LAPC2000型高效液相色谱仪验证方案(ACT)概要
液相计算机系统验证方案

D-2000 Elite HPLC System ManagerQualification ProtocolValidation No.:QC-V AL-151-01Suzhou Homesun Wanqing Pharmaceutical Co., LtdOct, 2010P r o t o c o l C o n f i r m a t i o nTABLE OF CONTENTS1.STATEMENT (1)2.SCOPE (1)3.REFREENCE DOCUMENTS (1)4.GLOSSARY (1)5.RESPONSIBILITIES (2)6.INSTALLATION QUALIFICATION (IQ) (3)7.OPERATION QUALIFICATION (OQ) (7)8. PERFORMANCE QUALIFICATION (PQ) (9)9. ATTACHMENT (11)1.StatementThe objective of this protocol is to define the main requirements and processes for qualification of D-2000 Elite HPLC System Manager used in the laboratory analysis. D-2000 Elite HPLC System Manager is a chromatography data system supported by Hitachi used to control the Techcomp LC2000 HPLC instrument which also supported from Hitachi, the software can execute analyzer’s input automatically including establishing analytical method, executing method program, collecting sample response, acquiring data, reprocessing data and initiating electronic records. Techcomp LC2000 HPLC located in laboratories (including QC laboratory and in-process control laboratory), have been completed IQ, OQ, PQ test and approved for use. In order to prove that D-2000 Elite HPLC System Manager is appropriate for GMP requirements on laboratory computerized system, QC department will execute the qualification.2.ScopeThis protocol applies to D-2000 Elite HPLC System Manager used to control Techcomp LC2000 HPLC.This protocol does not apply to LC2000 HPLC Qualification, it only contains applicable software qualification. The instrument qualifications are performed according to QC-V AL-105-01.Chromatography Data System Qualification and HPLC Qualification constitute good operation and maintenance in HPLC application.3.Reference Documents3.1GAMP Good Practice Guide: Validation of Laboratory Computerized Systems, 2005.3.2PDA technical report 31, ‘Validation and Qualification of Computerized Laboratory DataAc quisition Systems’, June 1999.3.3GMP Regulations (e.g. – 21 CFR 211.194, 91/356/EEC Ch 6, and ICH Q7A 6.6)3.4ICH Q7A – Good Manufacturing Practices for Active Pharmaceutical Ingredients, Chapter5.4 Computerized Systems4Glossary4.1 Computerized System: Computer system (hardware and software, including infrastructure)with associated functions and controlled process (equipment, users, procedures), in its operating environment4.2 Data backup: Copy of