影响凝血四项检测的因素

合集下载

常见影响凝血四项检测结果的因素

常见影响凝血四项检测结果的因素

常见影响凝血四项检测结果的因素目的分析影响因素与检测结果的关系,便于检测中更好的控制影响因素对试验的不良影响,提高检测数据的准确性和可靠性。

方法对2011年1月~2011年3月试验用血液标本360例检测不同因素对凝血酶原时间、凝血酶时间、部分凝血酶时间和纤维蛋白原的影响加以分析。

结果样本血量不足与过量对凝血酶原时间、凝血酶时间、部分凝血酶时间和纤维蛋白原四项均存在明显差异,溶血样本对凝血酶原时间、凝血酶时间和部分凝血酶时间三项均存在明显差异,室温放置8 h后,凝血酶原时间和凝血酶时间两项均存在明显差异,4 ℃环境温度存放8 h以内凝血四项检测结果无明显差异。

结论检测的质量控制是确保凝血四项检测结果准确的关键。

标签:凝血检测;血浆凝血酶原时间;部分凝血活酶时间;凝血酶时间;纤维蛋白原The common influencing factor on results of four blood coagulation index ZHAO Hong-ji.The Third People s Hospital of Luoyang,Luoyang 471002,China【Abstract】Objective To analyze relation between the influential factors and the results of detection,control effect factors,and improve the accuracy and reliability in the testing data.Methods Tested blood specimens of 360 cases from January 2011 to March 2011 of detection of different factors on the prothrombin time,thrombin time,partial thrombin time and fibrinogen effects.Results Blood deficiency and excess were obvious difference in PT,TT,APTT and FIB.Hemolytic samples were obvious difference in PT,TTand APTT.Samples placed after eight hours in room temperature were obvious difference in PT and TT.Samples placed in 4 ℃temperature for 8 hours had no obvious difference in four test results.Conclusion Examination quality control was a key to ensure the accuracy on detection results.【Key words】Clotting detection;Prothrombin time;Activated partial thromboplastin time;Thrombin time;Fibrinogen凝血试验是临床常用的检查手段,其准确的检查结果有助于提高临床治疗的正确性[1]。

