卧式灭菌器使用操作规程
卧式圆形压力蒸汽灭菌器的操作流程

卧式圆形压力蒸汽灭菌器的操作流程1.将压力蒸汽灭菌器放置在水平稳定的台面上。
Place the pressure steam sterilizer on a stable and level surface.2.打开蒸汽灭菌器的盖子,并确保内部干净整洁。
Open the lid of the steam sterilizer and ensure that the interior is clean and tidy.3.将待灭菌的物品放置在蒸汽灭菌器的合适位置上。
Place the items to be sterilized in the appropriate position within the steam sterilizer.4.确保所放置的物品之间相互之间不触碰。
Ensure that the placed items do not touch each other.5.加入适量清水至蒸汽灭菌器的水槽中。
Add an appropriate amount of clean water to the water tank of the steam sterilizer.6.关上蒸汽灭菌器的盖子,并确保盖子已经紧密关闭。
Close the lid of the steam sterilizer and ensure that it is tightly sealed.7.调节蒸汽灭菌器的压力阀和温度控制器至所需的设定数值。
Adjust the pressure valve and temperature controller of the steam sterilizer to the desired setting.8.打开蒸汽灭菌器的电源,启动设备。
Turn on the power of the steam sterilizer to start the device.9.等待蒸汽灭菌器内的水开始产生蒸汽,并加压。
卧式高压灭菌器的使用流程

卧式高压灭菌器的使用流程1. 准备工作在使用卧式高压灭菌器之前,需要进行一些准备工作,以确保其正常使用和灭菌效果。
•清洁灭菌器:首先,需要使用适当的清洁剂对卧式高压灭菌器进行清洁,包括内部和外部。
确保清洁剂不会对灭菌器产生腐蚀作用。
•准备灭菌器配件:检查并确保所有灭菌器配件的完整性和干净程度,如灭菌袋、灭菌器篮子等。
•准备需灭菌的物品:准备好需要进行灭菌的物品,并将其放置在灭菌器篮子中。
注意遵循物品的灭菌要求,不要超过灭菌器的负荷限制。
2. 设置和调试在开始灭菌程序之前,需要对卧式高压灭菌器进行设置和调试,以确保符合所需灭菌参数。
•设置温度和压力:根据灭菌要求,将温度和压力设置到适当的数值。
通常情况下,温度和压力的设置是提前确定好的,可以在灭菌器的控制面板上进行调节。
•调试操作:进行一次试运行以检查卧式高压灭菌器的正常工作状态。
确认灭菌器的门是否正常关闭、排气管道是否通畅等。
3. 进行灭菌在完成准备工作和调试后,可以开始进行卧式高压灭菌器的正式灭菌程序。
以下列点方式描述灭菌步骤:•将灭菌器篮子放置在灭菌器的固定位置,并确保灭菌篮子之间有足够的间隔,以便蒸汽可以顺畅流通。
•打开卧式高压灭菌器的门,将待灭菌的物品放入篮子中,并确保物品之间的间隔合适,以免影响灭菌效果。
•关闭灭菌器的门,并确保门已经完全关闭并且密封。
•在控制面板上选择合适的灭菌程序,包括温度、压力和时间等参数设定。
根据物品的灭菌要求,选择合适的程序。
•按下开始按钮,启动灭菌程序。
卧式高压灭菌器将开始加热、产生蒸汽,并保持一定的温度和压力,以达到灭菌要求。
•等待灭菌程序完成后,灭菌器会自动停止工作,并发出提示音。
此时,可以打开灭菌器的门,取出已灭菌的物品。
4. 灭菌后处理在取出已灭菌的物品之后,需要进行一些处理工作,以保持物品的无菌状态和安全性。
•检查物品:检查已灭菌的物品是否满足要求,如无菌包装完好、物品表面无明显污染等。
如发现问题,应及时进行处理。
卧式灭菌锅使用操作规程

卧式灭菌锅使用操作规程一、目的:建立灭菌锅正确使用操作程序,保证标准操作。
二、范围:适用于卧式灭菌锅的使用。
三、责任者:使用操作人员。
四、正文:1、依据:卧式灭菌锅使用说明书。
2、操作程序:2.1 将消毒物品放入灭菌锅内,要求不宜塞的过紧,不宜紧靠锅门。
2.2 将锅门关上,顺时针方向拨动紧锁手柄,使挡杆进入门内,旋紧八角转盘,关紧锅门。
2.3 开启锅上总进汽阀,再打开冷凝水排汽阀,待冷凝水排尽后,关闭此排汽阀,缓慢打开进汽阀,使蒸汽进入锅内,观察内外锅压力表,待内锅压力及温度达到所需温度时开始计时(升温时间应控制在10分钟左右),同时应打开灭菌锅底部排汽阀,并将安全阀臂提起数秒钟观察蒸汽是否畅通。
2.4 灭菌时操作人员应注意仔细观察内外锅压力,不得大于0.13MPa。
灭菌期间严禁离开工作现场(操作人员禁止正对锅门观察)。