the data on a different medium (for example floppy disk, cartridge,CD-Rom) to allow data retrieval in case of a failure of the main storage medium.4.3 Data archival: Removal of some system data onto more durable and high capacity storagemedia (tapes, optical digital disks), necessary when the data is no longer required for routine work and data volumes are too great for the main storage medium.4.4 Audit trail: An audit-trail is a computer -generated recording of changes to GMP recordsindicating who created or changed the data, when, and if needed, why.4.5 Electronic Record (or Electronic Data): Combination of information in digital form that ismaintained by a computerized system and used for a GMP regulated activity.4.6 Applicable program: A document folder registered in D-2000 Elite HSM used to establishand storage analytical method, sample table, data and report. For example, create an applicable program <Sample A>, then authorized analyst could establish analytical method, sample table in this program, Moreover the analyst could use the method and sample table to initial origin data and report,5Responsibilities5.1System Administrator: To install the software per the user requirements, draft the validationprotocol and implement the validation, maintain the system to correct errors. Complete the protocol and generate a paper report for review and approval.5.2QC management (QC supervisor or manager): To ensure that the system is in compliancewith regulatory requirements and that the validation process is kept consistent with the protocol.5.3QA management (QA supervisor or manager: To ensure that documentation (includingprotocol and report) are consistent with applicable GMP regulations and to verify that the process comply with regulatory requirement.6Installation Qualification (IQ)Installation Qualification (IQ) must be performed to assure that the hardware and software components of the system are installed properly, and that the required documentation is available.6.3Verification of sensorsFor controlling the HPLC instrument from the PC, it is required to provide the USB interface board (IFB) and install the IFB driver software. Install the Analog Input Device (AID) and connect it with PC and HPLC per the following figure (see page 4):6.4Verification of instrumentsMake sure HPLC has been performed with IQ, OQ, PQ. Qualification meets the acceptance criteria, all