血凝四项实验结果异常影响因素

血凝四项实验结果异常影响因素

血凝四项实验结果异常影响因素一、PT/APTT 时间延长影响因素A、标本采集:1、浊度法测定终点的自动化血凝仪,有溶血和较明显的黄疸及脂血的样本。

2、抗凝剂与血液比例不是1:9 。

3、抗凝剂配制错误,如枸橼酸钠含2 个结晶水仍按3.8%配制。

4、血细胞比容>55 %,抗凝剂仍用1 :9 ,引起抗凝剂比例过高。

5、标本陈旧,超过2 小时,甚至更长。

6、标本运输未加盖,致使CO2 ↑ PH ↑ 。

7、口服抗凝剂或凝血因子异常。

8、质控标本失效。

9、采血时,血浆与抗凝剂混合时振摇过于剧烈,使凝血因子破坏(轻轻混匀即可)。

10、血浆分离应2500g,10 分种。

B、复溶液:1、未用公司试剂盒同批号配制的复溶液。

2、用蒸馏水取代公司的复溶液。

3、复溶稳定时间不够。

C、操作:1、血浆与试剂比例不对(按照试剂说明书或仪器厂家的使用说明书)。

2、标本、试剂未经预温至37 ℃。

3、APTT 试剂与标本孵育小于3 分钟。

4、APTT混悬液加前未混匀。

5、试验测试温度不在37±0.5 ℃的范围。

D、试剂本身:1、APTT(鞣花酸)被冷冻(低于0 ℃)贮存或运输。

2、试剂贮存、运输温度过高,未注意冷藏运输。

3、试剂超过有效期。

4、购入了冒牌试剂。

5、质控品过期。

二、PT/APTT时间缩短影响因素A、标本采集:1、病人为高凝状态或血栓性疾病。

2、血细胞比容<25%,抗凝剂含量不够。

3、抗凝剂自配,配制错误,如对5个结晶水的枸橼酸钠仍按3.2%配制(应按3.8%配制)。

4、抗凝剂存放时间长:长霉、变质。

5、血浆于2-8 ℃运输,存放(冷凝集效应),或标本在室温中存放时间过长(超过2小时,甚至更长)。

6、血浆分离时,离心机转速不够,血浆内有较多血小板。

B、复溶液及材料:1、未用公司试剂盒内同批号配制的复溶液。

2、用蒸馏水取代公司的复溶液。

3、比色皿或试验器材污染。

C、试剂本身:1、鞣花酸和白陶土搞错,如原先用白陶土,后改用了鞣花酸,用白陶土正常值来衡量鞣花酸。

常见阻碍凝血四项检测结果的因素

常见阻碍凝血四项检测结果的因素

常见阻碍凝血四项检测结果的因素产前及手术前血栓及出血机制方面的查验已成为必查查验项目之一,自2000年卫生部明确规定淘汰手指法查出凝血时刻以来,我院引进普利生C-2000高性能血凝仪对凝血酶原(PT),活化部份凝血活酶时刻(APTT),纤维蛋白原含量(FIB)及凝血酶时刻(TT)进行测定,利用本机几年来,对凝血四项检测结果的阻碍因素有几点体会与大伙儿一起探讨。

C-2000型血凝仪采纳磁珠法原理,通过先进的双磁路系统测定钢珠在检测中振幅的转变,确信凝固终点,经计算处置得出检测结果。

尽管它是目前凝固法中,国际公认最先进的直接检测方式,有不受血液中黄疸,溶血,乳糜微粒,浑浊试剂本身浊度等缘故造成的干扰和不受原血浆粘度的阻碍优势外,但其有受人为因素较多:1机械方面机械与电源连接时,要注意插座的接地线是不是靠得住,严禁零地共用。

进行测试时,机械要先进行预热,显示窗达到37℃后方可正常利用以保证测量的准确度。

本机械属高精度测量仪器,应放置在平稳的工作台上利用(工作台不稳也会阻碍磁珠的左右运动)。

仪器和加珠器应远离电磁场干扰源,为保证测量精度应用普利生生产的-次性测试杯和一次性钢珠。

2试剂方面试剂盒质量采纳国际太平洋凝血制品公司生产的试剂盒,严把质量关,每批试剂都要做正常对照。

试剂配制严格依照试剂说明书要求,用蒸馏水或厂家特定溶液溶解,溶解加量要准确,干粉要完全溶解,轻轻转动容器摇匀。

新配制和复溶试剂应放置5-10分钟后方可利用,复溶试剂不可再冰冻,按规定的稳按期及有效期利用,一样时刻太长或反复利用污染的试剂会致使结果延长。

3样本方面样本搜集静脉抽血,穿刺要一针见血,幸免瘀血,气泡,溶血或组织液污染,采血量必需准确,不低于打算用血量的90%,采血后与抗凝剂充分混匀。

样本处置抗凝剂的应用草酸钠,乙二胺四乙酸,肝素不适合作抗凝剂,利用%或%(W/V)枸橼酸钠溶液作抗凝剂。

血样与抗凝剂的比例为9:1,若是血球压积&lt;20%或&gt;55%,那么需调整那个比例,方式是×血液毫升数×(100-压积)=抗凝剂毫升数。

凝血四项检测结果的影响因素研究

凝血四项检测结果的影响因素研究

凝血四项检测结果的影响因素研究张淑华【期刊名称】《中国继续医学教育》【年(卷),期】2015(000)005【摘要】目的:分析影响凝血四项检测结果的因素。

方法血液标本320例,观察其凝血酶原时间、凝血酶时间、活化部分凝血酶时间、纤维蛋白原的检测结果,分析不同因素对检测结果的影响。

结果采血量不足或过量、溶血、室温环境不同放置时间、不同离心时间均会对凝血四项检测结果产生影响(P<0.05)。

结论采血量、溶血、室温环境放置时间、离心时间都是检测结果的影响因素。

%Objective To analyze the test results of blood clotting factors. Methods 320 cases of hospital blood test specimens, observe the prothrombin time, thrombin time, activated partial thromboplastin time,fibrinogen test results, analyze the impact of different factors on the test results. Results The volume of blood deficiency or excess, hemolysis, room temperature stored for different time, different centrifugal time will affect the outcome of four items of blood coagulation detection (P<0.05). Conclusion Blood volume, hemolysis, room temperature storage time, centrifugation time are the factors affecting the test results.【总页数】2页(P175-176)【作者】张淑华【作者单位】157199黑龙江省海林市人民医院【正文语种】中文【中图分类】R446【相关文献】1.标本离心时间对凝血四项检测结果的影响 [J], 许军生;刘付明军;谭洪辉2.影响凝血四项检测结果异常原因的分析 [J], 陈勇;凌建;徐容;李倩3.凝血真空采血管和标本放置时间与温度对常规凝血四项检测结果的影响 [J], 周以华4.凝血四项检测结果的影响因素分析 [J], 赵羚;马聪;韩丁丁;高玉娟;苏雁华5.采集量、保存时间、溶血及离心时间与速度对凝血四项检测结果的影响 [J], 叶琳;蔡贤惠因版权原因,仅展示原文概要,查看原文内容请购买。