2.5 如压力上升过快,安全阀跳阀,可关小进汽阀,打开内外排汽阀;如温度显示计的温度无变化,应立即关闭进汽阀,请专业人员来维修。
2.6 灭菌至所需时间后,缓缓关紧进汽阀及总进汽阀,待压力降至零、温度低于90℃时,打开锅门,取出物品,将挡板上的水擦干。
3、灭菌锅应在正常状态下进行工作,若有问题应立即停止使用。
4、定期用留点温度计校正灭菌锅内不同点的温度,保证灭菌温度均一且达到灭菌要求;定期用生物指示剂做挑战实验,验证灭菌效果。
及时填写记录。
5、注意事项:5.1 每次使用前应完全排出锅内冷空气。
5.2 注意被灭菌物品温度是否与锅内温度一致。
5.3 升温速度不易过快,否则造成恒温缩短,必须缓慢升温(约控制在10分钟左右)。
5.4 加压和减压必须缓升缓降。
5.5 锅内压力没有降到“0”时不能开锅。
5.6 允许使用压力0.1Mp a。
5.7 灭菌时,操作人员不能正对锅门,应站在灭菌锅门侧面操作。
5.8 严禁灭菌锅带病运行。
5.9 灭菌锅内的余水必须擦干。
5.10 定期对灭菌锅进行检查,避免事故的发生。
卧式矩形压力蒸汽灭菌的使用清洁标准操作程序

卧式矩形压力蒸汽灭菌的使用清洁标准操作程序 The latest revision on November 22, 20201目的建立卧式矩形压力蒸汽灭菌器使用、清洁标准操作程序。
2范围注射液车间灭菌器的操作。
3责任灭菌器操作人员、工艺员、质监员。
4参考文件SOP文件之灭菌器的使用与清洁。
5内容使用操作规程开机前:应对灭菌器各部位检查良好正常,若有故障立即排除。
检查灭菌柜内部,灭菌车等内外部位,清洁应符合规定。
灭菌前开启蒸汽阀门,排除管道中的冷凝水至成汽状。
检查待灭菌物品如工衣、帽应洗净并包装严密符合要求。
通向灭菌物品洁净室一端的门应关闭。
开机。
关闭近身一端的门,开动蒸汽阀门。
按程序进行预热-灭菌-排气-干燥-关闭蒸汽阀。
蒸汽若有变动或灭菌阀门漏气时,应及时调整或停机检修,注意安全。
灭菌物品按工艺规定,进行定温定时灭菌,在操作过程中,做好原始记录。
工作完毕检查关闭好蒸汽阀,不使漏气。
灭菌完毕,近身门严禁打开,按规定由洁净室内人员打开室内的一端门取走灭菌的物品,然后再关闭。
上述工作完成后,才能打开近身的门,清洁灭菌器内外,符合规定。
挂上状态标志。
清洗操作规程清洁工具:丝光毛巾、刷子。
清洁部位:设备内外部位灭菌车。
清洁剂名称:水、注射用水、酒精、洗衣粉。
清洁周期:每批一次。
清洁程序灭菌器内部灭菌车在洁净室的门,藏垢处用刷子洗刷,再用注射用水冲净,以丝毛巾擦干,必要时用酒精擦拭干燥。
设备外部敝露部位用干净的丝光毛巾擦拭,符合要求。
清洁程序验证用一块白晴纶布擦拭设备被验处,看白布表面有无异色,如无则清洗合格。
清洁工具的清洗与存:用自来水冲洗,再用注射用水洗净毛巾晾干。
存放地点:指定的清洁工具箱内。
6培训培训对象:工艺员、质监员、灭菌岗位操作工。
培训时间:二小时。
卧式圆形压力蒸汽灭菌器的操作流程

卧式圆形压力蒸汽灭菌器的操作流程英文回答:The operating procedure of a horizontal cylindrical pressure steam sterilizer is as follows:1. Preparing the sterilizer: Ensure that the sterilizer is clean and free from any debris or residue. Check the water level and replenish if necessary. Ensure that the drain valve is closed.2. Loading the items: Place the items to be sterilized in the sterilization chamber, ensuring that they are properly arranged and spaced to allow for adequate steam penetration. Avoid overloading the chamber to ensure effective sterilization.3. Closing the door: Close the sterilizer door securely and ensure that it is properly sealed. This is important to maintain the pressure and temperature inside the chamberduring sterilization.4. Setting the parameters: Set the desiredsterilization parameters such as