instruments being used within the expiry dates.6.5Verification of software version6.6Verification of software installation procedureExecute D-2000 Elite HSM (HPLC System Manager) installation according to D-2000 Elite HSM Installation Manual, record the whole process to assure main sequence and detailed operation of installation maintaining consistency with vendor requirement.6.8Verify the configuration of backup systems6.9 IQ should be completed and approved for systems prior to commencing the OQ. Anydeviation should be recorded. A Summary Table should be created confirming that the IQ was performed according to plan and/or explaining any deviations discovered and their resolution or follow-up actions.Deviation description:Deviation resolution or follow-up action:IQ Conclusion:7 Operation Qualification (OQ)Operational Qualification (OQ) must be performed by User Representatives to assure testing of the system functions. OQ must focus on the GMP related function(s) of the system.7.17.3 Testing the defined user profiles“D-2000 Elite Group Security Protection ” Function has defined 3 groups for D-2000 Elite users with different levels. The function runs bind to <Microsoft Windows XP user and group management>Function, D-2000 user groups are created by Windows XP Administrator.7.4Testing the analytical method management functionsD-2000 Elite HSM could create a new analytical method in applicable program. In order to validate that the software could correctly execute the analyst’s setup. The checker will create a simulated analytical method in D-2000 Elite HSM, record the setup process and the related response. Document the execution procedure by pasting image in sequence to prove the validity for the analytical method setup process.7.5Testing the sample table management functionsD-2000 Elite HSM could create a new sample table in applicable program. In order to validate the software could correctly execute the analyst’s setup. The checker will create a simulated sample table in D-2000 Elite HSM, record the setup process and the related response. Document the execution procedure by pasting image in sequence to prove the validity for the sample table setup process.7.6 OQ should be completed and approved for systems prior to commencing the PQ. Anydeviation should be recorded. A Summary Table should be created confirming that the OQ was performed according to plan and/or explaining