分析前因素对凝血四项测定结果的影响分析

分析前因素对凝血四项测定结果的影响分析

分析前因素对凝血四项测定结果的影响分析目的分析影响凝血四项测定结果的前因素。

方法选取该院2013年6月—2014年5月不合格采血标本206份,与其重抽后合格标本,检测凝血四项指标;采集送检合格标本26份即刻、4°C放置8 h、室温下放置8 h再检测。

结果采血量不足组的PT、APTT、TT显著高于重抽合格标本组,差异有统计学意义(P <0.05);FIB显著低于重抽合格标本组,差异有统计学意义(P<0.05);采血量过多组的PT、APTT、TT显著高于重抽合格标本组,差异有统计学意义(P<0.05);FIB显著低于重抽合格标本组,差异有统计学意义(P<0.05)。

溶血组的TT显著高于重抽合格标本组,差异有统计学意义(P<0.05);PT、APTT显著低于重抽合格标本组,差异有统计学意义(P<0.05);两组的FIB差异无统计学意义(P>0.05);室温环境下即刻PT、APTT显著低于8 h后指标,差异有统计学意义(P<0.05);而即刻、8 h后TT、FIB指标差异无统计学意义(P<0.05);4°C 下即刻PT、APTT、TT、FIB指标与8 h后比较,差异无统计学意义(P>0.05)。

结论凝血四项检测结果分析前质量控制是确保实验室检查结果准确性的前提。

[Abstract] Objective Analysis of factors affecting four before the determination results of blood coagulation. Methods 206 copies of unqualified blood specimens in our hospital from June 2013 to May 2014 were selected,With the qualified samples,detection coagulation four indicators;26 inspection qualified samples immediately,4 ° C placed 8 h,placed to 8 h at room temperature. Results TT,APTT,PT of Insufficient blood group were significantly higher than that of resampling qualified samples group,the difference was statistically significant (P<0.05);While the FIB was significantly lower than that of resampling qualified specimens,the difference was statistically significant (P<0.05);TT,APTT,PT of much blood volume group were significantly higher than that of resampling qualified samples group,the difference was statistically significant (P<0.05);While the FIB was significantly lower than that of resampling qualified specimens,the difference was statistically significant (P<0.05);TT of Hemolytic group was significantly higher than that of resampling qualified samples group,the difference was statistically significant (P<0.05);APTT,PT were significantly lower than those of re sampling qualified samples group,the difference was statistically significant (P<0.05);FIB had no statistically significant difference between the two groups (P>0.05);room temperature environment of immediate PT,APTT was significantly lower than that of 8h index,the difference was statistically significant (P<0.05);and immediately after 8h,TT,FIB index had no statistical difference (P<0.05);4°C immediate PT,APTT,FIB,TT index and 8h after comparison,he difference has no statistically significant(P>0.05).Conclusion Pre analysis quality control of the results of four items of blood coagulation detection is the premise to ensure the accuracy of laboratory results.[Key words] Quality control;Pre analysis;Coagulation four indices人体的止血功能极其重要,当人意外受伤流血时,止血功能可以快速凝固血液,使伤口止血[1]。

影响凝血四项检测的因素

影响凝血四项检测的因素

影响凝血四项检测的因素【摘要】目的观察血液标本、采集、运送、放置时间对凝血四项测定结果的影响。

方法对我院2009年至2010年住院患者2149例标本,进行凝血酶原时间(PT),纤维蛋白原(FIB),活化部分凝血活酶时间(APTT)和凝血酶时间(TT)测定。

