室安全规定进行操作。

- 仪器的维护和保养,应按照仪器使用说明进行操作,避免损
坏仪器。

安捷伦气相色谱仪8860操作规程

安捷伦气相色谱仪8860操作规程

安捷伦气相色谱仪8860操作规程安捷伦气相色谱仪(Agilent 8860)是一种常用于化学分析实验室的仪器。

它具有高分辨率和灵敏度,可以进行定性和定量分析。

本文将为您介绍如何正确操作安捷伦气相色谱仪8860,以及相关实验步骤和注意事项。

第一步:准备工作在操作安捷伦气相色谱仪之前,必须确保所有操作员都具备基本的气相色谱仪原理和基础知识。

此外,还需要进行仪器的基本维护和日常保养。

1. 清洗色谱柱:色谱柱是色谱仪中最关键的部件之一,需定期进行清洗和保养以确保其正常运行。

首先,将适当的溶剂通过色谱柱,以去除残留的物质。

然后,用干燥的气体吹干色谱柱。

2. 检查气源:确保供气系统(如氢气、氦气和空气)正常运行,并确保气瓶有足够的气体供应。

3. 校准标准溶液:根据需要,准备相应的标准溶液,用于仪器的校准和定量分析。

4. 检查进样器:确保进样器之前和之后的连接处没有任何泄漏。

第二步:仪器设置在进行任何实验之前,必须进行仪器设置和调整。

下面是一些关键设置步骤。

1. 仪器开机:按下仪器的电源按钮,等待仪器自检完成并运行到待机状态。

2. 连接计算机:通过串口或网络接口将仪器连接到计算机,并确保与色谱数据处理软件的通信正常。

3. 设置温度:根据实验要求,设置色谱柱和进样器的温度。

注意,不同的样品和分析方法可能需要不同的温度条件。

4. 设置流速:调整气体的流速,以确保样品在色谱柱中的适当保留时间。

这可以通过调整流量计或气体压力来实现。

5. 校准仪器:使用已知浓度的标准样品进行仪器的校准。

这可以通过进样已知浓度的标准溶液并测量峰面积来完成。

第三步:样品处理和进样在进行样品分析之前,需要对样品进行适当的处理和准备。

1. 样品准备:根据实验要求,采用适当的方法对样品进行提取、纯化或预处理。

这可以包括固相萃取、液液萃取或其他样品处理方法。

2. 进样方式:根据样品类型和目标分析,选择适当的进样方式。

进样可以是液体进样、气体进样或固体进样。

气相色谱仪作业指导书(一)2024

气相色谱仪作业指导书(一)2024

气相色谱仪作业指导书(一)引言:气相色谱仪是一种常用的分析仪器,在化学、制药、环境等领域有广泛应用。

本文是《气相色谱仪作业指导书(一)》,旨在为初学者提供操作和维护气相色谱仪的指导。

本指导书将详细介绍气相色谱仪的原理、操作流程、常见故障及维护方法等内容。

一、气相色谱仪原理的介绍1. 气相色谱仪的基本原理2. 气相色谱仪的工作流程3. 色谱柱的选择及使用注意事项4. 色谱柱的保养与存储5. 色谱峰的分析与定性定量方法二、气相色谱仪操作流程1. 开机前的准备工作2. 样品的准备与进样操作3. 参考物质的选择和加入4. 仪器的操作设置5. 数据记录与结果分析三、气相色谱仪常见故障及排除1. 色谱峰异常及解决方法2. 柱子不流动怎么办3. 柱塞堵塞的处理办法4. 柱子中脂溶性物质残留的清洁方法5. 老化柱子的更换与维护四、气相色谱仪的维护与保养1. 仪器的日常检查与清洁2. 气源的检查与维护3. 柱子的更换与保养4. 检测器的校准与维护5. 仪器的存储与运输五、气相色谱仪操作注意事项1. 安全使用气相色谱仪的注意事项2. 操作步骤的注意事项3. 样品处理和进样操作的注意事项4. 仪器操作设置的注意事项5. 数据记录与结果分析的注意事项总结:本《气相色谱仪作业指导书(一)》详细介绍了气相色谱仪的原理、操作流程、常见故障及维护方法等内容。