any deviations discovered and their resolution or follow-up actions.Deviation description:Deviation resolution or follow-up action:OQ Conclusion:8Performance Qualification (PQ)Performance Qualification (PQ) must be performed by instrument users to verify the performance of the system. This activity may be performed in actual environment. PQ must focus on the GMP related steps of the supported process.8.1Verification of authorized users and their corresponding rightsAfter installation of D-2000 Elite HSM in windows, three levels of applicable individuals with different rights were authorized to ensure laboratory data security.In OQ test (section7.2), The users “A”, “B”, “C”, “D”, “E”, “F”were simulated set up.Define “A”, “D”as administrator, “B”, “E”as developer, “C”, “F”as operator to test the rights with different grades. Record the results in the following chart.8.2Testing of sample and analytical method preparation8.3Testing of data acquisitionPlace sample (prepared in Section 8.2) in sample rack, prime the column to allow the chromatographic system to be stabilized, then start sample injection and data acquisition.Paste the observed data acquisition image to demonstrate that the system could execute D-2000 Elite HSM well. No error or deviation occurs.8.4 Testing that an operator can not delete raw data.After the raw data is created by D-2000 Elite HSM, perform the challenge to delete rawdata. Paste the image to document the process and result.8.5 Testing the full traceability of performed analytical methods (audit trail).Change the analytical method and save the updated method. Paste the images to document the process and result.8.6 Testing the full traceability of data creation and reprocessing (audit trail). Select a raw data and reprocess the data, Paste the images to document the result.8.7 Testing or documenting the traceability of equipment used (audit trail).Paste the images to document the traceability information.8.8 Testing data backup and retrievalBackup data into another medium and perform the data retrieval. Paste the images todocument the process and results.8.9 Testing of results printout and comparison of the printout with the screen image.8.10 Testing that all critical parameters and data that can influence the result are printed out.8.11 Testing that the system will not allow any user to alter data in print and thus to printerroneous results.8.12 