结果在2149例标本中有273例有误差,误差率12.7%。

其中采血量比例不当者253例,占92.7%;标本与抗凝剂没充分混匀者11例,占4.02%;标本放置时间过长者8例,占2.93%;混有肝素者1例,占0.33%。

结论标本采集不当与送检时间是导致测定结果误差的主要因素。

【关键词】凝血四项;影响因素;解决方法凝血实验结果的准确性、可靠性除了靠实验室的质量控制外,护理人员对标本采集和运送是分析前质量控制的一个重要方面。

凝血四项是判断凝血系统病理变化、术前筛查凝血系统病和指导临床用药的重要指标。

Rod-gers等研究认为,Duke法测定出血时间不能作为凝血功能的预测指标[1]。

2000年卫生部明令停止使用术前的常规出、凝血时间检测,用凝血四项取代,使临床标本量激增。

丛玉隆等认为:血液离体即开始变化,随存放方式和时间的不同,凝血因子逐渐消耗而导致检验结果不同[2]。

本文分析了非疾病因素异常结果和标本、收集及放置时间的关系,探讨造成凝血四项误差的常见原因及解决方法。

1 材料与方法1.1 标本来源所有标本均来自于本院住院患者,由护士采集送到本科。

收集我院2009年至2010年2149例标本住院患者静脉血。

收集不合格标本273例(非病理因素引起结果异常者),其中血量比例不当者253例,标本与抗凝剂没充分混匀者11例,混有肝素者1例,标本放置时间过长者8例,重新抽血复查。

1.2 仪器SYSMEX CA510全自动血凝分析仪。

1.3 试剂SYSMEX原装试剂1.4 比色杯原厂比色杯。

1.5 试管杨州市华光医疗器械有限公司生产2 ml血凝管带盖。

1.6 方法检测项目为凝血酶原时间(PT)、凝血酶时间(TT)、活化部分凝血酶时间(APTT)、纤维蛋白原(FIB)。

凝血四项检测结果影响因素的分析

凝血四项检测结果影响因素的分析目的探讨凝血四项检测结果影响因素。

方法收集我院近年来进行静脉采血患者血标本300例,觀察其凝血四项检测结果,并对影响四项检测(纤维蛋白原、凝血酶时间、凝血酶原时间及活化部分凝血酶时间)结果的因素进行分析。

结果溶血、脂血、血液凝固不良、采血量不足、小凝块等均是导致凝血四项结果偏差的重要原因(P<0.05)。

结论凝血四项检测结果能够受到抗凝剂使用、采血时间等多种因素的影响,为提高检测准确性,相关人员要加强控制静脉采血过程及时间。

标签:凝血四项;检测结果;影响因素凝血四项是患者术前必须接受检查之一,目的在于对患者止血功能是否存在缺陷进行观察,减少术中大出血现象。

凝血四项是指纤维蛋白原、凝血酶时间、凝血酶原时间及活化部分凝血酶时间,然而在实际检测中常受到各类因素的影响,从而导致检测出现偏差,对手术造成严重不良后果,因此对影响凝血四项检测结果的因素进行分析有着重要的临床意义[1]。

笔者对我院收治的300例患者血液标本检测情况进行观察,具体报告如下。

1资料与方法1.1一般资料我院自2012年5月~2014年5月收治的300例患者血液标本,患者情况:男186例,女114例,年龄6~75岁,平均年龄(32.65±5.21)岁。

1.2方法使用CA1500全自动血凝仪(生产公司:日本sysmex公司),使用质控品与试剂均为进口配套产品。

患者晨起状态下抽取静脉血5mL,加入枸橼酸钠109mmol/L,抽出血液后,将之放入含有抗凝剂的真空采血管内,枸橼酸钠:血液为1:9。

查看血液样本体积,是否出现溶血、止血及血块等现象。

抽血完成后分别在1、2、4h随机抽取50份血液标本,严格按照操作说明书进行凝血四项检测。

将剩余150份血液标本随机分为三组,3000r/min离心机离心5、10、15min,取上层血浆进行凝血四项检测。

1.3统计学分析本次研究所有患者的临床资料均采用SPSS18.0统计学软件处理,计量资料采用均数加减标准差表示(x±s),计数资料采用t检验,组间对比采用x2检验,P<0.05为差异具有显著性,具有统计学意义。

影响凝血分析仪检测结果的六个因素

影响凝血分析仪检测结果的六个因素凝血分析仪是医院里常用的检验设备,用于凝血酶原时间(PT)、凝血酶时间(TT),活化部分凝血活酶时间(APTT)、纤维蛋白原(FIB)及各种凝血因子,是临床凝血、出血和血栓疾病的常见诊断方法,是术前检测的好帮手。