通过学习本指导书,读者将能够全面了解气相色谱仪的使用和维护,提高实验室工作效率,并保证实验数据的准确性和可靠性。

同时,读者在使用气相色谱仪时也需注意操作的安全和步骤的正确性,以确保实验顺利进行。

68-8860-5977B气相色谱-质谱联用仪作业指导书

68-8860-5977B气相色谱-质谱联用仪作业指导书

一、目的为了使本实验室人员更规范使用8860-5977B气相色谱-质谱联用仪,特编制仪器操作作业指导书。

本指导书仅供本实验室内部人员使用。

二、开机前准备工作开机前必须进行开机前各项准备工作,以保证仪器正常使用,延长仪器使用寿命:2.1 气路检漏仪器开机前需进行气路检漏测试,可通过皂液检漏等方式,确保载气及辅助气气路无泄漏,对于MS检测器,必须严格检漏防止空气进入,以对柱子及检测器造成不可修复的影响。

2.2 进样口及进样塔相关检查在仪器开机前,需对仪器进样口各个构件及进样塔进行检查,包括但不限于:隔垫是否需要更换;所用衬管是否与进样方式相对应;进样针推杆是否阻塞;进样塔中溶剂瓶及废液瓶是否需要更换或者清理;检查色谱柱是否正常连接等等。

三、开机分析样品前应按照正常顺序进行开机操作:3.1打开气路首先,将对应的载气钢瓶总阀打开,调节减压阀使其压力控制在0.4~0.6MPa左右。

3.2 开机打开GC左下角电源开关键,打开质谱仪开关,此时,仪器各部分开始进行自检,待样品盘、进样塔、仪器等本身自检通过后,打开桌面上的图标,等待软件启动成功。

仪器正常启动后,质谱仪前级泵开始运转,使用人员应定期对前级泵泵油进行检查。

在软件界面找到真空图标,单击进入真空设置,并开始进行抽真空循环。

注:一般情况下,对于真空度的控制是质谱分析的关键操作,长期不用或者关机后间隔时间比较长的再开机操作,一般在拧紧色谱柱接口后,至少抽真空4h,以保证仪器有满足分析的真空度要求。

3.3调谐当真空循环完成后出现下面对话框,此时各状态实际值均达到一定要求,关闭真空循环。

在仪器主界面找到“视图”“调谐和真空控制”,进入到调谐界面。

在此界面下拉调谐菜单,选择自动调谐,此时仪器开始进行自动调谐操作,完成后将自动打印调谐报告。

使用人应根据调谐报告及评估报告对仪器真空度是否达标及离子源是否干净做出判断。

在质谱真空度未达到使用要求时,禁止不进行维护直接进行样品分析,以防对质谱仪造成不可修复的损伤。

安捷伦气相色谱仪操作规程

安捷伦气相色谱仪操作规程

安捷伦气相色谱仪操作规程安捷伦气相色谱仪操作规程一、实验目的本实验旨在熟悉和掌握安捷伦气相色谱仪的操作方法,了解色谱仪的常见故障及其处理方法,以提高色谱分析的准确性和可靠性。