Manual re-calculation of automated resultsManually measure the T and W h/2, then calculate the theoretical plates. The manualcalculated results are compared to that generated by D-2000 Elite HSM.The number of theoretical plates, N, is a measure of column efficiency, it is calculated byequation:Where T is the retention time of the substance and W h/2 is the width of the peak at its halfheight.Deviation description:Deviation resolution:PQ Conclusion:9Attachment。
高效液相色谱仪验证方案

高效液相色谱仪验证方案引言高效液相色谱(HPLC)是一种常用的分析技术,它在药物分析、环境监测、食品安全等领域具有广泛的应用。
为了确保HPLC仪器的准确性和可靠性,以及测试结果的可信度,对HPLC仪器进行验证是非常重要的。
本文将介绍一种高效液相色谱仪验证方案,以确保仪器的正常运行和测试结果的准确性。
1. 验证目的HPLC仪器验证的主要目的是评估仪器是否满足预定的性能要求,包括准确性、精密度和线性范围等指标。
通过验证,可以确保仪器在正常使用过程中能够提供准确和可靠的测试结果,以满足相关的法规和质量标准要求。
2. 验证内容HPLC仪器验证的内容包括以下几个方面:2.1 仪器安装和传感器校准在验证之前,首先需要确保HPLC仪器已正确安装,并且各个传感器和检测器已进行校准。
校准过程应按照仪器的操作手册进行。
2.2 仪器性能参数验证仪器性能参数验证是验证HPLC仪器在运行过程中是否符合规定的性能要求。
主要包括以下几个方面:•准确性验证:通过添加已知浓度的标准溶液,并测定其浓度来评估仪器的准确性。
•精密度验证:通过重复测定同一样品,评估仪器的精密度。
可以使用相对标准偏差(RSD)来评估测量结果的一致性。
•线性范围验证:通过逐渐增加样品浓度,测定仪器的线性范围。
应选取不同浓度的标准溶液进行测试,并绘制曲线来评估仪器的线性关系。
2.3 方法验证方法验证是验证HPLC方法是否可用于定量分析的过程。
主要包括以下几个方面:•特异性验证:通过检测样品中其他成分的干扰来评估方法的特异性。
可以使用纯溶液和样品添加物进行测试。
•精密度和重复性验证:通过重复测定同一样品,评估方法的精密度和重复性。
可以使用RSD来评估测量结果的一致性。
•准确性验证:通过添加已知浓度的标准溶液,并测定其浓度来评估方法的准确性。
3. 验证计划为了有效进行HPLC仪器的验证,需要制定详细的验证计划。
验证计划应包括以下几个方面:3.1 验证目标和范围明确验证的目标和范围,确定需要验证的仪器性能参数和方法。
高效液相色谱仪确认方案

高效液相色谱仪确认方案
齐全,要求内容详尽,优秀的文档可获得好的分数。
一、引言
随着制药行业的发展,高效液相色谱仪确认方案已经在GMP(Good Manufacturing Practice)准则中得到重视,从而成为GMP的一部分。
GMP是一种以确保制剂质量、安全性和有效性为基础的生产质量管理系统,可以确保制药厂在生产过程中符合有关药品的质量标准,以保证其安全有
效性,符合法律和规章的要求。
二、高效液相色谱仪确认方案
1.目的和范围
本确认方案的目的是为了确保制药厂在使用高效液相色谱仪质量控制
和分析的过程中符合GMP要求。
本确认方案所涉及的范围是该设备的安装、校准、维护和使用。
2.建议
2.1设备
(1)设备必须符合有关国家有关质量标准;
(2)设备必须具有相应的技术特性,可用于质量控制和分析;
(3)设备必须具有可靠的记录。
2.2安全性
(1)高效液相色谱仪在使用过程中应确保安全;
(2)为了保护用户及周围环境,应遵循安全操作规程,并配备必要
的安全设施,包括使用压缩空气、保护眼睛、严格控制电源连接;
(3)操作人员应遵守严格的安全防护措施,并配备必要的安全用具,包括安全帽、安全衣、安全鞋等。
LC-2000液相验证方案

QC-V AL-106-01高效液相Waters 26953Q验证方案验证编号:QC-V AL-106-01苏州弘庆药业有限公司QC-V AL-106-01验证方案确定QC-V AL-106-01目录1.概述 (1)2.安装确认 (1)2.1购买定单确认 (1)2.2设备规格确认 (2)2.3软件版本确认 (3)2.4设备手册确认 (4)2.5部件清单确认 (4)2.6公用设施确认 (5)2.7硬件环境与安装确认 (5)2.8软件配置确认 (6)2.9维修服务确认 (6)3.运行确认 (7)3.1试验项目及认可的质量标准 (7)3.2试验方法及结果 (8)3.2.1 基线性能及波长范围 (8)3.2.2 波长准确度 (9)3.2.3 最低检测限 (10)3.2.4 泵的流量 (11)3.2.5 梯度洗脱比例 (12)3.2.6 重现性 (14)3.2.7 柱温箱稳定性及设定值误差 (15)3.2.8 线性 (16)4.性能确认 (17)5.标准操作规程确认 (18)6.确认周期 (18)7.偏差记录 (18)QC-V AL-106-011.概述LC2000高效液相色谱仪由LC2130输液泵(四元梯度通道),L-2200自动进样器,LC2030紫外可见检测器、LC2060柱温箱、计算机、打印机组成,样品经进样器在高压泵的作用下随流动相进入色谱柱,由于吸咐系数的不同,样品各组分被分离。