temperature, pressure, and time according to the specific requirements of the items being sterilized. These parameters may vary depending on the type of material and the level of sterilization required.5. Starting the sterilization cycle: Once the parameters are set, start the sterilization cycle. The sterilizer will begin to heat up, and the pressure will gradually increase to the desired level. The sterilization time will also start counting down.6. Sterilization phase: During this phase, thesterilizer will maintain the set temperature and pressure for the specified duration. The steam will penetrate the items, killing any microorganisms present and ensuring effective sterilization.7. Cooling phase: After the sterilization phase iscomplete, the sterilizer will enter the cooling phase. The pressure and temperature inside the chamber will gradually decrease. It is important to wait until the pressure reaches a safe level before opening the door.8. Opening the door: Once the pressure has reached a safe level, open the sterilizer door carefully. Use appropriate protective equipment such as gloves to avoid any potential burns from the hot items inside. Remove the sterilized items and handle them appropriately.9. Cleaning and maintenance: After each use, clean the sterilizer chamber and accessories thoroughly to remove any residue or contaminants. Regular maintenance andcalibration of the sterilizer should also be performed to ensure its proper functioning.中文回答:卧式圆形压力蒸汽灭菌器的操作流程如下:1. 准备灭菌器,确保灭菌器清洁无残留物。
卧式高压灭菌器操作流程

卧式高压灭菌器操作流程英文回答:Operating a horizontal autoclave sterilizer involves several steps to ensure effective sterilization. First, I need to prepare the items to be sterilized. This includes cleaning them thoroughly and removing any debris or organic matter. For example, if I am sterilizing surgical instruments, I would need to clean them with detergent and water before placing them in the autoclave.Next, I need to load the items into the autoclave chamber. It is important to arrange the items in a way that allows for proper steam circulation and heat