二、实验仪器1. 安捷伦气相色谱仪一台2. 气相色谱柱3. 气源和载气瓶4. 样品进样器5. 检测器三、实验步骤1. 打开色谱仪电源,等待仪器启动完成。

2. 检查气源和载气瓶,确保气源充足。

3. 将气相色谱柱固定在色谱仪柱座上,并将柱座连接至进样器。

4. 打开进样器,将待测试样品加入进样器中。

5. 设置进样器的进样时间和进样容量,确保进样量符合分析要求。

6. 在色谱仪的控制面板上设置分析方法,包括温度程序、流速和检测器参数等。

7. 开始进样,将样品注入气相色谱柱进行分离。

8. 监控检测器输出信号,记录峰形和峰面积等数据。

9. 分析结束后,关闭进样器和色谱仪电源,并及时清理进样器和色谱柱。

四、注意事项1. 在操作色谱仪前,务必佩戴防护眼镜和实验手套,确保个人安全。

2. 气源和载气瓶的气压要稳定,确保色谱分析的流速稳定。

3. 进样器要经常清理,以避免污染样品和柱床,影响分析结果。

4. 在设置分析方法时,要根据待测物的性质和要求进行合理设定。

5. 小心操作、轻放仪器,避免对仪器造成损坏。

6. 如遇到色谱仪故障,应及时关闭仪器并联系维修人员处理。

五、常见故障及处理方法1. 色谱柱压力过高:可能是气源压力过高或柱床堵塞,应检查气源压力和清理柱床。

2. 峰形异常:可能是进样量过大或柱床老化,应调整进样量或更换色谱柱。

3. 检测器信号异常:可能是检测器故障或进样器漏气,应检查检测器和进样器。

4. 柱床损坏或破裂:可能是柱温不合适或柱床老化,应修复或更换柱床。

六、实验安全注意事项1. 使用化学品时,注意避免接触皮肤和吸入,加强通风换气。

2. 处理废弃物时,应按规定分类处理。

3. 操作过程中严禁大声喧哗,以免影响他人安全。

4. 实验后要及时清理工作台和仪器设备。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

安捷伦8860气相色谱仪作业指导书
一、仪器介绍
安捷伦8860气相色谱仪是一款高效、高灵敏度的气相色谱仪,广泛应用于环境监测、石油化工、食品药品等领域。

该仪器采用微电子技术,具有自动化程度高、操作简便、维护方便等特点。

二、仪器组成
安捷伦8860气相色谱仪主要由进样器、色谱柱、检测器和数据处理系统组成。

进样器采用六通阀进样方式,可以满足不同样品的进样需求;色谱柱采用高效毛细管柱,可以分离复杂的样品组分;检测器采用电子俘获检测器(ECD)或火焰离子化检测器(FID),可以高灵敏度地检测样品中的目标化合物;数据处理系统采用安捷伦化学工作站(GC ChemStation),可以对实验数据进行自动化处理和分析。

三、仪器安装与调试
在进行仪器安装与调试之前,请仔细阅读本手册及仪器的使用说明书。

确保实验室电源、气体(如氮气、氢气等)和水的供应充足。

按照仪器说明书中的步骤进行安装与调试,确保仪器正常运行。

四、样品前处理
在进行样品分析前,需要对样品进行前处理。

根据不同的样品类型和目标化合物,选择合适的提取方法。

例如,对于气体样品,可以采用气体吸收瓶或吸附剂进行富集;对于液体样品,可以采用萃取法或固相萃取法进行分离和纯化。

处理后的样品应尽可能避免污染和挥发,以保证分析结果的准确性。

五、样品进样
将处理后的样品注入进样器中,注意注射针头不要卡住进样器内的硅胶垫。

在进样前,需要将注射针头在进样口处放置一段时间,以保证样品完全进入进样口。

进样量应根据样品类型和目标化合物含量进行调整,以保证分析结果的准确性。

六、数据分析
通过安捷伦化学工作站(GC ChemStation)对实验数据进行采集和处理。

根据需要选择合适的色谱条件和分析方法,例如归一化法、内标法等。

在数据处理过程中,需要注意峰识别、基线校正、定量计算等环节的准确性。

同时,需要对实验数据进行质量评估,如标准偏差、可信区间等。

七、清洗与维护
为了保证仪器的稳定性和分析结果的准确性,需要对仪器进行定期清洗和维护。

根据不同的部件和使用情况,选择合适的清洗方法和清洗剂。

例如,对于色谱柱的清洗,可以采用高温老化或溶剂冲洗等方法;对于进样器的清洗,可以采用酒精棉球擦拭或超声波清洗等方法。

同时,需要定期检查仪器的各项指标是否正常,如压力表读数、气体流量等。

八、常见故障排除
在仪器使用过程中,可能会遇到一些常见故障。

为了及时排除故障,需要了解常见故障的种类和原因。

例如,常见的故障包括基线不稳定、色谱峰形异常等。

这些故障的原因可能包括仪器部件损坏、试剂质量问题等。

在排除故障时,需要先对故障原因进行初步判断,然后采取相应的措施进行维修或更换部件。

九、安全注意事项
在使用仪器过程中,需要注意安全事项。

首先,要确保实验室通风良好,避免因试剂挥发而造成空气污染;其次,要正确使用气体(如氮气、氢气等),避免气体泄漏或爆炸等危险情况;最后,要正确操作仪器,避免因操作不当而造成意外伤害或损坏仪器。