在紫外可见检测器的检测下得到色谱图,通过数据处理系统将色谱图处理、贮存从而达到分析的目的。
LC2000所使用的T2000P软件能够把测量的数据转化为图形,并能够记录、测量和打印出来。
确认目的:通过对LC2000高效液相色谱仪的安装确认、运行确认、性能确认,保证该仪器在验证周期内符合检测要求。
2 安装确认2.1购买定单确认接受标准:采购订单存在,且所购仪器与订单一致。
记录人日期复核人日期第1 页共19 页QC-V AL-106-012.2 设备规格确认目的:确保仪器是存在的,并按照生产商和系统规格标准和设计的要求进行安装的。
高效液相色谱仪验证方案

高效液相色谱仪验证方案验证文件类别:技术标准编号:V-A-c-004-0部门:验证委员会页码:共1页,第1页版次:¨新订¨替代:起草:年月日部门审核:年月日审阅会签:(验证委员会)批准:年月日实施日期:年月日复印数:批准:分发至:目录1.设备基本情况1.1概述1.2基本情况3职责3.1验证委员会3.2工程处3.3Qc4验证内容4.1安装确认4.1.1安装确认所需文件资料 4.1.2关键性仪表及消耗性备品4.1.3评价仪器的安装是否符合GmP及供应商提议的要求4.1.4起草标准操作规程4.2校正4.2.1紫外检测器4.2.1.1固定波长紫外检测器波长示值误差的校正4.2.1.2可调波长紫外—可见光检测器波长示值误差和重复性误差的检定4.2.2高压泵4.3运行确认4.3.1紫外检测器4.3.2高压泵4.4性能确认4.4.1最小检测浓度4.4.2定性、定量测量的重复性4.5拟订再验证项目及周期4.6验证结果评定与结论1.设备基本情况1.1概述本高效液相色谱仪主要由一个二元泵、一个柱温箱、一个可变紫外检测器、一个带有20μL定量环的进样器和一台工作站组成,各部分由工作站软件控制。
本仪器主要用于分析原料和成品的有关物质和含量。
1.2基本情况设备编号:设备名称:高效液相色谱仪型号:生产厂家:供货厂家:安装日期:使用部门:工作间:保管员:2.验证目的为确认高效液相色谱仪测试数据准确可靠,性能稳定,特制订本验证方案,对高效液相色谱仪进行验证。
验证过程应严格按照本方案规定的内容进行,若因特殊原因确需变更时,应填写验证方案变更申请及批准书,报验证委员会批准。
3职责3.1验证委员会1.负责验证方案的审批。
2.负责验证的协调工作,以保证本验证方案规定项目的顺利实施。
3.负责验证数据及结果的审核。
4.负责验证报告的审批。
5.负责再验证周期的确定。
3.2工程处1.负责设备的安装、调试,并做好相应的记录。
LAPC2000型高效液相色谱仪验证方案(ACT)概要

验证文件类别:验证方案编号:STP-VMP-xxx 部门:质量保证部页码:共9页LA PC2000型高效液相色谱仪验证方案版次:第0版起草:年月日审核:年月日验证小组会签:年月日批准:年月日实施日期:年月日复印数:2份分发至:质量保证部、设备部关于成立LA PC2000型高效液相色谱仪验证小组的通知为了加速我企业GMP认证的步伐。
顺利的完成验证工作,特成立LA PC2000型高效液相色谱仪验证小组,小组成员如下:组长:组员:验证小组的职责:1、负责起草验证方案、报告,确定再验证周期,报质量保证部审核,总工程师批准。
2、组织质量保证部及有关部门对本方案进行实施。
本通知自下发之日起,各有关部门要积极配合验证小组的工作,争取早日圆满完成验证工作。
质量保证部2002年7月5日目录一、概述 (1)二、验证内容 (1)三、职责 (1)四、安装确认(IQ) (2)五、运行确认(OQ) (4)六、再验证 (7)七、验证结果 (7)八、LA PC2000型高效液相色谱仪验证报告(附表1) (8)九、验证合格证书(附表2) (9)一、概述1.1 设备概述:本仪器的输液泵为单泵恒流结构,泵压平稳,流速可在0-10ml/min之间任意选择,为触摸键盘输入式,为单片机控制。
检测器为单波长紫外可见检测器,波长可在190-800nm之间任意可调,面板为旋钮式和开关式控制。
有数码管显示窗,可观察氘灯的起辉情况和AU值的升降情况。
工作站运用的是美国LA公司提供的“ANASTAR”软件,此软件独立运行,热行处理色谱数据功能。
本仪器主要用于xxx的含量及杂质检测。
二、验证内容本方案主要包括LA型高效液相色谱仪的安装确认(IQ),运行确认(OQ)。
三、职责3.1 验证小组⑴起草验证方案及验证报告,确定再验证周期,报质量保证部审核、总工程师批准。
⑵组织、协调使用单位对验证方案进行实施。
3.2 质量保证部⑴根据验证对象成立验证小组。
⑵负责验证方案及报告的审核。
EC-2000型高效液相色谱仪操作说明.ppt

AUTO ZERO 自动校零
样品池换上样品溶液,放入样品架
START 测定
如需连续读取数次测定值,可连续按 START PRINT ,可将屏幕上内容打印出。
,再按
*作说明
1.3.1.选择主菜单[1]SPECTRUM,按1。
1.3.2.设定MEASURE RANGE,λS,λE,QVRLY/SEQ, SCAN SPEED,REPEAT(T,N),SLIT等参数,按 NEXT 。
1.3.3.按 START ,光谱扫描,扫描完毕按 NEXT 进入数据 程序,可根据要求对光谱进行数据处理,最后用X-Y绘图仪打出图谱 和数据。具体操作如下:
*
MAIN MENU 主菜单
EC-2000型高效液相色谱仪操作说明
进入光谱测定菜单
1
移动光标设定参数
MEASURE选择ABS,或T%,或EMERGY,或R%