distribution. Overloading the chamber can result in uneven sterilization. For instance, if I am sterilizing glassware, I would place them on racks or trays in a single layer, ensuring there is enough space between the items.Once the items are loaded, I need to close theautoclave door tightly to create a sealed environment. This is crucial to maintain the pressure required for sterilization. I would ensure that the door gasket is in good condition and free from any damage or debris that could compromise the seal.After closing the door, I can start the sterilization cycle. The specific settings may vary depending on the type of items being sterilized and the autoclave model. Generally, the cycle involves heating the chamber to a specific temperature, typically around 121-134 degrees Celsius, and maintaining that temperature for a set period of time, usually 15-30 minutes. This ensures that the items are exposed to enough heat to kill any microorganisms present.During the sterilization cycle, it is important to monitor the pressure and temperature inside the autoclave. Most autoclaves have built-in sensors and indicators to provide real-time feedback. I would keep an eye on these readings to ensure that the sterilization conditions are being maintained properly. If any abnormalities aredetected, such as a sudden drop in pressure or temperature, I would need to investigate and address the issue before continuing the cycle.Once the sterilization cycle is complete, I need to allow the autoclave to cool down before opening the door. This is to prevent any sudden release of steam or pressure, which could be dangerous. I would wait until the pressure inside the chamber has reached a safe level before slowly opening the door. It is important to be cautious and avoid any direct contact with the hot items or steam.Finally, I would carefully remove the sterilized items from the autoclave and store them in a clean and dry environment. It is essential to handle the items with clean gloves or tools to prevent recontamination. I would also label the items with the date of sterilization and anyother relevant information for proper tracking and usage.中文回答:操作卧式高压灭菌器需要几个步骤,以确保有效的灭菌。