在实际的检测应用中,凝血四项检测结果易受诸多因素的影响,现进行各方面影响因素的分析,以帮助用户在操作时可以避免这些错误。

标本的采集(1)采集血标本前要让患者处于空腹和平静状态,情绪激动、剧烈运动或神经紧张会引起血小板数增多、凝血和纤溶活性增强。

(2)采血时,止血带不应扎得太紧或时间太长,因长时间结扎会使因子Ⅷ和组织纤溶酶源激活剂(t-PA)释放和活化。

注入血液不可太用力,以免血细胞破碎激活凝血系统。

(3)要用一次性塑料注射器和塑料试管,以减少血小板和凝血因子的活化,进针要顺利,第1支注射器取实验用血,这样可以避免混入组织液,血液和抗凝剂要充分混匀,否则影响凝血检验的准确性。

标本的储存和运输(1)标本采集后注入塑料或硅化的试管内,并且立即塞紧盖子,否则会有二氧化碳挥发,pH值增高,影响APTT、PT延长。

(2)要及早分离血浆,最好在2h内分离出来,对于当前不能做的标本,可将血浆分离分装在带塞子的塑料管内。

冷冻在-20℃的冰箱内,将需要存放较长时间的标本冰冻存于-80℃的冰箱内,冰冻的血浆融化时不能静置于室温中融化,应放置37℃水中轻轻摇动,使其迅速融化。

(3)标本必须在室温下运送,因低温会损伤血小板、活化因子Ⅶ和因子Ⅺ,使PT、APTT结果缩短。

必须低温保存的标本要在4℃以下运送,以防止因子V 和因子Ⅱ降解。

标本存放不同的温度或时间对凝血因子的活性会有不同的影响。

抗凝剂的选择应用抗凝剂为0.109mol/L的柠檬酸钠,它能有效地阻止因子V和因子Ⅷ的活化,严格按照1:9的比例抗凝,当红细胞比容(HCT)>55%或<25%时,应按如下公式重新核算抗凝剂的用量,抗凝集量(mL)=0.00185×全血量(ml)×[100—HCT(%)],不应使用变质的枸橼酸盐溶液,否则会使PT、APTT试验结果偏低。

标本采集和处理对凝血四项检验结果的影响

标本采集和处理对凝血四项检验结果的影响临床实验室及时、可靠的检验结果,除了按实验室严格的质量控制外,还与医护人员正确采集检验标本有很大关系,一份送检标本合格与否,往往是影响检验结果的重要环节。

多数临床医生认为检验结果的及时性与可靠性由检验科来保证,而实际工作中导致检验结果质量误差的因素有很多方面,如患者个体的差异、仪器误差、操作者个体误差和采样送检是否正确等。

为了保证检验结果的及时性和可靠性,应注意标本检验分析前的质量控制。

1 、标本采集由于小孩、老人、烧伤、低血压休克等因素造成静脉血不易抽取,抽血时操作不规范,使标本中有肉眼可见的微小凝块,使PT、APTT、TT延长,FIB降低。

血液与抗凝剂未能充分混匀,也可以使标本中有肉眼可见的凝血块。

其原因为:往往由于工作忙,标本采集后未能及时处理,抽血时操作不规范,使标本溶血,造成PT、APTT缩短,TT延长。

血量与抗凝剂比例减少,造成PT、APTT显著延长。

长时间扎止血带、静脉穿刺不能一次成功造成血管损伤或组织因子进入血液。

造成检验结果波动的因素很多,而不规范标本采集送检是其最主要的原因之一,为了避免检验结果与临床诊断不符,造成不必要的抽血和复查,增加患者的负担,建议标本采集时,护士应加强责任心,以科学的态度采集及处理标本,减少不必要的误差。

2 、标本的送检在实际工作中,标本送到实验室检验时,为了避免实验室由于本身某些因素造成的结果误差,我们建议:(1)加强检验人员的个体素质,不断提高基础理论水平及现代的技术水平。

在日常工作中,检验人员必须严格遵守实验室的规章制度,严格执行操作规程,对不合格标本予以拒收,对检验的结果做到心中有数。

(2)加强与临床医护人员的联系,对检验中出现异常和某些阳性结果应主动及时与临床医生取得沟通,对临床有疑问的结果做到合理解释,尽职尽责地为临床抢救工作及手术等提供及时准确的检验结果。