*
EC-2000型高效液相色谱仪操作说明
2.2.在软盘驱动器上红灯亮时,不能关闭电源和取出操作软盘,否则 操作软盘内存将被破坏。
2.3.操作软盘插入和取出时,动作要轻,测定时操作软盘的保护开关 应在off位置,测定完毕后应将操作软盘保护开关放在on位置上,放在塑 料盒内保存。
2.4.操作软盘的保存 操作软盘应保持在远离磁场处,不能接触稀释剂氟利昂、酒精等有
EC-2000型高效液相色谱仪操作说明
起动测定
NEXT
2 PLOTOUT
进入数据处理屏幕,有ZOOM,PEAK PRINT CAL(+-*/log),T/A,DEAIV, ENT
等功能。 如需检出峰。
2=PEAK 输入合适的峰阀值(THRESHOLD)
打印出所需图谱及峰、谷波长以及吸收值
高效液相色谱仪验证方案

高效液相色谱仪验证方案××××××饮片企业GMP管理文件目的:此验证是为确认高效液相色谱仪的购置、安装、运行和各种性能指标是否满足规定要求。
范围:证明并保证经本仪器检验后,中间产品、成品符合质量要求。
责任:验证小组负责对此仪器的验证,操作员工进行调试。
内容:1 设备描述1.1 设备概述:高效液相色谱仪主要由主机、检测器、工作站等组成。
高效液相色谱法是用高压输液泵将具有不同极性的单一溶剂或不同比例的混合溶剂,缓冲液等流动相,用泵压入装有固定相的色谱柱,经过样阀注入供试品,由流动相带入柱内,在柱内各成分被分离后,依次进入检测器,色谱信号由工作站积分记录。
1.2 设备名称:高效液相色谱仪1.3 设备型号:Waters 26901.3 设备编号:#A02SM7041M1.4 制造商:×××1.5 地址:×××1.6 电话:×××2 预确认通过调研分析和对供应商进行资质审计,最终选定了美国沃特斯公司生产的Waters 2690高效液相色谱仪。
该仪器的主要技术参数为:3 安装确认本仪器主要用于药品的含量测定、有关物质检查、杂质限度检查和鉴别等。
3.1 文件资料:表8-19所列文件资料齐全,并符合GMP要求。
表8-19安装确认所需文件资料结论:确认人:复核人:日期3.2 维修服务服务单位名称:×××地址:×××联系人:×××电话:3.3 消耗品备件(表8-20)〕表8-20消耗品备件清单3.4 一般状况3.4.1 仪器的所有其他方面是否均符合采购质量标准?是否确认人:日期:3.4.2 仪器的安装是否符合要求?仪器周围有足够的空间可进行实验的操作,安装地点为平坦、稳固的实验台。
- 1、下载文档前请自行甄别文档内容的完整性,平台不提供额外的编辑、内容补充、找答案等附加服务。
- 2、"仅部分预览"的文档,不可在线预览部分如存在完整性等问题,可反馈申请退款(可完整预览的文档不适用该条件!)。
- 3、如文档侵犯您的权益,请联系客服反馈,我们会尽快为您处理(人工客服工作时间:9:00-18:30)。
验证文件
类别:验证方案编号:STP-VMP-xxx 部门:质量保证部页码:共9页
LA PC2000型高效液相色谱仪验证方案
版次:第0版
起草:年月日
审核:年月日
验证小组会签:
年月日
批准:年月日
实施日期:年月日
复印数:2份
分发至:质量保证部、设备部
关于成立LA PC2000型高效液相色谱仪验证小组的通知
为了加速我企业GMP认证的步伐。
顺利的完成验证工作,特成立LA PC2000型高效液相色谱仪验证小组,小组成员如下:
组长:
组员:
验证小组的职责:
1、负责起草验证方案、报告,确定再验证周期,报质量保证部审核,总工程师批准。
2、组织质量保证部及有关部门对本方案进行实施。
本通知自下发之日起,各有关部门要积极配合验证小组的工作,争取早日圆满完成验证工作。
质量保证部
2002年7月5日
目录
一、概述 (1)
二、验证内容 (1)
三、职责 (1)
四、安装确认(IQ) (2)
五、运行确认(OQ) (4)
六、再验证 (7)
七、验证结果 (7)
八、LA PC2000型高效液相色谱仪验证报告(附表1) (8)
九、验证合格证书(附表2) (9)
一、概述
1.1 设备概述:
本仪器的输液泵为单泵恒流结构,泵压平稳,流速可在0-10ml/min之间任意选择,为触摸键盘输入式,为单片机控制。
检测器为单波长紫外可见检测器,波长可在190-800nm之间任意可调,面板为旋钮式和开关式控制。
有数码管显示窗,可观察氘灯的起辉情况和AU值的升降情况。
工作站运用的是美国LA公司提供的“ANASTAR”软件,此软件独立运行,热行处理色谱数据功能。
本仪器主要用于xxx的含量及杂质检测。
二、验证内容
本方案主要包括LA型高效液相色谱仪的安装确认(IQ),运行确认(OQ)。
三、职责
3.1 验证小组
⑴起草验证方案及验证报告,确定再验证周期,报质量保证部审核、总工程师批准。
⑵组织、协调使用单位对验证方案进行实施。
3.2 质量保证部
⑴根据验证对象成立验证小组。
⑵负责验证方案及报告的审核。
⑶负责LA PC2000型高效液相色谱仪再验证周期的确认。
1/9
⑷负责发放验证合格证书。
⑸组织验证小组对参与验证工作的相关人员进行培训。
⑹负责设备操作、维护保养及清洁的具体实施。
3.3 设备部
⑴负责组织设备安装。
⑵负责LA PC2000型高效液相色谱仪的校正。
⑶负责检修方案的制订,建立设备操作、维护保养及清洁规程。
四、安装确认(IQ)
检查人:复核人:日期:
4.2 维修服务
服务单位名称:2/9 地址:
联系人:
电话号码:
4.3技术特性
检查人:复核人:日期:
检查人:复核人:日期:
3/9 4.4一般状况
4.4.1 仪器其它方面是否符合采购质量标准?
4.4.2 仪器的安装是否符合本公司GMP及供货商提议的要求?