卧式高压蒸汽灭菌器安全操作及保养规程
卧式高压蒸汽灭菌器安全操作及保养规程一、引言卧式高压蒸汽灭菌器是一种常用的医疗设备,用于灭菌医疗器械、器具等。
为了确保工作效果和操作安全,正确的操作和定期的保养非常重要。
本文将详细介绍卧式高压蒸汽灭菌器的安全操作规程和保养方法。
二、操作规程2.1 准备工作在使用卧式高压蒸汽灭菌器前,需要进行以下准备工作:•确保设备正常工作,无异常提示或故障;•检查灭菌室内是否干净,并清除任何杂物;•准备干净的灭菌器锅和其他操作所需物品。
2.2 开机操作1.将灭菌器电源插头插入电源插座;2.打开设备电源开关,并确认仪表板上的指示灯亮起;3.按照设备说明书操作,打开灭菌器室门。
2.3 操作流程1.将待灭菌的器械放置在灭菌室内的适当位置,并确保器械之间有足够的间隙;2.关闭灭菌室门,并确保密封良好;3.根据灭菌器型号和器械的灭菌要求,选择相应的灭菌程序;4.设置好灭菌程序后,按下启动按钮,开始灭菌过程;5.等待灭菌过程结束,并等待指示灯显示灭菌完成。
2.4 关机操作1.当指示灯显示灭菌完成时,首先按下停止按钮停止灭菌过程;2.关闭设备电源开关,断开电源;3.打开灭菌器室门,取出灭菌好的器械;4.清除灭菌室内的杂物,并做好清洁工作。
三、保养规程3.1 定期清洁定期清洁是确保卧式高压蒸汽灭菌器正常工作的重要环节。
建议按照以下步骤进行定期清洁:1.使用洁净的湿布擦拭设备的外表面,并确保无尘;2.温和地清洗灭菌器室内的各个零部件,确保除去任何污垢;3.使用清水冲洗并彻底清除洗涤剂残留;4.定期清洗和清理灭菌器过滤器,以确保正常的通风和气体流动。
3.2 定期维护除了定期清洁之外,卧式高压蒸汽灭菌器还需要进行定期维护,以确保其正常工作和延长使用寿命。
以下是常见的维护方法:1.确保灭菌器的供电电压稳定且符合设备要求;2.定期检查灭菌室的密封情况,如有损坏或老化应及时更换;3.检查仪表指示灯和控制面板的工作情况,如有异常应及时报修;4.根据设备说明书,定期检查和更换关键部件,如热交换器、传感器等。
卧式圆形压力蒸汽灭菌器操作规程
卧式圆形压力蒸汽灭菌器操作规程一、设备准备1.检查设备的电源线和插头是否完好无损,并接好电源。
2.检查设备的控制系统、显示屏和操作按钮是否正常工作。
二、设备预热1.打开设备的蒸汽开关,加热锅内的水,等待水温升至设定温度。
2.在设备预热过程中,检查设备的温度计和压力计是否正常显示。
三、操作准备1.检查待灭菌物品是否符合灭菌要求,如密封包装、无潮湿、无明显污染等。
2.检查设备内部的结构件和密封件是否完好无损,如有破损应及时更换。
3.准备好所需的灭菌指示物品和记录表格。
四、灭菌操作1.将待灭菌物品放置在灭菌器的内部篮子中,并确保物品之间有足够的间距。
2.关闭设备的揭门,将锁紧装置锁好,确保密封。
3.设置灭菌时间和温度,可根据不同物品的要求进行调整,一般为121℃、15分钟。
4.启动设备的灭菌程序,设备会开始加热蒸汽,同时控制温度和压力。
5.在灭菌过程中,密切观察设备的温度和压力变化,确保符合设定范围,如有异常应及时停止操作。
6.灭菌完成后,设备会自动停止加热,锁紧装置会解除,灭菌器内的压力逐渐下降。
7.等待设备内压力完全释放后,打开揭门,取出已灭菌物品。
五、灭菌后处理1.将已灭菌物品进行分类整理,如存放在干燥、洁净的区域内。
2.检查已灭菌物品的密封包装是否完好,如有破损应立即重新处理。
3.根据灭菌指示物品的变色情况,确认灭菌效果,并将结果记录在相关表格上。
4.将灭菌器的内部进行清洁,去除残留物和污垢,保持设备的清洁和卫生。
5.定期检查设备的温度计、压力计和密封件的工作状况,如有异常应及时维修或更换。
六、安全注意事项1.操作人员在操作设备时,必须穿戴好防护服和防护手套,以避免烫伤和感染。
2.操作人员要时刻注意设备的工作状态,如发现异常情况应立即停止操作,并报告相关人员处理。
3.操作人员离开设备时,应关闭设备的电源,确保设备的安全。
4.灭菌器内的蒸汽和压力具有一定的危险性,操作人员在打开揭门时要小心操作,避免受伤。
卧式圆形压力蒸汽灭菌器的操作规程
卧式圆形压力蒸汽灭菌器的操作规程使用前的准备工作检查各仪表是否正常,电源是否正确良好,备用阀门应处于关闭状态,工作前必需把水加到高水位线以上,并设定好灭菌温度和灭菌时间,操作规程、步骤1: 堆放灭菌器的装载量不得超出灭菌室内容量的80%。
应尽量将同类物品放在一起灭菌,若必需将不同类物品装放在一起,则以最难实现灭菌物品所需的温度和时间为准物品装放时,上下左右相互间均应间隔确定距离以利蒸汽置换空气,物品包体积不得超出200mmx200mmx250mm难于灭的大包放在上层,较易灭菌的小包放在下层,金属物品放下层,织物包放上层,物品装放不能死靠门和四壁,以防吸入较多的冷凝水,金属包应平放,盘、碟、碗等应处于树立的位置;纤维织物应使折叠的方向与水平面成垂直状态;玻璃瓶等应开口向下或侧放以利蒸汽进入和空气排出,启闭式筛孔容器,应将筛孔的盖打开。