凝血四项偏倚标准

凝血四项偏倚标准凝血四项是临床检验中常见的血液检查项目,用于评估一个人的凝血系统功能。

这四项测试包括:凝血酶原时间(PT)、部分凝血活酶时间(APTT)、纤维蛋白原(Fbg)、凝血酶时间(TT)。

每一项测试都各有其特定的参考范围,但也可能受到一些因素的影响,导致结果出现偏倚。

以下是这四项测试的偏倚标准:1. 凝血酶原时间(PT):正常值约为12-13秒。

延长可能表明外源性凝血途径的障碍,例如缺乏维生素K、严重肝功能衰竭或先天性凝血因子缺乏;缩短则可能表示血液高凝状态或血栓形成的风险增加。

2. 部分凝血活酶时间(APTT):正常值约为25-35秒。

延长可能表示内源性凝血途径的障碍,例如系统性红斑狼疮、类风湿性关节炎或肝脏疾病;缩短则可能表示血液高凝状态或血栓形成的风险增加。

3. 纤维蛋白原(Fbg):正常值约为2-4g/L。

增高可能表明共同凝血途径的功能增强,例如高脂血症、糖尿病或多发性硬化症;降低则可能表示共同凝血途径的功能减弱,例如急性肝炎或肝硬化。

4. 凝血酶时间(TT):正常值约为16-18秒。

延长可能表示纤维蛋白在纤溶系统纤溶酶原激活物作用下可降解为可溶性纤维蛋白,并可被纤溶酶降解为纤维蛋白降解产物,包括D-二聚体、FDP,故TT延长。

见于共同凝血途径的功能减弱,例如肝病或DIC;缩短则可能表示血液高凝状态或血栓形成的风险增加。

在进行凝血四项检查时,需注意以下几点以减少偏倚:1. 采血过程:确保采血过程规范,避免过度挤压或用力插管,以免激活血小板和凝血因子。

2. 标本处理:血液样本应尽快离心并分离出血浆,避免长时间放置或反复冻融,以免影响凝血因子活性。

3. 试剂质量:使用高质量的试剂盒,并按照说明书进行操作,以确保结果的准确性。

4. 仪器校准:定期对凝血分析仪进行校准和维护,以确保仪器的准确性和稳定性。

5. 人员培训:操作人员应接受专业培训,熟悉实验操作和仪器使用方法,以提高实验准确性和可靠性。

  1. 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
  2. 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
  3. 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。

影响凝血四项检测的因素
由于许多凝血因子是具有生物活性的蛋白,常温下易失活,而凝、止血又一个当复杂的病理过程,凝血四项检查在临床上对于出血性疾病的诊断、鉴别、病情观察治疗检测预后判断和术前检查具有重要的作用。

影响此实验的因素具有特殊性,而实验结果的准确性是临床应用的前提。

因此,为保证检测结果的准确,现将影响因素总结如下。

1样本因素
①样本采集超过或不带划线部分;②未使用规定的抗凝剂(枸橼酸钠与血1:9)抗凝剂的比例及用量不准确;③血样有凝块、溶血、黄疸、脂血;④血样本混匀不当,剧烈混匀产生气泡;⑤使用塑料试管来收集血样;⑥采血器或储血管不洁,或受到污染;⑦患者服用抗凝药物。

2检验人员因素
①对血样本离心时间和相对离心力选择不当;②未规定时间内对样本进行检测;③试剂复溶及预温时间不当;④试剂复溶时,稀释倍数不准;⑤实验人员对仪器操作方法不准确。

以上实验人员因素是引起实验结果不准确的重要来源。

3试剂因素
①试剂复溶时,稀释倍数不准或未使用I级试剂和厂家规定是稀释液;②试剂已被污染;③所用试剂超过了复溶后的稳定期。

不同厂家试剂的稳定性不同,尤其是凝血酶原时间检测,国际敏感指数(INI)的试剂影响较大。

以上试剂因素是影响检测结果的重要因素,未开启的试剂及干瓶复溶剂要严格按照使用说明书进行,确保试剂等符合测定要求。

冰冻保存的试剂不能反复冻溶,反复冻溶会使试剂中的因子失活而影响测定。

4仪器因素
主要是仪器的校正、保养和维护不当导致实验结果的不准,因此定时测试和校正仪器的性能,特别是对光源灯要定期检查、校正,避免因光源灯光学性能衰退对检测的影响,同时保持机械和光学附件的清洁。

以上影响凝血四项检测的主要影响因素。

因此为保证凝血四项检测结果的有效、稳定性及准确性,除克服上述因素外,各实验室还应用各自的方法,对一部分健康人进行检测,建立自己实验室的参考值,做好室内质控,同时提高检验人员的业务素质,要求检验人员对仪器的原理、性能、注意事项及影响因素等方面的知识
全面全面了解掌握,进一步保证检测结果准确性、可靠性。

相关文档
最新文档