4.5安装检查
检查人:复核人:日期:
五运行确认(OQ)
5.1目的:在不使用任何试品的前提下,确认该仪器达到设计要求。
4/9
检查人:复核人:日期:
5.3泵流量设定值误差S S和流量稳定性误差S R的测试方法及认可标准:
其中S S=(F m-F S)/F S×100% S R=(F max-F min)/F×100%
F m=(W2-W1)/ρt·t为流量实测值(ml/min) W2:容量瓶+流动相的重量(g) F S:流量设定值W1:容量瓶的重量ρt:实验温度下流动相的密度(g/cm3)
t:收集流动相的时间F max:同一组测量中流量最大值(ml/min)
F min:同一组测量中流量最小值(ml/min) F:同一组测量值的算术平均值(ml/min)
检查人:复核人:日期:
5.4 定性定量重复性误差RSD的测试方法和认可标准:
RSD= n 1
√〔∑(Xi-X)2〕/(n-1)××100%
i=1 X
公式中:Xi—第i次测得的保留时间或峰面积;
X—n次测得结果的算数平均值;
i—测量序号;
检查人:复核人:日期:6/9
5.5基线漂移和基线噪声
检测器波长为254nm,流动相为100%甲醇,流速为1.0μl/min,吸收度范围最灵敏,待基线稳定后,记录40分钟。
-4-3
检查人:复核人:日期:
5.6校正:
由技术监督部门进行外观、输液、柱恒温箱温度设定值误差△T、定性定量测量重复性,检测器性能的检定和校正。
六、再验证
在下述情况下对检验仪器进行再验证。
6.1 仪器更新或大修。
6.2 对检验结果作趋势分析,发现系统性偏差。
6.3 对检测条件发生和变更的。
6.4 使用时间两年进行再验证。
七、验证结果
7.1 验证小组负责测试记录,并起草验证报告(附表2),报质量保证部。
7.2 质量保证部负责对验证报告进行审核,发放验证合格证书(附表3),
确认验证的周期。
7.3验证过程应严格按照本方案规定的内容进行,若因特殊原因确需变更时,应报质量保证部审核,总工程师批准。
7/9 附表1:
8/9
附表2:
验证合格证书
★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★
验证方案名称:LA PC2000型高效液相色谱仪验证方案
验证方案编号:
该仪器按本方案规定项目进行了验证,验证合格批准使用。
验证报告名称:LA PC2000型高效液相色谱仪验证报告
验证报告编号:
有效期:
批准:
年月日备注:
1、该仪器适用于在当前验证条件下使用,当使用条件发生变更时,应报质量保证部审核,必要时重新验证。
2、当仪器进行检修时重新进行验证。
3、LA PC2000型高效液相色谱仪应在批准的操作、维护保养、清洁规程下进行操作、维护保养及清洁。
9/9
验证文件
类别:验证报告编号:STP-VMR-xxxx 部门:质量保证部页码:共7页
LA PC2000型高效液相色谱仪验证报告
版次:第0版
起草:年月日
审核:年月日
验证小组会签:
年月日
批准:年月日
目录
一LA PC2000型高效液相色谱仪验证报告 (1)
二、仪器基本情况确认记录 (2)
三、资料档案确认记录 (2)
四、维修服务确认记录 (2)
五、关键性仪表确认记录 (3)
六、消耗性备件确认记录 (3)
七、安装确认记录 (4)
八、测试项目的测试方法及认可标准确认记录 (5)
九、泵流量设定值误差S S和流量稳定性误差S R测试方法及认可标准确认记录 (6)
十、定性定量重复性误差RSD的测试方法和认可标准确认记录 (6)
十一、基线漂移和基线噪声确认记录 (7)
十二、检定确认: (7)
1/7
检查人:复核人:日期:
检查人:复核人:日期:
维修服务确认记录
服务单位名称:
地址:
联系人:
电话号码:
2/7
检查人:复核人:日期:
检查人:复核人:日期:
3/7
检查人:复核人:日期:
4/7 测试项目的测试方法及认可标准确认记录
检查人:复核人:日期:
5/7泵流量设定值误差S S和流量稳定性误差S R
检查人:复核人:日期:
检查人:复核人:日期:
6/7 基线漂移和基线噪声确认记录
检查人:复核人:日期:
检查人:复核人:日期:
7/7
验证合格证书
★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★★
验证方案名称:LA PC2000型高效液相色谱仪验证方案
验证方案编号:STP-VMP-xxxx
该仪器按本方案规定项目进行了验证,验证合格批准使用。
验证报告名称:LA PC2000型高效液相色谱仪验证报告
验证报告编号:STP-VMR-xxx
有效期:一年
批准:
年月日
备注:
1、该仪器适用于在当前验证条件下使用,当使用条件发生变更时,应报质量保证部审核,必要时重新验证。
2、当仪器进行检修时重新进行验证。
3、LA PC2000型高效液相色谱仪应在批准的操作、维护保养、清洁规程下进行操作、维护保养及清洁。