对溶液进行灭菌时,应灌注于硬质而热玻璃瓶中,以不超出其容积的314为宜,用棉龙纱布赛将瓶赛好,并里纱强于瓶颈,防止纱布赛落入瓶内,切勿使用未打孔的像胶或软木赛,最好将玻璃相放置于容积稍大的搪瓷或金属盘内,以防玻璃瓶爆裂时,溶液流失和污染灭菌室金属内壁。
2: 关门将门关上,按顺时针方向拨动锁紧手柄至中方位,使撑挡杆进入门圈内。
然后顺时针方向转动八角手轮,使门和橡胶垫圈密合。
3: 加水必需在操作前关团排水阀,开启进水阀,打开排空阀,使水进入水箱内,加至高水位上5—10m处为上,然后关团进水阀和排空阀。
每次灭菌前都应当检查败水,如有不足应予增补。
4: 加热关紧冷凝掌控阀,将电源开关拨至“开”,电源指示灯亮,水位指示灯亮,开启加热开关,加热指示灯亮,表示水箱已开始加热,产生蒸汽进入夹套。
当水箱水位低于规定时,能自动切断加热电源,水位指示灯熄灭,报警指示亮并发出峰鸣声,这时应关闭电源开关,待夹层蒸汽压力为季时再加水,加水后,方能线续加热。
当夹套压力超出夹套压力表设定上限值时,加热电源会自动切断,加热指示灯随即熄灭,以后电流时通时断,自动使压力保持稳定,当夫套压力超出压力掌控器的设定压力时,加热指示灯熄灭,加热停止。
卧式蒸汽灭菌器操作方法
卧式蒸汽灭菌器操作方法
卧式蒸汽灭菌器是一种常见的灭菌设备,用于对医疗器械、制药品等进行灭菌处理。
以下是卧式蒸汽灭菌器的操作方法:
1. 填充水箱:打开灭菌器的水箱盖,倒入适量清水,确保水箱中有足够的水。
2. 放置物品:将待灭菌的物品放置在灭菌室内,注意排列整齐,确保蒸汽能够充分接触到每个物品表面。
3. 设置灭菌程序:根据物品的特性和灭菌要求,在控制面板上选择合适的灭菌程序和参数,如温度、压力和时间等。
通常,灭菌程序会根据物品的不同而有所区别。
4. 启动灭菌器:按下启动按钮,蒸汽灭菌器开始工作。
设备会自动加热水箱内的水,产生蒸汽,并将蒸汽输入到灭菌室中。
5. 灭菌过程中的监控:在整个灭菌过程中,可以通过控制面板上的显示屏实时监控灭菌的进度和状态。
确保温度、压力等参数在设定范围内。
6. 灭菌完成提示:当灭菌程序完成后,灭菌器会发出声音提示或显示灯亮起。
此时,需要关闭蒸汽灭菌器,并等待一段时间,待灭菌室内的压力和温度下降后再打开设备门。
7. 取出物品:小心打开设备门,使用器械或工具将灭菌后的物品取出。
注意操作过程中的个人安全和卫生要求。
8. 清洁和维护:使用后,务必对蒸汽灭菌器进行清洁和维护工作。
清理蒸汽灭菌器内的水箱、灭菌室和排放管道,确保设备的卫生和功能正常。
以上是卧式蒸汽灭菌器的基本操作方法,根据不同品牌和型号,具体的操作步骤可能会有所不同,建议在使用前仔细阅读设备的使用说明书,并按照厂家提供的操作指导进行操作。
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
标准操作规程
Standard Operating Procedure
1.目的
建立卧式灭菌器使用操作规程,确保其操作规范化。
2.适用范围
适用于卧式圆形压力蒸汽灭菌器(电加热)的操作。
3.责任者
操作者
4.操作规程
4.1加水
灭菌器必须在操作前将仪表箱上总阀转至全排位置并打开进水阀,加水至水箱高水位标记处,然后将进水阀关闭。
每次灭菌前都应检查贮水量,如有不够,应予补充。
4.2加热
先将总阀旋钮转至“关闭”,然后将压力控制开关的旋钮旋至“开”处,按物品所需灭菌温度,将压力控制旋纽旋转至相应位置,此时电热指示灯亮,同时其对应的压力指示灯亮,表示加热开始。
4.3堆放
将待灭菌物品顺序地、保持适当间隙地放入灭菌室内。
同时将瓶盖旋开少许。
4.4密闭
将门关上,按顺时针方向拨动锁紧手柄中心盘转至中方位,使撑挡进入门圈内,然后按顺时针旋转八角转盘直至仪表箱面板上门指示灯亮。
4.5灭菌
当夹套内蒸汽压力达到所选择的控制压力时,同时门关闭指示灯亮,即将总阀旋转至“灭菌”位置,同时将冷凝水排出阀旋开适宜位置。
当灭菌室温度达到预选温度值时,开始记时。
注意调整冷凝水排出阀,使灭菌温度保持在预选温度值。
4.6排气
当消毒终了时,应将总阀旋钮旋至“慢排”或切断电源使灭菌室压力和温度自然降低,直至灭菌室压力表回降至“0”位,随后将总阀旋钮旋至“全排”档,排去套层和灭菌室内的蒸汽,并打开灭菌室门少许,散发灭菌室内的剩余蒸汽,使灭菌室内壁保持干燥。
4.7物品取出储存
当所灭菌物品冷却后,打开灭菌器门,将其取出放于冰箱存放。
(需旋紧瓶盖的先旋紧瓶盖后取出)
5.注意事项
5.1压力表经常检查是否灵敏,以保安全。
5.2灭菌室内有压力切不可打开灭菌器门,以防事故。
5.3不同类型物品最好不要同时进行灭菌。
5.4每次加热前须检查水位表。